USER AND MAINTENANCE GUIDE FOR
Samsoft Mini
PATIENT LIFT
9
undone, the upholstery and sling weft is not visible, and there
is no torn part.
Never use a defective sling.
The worn or malfunctioning parts should be replaced
with original parts manufactured by Drive DeVilbiss
Healthcare France.
Have some tries with a load before commissioning the patient
lift again.
9.5) Preventive maintenance
Every year or when necessary, and to be performed by a
competent sta, according to paragraphs 9.5.1, 9.5.2, and 9.5.3
below.
It is necessary to control the device every year.
The battery pack should be changed about every 18 months
depending on how it is used.
The crossbar suspension axle should be imperatively checked
yearly on each maintenance task.
The yearly maintenance should be performed by skilled technical
sta experienced with electromechanical device maintenance.
The maintenance tasks do not require particular tools; adequate
tools should be used according to the best industry practices.
Mechanical fastening, plays, and adjustments should be realised
according to the best industry practices.
9.5.1) Mechanical control :
Disassembly and systematic lubrication of the crossbar axle.
In case of any wear sign, the axle should IMPERATIVELY be
replaced.
Dismantling, cleaning, and lubrication of the front and
rear casters.
Control and lubrication after dismantling the mast and
side rail pivots.
Check the welding. Visual inspection.
Check the alignments. Visual inspection.
Control and lubrication of the cylinder fastenings.
Check the tightening of all the parts.
Dismantling, cleaning, and lubrication of the mast hinge.
Check the fastening, caster play, and brake performance. The
mechanical parts that have cracks, wear signs or geometrical
defects should be replaced with original components delivered
by Drive DeVilbiss Healthcare France.
9.5.2) Electrical inspection :
Check the remote control. Reminder: the remote control does
not enter the warranty field.
Check the charger.
Check the remote control box and the cylinder.
Check the lifting/lowering speed.
Check the emergency stop and lowering.
Condition of the electric cables (change the cables in case of
doubt).
RReplace the electric cables and/or the components to which
they are connected if there are signs of wear, sheath cut, outlet
damage or plastic and sheath darkening around the coupling
area. Replace the electric cables and/or the components to
which they are connected if there are signs of wear, sheath cut,
outlet damage or plastic and sheath darkening around the coupling
area.
Note : Never disassemble the protective housings of the cylinder,
battery pack, remote control and control box, and battery charger.
9.5.3) Charging test:
Charge the battery pack.
Check the battery pack power and autonomy.
Test the device when fully loaded (150 kg) and control the
power cut threshold (230 kg +/- 10 kg).
Tests.
Battery pack should be changed every 18 ùonths, depending
on use.
9.5.5) Service book:
Every maintenance operation or intervention should be recorded
in the service book attached to these instructions.
X SHIPPING, STORAGE, AND DISPOSAL
The patient lift should be returned or transported in its original
packaging.
Before storing the patient lift, clean it totally and check its
performance. Store the patient lift in a dry place away from dust
and humidity. The batteries should be totally charged every 3
months.
When commissioning, after storing, charge the battery pack,
and check its performance before using it again.
Storage temperature: between 0 and + 50°C Humidity between
10 and 80%
Service temperature: between +10 and + 40°C Humidity
between 30 and 80%
The disposal of defective parts, packaging
and battery pack should be performed by
approved waste treatment centres.
The involved system should not be disposed of with household
waste, but should be sent to an appropriate waste collection
point or to a distributor.
A financial contribution is given for collecting, dismantling, and
recycling the system in question.
By respecting these instructions, you respect the environment
and participate in protecting the natural resources and the
human health.
1) Check the axle is not distorted.
2) Check the collar is not worn.