manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Danfoss
  6. •
  7. Measuring Instrument
  8. •
  9. Danfoss SonoMeter 30 User manual

Danfoss SonoMeter 30 User manual

© Danfoss | 2019.10 VU.SH.I1.02 | 1
Ultrasonic heat meter SonoMeter 30
Test and calibration instruction
User guide
Table of Contents 1. General information ............................................................................................................... 2
2. Test instruction ....................................................................................................................... 2
2.1 Calculator energy measurement error (δE) test .................................................................................... 2
2.2 Volume measurement error (δV) test........................................................................................................ 2
3. Calibration instruction ........................................................................................................... 3
3.1 Calibration mode ............................................................................................................................................. 3
3.2 Program installation ....................................................................................................................................... 3
3.3 Program interface............................................................................................................................................ 3
3.4 „Manufacturer“ window................................................................................................................................ 3
3.5 Volume calibration .......................................................................................................................................... 4
User guide Ultrasonic heat meter SonoMeter 30
2 | VU.SH.I1.02 © Danfoss | 2019.10
1. General information This instruction designed for ultrasonic heat
meter SonoMeter 30 flow and energy verification
and calibration.
Heat meter energy measurement error (δEs)
assess the application of the partial verification
method: assesses the calculator thermal energy
measurement error (δE), the flow volume
measurement error (δV) and temperature sensor
temperature difference measurement error (δT).
Heat meter energy measurement error (δEs) are
all components of the arithmetic sum of the
errors:
Es = ׀
E׀
+ ׀
T׀
1) Energy pulse out terminals (Fig. 1) connected
to the pulse counter (not required when
reading the measurement results directly
from the display or via a digital interface)
Pin 53 - Energy Pulse output (Pulse)
Pin 52 - Energy Pulse output (GND)
2. Test instruction
2.1 Calculator energy
measurement error (δE) test
2) Temperature sensors placed in the
temperature testing bench (for temperature
difference simulation).
3) Verification mode is activated by connecting
TEST jumper contacts 1 (FIg. 1).
4) Determination errors:
- before each measurement run automatic
verification mode with long button clicking
on the meter (simulated flow pulses),
- When the LCD display stops blinking “TEST”,
energy and volume values can be read
directly from the LCD screen or through the
optical / digital interface. The pulses shall
begin calculated with the corresponding
nominal flow pulse values as follows:
Permanent flow qp
(m3/h)
Energy pulse value,
(Wh/imp)
0.6 0.1
1.0 0.2
1.5 0.2
2.5 0.5
3.5 0.5
6.0 1
10.0 2
15.0 5
- energy E (kWh) measurement error is
calculated by formula:
E = (E – E
0
) / E
0
• 100 %
there:
E0– the conventional true energy E value
(kWh), calculated in consideration of the
fluid temperature values simulated Θ1, Θ2
and simulated change in volume values:
Flow sensor mounting
place E0calculating formula
Supply pipe E0=V1 • ρ1 • (hT1-hT2)
Return pipe E0=V1 • ρ2 • (hT1-hT2)
There:
E0– the estimated true value of the energy component
(kWh);
V1 – simulated water volume value (m3);
ρ1…ρ2 – density of water, corresponding to the sumulated
temperature value Θ1,Θ2, (kg/m3);
hT1… hT2 – relative enthalphy of water, corresponding to the
simulated temperature value Θ1,Θ2
1) Volume pulse output terminals (Fig. 1)
connected to the pulse counter (not required
when the measurement results are directly
reading from the display or via a digital
interface)
Pin 51 - Volume Pulse output (Pulse)
Pin 50 - Volume Pulse output (GND)
2.2 Volume measurement
errors (δV) test
2) Switched on verification mode by connecting
TEST jumper contacts 1 (1 pav),
3) The meter mounted on the flow test bench,
flow are simulating.
Recommended minimum pulse content and
measurement time:
Flow Pulse content Measurement time
0.1 qp <q ≤ qsN≥1000 T> 1 (min)
qi ≤ q ≤ 0.1 qpN>500 T > 1 + 8 qi/q, (min)
Fig. 1
User guide Ultrasonic heat meter SonoMeter 30
VU.SH.I1.02 | 3© Danfoss | 2019.10
It must be maintained both minimum time and
minimum pulse quantities.
4) After the simulation of flow measured volume
V (m3) value is read directly from the display
(or through a digital interface) or calculated
pulse counter calculates the volume pulse
multiplied by the pulse value:
Permanent flow qp, (m3/h) Volume pulse value, (l/imp)
0.6 0.001
1.0 0.001
1.5 0.001
2.5 0.002
3.5 0.01
6.0 0.01
10.0 0.02
15.0 0.02
- volume V (m3) measurement error calculated
by formula:
V = (V – V0) / V0 • 100 %
there:
V0– device true value of simulated volume (m3).
3. Calibration instruction
3.1.1 Calibration mode
To enter calibration mode it is necessary to
connect the verification jumper 1, remove
metrological seal and short circuit calibration
contacts 2. (Fig.2).
3.2 Software installation The software “SonoMeter30Config.exe” is used
to read and configure SonoMeter 30 device via
OG-1-USB optical reader.
Meter configuration / calibration cab be made
also via M-Bus interface module then converter
COM/M-Bus (MB-1) can be used.
Open “SonoMeter30Config.exe”. Press “Next”.
Then, you can choose installation directory.
“SonoMeter30UserConfig” has five user interface
windows for different purposes:
• “Actual” – actual data window;
• “LCD Menu” – LCD configuration window.
• “Configuration” – configuration window for
pulse channels, archive and tariffs;
3.3 Program interface • “Archive” – window for information from
device archive;
• “Manufacturer” – device calibration, reset and
configuration window.
2.2 Volume measurement
errors (δV) test (continuous)
“SonoMeter30UserConfig” “Manufacturer”
window is shown in Fig. 3, information about
window parameters is in Table 1.
3.4 “Manufacturer” window
Fig. 2
1
2
3
4
5
6
7
89
10
Fig. 3
“Manufacturer” window
© Danfoss | DHS-SRMT/SI | 2019.104 | VU.SH.I1.02
User guide Ultrasonic heat meter SonoMeter 30
Measurement units, flow pipe tagging,
calibration can be configured in “Manufacturer”
window.
Select energy units option in “energy unit” field
3.5 Volume calibration and press “set” in order to configure what energy
units are used for measurements. If device is
successfully configured message “Operation
done” will pop up.
If device is going to be placed on flow pipe, tag
“Flow pipe”. If device is going to be placed on
return pipe, tag “Return pipe”. Press “set” to save
configuration into device. If device is successfully
3.5.1 Changing pipe type configured message “Operation done” will pop
up. If is it needed that device measure only
heating energy, “heating only” box should be
tagged.
Press “Read” and then press “Manufacturer Reset”.
Disconnect the battery from the device. Use a
jumper to short the contacts on the device for
3.5.2 Resetting device 3 seconds in place where need to be connect
the battery. Connect a battery to the device and
reconfigure the date and time parameters.
Place jumper under removed device cover
(see Fig. 4). Press “Read” and then write error
“Err %” and flow “Q” data. Press “set” to save
3.5.3 Calibration configuration into device. If device is successfully
configured message “Operation done” will pop
up.
Table 1 “Manufacturer” window parameters
No. Name Note
1Calculator
-2Flow pipe (T1)
3Return pipe (T2)
4Heating only Check this when device is only for heating energy measurements
5Calibration -
6set Press button to save configuration to device device
7Calibration parameters Write calibration parameters here
8Energy unit -
9set Press set to save energy unit parameter configuration
10 Manufacturer Reset Press to reset the device
Fig. 4
Jumper Position “TEST”

Other manuals for SonoMeter 30

3

Other Danfoss Measuring Instrument manuals

Danfoss Micromon User manual

Danfoss

Danfoss Micromon User manual

Danfoss MAGFLO MAG 1100 User manual

Danfoss

Danfoss MAGFLO MAG 1100 User manual

Danfoss PFM 100 User manual

Danfoss

Danfoss PFM 100 User manual

Danfoss MCH19AM1649 Manual

Danfoss

Danfoss MCH19AM1649 Manual

Danfoss AK-LM 350 Guide

Danfoss

Danfoss AK-LM 350 Guide

Danfoss 40 User manual

Danfoss

Danfoss 40 User manual

Danfoss AK-EM 100 User manual

Danfoss

Danfoss AK-EM 100 User manual

Danfoss Vickers Target-Pro 2 User manual

Danfoss

Danfoss Vickers Target-Pro 2 User manual

Danfoss PFM 1000 User manual

Danfoss

Danfoss PFM 1000 User manual

Danfoss AKS 4100 User manual

Danfoss

Danfoss AKS 4100 User manual

Danfoss SonoMeter 30 Operation manual

Danfoss

Danfoss SonoMeter 30 Operation manual

Danfoss SonoSelect User manual

Danfoss

Danfoss SonoSelect User manual

Danfoss AK-SM 350 User manual

Danfoss

Danfoss AK-SM 350 User manual

Danfoss SonoMeter 30 User manual

Danfoss

Danfoss SonoMeter 30 User manual

Danfoss AKS 4100 User manual

Danfoss

Danfoss AKS 4100 User manual

Danfoss AK-LM 330 Guide

Danfoss

Danfoss AK-LM 330 Guide

Danfoss SonoMeter 30 User manual

Danfoss

Danfoss SonoMeter 30 User manual

Danfoss PR-SC4K Modbus User manual

Danfoss

Danfoss PR-SC4K Modbus User manual

Danfoss EKC 201 User manual

Danfoss

Danfoss EKC 201 User manual

Danfoss PR-SC4K Modbus User manual

Danfoss

Danfoss PR-SC4K Modbus User manual

Danfoss PFM 5001 User manual

Danfoss

Danfoss PFM 5001 User manual

Danfoss SonoSensor 30 Operation manual

Danfoss

Danfoss SonoSensor 30 Operation manual

Danfoss SONO 3500 CT User manual

Danfoss

Danfoss SONO 3500 CT User manual

Danfoss Supercal 5 User manual

Danfoss

Danfoss Supercal 5 User manual

Popular Measuring Instrument manuals by other brands

Greystone Energy Systems CR3 Series installation instructions

Greystone Energy Systems

Greystone Energy Systems CR3 Series installation instructions

Mollet LF20 Series Operating	 instruction

Mollet

Mollet LF20 Series Operating instruction

Keysight PNA-X Series Configuration guide

Keysight

Keysight PNA-X Series Configuration guide

SIRETTA SNYPER-LTEM quick start guide

SIRETTA

SIRETTA SNYPER-LTEM quick start guide

Carlisle Ransburg 310-9000 Service manual

Carlisle

Carlisle Ransburg 310-9000 Service manual

feedback EW604 manual

feedback

feedback EW604 manual

Force-A DUALEX SCIENTIFIC+ user manual

Force-A

Force-A DUALEX SCIENTIFIC+ user manual

BTI Profiline BTI-KLL 20 operating instructions

BTI

BTI Profiline BTI-KLL 20 operating instructions

Magnescale DF800S Series instruction manual

Magnescale

Magnescale DF800S Series instruction manual

Siemens SITRANS FC330 operating instructions

Siemens

Siemens SITRANS FC330 operating instructions

Badger Meter Blancett QuickSert B131C-038 user manual

Badger Meter

Badger Meter Blancett QuickSert B131C-038 user manual

ECOTEST MKS-05 TERRA operating manual

ECOTEST

ECOTEST MKS-05 TERRA operating manual

PUIG 5401N quick start guide

PUIG

PUIG 5401N quick start guide

Tricor CLASSIC Series Service manual

Tricor

Tricor CLASSIC Series Service manual

Bühler technologies BCI 24-D Series Brief instructions

Bühler technologies

Bühler technologies BCI 24-D Series Brief instructions

PCB Piezotronics EX357E93/020NJ Service manual

PCB Piezotronics

PCB Piezotronics EX357E93/020NJ Service manual

LMI RFP Series instruction manual

LMI

LMI RFP Series instruction manual

Shodex RSpak DC-613 Operation manual

Shodex

Shodex RSpak DC-613 Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.