Section 4
Basic Operation
6
4
CooperSurgical, Inc.
Temperature Setting and Control
The display indicates the average temperature of the bottom of chamber. The displayed
temperature is updated 4 times every second.
Action Key
Press set-point key until set temperature is shown.
Press and hold set-point key.
Press arrow keys to raise or lower the temperature.
When the desired temperature is reached, let go of the set-point key.
Note: The set-point can be adjusted to a reading between ambient and 42.0˚C.
Combination Keys
Action Key Display
Press alarm and UP key. The keyboard is locked.
Press alarm and arrow UP key. The keyboard is unlocked.
Press Heat ON/OFF and set-point
key.
Switches between temperature and
time.
Temperature Alarm
The temperature alarm indicates too high or too low temperature. The alarm will be activated if
the temperature rises or falls more than 0.5°C from the set point. There will be an acoustic signal
and the red LED light in the alarm key will be activated.
Action Key
Press alarm key to mute the acoustic signal.
Press alarm key once again to bring the signal back on.
Note: The red LED light and the acoustic signal will turn o when the temperature stabilizes to
the set-point level
+
+
+