Daruifuno MCC200 User manual

Digital Multi-parameter controller
User Manual
Daruifuno
Model: MCC200
Version 1.0


The caution sign is shown above, it reminds the user that the
instrument may malfunction or be damaged
Caution:
Warning:
The warning sign is shown above, which warns the user might get hurt
☞
The note sign is shown on the left, warns the user of important
information about operation.
Useful signs
In addition to installation and operation information, the manual also includes warning
signs related to user safety, caution signs related to possible instrument failure, and
note signs related to important and useful operating instructions.
Power Wiring Note
1. Please install voltage stabilization and surge protection equipment that meets the
instrument parameters at the power supply terminal to ensure that the power supply
provided by the instrument is stable and reliable and meets the standards, otherwise
the instrument will be irreversibly damaged due to poor power supply.
2. Damage caused by incorrect power supply is not covered by our quality guarantee.
3. Detailed parameters see the technical parameters.
Important Safety Information
Please read and observe the following:
Please read this entire manual before unpacking, installation and operation, with particular
attention to all danger warnings and precautions. Failure to do so may result in damage to
the instrument or personal injury to the operator due to misuse. Do not install or use the
instrument by any means other than those specified in this manual.
•After opening the analyzer case, the user may touch the instrument supply voltage.
Please make sure to disconnect the power, before you open the analyzer cases. Analyzer
housing assembly only low voltage, the operation is safe.
•The analyzer must be installed in accordance with relevant local codes and the
professional and technical personnel, to comply with the analyzer's technical
specifications and input ratings. If you are not sure which of the main power lines is the
zero line, use a double-blade switch to disconnect the analyzer.
•As soon as a problem occurs with the analyzer, disconnect the analyzer from power to
prevent any unintentional operation that could result in damage to the instrument. For
example, it may be in an unsafe state when the following situations occur:
(1) The analyzer appears visibly damaged;
(2) Analyzer fails to operate properly or provide the intended measurements;
(3) Analyzer has been stored for a long period of time at temperatures over 70°C
environment.
•Wiring or repair should be done by professionals, and only the power-off analyzer
should be operated.

Guarantee
Our company warrants the instrument for one year (12 months) from the date of
delivery. Consumables and consumable parts in the equipment are not covered by
the warranty. The terms of this warranty shall not apply if damage to the instrument
occurs beyond the warranty period, or in the opinion of the company, the breakage or
destruction of the instrument is due to improper use, lack of maintenance, improper
installation, improper modification, abnormal environmental conditions, etc. The
Company's obligation under this warranty is limited to providing replacement or
repair of this product, as the case may be. The product must be thoroughly cleaned
to remove any contaminated chemicals before it is accepted for replacement or
repair. Our obligations shall not exceed the price of the product itself. In no event
shall the company be liable for damage caused by incidental or consequential
damages, whether to persons or objects. The Company shall not be liable for any
other loss, damage or expense of any kind, including economic loss resulting from
the installation, use or improper use of the product.
For details, please refer to the product's quality promise with the product, and
keep this manual and the quality promise properly.

7.1ProbeSet......................................................11
7.1.1 Scan probe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
7.1.2Probeoptions..............................................11
7.1.3 Remove probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
7.2Alarmset..................................................11
7.2.1 Wash relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
7.2.2 Alarm relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
7.3 Current set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
7.4Communicationset..........................................12
7 . 4 . 1 N e t
address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
7.4.2 Baud rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
7.4.3Verifybit..................................................12
7.4.4 Stop bit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
7.4.5 Format . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
7.4.6 Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Part 2 Installation
5 . 1 P a n e l
keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . .8
5.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . 9
Chapter 7 Menu Description
4.1 Power Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4.2 Wiring Terminals Definition. . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
C h a p t e r 4 E l e c t r i c a l
I n s t a l l a t i o n
1.1 Technical Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
3.1 Installation Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
3.2 Dimension . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3.3 Panel Mounting. . . . . . . . . . . . . . . . . . . .. . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . 3
3 . 4 W a l l
Mounting...........................................................4
3 . 5 P i p e
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
C h a p t e r 6 S y s t e m D e f a u l t
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . .. . . . . . . . . . .. . . . .10 2
Chapter 1 Overview
Part 1 Introduction
Chapter 5 User Interface
Part 3 Operation
Contents
Chapter 3 Mechanical Installation
Chapter 2 Unpacking........................................................2

Appendix
7.5SystemSet.....................................................12
7.5.1 Data/Time set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
7.5.2Backlight.................................................12
7.5.3 Contrast set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
7.5.4 Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
7.5.5 Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
7.6Historylogs....................................................13
7.6.1Memoryinform.............................................13
7.6.2Setinterval................................................13
7 . 6 . 3 P o i n t
Query . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
7 . 6 . 4 W a r n
Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
7.7 Servings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
7.7.1 Measure mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
7 . 7 . 2 O u t p u t
module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
7 . 7 . 3 R e s t o r e
defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . .13
A M o d b u s r e g i s t e r
information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
M e n u s t r u c t u r e
diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16

1
Relay Control
Channels Quantity
1
Control Type
Wash relay
Contact Type
SPST
Contact Capacity
3A 250VAC
Communication Interface
RS485 MODBUS-RTU
Memory Backup
All user settings are retained in EEPROM indefinitely
Historical Data
7800 storage points, the recording interval can be set
Operating Environment
0~+60℃, 0~95% relative humidity, no condensation
Storage Environment
-20~+70℃, 0~95% relative humidity, no condensation
Power Requirements
100~240VAC or 18~36VDC, 24W MAX
Installation Method
Panel / Wall / Pipe
Instrument Size
144*144*120 (unit mm)
Protection Class
IP66
Controller Weight
About 1600g
——— Chapter 1 ———
Overview
This product is a 4-channel digital universal controller, which realizes measurement by
connecting various types of digital sensors for detecting physical, chemical and biological
properties.
1.1 Technical Parameters
Part 1 Introduction

2
Note: Please read this operation manual thoroughly before installation to
avoid damage to the instrument due to wrong wiring.
——— Chapter 2 ———
Unpacking
After unpacking, it is recommended that the shipped cardboard boxes and packing
materials be saved for instrument storage or reshipment. Inspect equipment and packing
materials for signs of damage during shipment. If there are signs of damage, immediately
notify the person delivering the shipment.
Part 2 Installation
• Please choose a well-ventilated location to install the instrument to avoid direct
sunlight.
• The electrode signal transmission must use specific cables or cables provided by our
company, and cannot be replaced by random wires.
• When using AC 220V power supply, avoid using three-phase power supply to avoid
power spike interference (if power spike interference occurs, the power supply of the
controller can be separated from the power supply of the control device, so that the
controller uses a separate power supply, or the coil ends of all electromagnetic
switches and power control devices are connected to spike absorbers to eliminate
spikes, such as dosing machines, mixers, etc.).
• It is generally recommended that the controller be accessible to the electrode during
installation.
• Avoid corrosive liquids, or need to protect and isolate the environment.
3.1 Installation Location
Caution:
Analyzer directly exposed to the sun, its operating temperature may exceed its
specified limit temperature, and will reduce the visibility of the display.
☞
The installation location of the meter is:
Unaffected by weather (rain, snow, ice, dust, etc.)
• Clean, dry and with little or no vibration
• Keep away from corrosive liquids
• Within ambient temperature limits(32~140℉or 0~60℃)
——— Chapter 3 ———
Mechanical Installation

3
Figure 3-1 Schematic diagram of the appearance and dimensions
When the instrument is installed in a panel cabinet, use two fixing brackets to fix the
instrument. The installation diagram is as follows:
3.3 Panel mounting
Figure 3-2 Panel mounting
3.2 Dimension

4
3.4 Wall Mounting
Connect the gray back panel to the instrument, and then install it on the box or the
side wall of the wall. The installation dimensions and schematic diagram are as
follows:
Figure 3-3 Schematic diagram of wall-mounted installation
Figure 3-4 Wall-mounted installation hole size diagram

5
3.5 Pipe mounting
Connect the gray backplane to the instrument, and then fix the backplane to the
horizontal pipe or vertical pipe with a clamp. The installation diagram is as follows:
Figure 3-5 Schematic diagram of pipeline installation

6
———Chapter 4 ———
Electrical Installation
☞
Note: Before connecting AC power, be sure to cut off the power supply
4.1 Power Connection
Two types of power supply: AC power supply 100~240VAC and DC power supply 18~
36VDC.
After unscrewing the screws at the back of the meter and taking off the cover, you can see
two rows of terminals.
Figure 4-1 indicates the location and name of the access terminals for both power
supplies.
Figure 4-1 Two types of power terminals

7
POWER
AC IN
100-240V
L
AC power fire wire input
POWER
DC IN
18-36V
+
DC power positive
AC power ground wire
NC
Floating terminal
N
AC power zero line input
-
DC power negative
RS485
A
RS485 signal D+(A) end
V+
Electrode power + terminal
B
RS485 signal D-(B) end
V-
Electrode power-terminal
R1
Relay one contact
AS
Electrode communication terminal A
Relay one contact
BS
Electrode communication terminal B
4.2 Wiring Terminals Definition
The following table is the terminal definition:
Two kinds of power terminals are defined in Table 4-1:
The terminal locations and names are shown in Figure 4-2:
Figure 4-2 terminal blocks

8
Part 3 Operation
5.1 Panel keys
——— Chapter 5 ———
User Interface
The instrument panel consists of a display screen and eight buttons, namely IOT (Internet
of Things) button, menu (menu) button, back (return) button, enter (confirm) button, four
up, down, left and right Arrow keys.
Up key: In the menu option interface, move the cursor up;
In the parameter set interface, increase the data by 1 (change in the order of 0, 1, 2,... 9, 0).
Down key: In the menu option interface, move the cursor down;
In the parameter set interface, reduce the data by 1 (changes in the order of 9, 8, 7... 0, 9);
In the measurement interface, switch to display the relay status.
Left key: In the parameter set interface, move the cursor to the left.
In the menu interface, return to the previous menu.
Right key: In the parameter set interface, move the cursor to the right.
In the measurement interface, switch and display different measurement objects, and switch
between single-parameter and multi-parameter interfaces.
LOT key:In the measurement interface, short press this key to turn on or off the LOT
function, and long press to enter the LOT setting menu.
Note: The LOT key can only be used in instruments with hardware version C0 or above.
☞
Return key
Enter
LOT key
Up key
Left key
Menu
Right key
Down key

9
Date and time: Display the current date and time information.
Electrode type: Display the currently connected electrode measurement object
type.
Measured value: Display the current measurement value, and the unit changes
according to the measurement object.
Relay Status: Display the current switching status of the relay.
Menu button: In the measurement interface, enter the menu interface. In the menu
option interface, directly return to the measurement state. In the parameter set interface,
return to the previous menu.
Enter button: In the menu option interface, select the selected item. In the parameter set
interface, confirm and return to the previous menu.
Return button: Press this key to return to the previous menu. For example, in the
parameter setting interface, press this key to abandon the modification and return to the
previous menu.
After the instrument completes the scanning probe step, the instrument is powered
on again, and the corresponding scanning probes measurement interface will be
displayed, such as the following:
5.2 Display

10
Setting Type
Setting Options
Initial Value
Wash Relay
Clean interval
5min
Duration of clean
60sec
Wash state
Hold
Comm. Set
Net address
001
Baud Rate
9600bps
Verify Bit
None
Stop Bit
1 bit
Format
CDAB
Protocol
ModBus
System Set
Backlight
Scanty Power 01min
Contrast Set
25
Password
000000
History Logs
Set Interval
5min
Servings
Measure mode
Probe measure
——— Chapter 6 ———
System Default Settings

11
7.1.2 Probe options
The initial state of the meter is no measurement value display. After connecting the
probes, you need to scan the probes first, and then the corresponding measurement
value can be displayed.
7.1.1 Scan probe
The meter is connected to different probes, and the options are set according to the
probe type.
For details, please refer to the "Digital Universal Controller Electrode Setting
Instructions".
——— Chapter 7 ———
Menu Description
In the measurement interface, long press the button to enter the menu. The
following is a detailed introduction to the menu content.
7.1 Probe Set
7.1.3 Remove probe
The removed electrode will no longer be displayed in the measurement interface.
The instrument provides a relay output contact, and the user can set the function and
parameters of the relay.
7.2 Alarm set
7.2.1 Wash relay
The relay is used for wash control, and the cleaning time and cleaning interval can be
set to be closed periodically.
The relationship between the relay state and the cleaning time and interval is shown in
the following figure:
Wash state:
Select "hold", when cleaning, the displayed value remains the same as before the relay
action.
Select "constant", the measured value is the real-time value of continuous measurement.

12
The relay is used for alarm control and can only be used after installing an external
expansion module.
7.2.2 Alarm relay
To set the current signal output, it can only be used after installing an external
expansion module.
7.3 Current set
The instrument provides a digital communication interface, and this menu can set the
parameters of the communication interface.
7.4 Comm.set
Set the slave address used by the instrument in the communication network.
7.4.1 Net address
Select the baud rate used by the communication interface.
7.4.2 Baud rate
Select the verification method, the default is "none".
7.4.3 Verify bit
Select the number of stop bits, the default is "1bit".
7.4.4 Stop bit
Set the floating point format in communication data, the default is "CDAB".
7.4.5 Format
Select the communication protocol, the default is "ModBus".
7.4.6 Protocol
7.5 System set
Set the system time "Year/Month/Day" and "Hour/Minute/Second".
7.5.1 Data/Time set
Set the LCD backlight working mode. If it is set to "scanty power" mode, the backlight
will automatically go out without any key operation within the set time.
7.5.2 Backlight
Set the contrast of the LCD display, the left key decreases the contrast, the right key
increases the contrast.
7.5.3 Contrast set
View program version information.
7.5.4 Software
Used to set the verification password when entering the menu in measurement mode.
The password is set to "000000" when leaving the factory. At this time, the menu can
be directly entered without a password in the measurement mode. If the user sets
another password, the user will be prompted to enter the password when entering,
and the menu can be entered only after the password is verified.
7.5.5 Password

13
After entering the menu, you can enter the query date, and press the key to
display the historical data closest to the query point.
7.6 History logs
View meter history storage information.
7.6.1 Memory inform
Set the interval time for the instrument to record the measurement data, the setting
range is 0~999min. When the recording interval is set to 0, it means that the recording
function is turned off.
7.6.2 Set interval
7.6.3 Ponit query
Record electrode warn logs.
7.6.4 Warn log
The displayed measurement value can be set as "probe measure" or "sim measure".
Probe measure: The screen displays the measurement value as the value measured
by the electrode.
Sim measure: The measurement value displayed on the screen is the input value set
by the user.
7.7 Servings
7.7.1 Measure mode
After the instrument is connected to the expansion module, the module can be added
by scanning.
7.7.2 Output module
The instrument calibration parameters and all setting parameters can be
restored to the factory default state.
7.7.3 Restore defaults

14
1
2
3
4
5
6
7
8
Slave
address
Function
code
Register
start
Address
high 8 bit
Register
start
address
low 8 bit
Register
number
high 8 bit
Register
number
low 8 bit
CRC
low 8 bit
CRC
high 8 bit
Addr
04
xx
xx
xx
xx
CRC
CRC
This function enables the master (host computer) to obtain the real-time measurement
value of the slave (instrument). The value is specified as a single-precision
floating-point type (that is, occupying two consecutive register addresses), and the
corresponding parameters are marked with different register addresses;
The host can send a command frame to read a single or all data results. The data frame
format is as follows (the data is all in Hex format):
2. Function code details (function code 0×04):
The instrument acts as a slave on the network and supports Modbus RTU
communication protocol.
Data communication is initiated by the host, and the first byte of the transmitted
message is the target slave address. When the first byte is received by all the slaves on
the network, each slave will decode it to determine whether the message is sent to
itself.
The transmission of the RTU message frame must start with a pause interval of at least
3.5 characters. After the last character is transmitted, a pause of at least 3.5 characters
marks the end of the message frame. A new message can start after this pause. In the
transmission process, the entire message frame must be transmitted in a continuous
stream. If there is a pause interval of more than 1.5 characters before the completion of
the message frame transmission, the receiving device will refresh the incomplete
message and assume that the next byte is the beginning of a new message. Similarly, if
a new message starts after the previous frame in less than 3.5 characters, the receiving
device will consider it a continuation of the previous frame, and this will cause an error
because of the final CRC The value cannot be correct.
The meter uses the function code 0x04 to complete the output of the measurement
results and the output of two current values.The meter uses the function code 0x04 to
complete the output of the measurement result and the output of the two current values.
A Modbus register information
Appendix
1. About Modbus RTU overview:
Host send:
Table of contents
Other Daruifuno Controllers manuals