CORNET ED-25G User manual

CORNET Microsyslems ED-25G Electrosmog meter is a dual mode device for quick
measurement of both High frequency (RF) Electromagnetic wave field strength/power density level
and Low frequency (LF) Magnetic field (Gauss, Tesla) for living environments. lt is an excellent
device for individual or company with Electromagnetic wave safety concerns. lt has RF bandwidth
of 100MHz to 3GHz useful detection range, high sensitivity of 1 .Suw/m' to 0.58w/m2, and LF
bandwidth of 50Hz to 1SKHz, sensitivity of 0.1uT to 60uT(1 mG to 600mG). Also with very fast
response time.
Applications:
- High frequency RF Electromagnetic wave field strenglh and power density measurement
- Low frequency LF Magnetic field measurement (Gauss meter function)
- Mobile phone base station antenna radiation power density measurement
- Wireless communication applications (AM/FM, TDMA, cSM, DECT, CDMA)
- RF power measurement for transmitters
- Wireless LAN (Wi-Fi), Bluetooth, Ultra-wide-band detection, installation, optimization
- Spy camera, wireless bug finder
- Cellular/Cordless phone radiation safety level measurement
- AC power line, High voltage tower, power Transformer, motors and small appliance EMF
- Microwave oven leakage detection
- Personal living environment EMF safety evaluation
Usage guide:
Put the 9V battery in the ED25G, Handle the unit with right hand in vertical direction, and
push the power switch button to turn on the power, it will come up with RF meter mode.
The RF sensor is located in the left hand side of the ED25G; the LF sensor is located in the
right hand top ofthe ED25G, please do not cover the sensor area with hand or other objects.
There are two push button below the power switch: "Mode" button and "Hold" button.
"Mode" button is used to switch in between RF mode and LF mode. The "Hold" button is used
to freeze the data reading of the ED25G
ln RF mode, measured RF field strength/power density is shown on the digital LCD display
(with dBm and mw/m2). 8 LED lights. With Red, Yellow, and Green color on the right hand
side of LCD window is used for quick RF signal level indications.3 Red LEDs are used to
indicate the 3 safety range. The power level of each LED can be found in the table on the
ED25G front panel.
ln LF mode, measured LF magnetic field strength is shown on the digital LCD display (with uT
and mcauss). 8 color LED lights are used to show the relative strength of the magnelic fleld.
Histogram of previous 30 signal level readings are recorded and shown as moving graph
on the LCD display for both RF and LF mode.
"Hold" button can be used to halt the data measurement of the ED25G, a "HLD" mark
will be shown on the LCD screen to indicate the "Hold" condition. Push the "Hold" button
again ED25G will exit from the "Hold" condition.
ln LF mode, the Maximum value of the magnetic field measured since the last power-on
or mode change is shown on the LCD display, this can be used to record the maximum
magneticfield automatically inthe area.
(9) Most high frequency RF antenna such as Mobile phone base siation is vertical polarized
(in vertical direction), therefore while in RF mode, the ED25G is normally used in vertical
direction. ForLFmode,theLFsensorislocatedin theToprighthandoftheED25Gandthe
meter is normally used in Horizontal position in LF mode. Please rotate the meter to find the
maximum reading direction in either case. The maximum reading will also increase as you
approach the source. ED25G can be used to find the location of signal source.
(10) Most of moderh communication devices (Mobile phone, Wireless LAN, Wi-Fi, etc.,) use digital
communication technology with burst RF signals. When measuring this type of RF signals,
several LED lights will blinking at the same time. This is normal and it can be used as an
indication of burst type of RF signals. For continues waves (AM/FM) signals, the LED light will
stable. ED25G measures peak power density of signal with very fast response time. lt is more
accurate than the needle style of readout which only shown the average value of signal power
most of the time.
(1 1) While in LF mode magnetic field measurement, please hold the ED25G steady to get the
good stable reading, avoid fast moving of the ED25G to prevent the sudden change of the
reading caused by the Earth magnetic field.
(12) To turn-on/turn-off the LCD backlight, go into hold mode first, then toggle the mode switch to
turn-on/turn-off the LCD backlight. Remember to turn-off the the LCD backlight to reduce the
battery current consumption.
- the "SENSOR" mark on the bottom of the back enclosure is not used in ED25G
(1)
(2)
(3)
(4)
(5)
(b)
(7)
(8)
RF
tGn30t
aeta
LEO rignal
level lable
Field strength/power density readout:
ED25G use 8 high brightness LED to indicate the
indications. measured power density with 3 safety
@2007 CORNET Microsystems lnc., 1400 Coleman Ave #C28 Santa Clara
www.cornetmicLo-com
El*tEimog mctcr
fl.5+il
-3 tl.0 dEm
-l r-l ..1 I I
JLITTT
[]fi{,nEl
h'lf+H=45.3
LED
color Power
level
RF Power
density
LF
level Indication Aclion
RED3 -5 dBm
ltn 0.'18 w/m'? 30uT
lto Italy RF safety standard Caution!
RED2 -10 dBm 0.058 w/m'? 20uT Swiss RF satety standard
{0 04w/m2) Caution!
REDl '15 dBm 0.018 w/m'? 1 0uT Russian RF safety standard Caution!
YELLOW3 -20 dBm 5.8 milm2 5uT safe
YELLOW2 -25 dBm '1.8 mw/m? 2uI safe
YELLOWl -30 dBm 0.58 milm'z 0.5uT safe
GREEN3 -35 dBm 0.18 mvm'? 0.2uT WiFi Wireless LAN typically
in this ranoe safe
GREEN2 -40 dtsm
down 0.06 mw/m'? v,zu I
down Some signal source around safe
CA 95050 USA Tel: (408)9690205
ED25GV2.'1 12tO412008
* Electromagnetic wave field strength/power density reduces very fast with distance (distance
square), keep a good distance from the high frequency RF signal source can reduce the high
frequency radiation effect. Alumina foil or window sun reflector film (silver color) can be
used as a effective and low cost shielding material for most of RF radiations.
- ED25G is designed for quick living environment RF radiation evaluation and is for reference use
only. Official RF safety radiation measurement procedure is complicate and should be
handled by trained technical person with lab instruments. Safety range standard listed here is
for reference only. ED25G is not a medical instrument, Please do not use it in medical, legal or
other related applications.
- ED25G is designed for home and personal use, lt is not for commercial rental pu]Fose.
The European Community provided general guidelines in its Council Recommendation ofJuly 1999.1
ICNIRP published similarguidelines in April 1998.'Table I gives a sampling ofthe international and
nationalfield-strengthlimitvaluesforthegeneralpublicandcontinuousexposure (forReferenceonlyl)
9501\4hz 1850Mhz
lnternational Council Recommendation 1 999/51 g/EC 42Vlm
// 75\l//m2l 59 V/m
lQ 25Wm2\
lnternational ICNIRP Guidelines, April 1998 42Vlm
I I 7E\Al lm2\ 59 V/m
Iq ?5W/m?1
Aushia oNoRM S1120 49 V/m
/A 11M/im2\ 61 V/m
l'l0M/im2\
Belgium Belgisch Staatsblad F.2001 -1 365 21vlm
(1.18Wm'?) 30 V/m
t2 ?1Wim21
Germany 26. Deutsche Verordnung 42Vln
(4 75Wm'?) 59 Vim
/Q ?5\/V/m2\
Italy Decreto n.381, 1998
6 V/m
(0.1Wm'?)
20 V/m
0 Vim
(0.1Wm'?)
20 V/m
The Netherlands Health Council 51 V/m
/A Q?W/m2\ 83 V/m
l'18Wm2\
Switzerland Verordnung 1 999 4 V/m
I0 0dWm2\ 6 V/m
/0 1W/m2\
United States IEEE C95.1 49 V/m
(6.33Wm') 68 V/m
(2Wlnz\
China Draft: National Quality Technology
[,lnnifnrinn Rrrroerr 49 V/m
(6.33Wm'z) 61 V/m
/10W/m2\
Japan Radio-Radiation Protection Guidelines,
1 990 49 V/m
(6,33Wm'?) 61 V/m
(10Wm'z)
Specification:
Sensor type:
Frequency range:
RF Sensitivity:
RF Dynamic range:
RF Peak power measurement
LF Sensitivity:
Display type:
Unit of measurements:
LCD back light:
Display of data:
Safety standard indication:
Battery used:
Battery life:
Electric field sensor, Magnetic field sensor
RF:100MHz to 3GHz useful detection range
LF:50H2 to 1SKHz useful detection range
-55dBm to OdBm (25mvlm to 14.8v/m)
60 dB
: 1.5uw/m2 to 0.58w/m'
0.1uT to 60uT (1mG to 600mG)
digital LCD display
dBm, mw/m2, uT, mG
1 5 seconds auto-off
LCD 3 and 5 digit, 8 LED color segmeni,
Moving Histogram of 30 readings
3 safety range indication by 3 Red LED
9V alkaline battery, (not included)
>20 hours
(Gt
o
o
w
@2007 CORNET l\ilicrosystems lnc., 1400 Coleman Ave #C28 Santa clara
ww.cornelmbrolqm CA 95050 USA Tel: (408)9690205
ED25GV2j 1210412008
Electrosnrog Meter

Other CORNET Measuring Instrument manuals

CORNET ED-15C User manual

CORNET

CORNET ED-15C User manual

CORNET ED-88T User manual

CORNET

CORNET ED-88T User manual

CORNET ED-88TPlus User manual

CORNET

CORNET ED-88TPlus User manual

CORNET ED-85EXPlus User manual

CORNET

CORNET ED-85EXPlus User manual

CORNET ED-78S User manual

CORNET

CORNET ED-78S User manual

CORNET ED-88TPlus User manual

CORNET

CORNET ED-88TPlus User manual

CORNET iVDO Multi E+2 User manual

CORNET

CORNET iVDO Multi E+2 User manual

CORNET ED-88T User manual

CORNET

CORNET ED-88T User manual

CORNET ED-78SPlus User manual

CORNET

CORNET ED-78SPlus User manual

CORNET MD18 Owner's manual

CORNET

CORNET MD18 Owner's manual

CORNET ED-85EXPlus5 User manual

CORNET

CORNET ED-85EXPlus5 User manual

Popular Measuring Instrument manuals by other brands

Bender A-ISOMETER IRDH375 operating manual

Bender

Bender A-ISOMETER IRDH375 operating manual

dewhurst ULSP32HS Installation and setting instructions

dewhurst

dewhurst ULSP32HS Installation and setting instructions

Bender ISOMETER iso685W-D manual

Bender

Bender ISOMETER iso685W-D manual

Polar Electro 7700-0001 instruction manual

Polar Electro

Polar Electro 7700-0001 instruction manual

Würth 0964 764 009 Translation of the original operating instructions

Würth

Würth 0964 764 009 Translation of the original operating instructions

dymax ACCU-CAL 160 user guide

dymax

dymax ACCU-CAL 160 user guide

WIKA 2XX.34 Operating & installation instructions

WIKA

WIKA 2XX.34 Operating & installation instructions

GE EPM6000 instruction manual

GE

GE EPM6000 instruction manual

Hinds Instruments EXICOR 150AT user manual

Hinds Instruments

Hinds Instruments EXICOR 150AT user manual

SPI 18-010-9 Operation manual

SPI

SPI 18-010-9 Operation manual

KROHNE OPTISWIRL 4200 quick start

KROHNE

KROHNE OPTISWIRL 4200 quick start

SATO KEIRYOKI PC-53D instruction manual

SATO KEIRYOKI

SATO KEIRYOKI PC-53D instruction manual

Elesa DD51-E operating instructions

Elesa

Elesa DD51-E operating instructions

ICS Schneider Messtechnik GA35 operating instructions

ICS Schneider Messtechnik

ICS Schneider Messtechnik GA35 operating instructions

Dinel CLM-36 instruction manual

Dinel

Dinel CLM-36 instruction manual

Powerfix Profi Z22855 Operation and safety notes

Powerfix Profi

Powerfix Profi Z22855 Operation and safety notes

Sierra HART 600 Series instruction manual

Sierra

Sierra HART 600 Series instruction manual

Honeywell MagneW 3000 Plus Application guide

Honeywell

Honeywell MagneW 3000 Plus Application guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.