
LT-2023
10
MAINTENANCE
PERIODIC HOMECARE PROVIDER
PREVENTATIVE MAINTENANCE
Use only DeVilbiss concentrator replacement parts and accessories.
Every DeVilbiss Oxygen Concentrator is tested at the factory. To assure
continued trouble-free performance, the following preventative maintenance
should be performed by the homecare provider during periodic oxygen patient
visits not to exceed 3 years. Failure to properly maintain the unit will void the
warranty.
1. Check the oxygen concentration with an oxygen analyzer (part #R217P62)
every 3 years.
a. Calibrate the oxygen analyzer prior to checking the oxygen
concentration. The analyzer should be properly calibrated using the
manufacturer’s recommended procedure.
NOTE–Changes in temperature, altitude, or humidity may affect
the analyzer’s oxygen concentration reading. The analyzer should be
calibrated in similar conditions to the location of the concentrator.
b. Power the unit, set the ow meter to 5 LPM and connect the analyzer
to the unit’s oxygen outlet port.
c. Wait 20 minutes for the display to stabilize. The concentrator must
operate for a minimum of 20 minutes before checking the oxygen
concentration.
d. Record the reading.
2. Check the audible alert and indicator lights during every service. When the
power switch is turned “On,” listen for the audible alert and check to see if
the front panel indicator lights are operating.
3. Inspect cabinet air lter (part #303DZ-605) every PM check. Replace if lter
is torn or damaged. Units manufactured since October, 2020 do not have a
gross particle or cabinet air lter. See note under Air Filter on page 9.
4. Inspect intake bacteria lter (part #MC44D-605) during every PM Check.
Replace as necessary.
a. Open the lter door and replace lter as required.
5. Inspect the nal bacteria lter (part #PV5LD-651) during every compressor
service. Replace as necessary.
a. Use the Cabinet Removal instructions found under CABINET
REMOVAL in this manual to remove and attach the cabinets.
b. Remove the hose from each end of the lter and discard the lter.
c. Install the new nal bacteria lter with the “IN” tting toward the ow
meter.
6. Inspect the compressor lter during every compressor service. Replace as
necessary.
NOTE–A change was made to the type of compressor lter used in the 525
series concentrator. The unit may have one of two different types of compressor
lters depending on when the concentrator was manufactured. The original one
is a white plastic lter that is located near the compressor. The newer style is a
sintered bronze lter (525DD-626) that is installed in one end of the compressor
exhaust hose near the rotary valve. All models of the 525 series are now being
manufactured with the sintered bronze compressor lter. See gure below.
Sintered Bronze Compressor Filter
(525DD-626)
NOTE– This PM Schedule reects:
• 5000 hour usage equal to one year
• a normal, clean operating environment.
The homecare provider is responsible for:
• determining the condition of the concentrator operating environment.
• determining a preventative maintenance interval frequency* which takes
into consideration the specic operating environment.
* Standard intervals are noted below. Service interval may be more or
less frequent than stated below provided that the Home Care
Provider establishes and documents appropriate protocols.
7. Inspect AC power cord, power switch and circuit breaker every PM check
or between patients. Replace as necessary.
PROVIDER’S NOTES - Cleaning and
Disinfection When There is a Patient Change
DeVilbiss Healthcare recommends that at least the following procedures be
carried out by the manufacturer or a qualied third party between uses by
different patients.
NOTE – If the following described complete processing of the concentrator by
an appropriately trained individual is not possible, the device should not be used
by another patient.
NOTE – If preventive maintenance is due at this time, these procedures
should be carried out in addition to the servicing procedures.
1. Use disinfectants safely. Always read the label and product information
before use.
2. Always wear personal protective equipment when performing this
procedure. Use suitable gloves and safety glasses. Cover exposed skin on
arms to prevent accidental contact with bleach solution that has been
applied to the concentrator.
3. Dispose of all accessories that are not suitable for reuse. This includes but
may not be limited to the oxygen tubing, tubing connectors, nasal cannula
and/or mask, oxygen outlet connector, and humidier bottle.
4. Clean the exterior of the concentrator with a clean lint-free cloth. Heavy soil
should be removed with a clean lint-free cloth dampened with water. A soft
bristled brush dampened with water can be used to remove stubborn soil.
Dry the concentrator using a clean lint-free cloth if water was used to
remove soil.
5. Use 5.25% chlorine bleach (Clorox Regular Liquid Bleach or equivalent).
Mix one (1) part bleach with four (4) parts water in an appropriate clean
container. This ratio produces a one (1) part bleach to ve (5) total parts
solution (1:5). The total volume (amount) of solution required is determined
by the number of concentrators in need of disinfection. NOTE– An
alternate suitable disinfecting agent (e.g. Mikrobac® forte or Terralin®
Protect) may also be used. Follow disinfectant manufacturer’s instructions.
6. Apply the bleach solution in an even manner to the cabinet and power cord
using a clean lint-free cloth. The cloth should be dampened only and not
dripping of solution. Do not use a spray bottle to apply the solution. Do not
saturate the device with the solution. Take care that no solution enters the
vent areas on the concentrator base or the Auxiliary O2 tting area on the
back of the unit. Avoid over-saturating the cabinet seams so that no
solution residue builds up in these areas. Avoid the caster wells located on
the bottom of the unit.
7. Exposure time of the disinfectant solution should be 10 minutes minimum
to 15 minutes maximum.
8. After the recommended exposure time, all surfaces of the concentrator
should be wiped with a clean lint-free cloth dampened with drinking quality
water no warmer than room temperature. Dry the unit with a dry, clean lint-
free cloth. This is to remove residue that may stain or leave a lm on the
unit, especially after repeated disinfections.
9. Check the cord, the plug on the back of the device, the power switch, the
fuse holder, and the indicator lights for possible damage. Replace all
damaged or worn components.
10. Replace the cabinet air lter on the back of the device.
11. Check the oxygen concentration. If the device is within specication, the
extended life intake bacteria lter does not need to be replaced between
patients. If the oxygen concentration is not within specication, the provider
should refer to the service manual section on Troubleshooting.
12. OPTIONAL INSIDE CLEANING: The concentrator must be disconnected
from the power supply for this step: Open the concentrator and remove all