C.B.M. S.r.l. a socio unico - Medical Equipment –
Via Castello 10, 2603 Torre dé Picenardi (CR) - Italy
Rev. 06 – 18/02/2019 Manuale SST1700-2200 r06 CE-1282 EN page 3 of 34
WARNINGS
WARNING: Burn hazard: The inner parts of the chamber and door reach
high temperatures.
WARNING: Electrocution hazard: Before removing any part of the sterilizer,
switch off the appliance and disconnect the power cable.
Warning: Before opening the door always check that the pressure shown
on the manometer is zero
Read the operating instructions carefully before using the appliance; the operating
instructions must be kept near the appliance.
Use of this appliance is reserved to trained and qualified personnel.
The sterilizer must be used with demineralized water only, without additives.
In the event of a power loss, the writing “Alarm 98 Black Out” will appear on the display
once power is re-established and the cycle shall be considered null.
In any case the cycle shall be considered valid if at the end of the cycle the
message CYCLE COMPLETED appears on the display
Check that the material is sterilizable with the chosen sterilization cycle.
It is recommended that the autoclave be used in a clean place.
The appliance is not intended for the sterilization of liquids, pharmaceutical and
radioactive products, therefore it is advisable NOT to introduce such materials.
NEVER use the appliance without the anti-bacteria filter.
Use original C.B.M. spare parts only.
In the event the appliance should be returned for repair, servicing or any other reason,
please use the original package and empty the water out of the tanks. C.B.M. S.r.l.
declines responsibility for damages caused during transport by inadequate packing.
Remember that the drying function OFF must not be used with wrapped materials.
This appliance has been designed and built in compliance with the community directive
93/42/EEC "Medical Devices” and subsequent amendments. C.B.M. S.r.l. considers itself
responsible for the purposes of safety, reliability and performance provided that any
modifications or repairs are carried out by authorized personnel, that the premises’ wiring
system is compliant with IEC Standards and that the appliance is used in compliance with
these operating instructions.
C.B.M. S.r.l. declines all responsibilities for any damages to people or property deriving
from improper use or failure to connect to the earth cable.
USE RECOMMENDATIONS
For correct autoclave use follow the indications listed hereafter.
•It is recommended to clean and dry the instruments to be sterilized prior to introducing
them in the sterilizer. It is NOT advisable to use liquids containing chlorides which could
rapidly cause damage to the inner parts of the appliance.
•Materials not intended for immediate use should be sterilized inside a packaging (bags
and/or rigid containers for steam sterilization).
•Please be informed that the 134°C – 4 min. Flash cycle type S is suitable only for
sterilization of single pieces of material packed in bags or simple hollow item unpacked