Delton PC-80D User manual

User Manual for PC-80D Easy ECG Monitor
I
Instructions to User
Dear Users,
Thank you very much for purchasing our product. Please read the manual very carefully before using this device.
These instructions describe the operating procedures to be followed strictly. Failure to follow these instructions may cause
measuring abnormality, device damage, or personal injury.
No part of this manual may be photocopied, reproduced or translated into another language without the prior written consent.
Creative reserves the right to improve and amend it at any time without prior notice.
Version of This Manual: Ver1.1
All rights reserved.
3502-2430011

User Manual for PC-80D Easy ECG Monitor
II
WARMING
1. Check the device to make sure that there is no visible damage that may affect user’s safety and measurement performance. Stop using the unit,
when there is obvious damage.
2. Stop using the ECG cable if it is damaged, please change a new one.
3. Do NOT make diagnosis oneself by the measurement and measurement results, always consult the doctor if abnormal information is presented
frequently.
4. It is NOT designed or intended for medical diagnosis.
5. It is NOT designed or intended for therapy purpose.
6. Do NOT use the device in the bathroom or moist circumstance.
7. Do NOT operate in the environment where strong electro-magnetic interference exists.
8. Necessary service must be performed by qualified service engineers ONLY.
9. Clinical ECG examination is suggested for normal examination of your heart condition.
10. Connect/Disconnect data cable to the monitor gently and carefully, do NOT wring or shake the data cable plug with force to insert into or
unplug from data interface port, or the data interface port may be damaged.
11. When the ECG amplifier is overload or saturation due to lead off or poor contacts between electrode and the skin, the monitor will prompt
“Contact?” to remind the operator.
12. Do not touch the metal electrodes or place the electrodes of lead wire on patient for ECG measurement while the USB cable is connected to
PC, electrical hazard will be caused by the leakage current.
13. ECG electrodes, lead wires and cable are not allowed to contact any other conductive parts.
14. It is strongly recommended to use Ag /AgCl (Silver Chloride) electrode to make sure the ECG signal quality is good.
15. Dispose of the expired device and its accessories according to applicable local regulations.

User Manual for PC-80D Easy ECG Monitor
III
Table of contents
1 Overview..................................................................................................................................................................................... 1
1.1 Appearance............................................................................................................................................................................ 1
1.2 Name and Model ................................................................................................................................................................... 2
1.3 Structure................................................................................................................................................................................ 2
1.4 Features................................................................................................................................................................................. 2
1.5 Intended Use.......................................................................................................................................................................... 3
1.6 Operating Environment ......................................................................................................................................................... 3
2 Installation.................................................................................................................................................................................. 3
2.1 Batteries installation.............................................................................................................................................................. 3
3 Default menu screen................................................................................................................................................................... 4
3.1 Menu screen .......................................................................................................................................................................... 4
4 ECG measurements.................................................................................................................................................................... 5
4.1 Methods of ECG measurement.............................................................................................................................................. 5
4.2 Measuring procedure............................................................................................................................................................. 8
4.2.1 Preparing Phase................................................................................................................................................................ 8
4.2.2 Quick measurement by built-in electrodes........................................................................................................................ 8
4.3Long term measurement by external lead wires...................................................................................................................... 8
4.3.1 Chest lead measurement................................................................................................................................................... 9
4.3.2 Limb lead measurement.................................................................................................................................................... 9
4.4Data analysis and measurement Result Description...............................................................................................................11

User Manual for PC-80D Easy ECG Monitor
IV
5 Waveform Review..................................................................................................................................................................... 12
5.1 Record list ........................................................................................................................................................................... 12
5.2 Waveform Review............................................................................................................................................................... 13
6 Data Management.................................................................................................................................................................... 15
6.1 Storage status....................................................................................................................................................................... 15
6.2 Delete ID............................................................................................................................................................................. 15
6.3 Delete all ............................................................................................................................................................................. 16
6.4 Upload Data......................................................................................................................................................................... 16
7 Measurement option setting..................................................................................................................................................... 17
7.1Creating new ID ................................................................................................................................................................... 18
8 System Setup............................................................................................................................................................................. 19
8.1 Sound setting....................................................................................................................................................................... 19
8.2 Brightness Setting................................................................................................................................................................ 19
8.3 Time/Date Setting................................................................................................................................................................ 19
8.4Auto power off ..................................................................................................................................................................... 20
8.5 Power saving....................................................................................................................................................................... 20
8.6 Language............................................................................................................................................................................. 20
9 Help........................................................................................................................................................................................... 21
10 Technical Specifications ......................................................................................................................................................... 22
10.1 ECG measurement............................................................................................................................................................. 22
10.2 Power supply..................................................................................................................................................................... 22
10.3 Classification..................................................................................................................................................................... 23

User Manual for PC-80D Easy ECG Monitor
V
10.4 Display .............................................................................................................................................................................. 23
10.5 Data Memory..................................................................................................................................................................... 23
10.6 Overall Dimension and Weight.......................................................................................................................................... 23
10.7 Additional declaration........................................................................................................................................................ 23
11 Maintenance............................................................................................................................................................................ 24
12. Troubleshooting..................................................................................................................................................................... 25
13 Key of Symbols....................................................................................................................................................................... 26
Appendix 1 Additional measuring method................................................................................................................................ 27
Appendix2 Measurement Result Description............................................................................................................................ 29
Appendix3 General Knowledge of ECG.................................................................................................................................... 31

User Manual for PC-80D Easy ECG Monitor
1
1 Overview
1.1Appearance
1. Power: Long time press (about 2 seconds) makes the device power on or off; short time
press turns on/off the backlight.
2. Back: back to upper level of menu.
3. Metal Electrode I
4. Measurement: shortcut button for built-in electrodes measuring, press this button to start
measuring.
5.
▇OK:
confirm selection or modification.
6. Navigation key:
Left: move the cursor forwards/to the left, or adjust parameter values.
Right: move the cursor backwards/to the right, or adjust parameter values.
Up: move the cursor upward/forward, or adjust parameter values.
Down: move the cursor downwards /backward, or adjust parameter values.
7. Display screen: display ECG waveforms and data.
8. I/O cover
9. Lanyard hole
10.USB interface: USB data socket.
11.DC input socket
12.
:
Lead wire socket: for connecting to lead wire.
13.Speaker
14.Metal electrode II and III
15.Battery cover, nameplate
Figure1-1

User Manual for PC-80D Easy ECG Monitor
2
1.2 Name and Model
Name: Easy ECG Monitor
Model: PC-80D
1.3 Structure
PC-80D Easy ECG Monitor mainly consists of circuit boards, display screen and metal electrodes/lead wires.
1.4 Features
1. Small in size, light in weight, and easy to carry.
2. One-key measurement, easy to operate.
3. ECG waveform and interpretation of results are displayed clearly on color dot-matrix LCD screen.
4. UP to 19 types of measurement results can be presented.
5. High capacity of built-in memory, 24 hours records storage and review for single channel waveform; and 8 hours records storage and review
for 3 channel waveforms.
6. Quick measurement by built-in metal electrodes, external lead wires (including limb lead and chest lead) for long term measurements are
optional.
7. Voice prompt is available
8. With data management function, data records can be reviewed and deleted locally, uploaded to PC as well.
9. Auto power off time can be set.
10.With power saving feature.

User Manual for PC-80D Easy ECG Monitor
3
1.5 Intended Use
This Easy ECG Monitor is intended for measuring and recording theECG waveform and average heart rate of adult patient. It is applicable for use in
clinics and homes, and convenient to operate by the patients themselves.
This device is not an ECG monitor used in clinical institution or hospital, but is used for spot-check purpose ONLY. It can not be used to
substitute the normal ECG examination or real time monitoring. The measurement results are useful reference for doctor, but do not make direct
diagnostic or analytical decision based on the information provided by this device.
1.6 Operating Environment
Operating temperature 5℃~ 40℃
Operating humidity 30%~80%
Atmospheric pressure 70 kPa ~106kPa
2 Installation
2.1 Batteries installation
1.Open the battery cover by a screwdriver or a coin.
2. Insert four AAsize batteries directly (as shown in figure 2-1).
3.Close the batter cover and fasten it.
Do NOT insert batteries with their polarities reversed. Dispose of the used batteries according to applicable local regulations.
Note: Protecting films are attached to the electrodes of device, so please remove the protecting films before measurement.
Figure 2-1

User Manual for PC-80D Easy ECG Monitor
4
3 Default menu screen
3.1 Menu screen
Long time press the“ ”Power key (lasting for 2 seconds) until a beep sound is heard to indicate the unit is turned on, then the booting screen will
appear, as shown in the figure 3-1, after that the device enters into the default menu screen, as shown in figure 3-
2.
There are 6 entries, press navigation key to shift focus on them and press “” ok key to go into the selected entry. Each entry has an icon and it is
defined as following:
Measure: Detect ECG signal, mean while, display waveform, HR value, and analyze the ECG
waveform to prompt necessary information.
Review: Recall and view the ECG records stored in this device (including waveform data and
measurement result).
Memory: ECG records management including storage status check and record deletion.
Option: Measurement related setup including signal input selection, measuring time, bandwidth,
HR alarm limit setting and user ID (ID number).
Setup: System setup including date/time, language, sound and display etc.
Help: Provide on-line information about measuring methods and general knowledge of ECG.
Version:1.10
Figure 3-1
Figure 3-2

User Manual for PC-80D Easy ECG Monitor
5
4 ECG measurements
There are 2 measuring modes for the ECG measurement, which includes quick measurement by built-in electrodes and long term measurement by
external lead wires.
Quick measurement: press “ ” measure key to enter into the quick measurement directly; it can display single channel ECG signal measured by
built-in electrodes, the signal can be lead I, lead II or chest lead according to the electrode placement. This measurement takes 30 seconds.
Long term measurement: after selecting the signal input option in the measuring option menu, move the cursor to the “ ” measure icon, then press
the “”ok key to start the long term measurement.
For long term measurement, the external lead wires will be used in stead of built-in electrodes; there are two options for signal input.
Limb lead measurement: it measures 3 channels ECG signal for lead I, lead II, and lead III simultaneously and displays them;
Chest lead measurement: it measures single channel ECG signal with different chest lead according to the electrode placement. It is suitable for long
term measurement.
4.1 Methods of ECG measurement
It is recommended to clean your skin with SOAP and water to keep it moist without oily.
1. Measurement by built-in electrodes:
Easy and quick ECG measurement by built-in electrodes can be done by putting electrodes on hand, leg and chest
respectively.
1). Hand measurement
It is convenient to take ECG measurement by hands or palms. The ECG waveform detected by this method is
equivalent to Lead I of regular ECG examination.
A. Palm placement: Hold the device with right hand; make sure that the fingers touch the metal electrode II and III reliably.
Press the electrode I on the left palm, as shown in figure 4-1.
B. Wrist placement: Hold the device with right hand; make sure that the fingers touch the metal electrode II and III
reliably. Press the electrode I below the left wrist about 1 inch, as shown in figure 4-2. When the ECG signal at palm
placement is not good, try to measure with wrist placement.
Figure 4-1 palm placement
Figure 4-2 wrist placement

User Manual for PC-80D Easy ECG Monitor
6
Note: Keep at ease and make sure the skin contact well with the electrodes during ECG measurement. Please place the device on correct position, do
not shake it during the measurement. Dirty skin may cause poor signal quality, so clean your skin before measuring.
2). Leg Measurement
The ECG waveform detected by Leg Measurement is equivalent to Lead II of regular ECG examination. Keep at ease when
seating down, and then lift your left leg on a bench. Hold the device with the right hand when measuring and make sure that
the fingers touch the metal electrode II and III reliably, place the electrode I on the left leg above the ankle, as shown in
figure 4-3.
Note: Keep at ease and make sure the skin is clean and moist before measuring.
3). Chest Measurement
It is recommended to take ECG measurement on chest, since the signal detected by this way is stronger and with less
interference. The ECG waveform detected by Chest Measurement is similar to the Lead V4 or V5 of regular ECG
examination. Hold the device with the right hand, make sure that the fingers touch the metal electrode II and III
reliably, as shown in figure 4-4, place the electrode I on the left chest align with the midclavicular line (V4 position) or the
anterior axillary line(V5 position) and on the level of the 5th intercostal space.
Note: keep at ease and clean the electrode I by using the wet cloth before measuring. If the waveform baseline wanders or
noisy waveform appears, hold your breath for a moment to get stable and clean signal.
2. Measurement by external lead wires
The ECG signal which is detected by external lead wires is more stable than that by hand measurement. ECG
measurement by external lead wires can be done by placing the electrodes on limbs or chest respectively.
1). Electrodes placement on limbs
On default menu screen enter “ ” option setup menu to select “limb lead” for signal input. By this way, the 3 channels of
limb lead I, II and III can be measured and displayed simultaneously and continuously. Either adhesive electrodes or limb clip
type electrodes (optional) can be used to do the measurement. The location of the electrodes is shown in figure 4-5.
R (red): place on interior of right arm (between the wrist and the elbow);
L (yellow): place on interior of the left arm (between the wrist and the elbow);
F (green): place on interior of the left leg (between the knee and the ankle);
N (black): place on interior of the right leg (between the knee and the ankle)
Figure 4-4 Chest measurement
Figure 4-3 Leg measurement
Figure 4-5 the location of electrodes on limbs

User Manual for PC-80D Easy ECG Monitor
7
Operation: Connecting the lead wires after placing the electrodes well and then insert the connector onto the device; Power on the device to select
the “limb lead” for signal input and set the measuring time on option setup menu, then press “”ok key to start the measurement.
Note: keep at ease when seating or lying down. Clean the skin by using the wet cloth when selecting limb clip type electrodes (optional) to measure.
Direction of using disposable ECG electrodes: Tearing off the protective film to place it on the skin, and then connecting the electrodes with lead
wires. It is recommended to put the electrodes on chest for longtime monitoring, please refer appendix 1 for more details.
2). Electrodes placement for Chest Lead Measurement
On default menu screen enter “ ” option setup menu to select “chest lead” for signal input. ECG signal detected by this method is single channel
measurement, which is equivalent to the waveform of regular chest lead. It is intended for professional use (ie. by doctor). The location of the
electrodes is shown in figure 4-6.
R (red): place on interior of right arm (between the wrist and the elbow);
L (yellow): place on interior of the left arm (between the wrist and the elbow);
F (green): on the 5th intercostal space at the left anterior axillary line;
N (black): unused.
Operation: Connecting the lead wires after placing the electrodes well and then insert the connector onto the device; Power
on the device to select the “chest lead “for signal input and to set the measuring time on option
setup menu, then press “”ok key to start the measurement.
Note:
1. The Chest Lead Measurement can obtain the ECG signal of Chest Lead (V1-V6). Please refer to appendix I to select the location of F (green)
electrode when you want to get the different chest lead. At the same time, you should set the lead information (V1-V6) that you want on the option
setup menu.
2. Either adhesive electrode or sucking bulb electrode (optional) can be used to Chest Lead Measurement. Please refer to appendix I for more details
about measuring methods.
Figure 4-6 the location of chest lead measurement

User Manual for PC-80D Easy ECG Monitor
8
4.2 Measuring procedure
4.2.1 Preparing Phase
The quick measurement is as following for details:
At default menu screen, press “ ” measure key to enter into the quick measuring screen directly, as shown in figure 4-7, in addition, at the
beginning of measurement (including quick and long term measurement), the preparing phase shows on the screen with a heart beat icon and counting
down seconds, the beeper will sound with the counting down at the interval of one second.
During this period the device will check the signal quality and auto scale the display. Please adjust your holding position if the detected waveform is
not smooth or poor signal and it may prompt:“Check well-connection with lead wires.”
4.2.2 Quick measurement by built-in electrodes
Please refer to the section 4.1 “Hand measurement” to make the electrodes well-connection with skin. At default menu screen, press “ ” measure
key can enter into quick measurement preparing screen directly, after 10 seconds, it will enter into the measuring screen, as shown in Figure 4-8. The
measuring time lasts for about 30 seconds.
4.3Long term measurement by external lead wires
There are 2 options (limb lead and chest lead) for long term measurement by external electrodes. The detail information about the electrodes
placement is described in section 4.1.
Figure 4-9measuring(chest lead )
Figure 4-7 (quick measurement)
Figure 4-8(quick measurement)

User Manual for PC-80D Easy ECG Monitor
9
4.3.1 Chest lead measurement
Please refer to the section 4.1 “Chest lead measurement” to connect the lead wires; enter “ ” option setup menu to select “chest lead” for signal
input; return to default menu screen, move the cursor to the “ ” measure icon, press the “”ok key to enter into external electrodes measurement (3
lead wires) preparing screen, after 10 seconds, it will enter into the chest lead measuring screen, as shown in Figure 4-9. In this measuring mode, it
can measure and display the ECG signal of lead I, lead II, lead III or chest lead (V4) respectively according to electrodes placement.
4.3.2 Limb lead measurement
Please refer to the section 4.1 “Chest lead measurement” to connect the lead wires; enter “ ” option
setup menu to select “limb lead” for signal input; return to default menu screen, move the cursor to the
“” icon, press the “”ok key to enter into external electrodes measurement (4 lead wires) preparing
screen. After 10 seconds, it will enter into the limd lead measuring screen, as shown in Figure 4-10.
On-screen display:
(1 ) “00
:
00
:
01”: Measuring time
(2) “X1”
:
“X1”: ECG waveform amplitude scale; “X2”: twice size of the nominal scale; “X4”: four
times size of the nominal scale.
(3) “ ”: Heart beat indicator. It flashes while heart beats.
(4) “HR=65”: Current HR value. It is displayed during measurement only when external lead wires are used, if the ECG signal is poor, this value
could become “---”.
(5) : Enhanced filter mode indication. It shows up only when the bandwidth is set to enhanced mode.
(6)“”: Status of battery capacity
(7) “”: ECG waveform
(8)
“
”: Mark corresponding to 1mV amplitude
Figure 4-4 measuring
(
Ext.4
)
Figure 4-10 measuring(limb lead )

User Manual for PC-80D Easy ECG Monitor
10
(9) “”: Status bar of measurement progress.
Operation:
When in the quick measurement by built-in electrodes:
①Press “” key, the measuring display screen will be made a quarter turn counterclockwise;
②Press the “” ok key or “ ”return key to stop measuring and return to the default menu screen directly, the measuring data won’t be saved.
When in the external electrodes lead wire measurement:
①The device will stop measuring automatically if the measuring time is over;
②Press the “” ok key or “ ”return key to stop measuring and return to the default menu screen directly, the measuring data would be saved.
Note:
①If the storage is full, the device will stop to measure and it prompts “Fail to save this record because of full data memory”, at this moment, you
should delete some records to make sure enough storage space, then measuring again.
②If the battery is low, the battery icon “ ” flashes red, the device stop to measure and it prompts “low battery voltage, measuring is interrupted,
return key to exit! ”, at this time, connect to external power can continue to measure, or exchanging a new battery to measure again.
③During measuring mode, it will have regular pulse beep sound.
④During the measurement by external lead wires, if “HR alarm” option is set to “On” and the HR value is out of the alarm limits, the alarm sounds
will occur with 3 short beeps, and the HR value is reversely displayed.
⑤During the quick measurement, if the device does not check well contact with the electrodes or the signal is poor, it will not be calculate the
average HR value, the measuring result is “poor signal, measure again”, this data will not be saved.
⑥During the measurement, if the placement of electrodes loose contact with the electrodes, it will prompt “poor signal”. If it is external lead wire
measurement, the HR value shows “__”.

User Manual for PC-80D Easy ECG Monitor
11
4.4Data analysis and measurement Result Description
When the measuring is finished, the device will analyze the measurement result, as shown in figure 4-11/4-12/4-13; the measurement result
description refers to the index 2.
1. In the external lead wire measuring mode, the current measurement result will be saved acquiescently, as
shown in figure 4-12/4-13. Press the “” ok key or “” back to return the default menu screen after
reviewing the result.
2. After the quick measurement by built-in electrodes, the device will be saved the result automatically, and it
prompts “The result is saved successfully, as shown in figure 4-14. Press the “” ok key or “” back to
return the default menu screen. If without any key operation in 6 seconds, press “ ” back key, return to
default menu screen directly and the current measurement result won’t be saved
Notes:
1. Voice prompts for working status and measuring result.
2. The device can provide with 19 types ECG measurement result display when using the quick measurement, the
measurement result descriptions refer to the appendix 2 for detail.
3. Up to 8 hours records can be stored in built-in memory, when the memory is full, the device will prompt that "Memory full".
4. During the external lead wire measurement, the measurement result screen will not be displayed if you press to stop measuring.
Figure 4-12(no irregularity found in Ext.)
Figure 4-14
Figure4-11(no irregularity found Quick measurement)
Figure 4-13 (irregularity found in Ext.)

User Manual for PC-80D Easy ECG Monitor
12
5 Waveform Review
5.1 Record list
Select “ ” review key on the default menu screen, then press “” ok key to enter into the user ID list screen, as shown in Figure 5-1. Selecting a
folder which contains the data for a given ID,then press OK key,the display screen is as shown in Figure 5-2.
Legend of icons:
: Indicates the measured result without irregularity found.
: Indicates the measured result with irregularity found.
: Indicates that the waveform with much interference or lead off during measurement.
: Indicates the Limb measurement by external electrodes (With 4 lead wires)
: Indicates the Chest lead measurement by external electrodes (With 3 lead wires)
: Indicates the quick measurement (by built-in electrodes)
Note: the ID list only display the ID which contains data.
Figure 5-2 waveform review/record list
Figure 5-1 waveform review/ID list
Figure 5-3 Review(Quick measurement)

User Manual for PC-80D Easy ECG Monitor
13
5.2 Waveform Review
In the record list, selecting a record, then press “”ok key, the quick measurement record is shown in the
figure 5-3; the external lead wire measuring records is shown in figure 5-4.
On-screen display:
(1)“”: Measured ECG waveform
(2) “ ”: Mark corresponding to 1mV amplitude
(3) “ ”: ECG waveform snap-shot
(4) “ ”: Zooming box for ECG waveform
(5) “ ”: Mark of irregular segment of waveform. Please refer to appendix 2 for detailed information.
Notes: The Zooming box contains ECG waveform measured in about 3~4 seconds.
1. For quick measurement by built-in electrodes, the display will review this record automatically, press
“” ok key, the display will browse the recorded waveform automatically, Press it again to stop the
browsing operation; Press the “/” can scroll the waveform manually. When the review is finished,
press “” ok key to bring the measurement result, as shown in figure5-5, in the measuring result screen,
press “” ok key, it prompts “Are you sure to delete this record?”, as shown in figure 5-6 screen. You can
delete this record according the prompts.
2. For external lead wires measurement (limb lead and chest lead), the display shows 5 lines of
compressed waveform, every line contains 30 seconds of waveform data as shown in figure 5-4. Press
the “/” key to move the grey zooming box pane, press “” ok key to zoom in the waveform
display within the grey zooming box.
The chest lead review screen is shown in figure5-7; the limb lead review screen is shown in figure 5-8.
Figure 5-5
Figure 5-4 Review
(
External electrodes
)
Figure 5-6

User Manual for PC-80D Easy ECG Monitor
14
In the figure 5-7/5-8/5-9 screen, press the “” ok key can review the result description for this ID record, as
shown in figure 5-10/5-11. In the measure result screen, press the “” ok key can delete this record. Note:
①When in limb lead review screen, press “
,
” key can switch the three and single signal channel on
screen display; and if there is any irregular found in the review screen, it will display the measuring HR
value instantaneously, and there is an inverted triangle shown on the irregular segment, as shown in figure
5-9.
②For the external lead wire measurement record description, if there is no irregular found, it will display a
smile face and prompts “No irregularity found”. HR trend graph which shown below displays the HR
record in length of 20 minutes in one page, press “,” to review next page, as shown in figure 5-10;
if there is any irregularity found, the result is as shown in the figure 5-11, it will display the statistical
result of irregularity.
Figure 5-7 (The Chest lead)
Figure 5-10 No irregularity found (Ext.)
Figure 5-8 (no irregular found in Limb lead)
)
Figure5-11 (irregularity found in Ext.)
Figure 5-9 (irregular found in Limb
00:00:01 X1
V

User Manual for PC-80D Easy ECG Monitor
15
6 Data Management
Select “Memory” icon on the default menu, then press “OK” button to enter into data
management screen, as shown in Figure 6-1.
6.1 Storage status
On data management screen, select “Statistic”, then press “” ok key to enter into the Storage
status screen, as shown in Figure 6-2.
6.2 Delete ID
1. On data management screen, select “Delete” with Navigation Key, then press “” ok key to
enter into the ID list screen, as shown in figure 6-3.
2. Press the Navigation Key “
,
,
or
” to select the ID; when pressing “”ok key, there are
two different prompts for two different situations:
Situation 1: this ID does not contain any data record.
(1)The ID number of “00000” ~ “00004” is reserved, when pressing “” ok key, the device
prompts “This ID can’t be removed”.
(2)If it is a new ID, press the “”ok key to delete the selected the record according to the
prompt: “Are you sure to delete this ID”.
Situation 2: there are data records under this ID. Pressing “” ok key to enter into the record
list screen, as shown in figure 6-4. The record list screen display the date and time of all records for this ID.
Press the Navigation Key “
” key to perform select this ID(press it again to cancel selection), the selected record will be labeled “√” ;
Figure 6-3
Figure 6-1
Figure 6-2
Table of contents
Other Delton Medical Equipment manuals
Popular Medical Equipment manuals by other brands

GE
GE LOGIC C2 Series Technical publication

VISIOMED
VISIOMED BewellConnect MyTens Pro BW-TS1 manual

BD
BD PureWick Quick setup guide

Zimmer Biomet
Zimmer Biomet Headless Compressionand Twist-Off Screws SURGICAL TECHNIQUE

Flaem
Flaem EasyNeb 3 Instructions for use manual

Siemens
Siemens MOBILETT XP Maintenance Instruction

Sirona
Sirona teneo Installation requirements

Seers Medical
Seers Medical ST7700 operating instructions

ulrich medical
ulrich medical ADDplus anterior distraction device plus Instructions for use

Alu Rehab
Alu Rehab Netti Mini user manual

LionsGate Technologies
LionsGate Technologies Kenek O2 user manual

3B Medical
3B Medical LUNA II Quick operation manual