4
¸MakewiringbyfollowingSection3,TerminalWiring.
¹TurnthePOWERswitchon.Fortheexternalpowertype,supplytoitthepowerindicatedonthenameplate.
ºCheckto insurethatthe pressureindicationis zerowithno pressureapplied.The zeroindicationmay bechangedslightly withpostureofthe mainbody.If the
indicationisover5counts,makezeroadjustmentwiththeZEROadjustingtrimmer.
»MakeHIandLOsettingswiththeSETswitchandHIandLOsettingtrimmers(seeSection5,step5).
¼ApplyapressuretothePressureGauge.Thenitwillindicatethemeasuredvalue.
½Iftheswitchoutputison,thePressureGaugewillshowLorH.
¸Theproduct,whichisaprecisioninstrument,mustbetakenspecialcarenottobedamagedbyimpactnorbybeingdroppedwhenitistrasportedandinstorage.
¹Theproductmustbeavoidedinstoragewhereisdusty,drippingandvibrated.
CopalElectronicswarrantstheproductfortheperiodofoneyearafterthedateofthecustomersreceipt.Wewillrepairthetroubledproductscausedbyourimproper
designingand/orproductioncontrolatourcost.Ourwarrantyislimitedtotheproductsonly,notonanotherdamagethatiscausedbytheproductmalfunction.
Thefollowingdamage/replacementarerepaired/takenatthecustomerscharge;
¸Failureordamagecausedbyunsuitableorwrongusage,modificationorrepairbythecustomer.
¹FailureordamagecausedbydisasteroranActofGod.
ºReplacementofattachments(HalfUnionandBarbJoint).
¸Mountingtheattachmentforpaneltype
Iffightenigtheattachmenttothepressureport,thetighteningtorqueshouldbearound1N・m.
¹Pipingforstand-alonetype
Forpiping,usethehexagonalportionofthepressureportfordriving.
ºMediatobeused
ThePressureGaugeshouldbeusedforfluidsthatcannotcorrodethediaphragmofSUS316L,pressureportofSUS316andO-ringoffluororubber.
»Excessivepressure
The pressure to be measured is within the specified range. Measurement of a higher pressure than the maximum allowable one may cause performance
degradationordestruction.
¼Internalbattery
1.ThebatterytobeusedisanER6VMlithiumoneof3.6V.Tohaveit,orderaagencyfromwhichyoupurchasedthemainbody.
2.ThePressureGaugewillshowBATTifthebatterypowerdecreases.Thenreplaceitimmediatery.
3.Donotdisassemble,short-circuit,change,putintofire,norheatthebattery.
½Maintenance
IfthePressureGaugegetsdirty,wipeitoffwithhardlysqueezedclothcontaininganeutraldetergent.Donotuseanythinnerandbenzine.
¾Protectionagainstnoises
1.NoiseshavingmixedintheswitchoutputorpowerlinemaycausethePressureGaugetochangethepressureindication,malfunctionorbebroken.
TakesomemeasuretoputthePressureGaugeawayfrompowerlineoruseshieldedwire.ItiseffectivetogroundthePressureGauge.
2.Iftheswitchoutputhasaninductionloadsuchasrelayorsolenoidconnectedtoit,itshouldhaveasurgeabsorbercircuitputinit.Therelaycontactsshouldhave
acontactpreventivecircuitputinittopreventnoisesfrombeinggenerated.
¿Prohibitionofshort-circuitingtheswitchterminals
Donotshort-circuittheswitchoutputterminaltoanypowerterminal.Theinternalcircuitmaybebroken.
§
6. Howtouse
§
7. HandlingNote
§
8. TRANSPORTATION&STORAGE
§
9. WARRANTY