DOCTOR SMILE WISER 3 User manual

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 1
WISER 3 DIODE LASER
L A 12D 001.X
USER MANUAL

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 2
INDEX
INTRODUCTION 3
1.1. VERSIONS LIST 3
1.2. INTENDED USE 4
1.3. SYMBOLS USED 5
1.4. APPLIED PART 5
SAFETY 6
2.1. ESSENTIAL REQUIREMENTS 7
2.2. GENERAL SAFETY DISPOSITIONS 7
2.3. WORKING AREA 7
2.4. INDIVIDUAL SAFETY MEASURES 8
2.5. OCULAR RISK 9
CERTIFICATIONS 11
3.1. CE DECLARATION OF CONFORMITY 11
3.2. CE CONFORMITY MARKING 12
3.3. NAMEPLATE 12
3.4. WARNINGS 13
3.5. ACCEPTANCE OF RESIDUAL RISK 13
INSTALLATION 14
4.1. CHECK ON DELIVERY 14
4.2. ENVOIREMENT REQUIREMENTS 14
4.3. CONDITION OF USE 14
4.4. ELECTRIC REQUIREMENTS 15
4.5. DELIVERY AND STORAGE 15
OPERATION MODE 16
5.1. OVERVIEW 16
5.2. ACCESSORIES INCLUDED 17
5.3. LASER DEVICE PREPARATION 18
5.4. PEDAL INSTALLATION 19
5.5. HANDPIECE AND TIP INSTALLATION 20
CLEANING AND STERILIZATION 23
6.1. HANDPIECE MAINTENANCE 23
6.2. CLEANING AND STERILIZATION 24
SYSTEM OPERATION 25
7.1. FIRST ACTIVATION 25
7.2. DEVICE SWITCH-ON 27
7.3. TREATMENT SELECTION 27
7.4. LASER ACTIVATION 31
7.5. EMERGENCY STOP 31
7.6. VISUAL SIGNALS 31
7.7. SETTING MENU 32
7.8. INFORMATIONS 32
7.9. FAVOURITSE AND PATIENT LIST 33
7.10. SWITCHING OFF 34
7.11. WIRELESS MODE /BATTERY RECHARGE 34
7.12. EXTERNAL CONNECTION 35
MAINTENANCE 36
8.1. PERIODICAL MAINTENANCE 36
8.2. GENERAL CLEANING 36
8.3. CHECK AND CLEANING OF TIP AND HANDEPIECE 37
SYSTEM ERRORS 38
SPARE PARTS 39
SPECIFICATIONS 41
SAFETY LABELS 43
EMC REGULATIONS (ELECTROMAGNETIC COMPATIBILITY) 45
13.1. ESD (ELECTROSTATIC DISCHARGE) 49
13.2. DEVICE DISPOSAL 49
WARRANTY 50

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 3
INTRODUCTION
Diode laser LA12D001.x (commercially identified as WISER 3) is a medical device that uses a laser
source.
This user manual is issued with reference to all products listed at the following chapter which differ
each other in wavelength and power emission according to the table below. If not specified
otherwise, the generic code LA12D001.x refers to all devices.
DOCTORSMILE is the commercial brand related to all products manufactured by Lambda SpA.
1.1. VERSIONS LIST
L A 12D 001.1
Model
Peak Power
CW Power
Wavelength
Total peal power
Module1
16W
8W
976 nm
24,5 W Peak
Module2
8W
8W
808 nm
Module3
500mW
500mW
635 nm
L A 12D 001.2
Model
Peak Power
CW Power
Wavelength
Total peal power
Module1
16W
8W
915 nm
24,5 W Peak
Module2
8W
8W
808 nm
Module3
500mW
500mW
635 nm
L A 12D 001.3
Model
Peak Power
CW Power
Wavelength
Total peal power
Module1
16W
8W
976 nm
24,5 W Peak
Module2
8W
8W
808 nm
Module3
500mW
500mW
450 nm
L A 12D 001.4
Model
Peak Power
CW Power
Wavelength
Total peal power
Module1
16W
8W
976 nm
20,5 W Peak
Module2
4W
4W
450nm
Module3
500mW
500mW
635 nm
L A 12D 001.5
Model
Peak Power
CW Power
Wavelength
Total peal power
Module1
16W
8W
808 nm
20,5 W Peak
Module2
4W
4W
450nm
Module3
500mW
500mW
635 nm
L A 12D 001.6
Model
Peak Power
CW Power
Wavelength
Total peal power
Module1
16W
8W
976 nm
16,0 W Peak
L A 12D 001.7
Model
Peak Power
CW Power
Wavelength
Total peal power
Module1
16W
8W
808 nm
16,0 W Peak
L A 12D 001.8
Model
Peak Power
CW Power
Wavelength
Total peal power
Module1
4W
4W
450 nm
4,0 W Peak

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 4
1.2. INTENDED USE
This laser equipment named “DIODE LASER” is a medical device and its application fields are:
•DENTAL SURGERY
•DENTISTRY
•DENTAL THERAPY
1.2.1. Examples of use
Some application fields are orthodontics, endodontics, periodontology, herpes therapy, canker sores,
chelitis, conservatives, whitening, care and maintenance of implants-peri-implantitis.
The use is temporary, related only to the therapeutic application.
1.2.2. Intended users and place of use
The ability and technical knowledge of dentistry and of the doctor who carries out the therapeutic
application correctly guide the choice of the type of treatment
The device is designed to be only used by specialized medical personnel who have received specific
training on the use of the laser device.
It must be only used within medical and/or dental studios, hospitals or other adequate sanitary
structures.
1.2.3. Intended patients, clinical applications, practical
suggestions and possible side effects
The medical device “DIODE LASER” is intended for patients suffering from oral problems related to
the dental system, the temporomandibular system, the musculoskeletal system, therefore mainly for
adult patients, but there are no contraindications to the use of the "diode laser" also on patients of other
age groups.
There are no exclusions in subjects who can be treated with the device using the treatments
described in the document AMNII78 Clinical Protocols.
The bibliography currently available, at international level, does not reveal any contraindication in the
use of the diode laser in the applications field reported in the intended use.
The laser has many beneficial effects on human tissue given the correct values of power, frequency
and application time. With high values of power and other inappropriate parameters it may though
cause either undesired vaporization or necrosis in the radiated tissue.
Should necrosis be desired, as in cases such as photodynamic therapy or in the equivalent use of the
scalpel, it will be unavoidable that along with the targeted tissue, the closest surrounding tissue might
be damaged as well. The extent of such harm is essentially determined by the energy density provided
to that tissue and exposure time. In many cases the harm will turn out as light and tolerable compared
to the benefits.
The user should therefore very attentively check the following parameters, in order not to cause
undesired effects on the patient:
•Power
•Diameter of the tip
•Distance between the end of the tip and the tissue spot
•Continuous or pulsed laser emission
•Application time

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 5
In case of whitening activity, use in the presence of infections, gingival lesions or carious diseases on
the dental elements must be avoided.
ATTENTION
The improper use of this laser device might lead to unwanted, sometimes
dangerous effects.
1.3. SYMBOLS USED
Symbol “Attention”
The text enclosed in this space and described by the "Attention" symbol indicates a
dangerous situation that may result in materials damages or minor to moderate
injuries.
Symbol “Danger”
The text enclosed in this space and described by the symbol "Danger" indicates a
dangerous situation that can lead to a situation directly connected to serious or fatal
injuries.
Symbol “Information”
The text contained in this space and described by the "Information" symbol indicates
useful information on the device for the user.
1.4. APPLIED PART
The device is provided with applied parts TYPE B:
FIBER
HANDPIECE
TIP
FLAT TOP
HANDPIECE
WHITENING TOOL
REVIVE
HANDPIECE
BIOTIP
(for biostimulation)

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 6
SAFETY
DANGER
The device must be installed in environments with electrical systems compliant with
the regulations in force in your country.
DANGER
Do not allow the use of the device to non-professional operators or who have not
read the instruction manual.
Always check the device is in good condition.
ATTENTION
Do not simultaneously touch the accessible contacts of the connectors on the device
(fiber connector and USB connector) and the patient
ATTENTION
Do not use the device if any part of it is defective or worn. In this case, contact
authorized technicians.
ATTENTION
Replace the defective or worn parts only with original and guaranteed spare parts.
DANGER
The device is not suitable for use in the presence of a flammable anaesthetic mixture
with air or oxygen or with nitrous oxide.
ATTENTION
Do not use the device in the presence of liquids on the floor.
DANGER
Do not modify the device without the manufacturer's authorization, the use of
unapproved accessories and/or unauthorized modifications may cause imminent
dangers, injury or damages to persons and materials.
ATTENTION
Any installation, modification and maintenance operations must be carried out with
the device switched off, in the absence of the patient and exclusively by qualified
personnel authorized by the manufacturer.

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 7
2.1. ESSENTIAL REQUIREMENTS
The device within its casing and in the test conditions described maintains operation and emission
stability within the standard tolerance of ± 20%.
2.2. GENERAL SAFETY DISPOSITIONS
During its normal use, the laser device laser exposes the human body to laser light radiation; therefore
it is important to scrupulously read and follow all the safety dispositions listed in this chapter.
•Never leave the fiber end diode opening on the laser without protection cap.
•Absolutely avoid putting fingers inside the laser aperture or looking directly inside it.
•Do not simultaneously touch the accessible contacts of the connectors on the device (fiber
connector and USB connector) and the patient to avoid unexpectable high contact current can
flow through the patient.
Any serious injury related to this medical device must be reported to the manufacturer and to the
competent authority of the Member State in which the user and/or the patient is established.
ATTENTION
The beam delivery system of the device is the optical fiber, the use of the laser
beam with the patient must be done exclusively using the fiber and tip. The device
cannot be used without fiber connected.
Laser aperture and fiber connection are marked with specific labels
Mobile communication devices, radio frequencies (including antenna cables and external
antennas) must not be used closer than 30 cm (12 inches) to any part of the device, including
the cables specified by the manufacturer. Otherwise the performance of the device may be
degraded.
The following device is not compatible in simultaneous operation with other high frequency
surgical equipment
2.3. WORKING AREA
After the device has been delivered and its content checked it is necessary to
prepare the place where the laser will be used. Doors must limit the working area,
and each of these doors must visibly carry a safety label like the one shown in this
figure.
No one other than the authorized personnel can have access to the laser system
area of use.
All personnel present must scrupulously follow all the individual safety measures.
The medical laser is protected by unauthorized use through a digital key (password) and an interlock
safety system, which is managed by connecting the interlock connector to the device.

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 8
It is possible to connect the interlock with a remote switch as shown in the picture below.
2.4. INDIVIDUAL SAFETY MEASURES
All the safety measures here described must be scrupulously followed in order to avoid
accidental exposure to laser radiation.
•The personnel authorised to work inside the laser working area
must wear protective eyewear.
•Never direct the laser beam towards the eyes.
•Never look into the fibre connector.
•The fibre connector must always be covered either by the fibre
or the protection cap.
•Eliminate from the operative area all reflecting and metallic
objects, including personal belongings such as watches and rings since these objects risk
reflecting the laser beam.
•In case of danger immediately press the emergency button.
•Turn off the main switch when the laser is not in use.
•The intrinsic characteristics of the diode laser ray, if not correctly used, could set some non-
metallic material on fire. It is therefore advisable to follow these simple rules very carefully:
•Do not point the laser ray towards any clothing.
•We recommend that only appropriate light coloured, and completely dry clothing be worn.
•Remove all potentially flammable materials such as paper, wood, or plastic.
•Never use flammable gas during laser use.
•Any solvent or inflammable solutions must be allowed to completely evaporate before using the
laser.
•Avoid using any potentially inflammable anesthetic or gases such as oxygen or nitrous oxide.
The high temperature produced in normal use of the laser equipment may ignite some materials
such as cotton or wadding when saturated with oxygen. It is also important that all inflammable
solutions normally used to disinfect should be allowed to evaporate before using the laser
equipment. Attention should be drawn to the danger of ignition of dangerous gases.

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 9
2.5. OCULAR RISK
Eyes can be seriously damaged in case of unprotected exposure to laser light. For this reason it is
compulsory to wear protection glasses both for the operator and for the people present in the work
area.
Use only glasses with the same specifications as those supplied. In case of breakages or
other necessities, it is advisable to contact the supplier to request the same or similar
glasses with the same characteristics as those supplied.
According to standard EN 60825 CEI 76-2 II ed. the following values of OD optical Density and
DNRO Nominal Distance of Optical Risk have been calculated.
The supplied glasses conform to the European standard EN 207 and have an Optical Density 5 or
greater, at the diode emission wavelength.
Model
LA12D001.1
LA12D001.2
LA12D001.3
Wavelength
976
808
635
915
808
635
976
808
450
Laser
Diode
Diode
Diode
Emission
Continue/Pulsed
Continue/Pulsed
Continue/Pulsed
Power
16
8
0.5
16
8
0.5
16
8
0.5
Divergence
220 mrad
220 mrad
220 mrad
Diameter
0.2mm
0.2mm
0.2mm
Exposure time
5 s
5 s
5 s
Observation
Direct Light
Direct Light
Direct Light
DO
(at 0.1 m)
3.1
3.2
3.2
3.2
2.2
2.2
2.2
3.1
3.2
DNRO
[m]
2,770
2,806
2,770

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 10
Model
LA12D001.4
LA12D001.5
Wavelength
976
450
635
808
450
635
Laser
Diode
Diode
Emission
Continue/Pulsed
Continue/Pulsed
Power
16
4
0.5
16
4
0.5
Divergence
220 mrad
220 mrad
Diameter
0.2mm
0.2mm
Exposure time
5 s
5 s
Observation
Direct Light
Direct Light
DO
(at 0.1 m)
3.1
3.1
2.2
3.1
3.4
2.2
DNRO
[m]
2,510
3.,590
Model
LA12D001.6
LA12D001.7
LA12D001.8
Wavelength
976
808
450
Laser
Diode
Diode
Diode
Emission
Continue/Pulsed
Continue/Pulsed
Continue/Pulsed
Power
16
16
4
Divergence
220 mrad
220 mrad
220 mrad
Diameter
0.2mm
0.2mm
0.2mm
Exposure time
5 s
5 s
5 s
Observation
Direct Light
Direct Light
Direct Light
DO
(at 0.1 m)
3.1
3.4
3.1
DNRO
[m]
2,440
3,590
2,510

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 11
CERTIFICATIONS
3.1. CE DECLARATION OF CONFORMITY
For MEDICAL LASER device named
L A 12D 001.x S/n: ____
Manufactured by
LAMBDA SpA
Under Annex II of directive 93/42/CEE, except point 4, transposed from DL 46/97 and
its integration with standard 2007/47/CE issued with Law Decree 37 of 25/01/2010
and with the application of standard 99/05/CE.
The writer LAMBDA S.p.A. company, situates in Via Dell’Impresa 1, Brendola (VI), Italy,
manufacturer of the device reported, declares under sole responsibility that:
this device satisfies all the essential requirements required by Annex of Directive 93/42/CEE
on Medical Devices and its integration with standard 2007/47/CE; design, construction and
final checks are carried out as described by the Complete System of Quality Warranty
approved
in date 20/01/2020 da IMQ S.p.A., Via Quintiliano n.43 – 20138 Milano (Italy)
under the requirements of Directive 93/42/CEE at Annex II.
The company also ensures and declares under its responsibility:
1. that the device is considered belong to Class IIb.
2. that the manufacturer agrees to maintain and make available to the competent Authority, the
following technical documents, specified in Annex II, at point 6.1, of Directive 93/42/CEE for a
period of ten years from the last product manufacture date:
▪this Conformity Declaration;
▪the documentation of the complete system of Quality Warranty;
▪the communications at Notified Body relating to any adjustment of complete system of Quality Warranty;
▪the description of design, manufacture, sterilization and product performance;
▪decisions and reports of Notified Body relating to revision of complete system of Quality Warranty;
▪decisions and reports of Notified Body relating to product design testing
▪decisions and reports of Notified Body relating to any design changes
▪decisions and reports of Notified Body relating to periodic inspection conducted at the company;
▪decisions and reports of Notified Body relating to unforeseen inspection conducted at the company
Timbro e firma dell’amministratore delegato:

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 12
3.2. CE CONFORMITY MARKING
This productis marked with the CE label according to theEuropean standard applicable
for medical devices: CEE 93/42.
The manufacturer of this product is:
LAMBDA SpA
via dell'Impresa 1
36040 Brendola (VI)- Italy
+39 0444 349165
3.3. NAMEPLATE
The nameplate label reports the following data:
a. Manufacturer (Name and address)
b. Device code (model)
c. Device Serial Nr.
d. Manufacturing Date
e. Power supply voltage
f. Laser class
g. Medical class (and applied part)
h. Insulation class
i. Discontinuous operation mode
j. Warning symbol: “Read carefully the instruction manual before using the device"
k. CE symbol
l. Disposal symbol: “the device at the end of its life must be disposed separately as a special
product”
m. “Medical device”
n. “MADE IN ITALY”
a.
b
.
c.
d
.
e.
f.
g.
i.
j.
m
.
k.
n
.
h.
l.

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 13
3.4. WARNINGS
The device works discontinuously as per data reported in the nameplate and at the chapter
“TECHNICAL SPECIFICATIONS”.
The manufacturer is not responsible for the direct and indirect effects due to the use of the device.
These effects remain under the direct responsibility of the medical personnel performing the
treatments.
The manufacturer, the assembler, the installer and the importer do not consider themselves
responsible for the safety, reliability and performance of the device, unless the points mentioned above
are respected.
The manufacturer makes the technical details of design outlines and test instructions available, prior
to written request, so that the qualified personnel authorized by LAMBDA SPA will be enabled to
repair or maintain those parts of the system that the manufacturer consider as possible to repair.
The device has been tested in accordance with ESD test - criterion B - and so there may be conditions
of transient artifacts on the display, always momentary and due to indirect electric shock events. The
transitory is temporary, the equipmentis always operative andthe operator does not have to implement
any intervention.
3.5. ACCEPTANCE OF RESIDUAL RISK
LAMBDA SpA has implemented all the provisions required by the regulations to make the device
compliant with the standards and the level of operational risk is acceptable. The user is obliged to
strictly follow all the safety instructions contained in this manual. By using the device, the user accepts
the residual electrical, thermal and optical risks inherent to the device.
Since the tests are mostly "LINE TEST", but some are necessarily "TYPE TEST", by using the device
the user accepts the residual electrical, thermal and optical risks inherent in the device.

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 14
INSTALLATION
4.1. CHECK ON DELIVERY
Upon arrival of the goods and in the presence of the carrier, it is important to pay accurate attention
that the shipped material is correct and intact. In particular:
•The number of parcels and corresponding codes.
•The external packaging conditions and inside for damaged parts.
•Compare the contents of the package with the attached packing list.
It is important to immediately notify the carrier of all possible non-conformities found during checking.
The manufacturer states that in accordance with national and international laws, the customer
always takes full responsibility for the shipped goods. Unless previously specified, the goods are
always shipped without insurance.
4.2. ENVOIREMENT REQUIREMENTS
Keep the laser in a dry and clean ambient with the following requirement:
•TEMPERATURE: 10°C - 30°C
•HUMUDITY: 30% - 70%
•ATM. PRESSURE: 700 –1060 hPa.
4.3. CONDITION OF USE
ATTENTION
INSTALLATION MUST BE CARRIED OUT BY AUTHORIZED PERSONNEL
ONLY.
Unauthorized installation brings immediately to the warranty interruption.
In the room previously prepared for the laser use, remove all unnecessary inflammable material and
verify that the electric power panel complies with the current safety norms. Check the electric power
to see if it matches the laser system's electric requirements.
•Place the device in a flat stable position.
•Be sure that the device has been maintained in these environment conditions for at least 2 hours
before turning the device on.
•Protect the laser from direct sun light to avoid possible system overheating.
•Do not place the laser next to walls (minimum distance of 10 cm) or other locations that could
decrease air exchange. Do not cover the air vents even partially during use.
•Place the device so that it is easy to disconnect the device from the network plug
•Place the laser device at a safe distance from other machinery, to avoid possible
electromagnetic interference.
•While working do not cover the machine with things or clothes.

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 15
4.4. ELECTRIC REQUIREMENTS
It is important to verify that the power cable is not damaged before using the laser system. In
particular, the cable plug must be compatible with the powering network socket. Do not use adapters,
extensions or multiple sockets of any type.
The power supply complies with the CEI EN 60601-1 standard and has the following features::
Model
SINPRO MPU101-106
Input Voltage:
100 ÷ 240 V
Frequency:
50 ÷ 60 Hz
Output voltage:
15 V
Max absorbed current:
6.66 A
The external power supply is an integral and indissoluble part of the medical device.
ATTENTION
Do not use power supplies other than those supplied.
If replacement is necessary, contact LAMBDA SpA or an authorized dealer for
original spare parts.
4.5. DELIVERY AND STORAGE
ATTENTION
LAMBDA SPA is not responsible in any case for losses and / or indirect damages of
any kind that the goods may suffer due to transport that does not comply with the
above specifications.
In case of the device must be moved from different rooms of the same building, disconnect the
power supply and move the device (it is not necessary to disconnect the fiber). Take care not to
damage the fiber.
Should you need to move or ship the laser between different buildings, follow these indications:
•Always contact the dealer or LAMBDA SPA before any shipping.
•Always use packaging material supplied on delivery for transport. If not available it is
necessary to apply for a new packaging.
•Disconnect all cables accessories (optical
fiber, power supply interlock, footswitch)
•If fiber is not connected, THE LASER
APERTURE AND THE FIBER
CONNECTOR MUST ALWAYS BE
COVERED WITH THE PROTECTIVE CAP
During storage, follow the environmental requirements reported in the "TECHNICAL
SPECIFICATIONS" chapter of the manual and reported on a specific label both on the casing and on
the packaging of the device.

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 16
OPERATION MODE
5.1. OVERVIEW
EMERGENCY STOP
ON/OFF SWITCH
CABLE PEDAL INPUT
POWER SUPPLY INPUT
INTERLOCK
USB INPUT
HANDPIECE HOLDER
BATTERY
FIBER INPUT
DISPLAY /
TOUCH SCREEN

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 17
ACCESSORIES INCLUDED
OPTICAL FIBER BOX
INCLUDES:
FIBER/HANDPIECE
INTERLOCK
SWAB
TIP
5 BOXES
01 tip ENDO (Blue - box 04)
01 tip PERIO (Yellow - box 04)
01 tip SURGERY ( Green - box 04)
01 tip IMPLANT (White - box 04)
01 tip THERA (Black - box 04)
ACCESSORIES BOX
POWER SUPPLY WITH CABLE
BLUETOOTH PEDAL
(WITH ADDITIONAL CABLE)
PROTECTIVE GOGGLES (N.3 PZ)
LASER RADIATION LABLELS (2 PZ)
USER MANUAL
PROTOCOLS
CE CERTIFICATE
PIEGA TIP (30° - 60° - 90°)
MANIPOLO SBIANCAM.
TIP BIOSTIMOLAZIONE

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 18
5.3. LASER DEVICE PREPARATION
1. Insert the fiber holder (A) into the
dedicated hole.
2. Remove the protective cap of the diode
output.
3. Remove the protective cap of the fiber
4. Insert and screw the fiber (A), connect
interlock (B) and power supply (C).
D connector is for the cable of the pedal
5. Insert the optical fiber inside the ring of
the fiber holder
6. Put the handpiece in the handpiece holder
when not in use
A
B
C
D

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 19
5.4. PEDAL INSTALLATION
The Bluetooth pedal is powered by N.3 non-rechargeable 1.5V AA batteries. Make sure that they are
present in the battery compartment before proceeding.
Turn on the pedal using the switch on the bottom.
The flashing red LED indicates that batteries are running low. Replace the batteries.
The green LED is on when the pedal is pressed.
The blue LED indicates the process of association of the pedal to the laser.
-On fixed: active connection
-Flashing slow: not connected
-Fast flashing: in association
To associate the pedal with the laser, enter the settings menu and select Bluetooth/pedal. Press the
red button on the bottom of the laser while starting the association process from the laser screen.
In case of malfunctioning or if the “LOW BATTERIES” led switches on, replace the batteries using
only equivalent batteries.
Replacement of the pedal batteries can be done by the user; remove the cover placed in front of the
pedal and replace batteries.
If the problem remains contact Service of Lambda Spa or your dealer.
At delivery, the pedal is provided already synchronized.
The above procedure is required in case of pedal replacement.
BATTERIES
PACK
LED
SYNCHRONIZATION
BUTTON
ON/OFF
ACTIVATION
SWTICH

DIODE LASER LA12D001.X - USER MANUAL LI12D001.1_EN 2021-01-12 20
5.5. HANDPIECE AND TIP INSTALLATION
On delivery of the device the tips supplied in the packaging have not been sterilised
Different tip sizes are available for different
applications. They are colour-coded for easy
identification. The tips include a screw on ring.
APPLICATION
COLOUR CODE
Size
Ø Tip
ENDODONTICS
BLUE
A 25mm
B 3mm
C 15mm
200µm
PERIODONTICS
YELLOW
A 25mm
B 3mm
C 10mm
400µm
SURGERY
GREEN
A 25mm
B 3mm
C 5mm
300µm
IMPLANT
WHITE
A 25mm
B 3mm
C 8mm
300µm
THERAPY
BLACK
A 25mm
B 3mm
C 5mm
400µm
WHITENING
GREY
BIOTIP
GREY
FLAT TOP
GREY
REVIVE
GREY
This manual suits for next models
8
Table of contents
Popular Measuring Instrument manuals by other brands

Protimeter
Protimeter Grainmaster i instruction manual

THORLABS
THORLABS PDA20H user guide

Hanna Instruments
Hanna Instruments HI 9214 instruction manual

Ametek
Ametek LAND SPOT+ Series user guide

Hanna Instruments
Hanna Instruments HI96702C instruction manual

Thornwave Labs
Thornwave Labs BT-DCPM quick start guide