CWE SAR-1000 User manual

SAR-1000
Advanced Small Animal Ventilator
©CWE 2023
Read instructions carefully before operating this device.
This device is not to be used for Human Life Support applications.
To avoid possible electrical shock, do not operate this device if is wet or has had liquids spilled onto it.
Service or calibration procedures should only be performed by qualified personnel familiar with the
electrical hazards of line-powered devices.
Instruction Manual
www.cwe-inc.com

SAR-1000 Instruction Manual 2
© 2023 CWE, Inc.
STATEMENT OF WARRANTY
IF THIS INSTRUMENT FAILS WITHIN A PERIOD OF ONE YEAR FROM THE DATE
OF DELIVERY OR INSTALLATION, CWE, INC. WILL, AT ITS OPTION, REPAIR OR
REPLACE IT FREE OF CHARGE. THIS WARRANTY EXCLUDES DAMAGE
INCURRED THROUGH MISUSE OR ACCIDENT. CWE, INC. DOES NOT ASSUME
ANY LIABILITY FOR ANY CONSEQUENTIAL DAMAGES RESULTING FROM THE
USE OF THIS INSTRUMENT.
DEFECTIVE UNITS SHOULD BE RETURNED TO THE FACTORY ALONG WITH A
NOTE DESCRIBING THE NATURE OF THE FAULT. THIS WARRANTY IS
APPLICABLE TO THE ORIGINAL PURCHASER OF THE INSTRUMENT ONLY, AND
IS NOT TRANSFERABLE.
FACTORY SERVICE
Out of warranty or damaged instruments may be returned to the factory freight prepaid
for service at prevailing rates. Upon request, a written or verbal quotation for such
service will be issued after examination of the unit but prior to commencing repairs or
service. Address requests for service or technical information to:
CWE, Incorporated
Technical Support
Ardmore, PA 19003 U.S.A.
(610)642-7719
LIFE SUPPORT POLICY
Instruments manufactured by CWE, Incorporated are not authorized for use as critical
components in human life support devices or systems. "Life support devices or
systems", as used herein, are devices or systems whose failure to perform, whether
through misuse, failure, or proper operation, can reasonably be expected to result in
significant injury to the operator or subject persons.
This document was last updated May 5, 2023

SAR-1000 Instruction Manual 3
© 2023 CWE, Inc.
TABLE OF CONTENTS
1.0 INTRODUCTION........................................................................................................ 4
2.0 BASIC VENTILATOR CONNECTIONS..................................................................... 5
3.0 VENTILATOR OPERATION...................................................................................... 6
4.0 TUBING CONNECTIONS.......................................................................................... 9
5.0 REAR PANEL INPUTS AND OUTPUTS................................................................. 10
6.0 REMOTE CONTROL USING THE USB PORT ....................................................... 13
7.0 USING ANESTHESIA WITH THE SAR-1000.......................................................... 15
8.0 SPECIFICATIONS................................................................................................... 17
9.0 ORDERING INFORMATION.................................................................................... 17
Figure 1: Model SAR-1000 Front Panel ....................................................................... 4
Figure 2: SAR-1000 LCD Display Panel....................................................................... 6
Figure 3: Top trace: Airway pressure, bottom trace: Sync Out ................................ 7
Figure 4: Auto-Sigh setup screen................................................................................. 8
Figure 5: SAR-1000 Rear Panel.................................................................................... 9
Figure 6: Connecting an external valve set .............................................................. 12
Figure 7: SAR-1000 anesthesia connections............................................................. 16
APPENDIX 1: TIDAL VOLUME VS. BODY WEIGHT CHARTS.................................... 18
APPENDIX 2: TYPICAL VENTILATOR SETTINGS FOR RODENTS.......................... 19

SAR-1000 Instruction Manual 4
© 2023 CWE, Inc.
1.0 INTRODUCTION
The SAR-1000 Small Animal Ventilator is designed to provide safe and reliable
mechanical ventilation for mice, rats, and other small laboratory animals.
This ventilator operates on the flow-time principle, where a measured airflow is gated
into the animal for a set time, thus producing a known tidal volume. In addition, the SAR-
1000 can operate in pressure-cycled mode, where the airway pressure is automatically
limited to a user-set value. This approach to ventilator design permits great flexibility,
since airflow, timing, and pressures are all independently adjustable over a wide range.
Special sigh breaths can be manually initiated or programmed to occur automatically at
user determined intervals.
The SAR-1000 is completely self-contained, with a built-in air pump to provide airflow. All
settings are made from the front panel and are displayed on an LCD display along with
computed measurements.
Besides being a stand-alone ventilator, the SAR-1000 has a built-in USB interface for
remote control applications. In addition, there is a DATA PORT for external logic signals,
which maintains compatibility with the SAR-830 series ventilators.
The SAR-1000 is compatible with anesthetic gasses, and complete anesthesia setups
are available. To ventilate multiple, or larger, animals, it can be used with any of the
CTP-VA series external valve assemblies. These accessories are available for a range
of animals (see: Ordering Information).
The following sections of this Manual explain how the ventilator works, and how to set it
up to suit your particular experimental needs. We encourage you to read through this
Manual and welcome you to the next generation of CWE ventilators.
Figure 1: Model SAR-1000 Front Panel

SAR-1000 Instruction Manual 5
© 2023 CWE, Inc.
2.0 BASIC VENTILATOR CONNECTIONS
Setting up the SAR-1000 is straightforward. For basic operation, only a mains power
cord and a cannula tubing set need be connected.
Power connections
Plug an appropriate mains cord into the MAINS INLET receptacle on the right of the rear
panel. This instrument uses a universal voltage input, so no mains voltage switch is
required.
CAUTION! To maintain electrical safety, use only a
3-wire grounded power cord. Operating this device from an
un-grounded receptacle could result in a dangerous electrical
shock hazard.
Cannula Tubing Set connections
The cannula tubing set connects the front panel tubing ports to the animal’s
endotracheal tube. For rats and similar size animals, the INSP and EXP ports connect to
the distal “W” connector using 0.125” ID x 0.250” OD (3 x 6mm) Tygon or similar flexible
tubing. For mice, use 0.063” x 0.125” (1.5 x 3mm) tubing and appropriate barbed tube
fittings. The PRESSURE MONITOR port is a female Luer connection and connects to the
center tap of the distal “W” connector (rat), or to a miniature “Y” connector (mouse). Both
rat and mouse tubing sets are provided with the instrument.
To reduce compression effects of large tubing volumes, these tubes should be as short
as is convenient, but their length does not affect dead-space, since flow is unidirectional
from the CANNULA I →tracheal tube →CANNULA E port. The only dead-space is the
volume from the distal end of the “W” connector to the tracheal tube.
It is highly recommended that the new user become familiarized with how the ventilator
operates before connecting an experimental animal. A test balloon is provided for this
purpose.

SAR-1000 Instruction Manual 6
© 2023 CWE, Inc.
3.0 VENTILATOR OPERATION
Volume mode
Once the ventilator is set up and operational, a few important settings need to be
determined. The appropriate respiratory rate (RR) and tidal volume (VT) are mainly
dependent on body weight. Appendix 1 contains tidal volume and rate vs. body weight
charts. For example, if we have a 250g rat, the chart shows a suitable VTof about 2.0ml
at a RR of 55 breaths/min.
To set up the SAR-1000 for this example, first adjust the RESP RATE knob until 55 bpm is
displayed. Adjust the %INSP knob until 50% Insp is displayed. Now adjust the INSP FLOW
knob until VT2.0 ml is displayed.
Note that you are setting the tidal volume by adjusting the inspiratory airflow, rather than
just dialing in VTdirectly. This may at first seem unintuitive, but it follows from the basic
flow-time ventilator principle: VT= flow x time. The time (TI) was set as a fraction (%Insp)
of the total respiratory cycle time (1/RR). VTis then dependent on the flow rate setting.
All the calculations are performed by a microprocessor so you don’t have to!
Figure 2: SAR-1000 LCD Display Panel
Pressure mode
Pressure mode operation is similar to the above description, except that inspiration is
terminated when a certain airway pressure (PP) is reached, rather than when the
inspiration time (TI) has elapsed.
This pressure is set using the PRESSURE knob, and is displayed, for example, as
PP 10.0 cm. Now, inspiratory flow will cease on each breath when the set PPpressure is
reached.
Inspiratory hold
Note that the SAR-1000 has another important feature that applies to pressure mode: If
PPis reached before the set TIhas elapsed, the lung volume will be clamped for the
duration of TI. This inspiratory hold feature is known to improve the efficiency of
ventilation and gas exchange by maintaining the end-inspiratory pressure, instead of
simply releasing the lung volume immediately when PPis reached. You can observe this
via the INSP, EXP, and INSP HOLD LED’s, which indicate exactly what is happening during
the respiratory cycle. Figure 3 illustrates this by observing the pressure signal.

SAR-1000 Instruction Manual 7
© 2023 CWE, Inc.
NOTE: By default, the INSP HOLD function is enabled. You can disable it by pressing the
SIGH pushbutton while turning POWER ON. You will hear two beeps to indicate the
function is disabled. Repeat this to re-enable INSP HOLD and you will hear four beeps to
indicate that the function is active. This setting is stored and will apply from now on until
the user changes it.
Figure 3: Top trace: Airway pressure, bottom trace: Sync Out
Manual Sigh function
Sighs, or periodic delivery of larger tidal volumes, are often used to maintain good lung
ventilation and gas exchange. The SAR-1000 has a manual sigh function actuated by
pressing the SIGH pushbutton. When this pushbutton is pressed, the ventilator is armed
to deliver a sigh breath on the next inspiration. A short beep will indicate that a sigh
breath is about to be delivered.
During a sigh breath, the inspiration will continue until the sigh pressure (PS) is reached.
This function is available regardless of whether the ventilator is in VOLUME or PRESSURE
mode. On the sigh breath, the lung volume will be clamped once the PSis reached, and
held for the programmed sigh time (see Programming the Auto Sigh function).
To set the sigh pressure, press and hold the SET SIGH PRES pushbutton while adjusting
the PRESSURE knob. The LCD display indicates the PSsetting in cmH2O.
NOTE: The sigh pressure (PS) setting is normally several cmH2O
greater than the pressure limit (PP) setting. However, it is possible to
set the sigh pressure to a lower setting if desired. In this case, the
sigh breath will deliver a lower than normal end-inspiratory
pressure.
Programming the Auto Sigh function
As described above, a single sigh breath can be initiated simply by pressing the SIGH
pushbutton. You can also automate the delivery of periodic sigh breaths by using the
auto sigh setup routine. You enter this routine by simultaneously pressing the SIGH and
INSPIRATORY HOLD
INFLATION
DEFLATION
SYNC:
Airflow
going
To animal

SAR-1000 Instruction Manual 8
© 2023 CWE, Inc.
SET SIGH PRES pushbuttons and holding for about ½ second. The setup screen shown in
Figure 4 will then be displayed:
Figure 4: Auto-Sigh setup screen
The two settings you can change here are the sigh frequency and sigh ratio values. In
the example shown, sigh frequency is set to 1:050, or a sigh every 50 normal breaths.
The sigh ratio is set to 250%, so the inspiratory time (TI) of the sigh breath will be 2.5X
(or 250%) of the normal TI. The other sigh variable, sigh pressure (PS) is set from the
main display screen as described earlier.
To change the sigh frequency setting, use the RESP RATE knob. To change the sigh ratio,
use the %INSP knob. (These controls are shown in brackets on the right side of the
display as a mnemonic aid.)
Any changes to the auto sigh settings take effect immediately. To exit the setup screen,
press the SET SIGH PRES pushbutton. After 15 seconds of inactivity, the screen will
automatically timeout and return to the main operating screen.
When Auto Sigh is in effect, the symbol ASwill be displayed on the bottom line of the
main operating screen.
Minute Ventilation
Minute ventilation (MV) is the sum of the inspired volumes delivered to the animal per
minute. This measurement is a useful indicator of adequate ventilation and is normally
proportional to body weight. MV is computed automatically and reported on the bottom
line of the LCD display.

SAR-1000 Instruction Manual 9
© 2023 CWE, Inc.
4.0 TUBING CONNECTIONS
Figure 5: SAR-1000 Rear Panel
Pump In / Pump Out ports
The SAR-1000 is equipped with an internal air pump to provide compressed gas for use
as the inspiratory flow. The inlet and outlet connections of this pump are brought out to
the rear panel as the PUMP IN and PUMP OUT ports, respectively.
The air pump has a capacity of approximately 2.5 lpm. For most setups, a tubing jumper
should be placed between the PUMP OUT and the GAS INLET ports.
NOTE: Never apply a pressure to the PUMP IN port. A collapsible bag
with a gas mixture can, however, be connected to this port.
Gas Inlet port
This port receives the compressed gas used for generating the controlled inspiratory
flow. Although normally this will connect to the PUMP OUT port, in some circumstances it
is desirable to connect a gas source directly to the GAS INLET port, thus bypassing the
internal air pump.

SAR-1000 Instruction Manual 10
© 2023 CWE, Inc.
Expired Air port
The EXP AIR port is connected directly to the lungs during expiration. In most cases, this
port should be left unconnected. However, there are some circumstances where a
connection is made:
PEEP (Positive End-Expiratory Pressure) --- PEEP is a commonly used
technique to maintain a small pressure in the lungs during expiration. For
example, in open-chest preparations, PEEP is essential to keep the lungs from
collapsing. To apply a PEEP, connect a flexible tubing to the EXP AIR port, and
submerge it under a few cm of water (the depth under water = the PEEP
pressure). This is illustrated in Figure 7.
Anesthesia --- If inhalational anesthesia is used, it is necessary to collect and
evacuate the gas from the exp air port. A tube connected to this port can be
directed to an anesthetic absorber or exhaust system.
NOTE: Never connect a vacuum source directly to this port,
as it will cause serious injury to the animal!
In Vent port
During the inspiration phase of ventilation, the inspiratory flow is directed to the animal.
During expiration, this same flow is gated out the IN VENT port. Thus, this is clean
inspiratory gas. In anesthesia setups, this gas can be sent to an anesthetic evacuation
system (see Figure7).
5.0 REAR PANEL INPUTS AND OUTPUTS
PRESSURE OUT
This BNC jack provides a continuous analog voltage signal corresponding to
airway pressure. The scaling of this signal is 50mV/cmH2O.
SYNC OUT
This BNC jack provides a TTL-compatible signal corresponding to the phase of
respiration. During inspiration, the output is HIGH (+5V); during expiration the
output is logic LOW (0V).
GROUND
This binding post is connected to chassis ground, or earth. If required, this post
can be connected to a system-wide ground connection.

SAR-1000 Instruction Manual 11
© 2023 CWE, Inc.
DATA PORT
The DATA PORT connector serves as a unified connection for several input and
output functions. These are listed below, with a description of their functions. The
connector required is a DB9M (9-pin DSUB male plug):
PIN 1
Ground
PIN 2
HOLD input
PIN 3
MAN input
PIN 4
SYNC out
PIN 5
INSP out
PIN 6
EXP out
PIN 7
PRESSURE out
PIN 8
RATE out
PIN 9
+5V
HOLD input –When a logic LOW (0V) is applied to this pin, the ventilator
will stop cycling and enter a HOLD mode, with the valves set for
expiration.
MAN input –When a logic LOW (0V) is applied to this pin, the ventilator
valves will be set to inspiration, and remain so as long as the pin is held
LOW.
SYNC out –This is a TTL-compatible logic signal corresponding to the
phase of respiration. Logic HIGH = inspiration, and logic LOW =
expiration.
INSP out –This is a TTL-compatible logic signal that is logic HIGH during
inspiration.
EXP out –This is a TTL-compatible logic signal that is logic HIGH during
expiration.
PRESSURE out –This is an analog voltage corresponding to the current
airway pressure reading (same as the PRESSURE OUT BNC jack).
RATE out –This is an analog voltage corresponding to the current
respiratory rate (RR) setting. The scale factor is 10mV/bpm.
The data port connections HOLD and MAN can be used to provide a powerful
remote-control function. By connecting these two TTL inputs to a data acquisition
system with I/O capabilities, it is possible to create a ventilator system with many
possibilities, such as constant-CO2operation, neural-controlled ventilation,
ventilation synchronized to MRI or CT scanners, etc. If this is of interest, please
contact CWE, Inc. for details.

SAR-1000 Instruction Manual 12
© 2023 CWE, Inc.
EXTERNAL VALVES CONNECTION
This 4-pin jack is used to connect one or more External Valve Assemblies to the
SAR-1000. By connecting an external valve set (e.g., CTP-VA-1), you allow an
additional animal to be ventilated. The additional ventilator station will operate
synchronously with the primary ventilator, but the tidal volume (VT) can be
independently set using an external flowmeter (see: Figure 6).
Figure 6: Connecting an external valve set
It is also possible to ventilate up to four rodents simultaneously with just one SAR-1000
as the control unit. The valve assemblies will all operate synchronously, but individual
tidal volumes can be adjusted independently by varying the flow rate to each animal.
When adding more than one external valve assembly, an electrical cable adapter is
required (Part No. MVA-4). This part provides four expansion jacks for adding the valve
assemblies. With multiple animal setups, a flowmeter manifold (Part No. EFM-4-x) is
useful. With this manifold, one air source can be used for all the animals being
ventilated. The required accessories are listed below:
Valve Assembly
Flowmeter
Animal
CTP-VA-1
EFM-1-1 (rat), EFM-1-100 (mouse)
Mouse, Rat, Guinea pig
CTP-VA-3
EFM-1-2.5
Cat, Rabbit, Monkey
CAUT ION!
USE 3 -W IRE
GROUNDED POW ER
CORD ONL Y.
SYNC OUT
PRESSURE
IN VENT
PUM P IN
DATA PORT
USB
GAS INL ET
PUM P OUT
PUM P ON
PUM P OFF
EXP AIR
EXTERNAL
VALVES
(Usemultipleflowmeters
asmixingmanifold)
I E FLOW IN
CTP-VA-x ValveAssembly
Flowmeter
Valve assembly cable
Tracheal
Tube
Animal#2

SAR-1000 Instruction Manual 13
© 2023 CWE, Inc.
6.0 REMOTE CONTROL USING THE USB PORT
The SAR-1000 Ventilator has a powerful new capability: remote control via a
USB connection. This allows an external computer to remotely change the
ventilator settings, exactly as they are done using the front-panel controls. In
addition, a set of primitive, low-level commands allow direct control of ventilator
internal functions. These functions are described below.
Setup for USB connection
Before connecting the USB cable to the SAR-1000, you must install the software
drivers from the supplied CD-ROM. Locate and execute the following program:
CDM20600.EXE. This program will install Virtual Com Port (VCP) drivers,
creating a new serial port connection on your computer (for example, COM5).
Connect the USB cable and power on the SAR-1000. Locate the new COM port
by: Start / Control Panel / System / Hardware / Device
Manager / Ports. This port will then be used by the optional SAR-COMM
software or the supplied terminal (terminal.exe) program to communicate with
the ventilator.
The serial port settings used are 9600,N,8,1. All commands should be sent as
simple ASCII text character strings. Besides the optional SAR-COMM program,
any control program (such as Labview) that can generate these commands can
be used. Note that the syntax must be exactly as shown below, and all
characters should be sent as a string, including leading zeros as necessary.
Alphabetic characters may be upper or lower case.
Ventilator setup commands
RRnnn –Set Respiratory Rate, nnn = 010 –200 (bpm)
FLnnnn –Set Insp. Flow rate, nnnn = 0010 –1100 (ml/min)
IEnn -- %Inspiration, 10 –90 (percent)
PPnnn –Pressure limit, 10 –500 (mmH2O)
PSnnn –Sigh pressure, 10 –500 (mmH2O)
SFnnn –Sigh frequency, 0 –255 (1 sigh per nnn breaths)
SRnnn –Sigh ratio, 100 –999 (percent of normal TI)

SAR-1000 Instruction Manual 14
© 2023 CWE, Inc.
Multiple settings commands
GS –Sends a full set of machine settings back to host computer in the format:
aaa,bbbb,ccc,ddd,ee,fff,ggg where:
aaa = Respiratory Rate
bbbb = Inspiratory Flow
ccc = Pressure Limit
ddd = Sigh Pressure
ee = %Inspiration
fff = Sigh Frequency
ggg = Sigh Ratio
ISaaa,bbbb,ccc,ddd,ee,fff,ggg –Downloads a full instruction set of machine
settings into the ventilator. The description and formatting of the individual
elements of this command string is the same as for the GS command described
above.
Note: This above command strings are sent (or received) as simple
ASCII characters with comma delimiting between values. Numbers are all
integers, and leading zeros are required.
Primitive commands
RC –Put ventilator in remote-control mode (valves set to expiration)
LC –Return to local control
I1 –Inspiration valve ON
I0 –Inspiration valve OFF
E1 –Expiration valve ON
E0 –Expiration valve OFF
A1 –Auxiliary valve ON
A0 –Auxiliary valve OFF
IN –Set valves for inspiration
EX –Set valves for expiration
BH –Set valves for breath-hold (inspiratory hold)
Note: The latency time after a command is sent before another one can
be received is approximately 5mS. If commands are received too quickly,
they will be ignored or possibly corrupted.

SAR-1000 Instruction Manual 15
© 2023 CWE, Inc.
7.0 USING ANESTHESIA WITH THE SAR-1000
The SAR-1000 Ventilator can be used with Isoflurane or other inhalational
anesthetic systems. Figure 7 shows how to do this.
Principle of operation
The preferred method of interfacing an anesthesia machine (gas source +
vaporizer) to the ventilator is by using a blow-by system, as shown. If you
examine the flow path from the vaporizer, you will see that the anesthetic flow
passes across the PUMP IN port of the ventilator, and then passes to the absorber
canister or evacuation system. The ventilator draws in the gas it needs from this
stream. This setup prevents harmful back-pressure on the vaporizer, and also
avoids applying pressure to the internal air pump on the ventilator. In operation,
the gas flow from the vaporizer is adjusted to be a little greater than the
INSPIRATORY FLOW shown on the ventilator LCD display.
Mandatory anesthetic absorber or evacuation connections
As shown on the diagram, the following port connections must be sent to an
absorber canister or evacuation system when using anesthesia:
IN VENT --- The gas from this port is clean inspiratory gas, which comes out of this
port when it is not being directed to the animal.
EXPIRED AIR --- This port is in direct communication with the animal’s airways
during the expiration phase of ventilation. Therefore, it is important not to apply
suction or any restrictions on this connection.

SAR-1000 Instruction Manual 16
© 2023 CWE, Inc.
Figure 7: SAR-1000 anesthesia connections with optional PEEP.

SAR-1000 Instruction Manual 17
© 2023 CWE, Inc.
8.0 SPECIFICATIONS
Respiratory rate range............................................................................. 5 - 200 breaths/min
Inspiratory flow range .................................................................................. 10 - 1100ml/min
Inspiration/Expiration (I/E ratio) range...................................................................... 10 - 90%
Tidal volume range (see Note 1) ............................................................................ 0.1 - 30ml
LCD panel.......................................................................................4 lines x 20 chars, backlit
Indicator lights................................................................................. INSP, EXP, INSP HOLD
Internal airpump capacity ............................................................................................. 2 lpm
Port connections .................................................................... 0.125" (3mm) ID Tygon tubing
PRESSURE MONITOR port connection............................................................. Luer female
SYNC OUT connector.....................................................................................................BNC
PRESSURE OUT connector ...........................................................................................BNC
PRESSURE OUT scaling................................................................................ 50mV/cmH2O
RATE OUT scaling (via DATA PORT jack)........................................................... 10mV/bpm
DATA PORT connector................................................................................................ DB-9S
Dimensions .................................................................... 9W x 4H x 9D in. (23 x 10 x 23cm)
Power requirements............................................. universal input,120/240V 50/60Hz, 100VA
Note 1: Volume is dependent on respiratory rate and airflow rate settings. Large volumes can be delivered at
slow respiratory rates only.
9.0 ORDERING INFORMATION
PART No
MODEL
DESCRIPTION
12-07000
SAR-1000
Ventilator with mouse & rat tubing sets, SAR-COMM software, USB cable, 115/230V
12-04000
CTP-VA-1
Solenoid Valve Assembly, mouse to Guinea pig size
12-05000
CTP-VA-3
Solenoid Valve Assembly, cat to small monkey size
12-10010
MVA-4
Electrical adapter for adding up to 4 external valve assemblies
12-10020
EFM-1
Flowmeter/regulator, with stand, specify range: 0.1, 1, 2.5, 5 l/min
12-10021
EFM-4
Flowmeter/regulator manifold w/ 4 flowmeters, with stand, specify range
12-10032
TRACH-1
Endotracheal tube for mice, flexible Teflon with Luer connector, 20 gauge, pack of 5
12-21032
TRACH-2
Endotracheal tube for rats, flexible Teflon with Luer connector, 14 gauge, pack of 5
11-10000
CAPSTAR-100
End-tidal CO2Analyzer, complete with accessory pack, 115/230V (rat and larger)
15-10000
MicroCapStar
End-tidal CO2Analyzer, complete with accessory pack, 115/230V (mouse and larger)

SAR-1000 Instruction Manual 18
© 2023 CWE, Inc.
APPENDIX 1: TIDAL VOLUME VS. BODY WEIGHT CHARTS
0.1
1
10
Tidal Volume (ml)
10 100 1000
Body Weight (g)
Tidal Volume Chart
100
90
70
60
50
40
Breaths per minute
100
90
70
60
50
40
20 30 40 50 70 200 300 400 500 700
0.2
0.3
0.4
0.5
0.7
2
3
4
5
7
1
10
100
1000
Tidal Volume (ml)
1 10 100
Body Weight (kg)
Tidal Volume Chart
10
15
20
25
30
40
Breaths per minute
2
3
5
7
20
30
50
70
200
300
500
700
2 3 45678920 30 40 50

SAR-1000 Instruction Manual 19
© 2023 CWE, Inc.
APPENDIX 2: TYPICAL VENTILATOR SETTINGS FOR RODENTS
Body
Tidal
Resp
Insp
I/E
Weight
Volume
Rate
Flow
Ratio
(g)
(ml)
(br/min)
(ml/min)
(%)
25
0.25
90
56
40
100
0.90
60
135
40
200
1.80
55
250
40
300
2.40
50
300
40
500
4.00
40
400
40
Note: Settings above are suggested guidelines. You will probably need to adjust the settings for
your particular animal, depending on its condition, anesthesia, or other factors.
Table of contents
Popular Fan manuals by other brands

Progress Lighting
Progress Lighting AirPro Willacy P250063 installation manual

Elta
Elta Skyflow SSMF Installation and maintenance instructions

Elington
Elington ANT54AV5WCR installation guide

CARNES
CARNES VFBB Installation, operating and maintenance instructions

AEG
AEG VL 5617 M instruction manual

Omega Altise
Omega Altise AP40WA instruction manual