Copal PG-200 User manual

DIGITALPRESSUREGAUGE
INSTRUCTION MANUALVer.5.1
PG-200
1
¸Model
¹Measurement……………Gaugepressure.(102AP:absolute)
ºIndication………………3-1/2digits,000〜1999Max,digitalLCDdisplay.
»Indication………………Around3times/sec.
¼Accuracy…………………±0.5%F.S.±2digit(at25℃±5℃)
½Temperaturecharacteristics
atzeropoint;±0.05%F.S./℃±2digit(102GP、103GP、103GMP、102VP)
±0.1%F.S./℃ ±2digit(101GP、102AP)
inSPAN;±0.05%Reading/℃ ±2digit
¾Switchingfeatures
Numberofsetpoints…2,HIandLO,activatidatupperlimit.
Setrangedivisions……000〜1000,HIandLO.
Settingmethod…………WithtrimmereachforHIandLO.
Setvalueindicationcanbemadetotimewithsettingswitch.
Operatingaccuracy
……Within±3divisionsfromindicationvalue.
Difference
………………Within4divisions,fixed.
Outputsystem…………NPNopencollectorof30Vdcand40mAatmax.
Operatingindication……HinHImodeorLinLOmodeatONoutput.
¿Operatingtemperature…0〜50℃
humidity…35〜85%RH,nosweatingallowed.
ÀStoragetemperature…−20〜70℃,withlowerhumiditythan65%RH.
ÁNetweight………………Netweight,paneltype:Around130g.
Stand-alonetype:Around330g.
ÂPressureport……………Paneltype:M5female.
Stand-alonetype:R1/4(PT1/4).
ÃAttachments
……………Onetouchjoint(foratubeof6mmO.D.)andMiniaturefitting(foratubeof4mmI.D.)
(Paneltypeonly)
Seriesname
Pressurerange(Table1)Type(Table2)
PG−200− −
Nishi-ShinjukuKimurayaBidg.,7-5-25
Nishi-ShinjukuShinjuku-kuTokyo160-0023,JapanPhone.:(03)3364-7055
Formoredetailedinformationpleaseaskforthe
nearestdistributororthefollowingsalescenter.
§
1. Specification
Thankyouforpurchasinga
NIDECCOPALELECTRONICSCORP.product.
Forproperandoptimaluseoftheproduct,pleasereadthis
manualthoroughlybeforeusing.
Keepthismanualforfuturereference.
*1:102AP…kPaabs *2:103GMP…MPa
*See
§3TerminalWiring
PRESSURERANGE MAX
PRESSURE
Break-down
Pressure
RESOLUTION
(kpa)
102VP
102AP
102GP
103GP
103GMP
101GP
0〜−100.0
0〜100.0
0〜100.0
0〜1000
0〜1.000
0〜10.00
200
200
200
1500
1.5
20
500
500
500
2000
2.0
50
0.1
0.1
0.1
1
0.001
0.01
Table1
※1
※2
*
Panel
Stand
−P
−1−P
−3−P
−S
−1−S
−3−S
HI、LO
HI、LO
HI
HI、LO
HI、LO
HI、LO
−
−
Voltage
−
−
Voltage
Internallithiumbattery
External5〜24Vdc
Sameasabove
Internallithiumbattery
External5〜24Vdc
Sameasabove
Table2
TYPE POWERSOURCE
SWITCH
OUTPUT
OPTIONAL
OUTPUT
*

4
¸MakewiringbyfollowingSection3,TerminalWiring.
¹TurnthePOWERswitchon.Fortheexternalpowertype,supplytoitthepowerindicatedonthenameplate.
ºCheckto insurethatthe pressureindicationis zerowithno pressureapplied.The zeroindicationmay bechangedslightly withpostureofthe mainbody.If the
indicationisover5counts,makezeroadjustmentwiththeZEROadjustingtrimmer.
»MakeHIandLOsettingswiththeSETswitchandHIandLOsettingtrimmers(seeSection5,step5).
¼ApplyapressuretothePressureGauge.Thenitwillindicatethemeasuredvalue.
½Iftheswitchoutputison,thePressureGaugewillshowLorH.
¸Theproduct,whichisaprecisioninstrument,mustbetakenspecialcarenottobedamagedbyimpactnorbybeingdroppedwhenitistrasportedandinstorage.
¹Theproductmustbeavoidedinstoragewhereisdusty,drippingandvibrated.
CopalElectronicswarrantstheproductfortheperiodofoneyearafterthedateofthecustomersreceipt.Wewillrepairthetroubledproductscausedbyourimproper
designingand/orproductioncontrolatourcost.Ourwarrantyislimitedtotheproductsonly,notonanotherdamagethatiscausedbytheproductmalfunction.
Thefollowingdamage/replacementarerepaired/takenatthecustomerscharge;
¸Failureordamagecausedbyunsuitableorwrongusage,modificationorrepairbythecustomer.
¹FailureordamagecausedbydisasteroranActofGod.
ºReplacementofattachments(HalfUnionandBarbJoint).
¸Mountingtheattachmentforpaneltype
Iffightenigtheattachmenttothepressureport,thetighteningtorqueshouldbearound1N・m.
¹Pipingforstand-alonetype
Forpiping,usethehexagonalportionofthepressureportfordriving.
ºMediatobeused
ThePressureGaugeshouldbeusedforfluidsthatcannotcorrodethediaphragmofSUS316L,pressureportofSUS316andO-ringoffluororubber.
»Excessivepressure
The pressure to be measured is within the specified range. Measurement of a higher pressure than the maximum allowable one may cause performance
degradationordestruction.
¼Internalbattery
1.ThebatterytobeusedisanER6VMlithiumoneof3.6V.Tohaveit,orderaagencyfromwhichyoupurchasedthemainbody.
2.ThePressureGaugewillshowBATTifthebatterypowerdecreases.Thenreplaceitimmediatery.
3.Donotdisassemble,short-circuit,change,putintofire,norheatthebattery.
½Maintenance
IfthePressureGaugegetsdirty,wipeitoffwithhardlysqueezedclothcontaininganeutraldetergent.Donotuseanythinnerandbenzine.
¾Protectionagainstnoises
1.NoiseshavingmixedintheswitchoutputorpowerlinemaycausethePressureGaugetochangethepressureindication,malfunctionorbebroken.
TakesomemeasuretoputthePressureGaugeawayfrompowerlineoruseshieldedwire.ItiseffectivetogroundthePressureGauge.
2.Iftheswitchoutputhasaninductionloadsuchasrelayorsolenoidconnectedtoit,itshouldhaveasurgeabsorbercircuitputinit.Therelaycontactsshouldhave
acontactpreventivecircuitputinittopreventnoisesfrombeinggenerated.
¿Prohibitionofshort-circuitingtheswitchterminals
Donotshort-circuittheswitchoutputterminaltoanypowerterminal.Theinternalcircuitmaybebroken.
§
6. Howtouse
§
7. HandlingNote
§
8. TRANSPORTATION&STORAGE
§
9. WARRANTY

3
Switchoutputofnpnopencollectortypewith30Vdcand40mAatmaximum.
¸Frontpanelcontrols
¹Descriptionofthefeatures
1.
2.
3.
4.
5.
6.
7.
8.
9.
Pressureindicator
Polarityindicator
Pressureport
POWERswitch
SETswitch
HIsettingtrimmer
LOsettingtrimmer
SPANadjustingtrimmer
ZEROadjustingtrimmer
Showsapressurevalueenteredtothepressureport.Alsoshowsasetvaluegivenwiththesetswitch.
Showsthenegative(−)uponnegativepressure.
Hasapressureapplied.
TurnsonoroffthepowertothePressureGauge.
SelectsamodetocheckaHIorLOsetvalue.IfyoupresstheSETswitchonce,theindicationwillbeautomaticallychangedin
thesequenceshownbelow.TochecktheLOsetvalue,waitforaround20secondsafterpressingtheSETswitch.
Ifsettingendsortointerruptsettinginthecourse,presstheSETswitchagain.Thenthepressurewillbeindicated.
UsedtoadjustpressuretotheHIsetvaluethatisindicatedwiththeSETswitch.
UsedtoadjustpressuretotheLOsetvaluethatisindicatedaftertheHIsettingtimehaselapsed.
Usedtoadjustthesensitivitysothatthepressurecanindicatethefullscalewiththeratedpressureappliedtothepressureport.
Asthetrimmeriscalibrated,itissealedup.
Usedtoadjustthesensortozero.Thepressureindicationshouldbeadjustedto000withnopressureapplied.
§
4. OutputType
§
5. FrontPanelControlsandFeatures
Powersource(+)
Power(−)source
Transistorprotectivediode
BC
E−
+
Exampleofrelay
connection
Frontcontrols
Pressureindicator
Batterychange
indicator
Polarityindicator
Switchingindicator
Decimalpoint
SETswitch LOsetting
trimmer HIsetting
trimmer
POWERswitch SPANadjusting
trimmer
ZEROadjusting
trimmer
Indicationmode
Settingmode
Around20sec Around20sec
NO
YES
SET
Pressure
indicaton HIsetting LOsetting

[Stand-alonetype]
[Paneltype] [Stand-alonetype]
ProceedonthePCboadinsidethemainbody.
The external power models have
powervoltageof10-24Vdcsetat
default. To operate your Pressure
Gaugewith 5-10 Vdc, change the
internalterminalblockwiring.
・COM.
・SW.OUTHI
・COM.
・SW.OUTLO
・5〜10V
・COM.
・10〜24V
Externalpowermodels
}
+
−
Select one of the terminal blocks for connection depending on the power
voltage.Donotapplyanyvoltageexceedingthevoltagerange.Foroutput
wiring,usethinnerwirethanAWG26.
−
−
+
2
[Paneltype]
ÄOptions(seeTable2)……
a.Internalbattery
Type…Lithiumbattery,ER6VM.
Serviceabletime…Oneyear(9000hours),continuouslyoperable.
HasabatterychangeindicatorBATT.
b.Externalpowersource
Inputvoltage
…………5〜10or10〜24Vdc(seeTable2).
Currentconsumption…Lessthan20mA.
c.Analogvoltageoutput
Outputvoltage
………0〜1V,notisolated.
Accuracy ………………±5%F.S.,whichisaddedtomainaccuracy.
Loadresistance
………Higherthan1kohm.
¸Mountingontopanel
Cuta hole out as shown in thepanel cut drawing above.Forcibly put the
mainbodyintothecuthole.Thehookwillfixthemainbodytothepanel.
¹Removingoutofpanel
Drawthemainbodyoutofthepanelwhilenippingthehookbetweenyour
thumbandmiddlefingers.
Atdefault
Black Red
Battery
(−P)
Exterminalpower
(−3−P)
1
SwitchoutputHI
Voltageoutput
SwitchoutputLO
SwitchoutputHI
Powersource
23456
Exterminalpower
(−1−P)
SwitchoutputHI SwitchoutputLO
Powersource
+−+−
+−+−+−
+−+−+−
5〜10V 10〜24V
−
COM.
Terminalblockof6pins
・COM.
・Outputvoltage
−
+
§
2. ExterminalDimensions tolerance(±0.5mm)
§
3. TerminalWiring
72
40
69.5 +0.5
0
68.5 +0.5
0
Unitofdisplay
Pressureport
M5,depth5
Panelcut
Panelthickness
(1〜2.5mm)
Panelmontnghook
13
72
72
40
13.5
R1/4
(PT1/4)
M5(P=0.8)depth6
Unitofdisplay
11.5
7222
13.5
28.5