DIGISYSTEM LABORATORY INSTRUMENTS INC. www.digisystemlab.com R EV.1-2017
INSTRUCTION MANUAL FOR PIPETTE SHAKER PS-600T
Warnings:
* Do not remove the shaker during operation.
* Place the shaker securely away from the edge of bench to avoid the rotator mixing falling off the bench
top.
* Do not disassemble the shaker by yourself. In the event of a malfunction, the shaker should be repaired by
a qualified technician. Before repairing, turn off the power.
1. Place the pipette shaker on a flat and stable bench top. Plug it into a properly grounded outlet which
corresponds with the electrical requirements on the label back on machine.
2. Place blood diluting pipettes on the ‘Holder’.
3. Select the switch model. You can place the ‘Switch’ at ‘TIMER’, ‘OFF’, or ‘ON’ position. When the ‘Switch’
is at ‘TIMER’ or ‘ON’ position, the ‘Power indicator’ lights.
3.1 If the Timer mode is selected, you can turn the ‘Timer’ to a required scale. There is a bell sound to
be heard when the setting time is end.
* How to set the ‘Timer’ correctly?
Turn the ‘Timer’ over the required scale, then return the time calibration as a required scale. (For
instance, you need 5 minutes for rotation, but you need to wind to 7 or 8 minute scale at first, then
return back to 5 minute scale.)
3.2 If the ‘ON’ mode is selected, the rotator will rotate continuously.
3.3 If the ‘OFF’ mode is selected, the power will be cut off.
Maintenance and care:
Keep the machine clean all the time, so no routine maintenance is required. Cleaning can be done with a
damp cloth. Avoid the use of solvents as they may damage the machine housing.
Holder
Power
indicator
Switch
Timer
DIGISYSTEM LABORATORY INSTRUMENTS INC. www.digisystemlab.com R EV. 1-2017
Circuit Diagram
Troubleshooting Guide
Caution: Always disconnect the power cord before troubleshooting.
Trouble
Cause
Remedy
Instrument inoperative.
Power cord not connected to outlet.
Plug in the power cord.
Dead power output.
Change to different output.
Holder doesn’t shake.
Power is not on.
Place the switch at ‘Timer’ or ‘ON’ position.
Belt is broken
Contact your distributor for repair.
Motor is broken. Contact your distributor for repair.
Power indicator can not light
up. The power indicator is broken. Contact your distributor for repair.
Timer is not accurate Timer is not set by the correct procedures. Read the operation manual and set it again.
Timer is broken. Contact your distributor for repair.

Other DIGISYSTEM Laboratory Equipment manuals

DIGISYSTEM DSC-103SD User manual

DIGISYSTEM

DIGISYSTEM DSC-103SD User manual

DIGISYSTEM DSB-500 User manual

DIGISYSTEM

DIGISYSTEM DSB-500 User manual

DIGISYSTEM WB-500D User manual

DIGISYSTEM

DIGISYSTEM WB-500D User manual

DIGISYSTEM DSB-500D User manual

DIGISYSTEM

DIGISYSTEM DSB-500D User manual

DIGISYSTEM RM-501 User manual

DIGISYSTEM

DIGISYSTEM RM-501 User manual

DIGISYSTEM DSO-D Series User manual

DIGISYSTEM

DIGISYSTEM DSO-D Series User manual

DIGISYSTEM DSD-100 User manual

DIGISYSTEM

DIGISYSTEM DSD-100 User manual

DIGISYSTEM DSC-302SMD User manual

DIGISYSTEM

DIGISYSTEM DSC-302SMD User manual

DIGISYSTEM VM-2000 User manual

DIGISYSTEM

DIGISYSTEM VM-2000 User manual

DIGISYSTEM DSD-100D User manual

DIGISYSTEM

DIGISYSTEM DSD-100D User manual

Popular Laboratory Equipment manuals by other brands

BioCare Medical Decloaking Chamber Pro manual

BioCare Medical

BioCare Medical Decloaking Chamber Pro manual

Bibby Sterilin Techne TC-3000 Operator's manual

Bibby Sterilin

Bibby Sterilin Techne TC-3000 Operator's manual

NoiseKen EPS-02Hv2 instruction manual

NoiseKen

NoiseKen EPS-02Hv2 instruction manual

OLS CASY TT Operator's guide

OLS

OLS CASY TT Operator's guide

Biobase BK-AutoHS96 user manual

Biobase

Biobase BK-AutoHS96 user manual

Reichert Darkfield Quebec 3325 manual

Reichert

Reichert Darkfield Quebec 3325 manual

Tuttnauer 1730E Operation & maintenance manual

Tuttnauer

Tuttnauer 1730E Operation & maintenance manual

NEW BRUNSWICK SCIENTIFIC innova 4340 Guide to operations

NEW BRUNSWICK SCIENTIFIC

NEW BRUNSWICK SCIENTIFIC innova 4340 Guide to operations

Speakman SE-925 Installation, Maintenance & Operation Instructions

Speakman

Speakman SE-925 Installation, Maintenance & Operation Instructions

Ecolab DG 1-II Operating	 instruction

Ecolab

Ecolab DG 1-II Operating instruction

Pall USD3295 Instructions for use

Pall

Pall USD3295 Instructions for use

Scilogex HB120-S user manual

Scilogex

Scilogex HB120-S user manual

Phoenix CAC3 Operating and service manual

Phoenix

Phoenix CAC3 Operating and service manual

Metrohm 860 KF manual

Metrohm

Metrohm 860 KF manual

Thermo Scientific SimpliAmp Installation and operational manual

Thermo Scientific

Thermo Scientific SimpliAmp Installation and operational manual

Merck Millipore Milliflex user guide

Merck

Merck Millipore Milliflex user guide

CBM SST 1700B instruction manual

CBM

CBM SST 1700B instruction manual

Labconco 44003 Series Service – technical manual

Labconco

Labconco 44003 Series Service – technical manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.