DISTELL Torrymeter User manual

USERMANUAL
DISTELLFISHFRESHNESSMETER
ModelTorrymeter
©DISTELL.Com,December2010(Version2.8)–forusewithv1firmware

distell‐ing.pres.part.verb.1.todistell.2.to
rapidlyanalysefatcontentoffoodproductsusing
microwavetechnology.

...theartofdistell‐ing...
FreshnessMeterUserManualPage3
Contents
Introduction........................................................................................................................................5
Freshnessoffishanditsmeasurement.......................................................................................................................5
FreshnessMeterbenefits............................................................................................................................................5
Generaldescription.....................................................................................................................................................6
Principleofoperation..................................................................................................................................................6
Technicaldata.............................................................................................................................................................7
Step‐by‐stepguide..............................................................................................................................8
TheFreshnessMeterkit..............................................................................................................................................8
MeterUnit..................................................................................................................................................................8
Data/powerSocket.............................................................................................................................................8
Sensor..................................................................................................................................................................8
ReadButton(Yes)................................................................................................................................................8
ResetButton(No)................................................................................................................................................8
Datacable...................................................................................................................................................................8
CD‐ROM......................................................................................................................................................................9
Usermanual................................................................................................................................................................9
Carrycase...................................................................................................................................................................9
Powersupply/charger...............................................................................................................................................9
ChargingtheMeter..............................................................................................................................................9
Gettingstarted..................................................................................................................................10
Switchon..................................................................................................................................................................10
Measuringproductsamples......................................................................................................................................10
Comparingthemeterwithlaboratoryanalysis.........................................................................................................11
MenuStructure.........................................................................................................................................................12
Productmenu....................................................................................................................................................12
Researchmenu..................................................................................................................................................13
Samplesmenu....................................................................................................................................................13
Downloadmenu.................................................................................................................................................13
AddProductoption............................................................................................................................................14
Date&Timemenu.............................................................................................................................................14
Usagerecommendations...................................................................................................................15
Usingthemeterinqualitycontrol.............................................................................................................................15
Sampling............................................................................................................................................................15
FreshnessMeterperformance...................................................................................................................................15
Therelationshipbetweenmeterreadingsandothermeasuresoffreshness.....................................................15
Effectsofhandlingandprocessing.....................................................................................................................17
Effectsoffreezingandbrining............................................................................................................................17
Metercareandservice..............................................................................................................................................18
Cleaningandgeneralcare..................................................................................................................................18
Beforecallingforservice....................................................................................................................................18
Warrantyandservice.........................................................................................................................................18
DataManagementSystem................................................................................................................19
Installation................................................................................................................................................................19
Requirements.....................................................................................................................................................19
InstallingtheDMS..............................................................................................................................................19
RunningtheDMS......................................................................................................................................................19
HistoricalDataDownload.........................................................................................................................................19
Step1:ConnectingtheMeter............................................................................................................................19
Step2:ConfiguretheDMS.................................................................................................................................19
Step3:Initiatethedownloadfromthemeter....................................................................................................20
RealTimeDataDownload.........................................................................................................................................20
Enablingreal‐timedatadownload.....................................................................................................................21

...theartofdistell‐ing...
FreshnessMeterUserManualPage4
Annotatingthedata.................................................................................................................................................21
DMSdatafiles..........................................................................................................................................................21
Printingdata.............................................................................................................................................................21
Exportingdata..........................................................................................................................................................21
DataManagementsoftwareHelp............................................................................................................................21
Advancedtopics.......................................................................................................................................................21
Organolepticcharts..........................................................................................................................22
Cod,boxed................................................................................................................................................................23
Cod,normalhandling...............................................................................................................................................24
Codfillets,skin‐on(skinside)....................................................................................................................................25
Codfillets,boneside.................................................................................................................................................26
Haddock&Whiting,boxed.......................................................................................................................................27
Haddock&Whiting,normalhandling.......................................................................................................................28
Saithe,boxed............................................................................................................................................................29
Saithe,normalhandling............................................................................................................................................30
Plaice,boxed............................................................................................................................................................31
Place,normalhandling.............................................................................................................................................32
Redfish,boxed..........................................................................................................................................................33
Redfish,normalhandling..........................................................................................................................................34
Herring,5%fat,boxed,unhandled...........................................................................................................................35
Herring,10%fat,boxed,unhandled.........................................................................................................................36
Herring,15%fat,boxed,unhandled.........................................................................................................................37
Herring,20%fat,boxed,unhandled.........................................................................................................................38
Herring,25%fat,boxed,unhandled.........................................................................................................................39
Herring,30%fat,boxed,unhandled.........................................................................................................................40
Mackerel,8%fat,boxed,unhandled.........................................................................................................................41
Mackerel,13%fat,boxed,unhandled.......................................................................................................................42
Salmon,normalhandling..........................................................................................................................................43
Bluewhiting,boxed..................................................................................................................................................44
Bluewhiting,normalhandling..................................................................................................................................45
Chisawasawa,normalhandling................................................................................................................................46
Ladyfish,normalhandling........................................................................................................................................47
Seabream(dentexspp.),normalhandling...............................................................................................................48
Seabream(pagellusspp.),normalhandling.............................................................................................................49
Spanishmackerel,normalhandling..........................................................................................................................50
Tilapia,normalhandling...........................................................................................................................................51
Guidanceforanglingclubcompetitions............................................................................................52
ContactDetails.................................................................................................................................53

...theartofdistell‐ing...
FreshnessMeterUserManualPage5
Introduction
Freshnessoffishanditsmeasurement
Fishisahighlyperishablecommodity.Whennewlycaughtithaspleasant,sweetflavoursthatarehighly
acceptable.Onstoragetheseflavoursareprogressivelylostandthoughthefishisacceptable,itisnot
ofsuchhighquality.Furtherstorageproducesoffflavours,andultimatelythesebecomesostrongthat
thefishisunfittoeat.
Fishisgenerallystoredandtransportedinice,and,undertheseconditionsspeciesfromarcticor
temperatewaterswillreachtheunfitstatewithinabouttwoweeks.Tropicalfishcanbeheldsomewhat
longer.Spoilageismuchfasterifthefishisnotheldatchilltemperatures.Itisthereforeimportantfor
anyoneconcernedwithqualityoffishtobeabletomeasureitsfreshnessi.e.theamountof
deteriorationthathasoccurredsinceitwascaught.
Traditionally,freshnesshasbeenassessedbysensorymethods;usingthehumansensesofsight,smell
andtouch.Thisapproachgivesanimmediateappraisalofqualitywithoutdamagetothefish.
However,therearesomedifficulties.Trainedandexperiencedstaffisrequired,andthistrainingtakesa
longtime.Itisnoteasytomakesensoryassessmentquantitative.Thechangesoccurringduring
spoilagecanbedescribedandcodified,butindividualsmayinterpretthesedescriptionssomewhat
differently.Thoughasinglejudgemaybeconsistentwithinhimself,differentjudgeswillvaryamong
themselves.
Thismakesitdifficulttoestablishstandardsthatcanbeappliedindifferentlocations,andbydifferent
assessors.
Thereisthereforeaneedforobjectivemethodsformeasuringfreshnessthatdonotdependonthe
subjectiveopinionofhumanjudges.Severalchemicalproceduresareinuseforthispurpose,butall
sufferfromcommondisadvantagesinrespecttoqualitycontrolwithintheindustrialenvironment,in
thattheyaresomewhatslowtocarryout,whereasqualitycontrollersandinspectorsusuallyrequirean
immediateanswer.Theanalysisrequireslaboratoryfacilitiesandappropriatelyqualifiedstaff,afeature
whichrendersthemethodsunsuitableforfieldwork,andexpensiveforfactoryuse.Themethodsare
destructiveandthedamagedfishsamplesarenolongerfitforprocessing.
Generally,aqualitycontrollerorinspectorneedstoevaluatethefreshnessofabatchoffishratherthan
oneindividualsample.Anappropriatenumberoffisharetaken,thefishevaluatedseparatelyandthe
averageofthebatchcalculated.Therecordingandcalculationrequired,thoughsimple,are
inconvenienttocarryoutatthemarketorinthefactory.
Thereistherefore,arequirementforarapidandobjectivemethodformeasuringfishfreshnessina
widerangeofconditions.Themethodshouldbeaccessibletountrainedstaffandnon‐destructivein
use.Preferablyitshouldalsoallowbatchgradingwithaminimumofcomputationandrecordkeeping.
TheFishFreshnessMeter,orTorrymeter,hasbeendevelopedtomeettheserequirements.
FreshnessMeterbenefits
TheFishFreshnessMeter,whenusedwithorganolepticcharts,eliminatestheuncertainty,delayand
costpreviouslyincurredbysendingfishsamplestoatestlaboratory.Itcanbeusedoutoftheboxon
known,calibratedfishspecies,andafterasimplecalibrationprocedurefornewspecies.
Thefollowingsectionsofthemanualexplainfullyhowtousethemeterandorganolepticcharts.Before
these,itwillbeusefultohaveanintroductiontotheirunderlyingconcepts.

...theartofdistell‐ing...
FreshnessMeterUserManualPage6
Forknownspecies–oneforwhichanorganolepticchartexists–theprocedureissimple.Takeameter
readingfromasampleofyourrequiredspecies;theresult,measuredontheTorryFreshnessScale,
shouldbebetween16forfreshand1forspoiled.Thenrefertotherelevantorganolepticchartforan
interpretationofthemeterreadinganditsimplicationforthefreshnessofyoursample.
Forexample,ifyouwishtosamplecodwhichisboxedbetweencatchandport,refertothefirstchartin
‘Organolepticcharts’.Thischartshowsthatameterreadingof16representsafishoflessthan2days
old.Areadingof8wouldindicatethatthefishwas10daysoldandareadingofbelow3wouldindicate
thatthefishwasmorethan18daysold,andspoiled.
Ifthespeciesisnotknownandhasnoorganolepticchart,youcanperformacustomcalibrationexercise
inwhichyougenerateachartforfuturereference.Todothis,takeasetofreadingsforthefishasit
decaysfromfreshtospoiled.Thismaygiveareasonablesetofreadingsfromsomewherenear16
(fresh)to3orbelowwhenspoiled.Ifthisisinfactthecase,thereadingscanbeusedtowriteintoan
organolepticchartwhichbecomesthereferenceforthenewspecies.
Alternativelyhoweveryoumayfindthattherangeofreadingsasthefishdecaysfromfreshtoputridis
toonarrowforusefulinterpretation.Forexampleyoumayseeaninitialreadingof12whenfreshand8
whenspoiled.Theansweristousethefullinternalrangeofthemeter,whichhasmuchmoreresolution
availablethanthescalesavailablewiththeTORRY‐1(0.1to18.5)orTORRY‐STD(1to18)scales.The
internalrangeisaccessedusingtheRESEARCH‐1calibration,whichoffersarangeof0.1toabove99.9.
Bytakingasecondsetofreadingsusingthishigherresolutionyouwillbeabletoobtainasetof
meaningfullydifferentiatedreadingsoverthenarrowrangeofthe12–8exampleabove.Having
obtainedareferencedatasetofsuchreadingsovertheperiodfromfreshtoputridyoucan:
a. Simplycontinueusingthisdatasetasareferenceforfuturemeasurementsofthesame
species,or
b. Loadthedataintoaspreadsheetforfurtheranalysisorreference,or
c. GenerateanAnalysisDatasetofequaldailydecrementsfrom18to1,andusethiswithinthe
DataManagementSystemtoscaletheReferenceDatasetreadings.Thescaledoutput
datasetbecomesthenewcustomcalibrationforthenewspecies;itcanbestoredinthe
meterandenteredintoanewOrganolepticChart.
YoucanrefertotheFreshnessMeterTechnicalManual,ontheCD‐ROMandontheDistellwebsite,for
moreinformationoncustomcalibrations.
Generaldescription
ThislatestmodeloftheFishFreshnessMeteriscompact,robust,fullyportableandsuitableforany
normalapplicationwithinfishprocessingindustrymarkets,processingfactoriesorqualitycontrol
laboratories.Itprovidesrapidmeasurementsandissimpletouse.
ThefreshnessofthefishisindicatedontheLCDdisplay.
Themetercanbesettomeasurefrom1‐16fish.Themenudrivensoftwarewillsumthereadingsofall
thesamplesandthendisplaytheiraveragevalue.Themeterismostaccurateinthismode.
Generallythemeasurementscanbetakenwithoutdisturbingthefishintheircontainers,andthetesting
doesnotdamageormarkthesamplesinanyway.
Principleofoperation
TheoriginalresearchleadingtothedevelopmentoftheFishFreshnessMeterwascarriedoutatTorry
ResearchStation,inAberdeen,Scotland.

...theartofdistell‐ing...
FreshnessMeterUserManualPage7
Itwasfoundthatcertaindielectricpropertiesofthefishskinandmusclealterinasystematicwayduring
spoilage,astissuecomponentsdegrade.Thesealterations,occurringatmicroscopiclevel,arestrongly
associatedwiththegrosschangesinappearance,odour,texture,andflavourwhichtakeplaceduring
spoilageandwhicharenormallyusedtojudgefreshness.Hence,determinationoftheappropriate
dielectricpropertiesgivesameasurementofthefreshnessofthefish.
Thebaseoftheinstrument(sensinghead)hastwopairsofconcentricallyarrangedelectrodes.This
sensingheadisapplieddirectlyontotheskinofthefish.Analternatingcurrentispassedthroughthe
fish,betweentheouterpairsofelectrodesandtheresultingvoltagesensedbytheinnerpair.The
phaseanglebetweenthecurrentandvoltageismeasuredandconvertedelectronicallytoallowdigital
displayonaconvenientscaleintherange0‐18.Thephaseangleandhencethemeterreading
decreaseswithspoilage.Thecurrentpassedthroughthefishisapproximately1milliampandsocannot
harmtheoperatororaffectthefish.
Betweenthemeasuringelectrodesaretwoauxiliaryelectrodes.Theseelectrodes,inconjunctionwith
oneofthecarbonelectrodes,sensewhetherthereispropercontactwiththefish.Thefishsensing
devicepreventsreadingsbeingtakenunderinappropriateconditionse.g.inair,orinice.Sincethe
phaseangledependsontemperatureaswellasfreshness,thereadingdisplayedbytheinstrumentmust
becorrectedtothevalueitwouldshowatareferencetemperatureof0°C.
Themeterispoweredbyrechargeablebatterieswithintheinstrument.Theyhavesufficientcapacityto
allowafullworkingday’soperation.
Technicaldata
ThisUserManualdoesnotcoverthetechnicalsideoftheFreshnessMeteringreatdepth.Youcanfind
acopyofourTechnicalManualontheCD‐ROMthatissuppliedwiththemeter,oryoucandownloada
copyfromourwebsite.ThiswillhelpyouexplorethetechnicalaspectsoftheFreshnessMeter’s
operation,aswellashowtocreatecustomcalibrations.

...theartofdistell‐ing...
FreshnessMeterUserManualPage8
Step‐by‐stepguide
TheFreshnessMeterkit
YourFreshnessMeterkitcomprisesthefollowingitems:
•Meterunit
•Powersupply/charger
•USBdatacable
•CD‐ROM
•Usermanual
•Carrycase
Pleasetakegoodcareofthemeter.Itisameasurement
instrumentandshouldbehandledcarefully.Themetershouldbecleanedandstoredinitscaseafter
use.
MeterUnit
Data/powerSocket
The7‐waysocketonthesideofthemeter
servesthreefunctions:
•Datadownloadsocket,for
transferringdatatoPC,usingcable
supplied.
•Allowsthemetertorunonexternal
power
•Usedforcharginginternalbatteries
Sensor
Thesensorhastwooutercarbonelectrodesfor
measuringfreshness,andtwoinnermetal
electrodesforsensingcontactwiththeskinof
thefish.
Itisagoodideatowipethesensorclean
betweenreadings,sinceitiseasyforfishscalestosticktothesensorandaffectreadings.
ReadButton(Yes)
•Pressandholdthisbuttontotakeareading;
•Presstoanswer‘Yes’toanyYes/Noquestiononthedisplay.
ResetButton(No)
•Presstoresetthemeasurementroutine,ortoabortanincompleteseriesofreadings;
•Presstoanswer‘No’toanyYes/Noquestiononthedisplay.
Datacable
Alwaysusethedatacablesuppliedfordownloadingtocomputer.Connectthe7‐wayplugtooutleton
meterandtheotherenddirectlyintoaUSBportonyourcomputer.
Sensor
Data/powersocket
Powerswitch
LCDDisplay
Read/YesbuttonReset/Nobutton

...theartofdistell‐ing...
FreshnessMeterUserManualPage9
CD‐ROM
TheCDcontainstheusermanuals,DataManagementSystem,calibrationcharts,andotheruseful
information.
Usermanual
Ahardcopyoftheusermanualandthecalibrationcertificateisincluded.
Carrycase
TheFreshnessMeterkitissuppliedinarobustcarrycase,andcomescompletewithtwokeysfor
securityofyourFreshnessMeter.
Powersupply/charger
Onlytheunitsuppliedwithyourkitshouldbeusedtocharge
thebatterypack,orpowerthemeter.Thisisconnectedtothe
7‐wayoutletonthemeter.Thechargersupplyunitsimplyplugs
intothemainssocket,acceptinginputmainsvoltagefrom
110v–240vAC,50–60Hz.Thechargerunitcomescomplete
withvariousplugconfigurationsforusearoundtheworld.
AredLEDwillilluminatewhencharger/powersupplyis
switchedon.Thebatterypackshouldbefullychargedaftera
periodof12hours.Whenchargingiscomplete,firstdisconnect
thechargerfromthemainsandthenfromthemeter.
Thepowersupply/chargercanalsobeusedtopowerthe
meterduringusewhilealsochargingtheinternalbatteriesatareducedrate.
ChargingtheMeter
NotesonchargingtheMeter:
•Insertpowerunitconnectorplugcarefullyinto
metersocket(toinsert,turnpluggentlyuntilthe
plugslidesintotheconnector.
•Metershouldbechargedforafull12hoursbefore
useifpossible.
•Alwaysensurethatmeterisswitched“OFF”before
charging.
•Whenchargingiscomplete,firstdisconnectthe
powerunitfromthemainsandthenfromthe
meter.
•Alwaysfitdust/moisturecapinplacebefore
commencingwork,especiallyinwetenvironments.
Note:UseonlythisDistellpowerunitforbatterychargingormeterpower.Donotuseexcessiveforce
whenpluggingthepowerunitintothemetersocket.

...theartofdistell‐ing...
FreshnessMeterUserManualPage10
Gettingstarted
Verifythatbatteryhasbeencharged.
Switchon
SwitchontheMeterandallowthemetertocyclethroughitsstartupprocedure.
Partofthestartupprocedureistoshowhowmanyreadingsarecurrentlystoredinthemeter.Youmay
wishtodownloadanderaseanystoredreadingsbeforetakingnewmeasurements.Youcanfindout
aboutthedownloadprocedureinthe‘DataManagementSystem’section.
Afteraround15secondsthemeterisreadyforuse.Thedisplaywillshowthelastcalibrationsetting
used,andwilllooksomethingliketheexamplebelow:
CalibrationSetting
TORRY-1
Y SN=3 N
Yes Numberofsamples used
tocalculateaverage
No
Measuringproductsamples
Haveacopyoftheorganolepticchartsathand.Forbestaccuracyalwaysfollowthemeasurement
recommendationsinthismanual.
Beforestartingwerecommendyoucarryoutsomechecks:
1. Checkthatyouhavesetthecorrectnumberofsamples,from1to16,fromtheSamplesmenu.
2. Checkthatyouhaveselectedthecorrectcalibrationsettingonthemeterforthefishbeing
sampled.
3. Checktoensurethatyouhavepreparedorhavebeensuppliedwithanorganolepticchartsuitable
forthisfishspecies.
4. Selectthefishsamplesatrandomfromthebatch.Whereyouaretakingfishsamplesfromavaried
batch,thesamplesshouldbegroupedaccordingtosize(length/weight).
5. Checkthattemperatureofthefishtobemeasuredisbetween0and10°C,withnoicecrystals
presentinthesamples.
Themeasurementprocedureisoutlinedasfollows:
1. Ensurethatthesensorisplacedfirmlyagainsttheskinofthefish,thusensuringthatthereareno
airpocketsbetweenthesensorandthesampletobemeasured.PresstheRead/Yesbutton;the
Meterwillconfirmcontactwiththefish,takethemeasurement,displayandstorethevalue.
2. Ifyousee‘NoContact’displayedonthemeterthencheckthatthesensorismakinggoodcontact
withtheskinofthefish.Themeterwillnottakeanyreadingsuntilitdetectsgoodskincontact.
3. Forbestresultsbeconsistent.Alwaysplacethesensoronthesamepartofeachfish.Thiswill
ensurethatyouachieveoptimum,comparableresults.Itisnormaltoexperiencevariabilityinthe
individualreadingswhentakenatdifferentpartsofthefish.Notealsothatdifferentpartsofthe
fishwillspoilatdifferentrates,dependinguponthelevelofbacteriologicalactivitypresentwithin
thefish.
4. Onesetofreadingsshouldbetakenfromeachfish,attherecommendedmeasurementsites.The
resultswillbestoredintheFreshnessMeterforlaterdownloading.

...theartofdistell‐ing...
FreshnessMeterUserManualPage11
5. Placethebaseofthemeterfirmlyonthefishsothatitliesflatagainstthesurfaceandparallelto
thelaterallineatathick,fleshypartofthefish.Thebestpositionisattheshoulderareawithinthe
firstthirdofthefish,intheregionofthelateralline,asshownintheillustrationbelow.Avoidthe
headandbellycavityregions.
6. Ifthisisthefirstreadingafterthemeterhasbeeninawarmplace,holdtheprobeheadagainstthe
fishfor10‐20seconds,toallowthesensortocooldownbeforepressingtheRead/Yesbutton.
7. Tomeasureindividualfish,settheFreshnessMetertoread3samples(seethe‘MenuStructure’
sectionfordetails),placethemeterattherecommendedmeasurementsite,andtake
measurementsthreetimes.Theindividualvaluesandtheaveragevaluewillbestoredinthemeter
fordownloading.WhentheaveragevalueisshownonthedisplaypressRead/Yestotakemore
readingsonadifferentfishorpressReset/Notofinish.
8. AvoidpressingtheREADbuttonwhenthemeterisnotincontactwithafish.Ifthereisathicklayer
ofslimeontheprobeheaditmayconductenoughcurrenttoactasiftheelectrodeswerein
contactwithafishandameasurementwillbemade.
9. Switchthemeteroffafterusetoretainthechargeinthebattery.
Normallythreereadingsarerecommended.Wehavefoundthatthisgivesthebestbalancebetween
accuracyandspeed.Youcanchangethenumberofreadingsthatthemeterwilltakebeforeaveraging
byusingthe‘Samples’menuoption.Seethe‘MenuStructure’sectionformoredetails.
TheFreshnessMetercanstoreupto1000setsofreadings.ThesecanbedownloadedtoaPCrunning
theDataManagementSystem.Seethe‘DataManagementSystem’sectionformoreinformationon
howtodothis.Thereisalsoanoptiontodownloadthedatainrealtime,i.e.asthereadingsaretaken,
andthisiscoveredinthesamesection.
Comparingthemeterwithlaboratoryanalysis
IfyouwishtocomparetheresultsachievedusingtheFreshnessMeterwithalaboratoryprocedure,itis
importantthattheentirefishcarcasssampleispackagedandsenttothelaboratoryfortheanalysis.
Thesampleshouldbestoredinasealedpolythenebagimmediatelyaftermeasurementandthe
temperatureofthesampleshouldbemaintainedascloseto0°Caspossible.Thiswillensurethat
spoilageofthesampleisreducedasfaraspossible.
Important:Donotallowthesampletofreeze.
Thewholecarcassofthefish/filletshouldbereceivedatthelaboratoryproperlypackedandsealed,
andthecurrent‘agefromdeath’ofthesampleshouldbeascertainedusingstandardlaboratory
methods,e.g.TVBN,etc.
IftheresultsfromtheLaboratorydonotcomparefavourablywithyourmetercheckthemeter‘suseas
follows:
•IstheoperatorproficientintheuseoftheFishFreshnessMeter?
•IsthefishspeciesthesameasthatquotedonyourOrganolepticChart?

...theartofdistell‐ing...
FreshnessMeterUserManualPage12
•Checktheproductforspeciesidentification,measurementtechnique,sizeandpreparation.
Checkthelaboratoryprocedureasfollows:
•Verifytheanalyticalmethodbeingused
•CheckthatthesamplepreparationwasOK
ThisisjustashortchecklistonhowtouseandcomparetheFreshnessMeterwithothermethods.
RememberthattheprimarypurposeinusingtheFreshnessMeteristoscreenalotofproductfor
uniformity,specificationandshelflife,andtoidentifyfishthathasbeenfrozenorirradiated.
Duringthemeasurementprocess,finiteaccuracyisnottheobjective.Theobjectistotryandensure
thatthebulkoftheday'srawmaterialsandprocessedproductionmeetswithyourspecificationstime
aftertime.
If,afterthesechecksthereisstillasignificantdifference,pleasecontactDistellforadviceandhelp.
NotethattheDistellwebsitehasacomprehensiveFrequentlyAskedQuestionssectionaswellas
technicalinformationontheFreshnessMeter.
MenuStructure
Whenthestandardmeasurementdisplayisshown(seetheexampleatthestartofthischapter)youcan
accessthemenubypressingtheReset/Nobutton.TheChoicesmenuscreenwillappearintheorder
shownbelow.Therearesixoptions.CyclethroughtheoptionsbypressingReset/Nountilthecorrect
oneisdisplayedthenpressRead/Yes.
Productmenu
IntheProductmenuyoucanchoosewhichoftheproductcalibrationsettingsyouwishtouse.Simply
choosethecalibrationthatcorrespondswiththetypeofmeasurementyouwishtouse.
CHOICES MENU
Y PRODUCT N
TORRY‐1TORRYSTD OTHER
ChoicesMenu
PRODUCTRESEARCHSAMPLESSETTIMEDOWNLOADADDPRODUCT
TORRY STD
Y SN=3 N

...theartofdistell‐ing...
FreshnessMeterUserManualPage13
Researchmenu
IntheResearchmenuyoucanchoosewhichoftheresearchcalibrationsettingsyouwishtouse.The
FreshnessMeterisnormallysuppliedwiththefollowingsettings:RESEARCH‐1,FISH‐1,FISH‐2...FISH‐5.
TheRESEARCH‐1calibrationsettingisusedtocreateanewcalibrationsetting.
SettingsFISH‐1toFISH‐5canbeprogrammedwithyourcustomcalibrationdatawhererequired.Full
detailsareavailableinourTechnicalGuide,whichisavailableontheCD‐ROMandfromourwebsite.
Samplesmenu
Forgreatestaccuracyyoushoulduseeightsamplesperfishandmeasureeightfishfromabatch.For
singlefishwerecommendthattheFreshnessMeterbesetupforthreesamplesperfish.Inbothcases,
allthemeasurementsshouldbetakenatthesamesiteonthefish.Youcanchoosebetween1and16
sampleshere.
Note:ifyouchoose16samplesthentheindividualsamplevalueswillnotbereportedastheyaretaken,
i.e.thesamplingisdone‘blind’.Theseindividualreadingsarestoredinthemeterandcanbeseen
whendownloadedtotheDataManagementSystem.
Downloadmenu
TheDownloadmenuallowsyoutosendanystoredsampledatatoaconnectedPCrunningtheData
ManagementSystem.TheDMSisdocumentedlaterinthisguide,andthedownloadprocedureis
coveredthere.
CHOICES MENU
Y DOWNLOAD N
(Ifnodatahas
beenrecorded
)
(Ifdataisready
fordownload
)
NoDataStandardD/L
CHOICES MENU
Y SAMPLES N
Samples=1Samples=2 Samples=3 ...Samples=16
CHOICES MENU
Y RESEARCH N
FISH‐5RESEARCH‐1FISH‐1FISH‐2FISH‐3FISH‐4

...theartofdistell‐ing...
FreshnessMeterUserManualPage14
AddProductoption
Usingthismenuprovidestheabilitytoprogrammethemeterwithadditionalcalibrations.Thisfacilityis
especiallyusefulforthosewhowishtorefineexistingmetercalibrationsorcreateanewcalibration
settingbasedontheirowndata.
FullinstructionsonhowtousethisfacilityareintheTechnicalManual,availableontheCD‐ROMand
fromtheDistellwebsite.
Date&Timemenu
Youshouldcheckthedateandtimewhenyoufirstreceivethemeter,andupdateitifnecessary.Each
samplethatisrecordedbythemeterhasadateandtimestamp,soitisrecommendedthatthedate
andtimebekeptaccurate.
Thiswillthenensurethatalldownloadsreflectthecorrectdateandtime.
Notethatifameterisconfiguredasarentalmeterthenthisoptionwillnotbeavailable.Youcanread
moreaboutrentalkeysathttp://www.distell.com/downloads/rentalkey.pdf.
CHOICES MENU
Y SET TIME N
Minutes
Hours
Day
Month
Year
CHOICES MENU
Y ADD PRODUCT N
Waitingdata

...theartofdistell‐ing...
FreshnessMeterUserManualPage15
Usagerecommendations
Usingthemeterinqualitycontrol
Sampling
TheFishFreshnessMeterwasdesignedwiththeneedsofthequalitycontrollerinmind.Theterm
‘qualitycontroller’ismeanttoincludeofficialsresponsibleforpublichealthinspectionorenforcement
ofmandatoryregulationsconcerningquality,aswellaspersonsconcernedwithqualitycontrolin
industryandcommerce.Inatypicalsituation,acontrollerisrequiredtotestbatchesoffishthatare
beinginspected,soldorprocessedasaunit,andforthisasuitablesamplemustbetaken.Themeter
simplifiesthisprocessofassessingabatchbysampling.
Ifthefreshnessofindividualfishwithinabatchismeasuredwiththemeter,itwillbefoundtovary.
Thisfish‐to‐fishvariationinmeterreadingsiscausedbytwofactors:‐
a. Variationinintrinsicfreshnessamongindividualfish.
b. Variationofmeterreadingsamongfishofthesamefreshness.
Bothareexpressionsofthevariabilityofbiologicalorganisms.Abatchoffishcaughtatthesametime
andhandledandstoredidenticallywillspoilatslightlydifferentratesbecauseofvariationsinchemical
constitutionandbacterialactivity.Hencetheindividualfishinthebatchwillhavearangeoffreshness;
arangethatwilltendtoincreaseasspoilageincreases.
Themeterreadingsthemselvesarenotdirectmeasuresoffreshness,asdefinedinsensoryterms,but
arestronglyassociatedwithit.Therelationshipbetweenmeterreadingsandfreshnessasmeasuredby
asensorypanelwillonceagainbedifferentforeachindividualfish.Forthisreasonitisnot
recommendedthatthemeterbeusedtoassessthefreshnessofsinglefish,exceptwithinratherwide
limits.
Thesedeviationsfromaveragebehaviourtendtocanceloutwhenthemeansofsamplesfrombatches
aretaken,andthebiggerthesamplethebetterthecorrelationbetweenthemeanmeterreadingand
freshnessscore.Itfollowsfromsamplingtheorythatthenumberoffishwhichoughttobesampled
fromabatchdependsmainlyonthedesiredprecisionofthemeanscore.Italsofollowsthat,provided
thebatchislargeenough,therequiredsamplesizedoesnotdependonthesizeofthebatch.From
observationsmade,ithasbeendecidedtouseasamplesizeofeightfishasthebasisfortheaveraging
intheFishFreshnessMeter.
Inmostmarketinganddistributionsystemsthefishareheldincontainerswithacapacityof50Kg.
Severalcontainersconstituteabatch.Inthissituationfishfromasmanycontainersaspossibleshould
besampledi.e.onefromeachcontainerifthebatchhaseightormore,fourfromeachcontainerifthat
isallthereisinthebatch.Ifthecontrollerorinspectorwantsgreaterconfidenceinthebatchaverage,
morethanonesetofeightfishcanbesampled,andagrandaverageoftheresultscalculated.
Notethattheprecisionofanaveragedresultincreasesonlyasthesquarerootofthesamplesize.An
averageof64readingsisonlytwiceaspreciseasanaverageof8readings.
FreshnessMeterperformance
Therelationshipbetweenmeterreadingsandothermeasuresoffreshness
Therearemanywaysofexpressingthefreshnessoffish.Perhapsthesimplestforfishheldiniceisto
quotethestoragetime.However,asiswellknown,theinitialqualityofmanyspeciesoffishvarieswith
theseasonsoftheyear,duetosuchfactorsasthespawningcycleandtheavailabilityoffood.These
factorsalsoaffecttherateofspoilage,sothatthetimeiniceisnotnecessarilyagoodmeasureof

...theartofdistell‐ing...
FreshnessMeterUserManualPage16
freshnessunlessrelatedtothetimeoftheyear.Ithasbeenfoundthatabetterwayofexpressing
freshnessistoconstructasystemofnumericalscalesbaseduponmethodsofobjectivesensory
assessmentbyatrainedtastepanel.Suchasystemisbetterbecauseitisdirectlyrelatedtoeating
qualityratherthandependingonknowledgeofthestorageconditions.
Oftenthefishisallocatedtooneofafewqualitygradesbasedondefinitionsinsensoryterms.An
exampleofthisistheEuropeanEconomicCommunity’ssystemofgradingfishformarketingpurposes.
Anextremeformofthistypeisthepublichealthinspector’sgradingintofitandunfit.Aquality
controllermaynotevenconsciouslyputascoreoragradetoabatch,butmustacceptorrejectitfor
thepurposeinmind.
Inpracticethequalitycontrollerinindustryadoptsthemostconvenientsystemforhispurpose,butan
officialinspectorwillhavetousewhateverislaiddownintheregulations.Itisquitepossibletodefine
grades,boundariesoracceptance/rejectionrangesoffreshnessforvariousoutlets,etc,intermsof
meterreadingsandindeedthisisagoodwayofusingtheinstrument.However,untiltheuserisfamiliar
withtheinstrument,itisbesttoemployitinparallelwiththeaccustomedmethodofassessing
freshness.Aftersomeexperience,therelationshipbetweenthequalitylevelstheuserisfamiliarwith
andthemeterreadingscanbeestablished.Nevertheless,themetercanbeusedbyanunskilled
operatorafterverylittleinstruction.
AsaguidetothevaluesthatcanbeexpectedfromsomespeciesofinteresttoEuropeanFish
Processors,averagereadingsfordifferentlevelsoffreshnessareshowninthesetoforganolepticcharts
inthe‘Organolepticcharts’section.TheseresultswereobtainedbyTorryResearchStationattheir
laboratoriesinAberdeenandHull,andatvariousfishingportsintheUK.Thesensoryscoresarebased
onthescalesusedatTorryResearchStation,andhavebeendescribedintheirpublications.
Thesescaleswereinterpretedasfollows:
•Perfectlyfreshfishisgivenascoreof10.
•Goodqualityfishhasascoreof6ormore.
•Belowascoreof4,thefishareconsideredunfittoeat.
Thesefreshnessscoresareonlyaroughguide.Usersarestronglyadvisedtoobtainequivalentlevels
fromscoresgivenbytheirowntrainedpanel,asthescoresmaybebasedupondifferentcriteria,or
obtainedunderdifferentconditions,fromthoseobtainedbyTorryResearchStationandDistell.
Forcomparison,theapproximatenumberofdaysofstorageinicewhichwouldcausethefishtoreacha
givenfreshnessscoreisalsoincludedinthecharts.Anexplanationofthedifferentcolumnsofmeter
readingsisgiveninthenextsection.
Theexamplesgivenaretypicaldemersalfish.Thesituationwithfishofhighandvariablefatcontentis
rathermorecomplex,andtherelationshipsbetweenthemeterreadingsandsensoryscoreshavenot
yetbeenmadeclear.However,extensiveworkcarriedoutbythestaffofTorryResearchStationwith
herringhasestablishedthatthemannerinwhichmeterreadingsvarywithtimeofstorageiniceisvery
dependenton:
a. Thefatcontentofthefish.
b. Whetherornotitisicedimmediatelyaftercatching.
c. Theextenttowhichitissubjectedtohandling.
Theorganolepticchartsshowtheapproximateageinicecorrespondingtogivenmeterreadingsfor
boxedherringofvariousfatcontents.Theage‐in‐iceisvalidonlyforfishwhichhasbeenboxedandiced
immediatelyaftercatching.Delayedicingalwaysgivesrisetolowermeterreadingsduringsubsequent
storageinicecomparedwiththoseshowninthechart.

...theartofdistell‐ing...
FreshnessMeterUserManualPage17
Theextentoftheeffectvarieswithfatcontentandtemperatureduringstoragebeforeicing.For
example,leanfishheldfor8hoursatatemperatureofabout5°Cbeforeicingwillgenerallyshowmeter
readingsaboutoneunitlowerthantheappropriateorganolepticchartwouldsuggest.Withfattyfish
theeffectcanbeuptotwiceasgreatforthesametimeandtemperaturebeforeicing.Thereis
evidencetoshowthatthehigherinitialtemperaturescaneasilydoublethiseffect.
Mackerelgenerallybehaveinmuchthesamewayasdoherring,althoughdetailedinformationisnotat
presentavailable.Theorganolepticchartsformackerel(pages41and42)givesomeideaofreadingsto
beexpectedfromcarefullyhandledfish,boxedandiced.
Althoughlessisknownoftherelationshipbetweenobjectivesensoryscoresandmeterreadingsthan
forwhitefish,itisbelievedthatmeterreadingsgiveafairindicationoftheintrinsicqualityof
unhandled,boxedfattyfish,irrespectiveoffatcontent.
Thechartsonpages46to51showrelationshipsbetweentheFishFreshnessMeterreadingsandtimeof
storageiniceforafewspeciesofAfricanfish.ThedatahasbeensuppliedbytheTropicalProducts
Institute,London.
Effectsofhandlingandprocessing
TheFishFreshnessMetermeasurescertainpropertiesoffishmuscleandskinthatchangeina
systematicwayduringstorageinthewetstate.Anyotherprocessthataffectsthestructureofthe
muscleatthecellularlevelwillalsoaffectthemeasurements,almostinvariablytolowerthem.Handling
isanimportantfactor.
AcommonmethodofhandlingfishinEuropeistostoreit,mixedwithice,inbulkaboardthevessel,
thenunloaditontothequaysideandsortthefishintocontainers.Thepressureitissubjectedtointhe
hold,thehandlingduringdischarge,andthesortingprocesstendtolowerthemeterreadingswhen
comparedwithfishthathavebeenstowedcarefullyinboxeswithiceandkeptundisturbeduntil
measured.Thedifferenceisindicatedinthechartsinthe‘Organolepticcharts’section.
Themetercanalsobeusedtogradefillets.Forskin‐onfilletsmakethemeasurementsontheskinside
intheusualway.
Themetermeasurespropertiesofbothskinandmuscle,butitcanstillbeusedwithskin‐offfillets.In
thiscasethemetershouldbeappliedtothebonesideofthefillet.Thereadingsthough,aremuch
lowerthanforwholefishorskin‐onfilletsofequivalentfreshness.Alsodiscriminationinfishmaterial
withafreshnessscorebelow6isnotpossible.Typicalvaluesforcodareshowninthechartsbeginning
onpage23.
Thequalityofherringandmackerelismostsensitivetohandling,andineachcasethisisreflectedby
thereadingsobtainedwiththemeter.Evencarefulhandlingofthefishcanlowertheinstrument
readingsby1or2units.Normalcommercialhandlingreducesthereadingsstillfurther.
Effectsoffreezingandbrining
Freezinghasadrasticeffectonthecellularconstituentsofmuscle.Itisnotpossibletodeterminethe
originalfreshnessofthawedfishsincemeterreadingsareinvariablyintherangeof0–3,whateverthe
qualitybeforefreezing.Infact,thispropertycanbeusedinmanycasestodeterminewhetherfishhas
beenfrozenatsometimeinitshistory.Ifthesample,judgingbyodour,appearanceandflavour
appearstobefreshi.e.shouldhaveahighmeterscore,butinfactgivesalowreadingthenthereare
groundsforsuspectingithasbeenfrozenatsometime.Onceagainjudgmentshouldnotbebasedon
onefishonly;severalfishshouldbeexamined.
Briningalsolowersthereadingthoughnotquiteasmuchasfreezing.Itisthereforenotpossibleto
checkthefreshnessofbrinedfish.

...theartofdistell‐ing...
FreshnessMeterUserManualPage18
Metercareandservice
Cleaningandgeneralcare
TheFreshnessMeterisasealedunitandiswaterproofedpriortoleavingthefactory.However,wedo
notrecommendthattheunitbeimmersedinwateroranyotherfluid.Anyonetamperingwiththeunit,
openingtheseals,etc.,willimmediatelyinvalidatethewarranty.
ThemeterisencasedinimpactresistantABSplastic,withthesensingprobesmadefromstainlesssteel
andcarbonmaterials.Itmaybecleanedwithadampcloth,usingamildsolutionofsoapordetergentin
water.Beawarethatsomehouseholdcleanersmaycontainchemicalswhichcouldseriouslydamage
theplastichousing.Avoidtheuseofanypetroleumbasedsolventcleaners.Donotletslimedryouton
theelectrodesurfaces.Donotattempttocleantheelectrodesurfaceswithabrasives.
Theunitshouldbechargedregularlyandthemeterretainedwithinitscasewhennotinuse,for
safekeeping.
TheFreshnessMeterisaprecisioninstrument.Ensurethattheunitisstoredsecurelyintransit.Donot
droporotherwisemisusetheequipment,asthismayinvalidatethewarranty.
Beforecallingforservice
Ifyouhavereasontobelievethatafaulthasoccurred,pleasefirstofallcheckthefollowing:
•Istheunitswitchedon?
•Hasthebatterypackbeencharged?
IfnoobviousfaultisapparentpleasetelephoneourHelpDeskforadviceandtoreportthefault.
Warrantyandservice
YourTorrymeterisfullywarrantedagainstmanufacturingdefectforaperiodof12months.The
warrantyandterms&conditionsareexplainedinourTechnicalManual,availableontheCD‐ROMor
fromtheDistellwebsite.
Ifyouneedtoreturntheunitforrepair,pleasecontactyoursupplierforadviceonhowtodoso.The
metershouldbecarefullypacked,togetherwithanexplanatorynoteastothenatureofthefault.
ShouldyouhavedifficultyinreturningyourFreshnessMetertoyoursupplierforservicing,youcan
alwayscontactDistelldirectlyusingtheinformationinthe‘ContactDetails’section.Pleasequotethe
dateofpurchaseandgiveafulldescriptionofthefault.

...theartofdistell‐ing...
FreshnessMeterUserManualPage19
DataManagementSystem
Installation
Requirements
TheDataManagementSystem,referredtoasthe‘DMS’,requiresaPCrunningWindowsXPorlaterand
afreeUSBsocketfordatadownloads.
InstallingtheDMS
Thissectiontellsyouhowtoinstalltheprogramonyoursystemandhowtoruntheprogram.
Itisbesttocloseallprogramsbeforeinstallingthesoftware.Youshouldhaveatleast5Mbofharddisk
spaceavailablefortheinstallation.
•InserttheCDinthecomputer;
•Findthe‘Installationfiles’folderontheCDthenopenthe‘DataManagementSoftware
(DMS)’folder;
•RuntheDMSInstaller.exefileandfollowthepromptstocompletetheinstallation
RunningtheDMS
ClicktheWindowsStartbuttononyourdesktop,selectprograms,selectDistellthenlocatetheDMS
iconfromthefileslistedthere.
Youcanusetheprogramstraightawayafterinstallation.Theprogramenablesyoutodownload,print,
save,andexportdata,quicklyandefficiently.TheDMSalsoallowsyoutocreateanduploadyourown
customcalibrationstotheFreshnessMeter(seetheTechnicalManualontheCD‐ROMandontheDistell
website).Allprogramfunctionsareaccessedusingthemenuandtoolbaroptions.Thereisa
comprehensiveon‐linehelpfacility,whichisavailablebyclickingontheHelpmenuoption.
HistoricalDataDownload
ThisfunctionenablesyoutodownloadreadingsthatarestoredintheFreshnessMeter.
Thedataincludesthedateandtimeofthesample,thecalibrationsettingthatwasused,thenumber
andvaluesoftheindividualsamplestaken,andtheaveragevalueofthesamples.
Toillustratetheprocedure,we'llassumethatyouhavetakensomesamplereadingswiththeFreshness
MeterandarereadytodownloadthemtoyourPC.
Step1:ConnectingtheMeter
ConnectthemetertoaUSBportonyourPC,usingthedatacablesuppliedwiththemeter,thenturnthe
meteron.WhenthecableisconnectedtheUSBdriversoftwareinstalledonthePCwillcreatetwo
‘virtualserialports’.NormallytheseareCOM3andCOM4,butthismaychangeifyouhaveotherUSB
virtualserialdevicesinstalled.YourcomputerwillnormallyreserveCOM1andCOM2forbuilt‐in
hardware,orforlegacypurposes
Step2:ConfiguretheDMS
StarttheDMSthenusethePreferencesoptionundertheEditmenutoselecttheCOMporttouse.If
theUSBdrivercreatedCOM3andCOM4thenyouwouldchooseCOM4here.Ingeneralyoushould
alwaysselectthevirtualCOMportwiththehighernumber.Ifyouseeanerrormessagewhenyou
selecttheCOMportthenensurethatthemeterisconnectedproperlyandturnedon.Ifitis,try
selectingadifferentCOMport.

...theartofdistell‐ing...
FreshnessMeterUserManualPage20
Onceyouhaveconnectedthemeterandselectedthecorrectcommunicationport,selectDownload
HistoricalDatafromtheDownloadmenu.TheDMSwillnowmonitortheCOMportforincomingdata
fromthemeter.
Step3:Initiatethedownloadfromthemeter
OncetheDMSisreadyandwaitingfordata,themeterneedstobetoldtobeginthedownload.
Findthe‘Download’optioninthechoicesmenu(see‘MenuStructure’).Atthe‘StandardD/L’option,
presstheRead/YesbuttonthenlookforthedataarrivingattheDMS.
IfnodataisseenattheDMSthenyoumayhaveselectedthewrongCOMport.Gobacktostep2and
tryanalternativeport.
Oncethedatahasbeensafelydownloadedthenyoucan,ifyouwish,erasethedatafromthemeter.
Youcan,alternatively,allowdatatobuildupinthemeteruptothemaximum1000readings.Ifyou
wishtoclearthedatafromthemeteratalaterstagethenyoucangothroughthedownloadprocess
butwithoutconnectingthemetertothePC,thenreplyYestothe‘Erase?’question.
Whenthedatahasbeensuccessfullydownloaded,itwillappearintheDMSinaspreadsheet/grid
form.WiththedataintheDMSyoucannow:
•Printthedata
•Addextracolumnsofinformation
•Savethedatainnativeformasan.FMDfile.
•ExportthedataasaCommaSeparatedValue(.CSV)fileforexportingintootherapplications
suchasExcel,Access,etc.
RealTimeDataDownload
Youcanusethisfunctiontodownloaddatainrealtime,samplebysample,eachtimethatanaverage
readingiscreatedbythemeter.Therealtimedatawillappearonscreenlinebylineasyouusethe
meter.
CHOICES MENU
Y DOWNLOAD N
NoDataStandardD/L
Downloading
Download
Complete
Erase?
YN
IfYesthendataiserased
fromthemeter‘smemor
y
IfNothendataisretained
inthemeter‘smemor
y
Table of contents
Other DISTELL Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Endress+Hauser
Endress+Hauser FTL62 Series Functional safety manual

HP
HP 5061B Operating and service manual

Apera
Apera DO850 instruction manual

Keit
Keit IRmadillo DOC1048R user manual

PCB Piezotronics
PCB Piezotronics IMI SENSORS 646B11 Installation and operating manual

Myron L
Myron L Ultrameter III 9P quick start guide