manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Durham
  6. •
  7. Measuring Instrument
  8. •
  9. Durham 9840 User instructions

Durham 9840 User instructions

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Setting up the
9840 Load Cell Indicator
To turn on the 9840 Load Cell
Connector, reach behind in the back
corner and ip the switch up.
Before starting anything, be sure to rst
plug in the sensor(s) you will be using.
In starting out, you will see
“>User Data Entry” on the screen.
To enter setup mode, press the two end
buttons at the same time (◄and ►).
(1) (2)
(3)
As it starts up, you will see a series of
screens introducing you to the 9840 Load
Cell Connector and displays of the
settings from its last use.
(4)
7418 East Helm Drive • Scosdale, Arizona 85260 • 480.948.5555 • www.interfaceforce.com
Page 1 of 7
Setup Guide 9840 v1.1 09-18-2018
(5)
(6)
Enter the password “888” by pressing
+or —to select the numbers while
pressing ◄ to move to the next digit.
Then press ENTER.
After being brought to the next screen,
press ►‘till you get to the
“>Calibration” option and
then press ENTER.
Before you can enter the
calibration settings, you will be
required to enter a password.
7418 East Helm Drive • Scosdale, Arizona 85260 • 480.948.5555 • www.interfaceforce.com
Setup Guide 9840 v1.1 09-18-2018
Page 2of 7
(7) (8)
(9)
Press ►until you get to the
“>>Cal by mV/volt” option
and then press ENTER.
Using ►, select
“>>>2-Point mV/V Cal”
and then press ENTER.
Select the measurement cell type that you
are using. For this example, we are using a
load cell, so we select “Load”.
Then press ENTER.
(10)
Enter the cell’s serial number by using the
+or —signs then ◄for the next digit.
ENTER.
7418 East Helm Drive • Scosdale, Arizona 85260 • 480.948.5555 • www.interfaceforce.com
Setup Guide 9840 v1.1 09-18-2018
Page 3of 7
Enter the month using the +or —and
then ◄ for the next digit(s). Do the same
for the day and year, and press ENTER
after each one.
Select the necessary unit of measure
and press ENTER.
(11)
(12)
7418 East Helm Drive • Scosdale, Arizona 85260 • 480.948.5555 • www.interfaceforce.com
Setup Guide 9840 v1.1 09-18-2018
Page 4of 7
Next, you will need to refer to the sensor’s Calibration Certicate. In it, identify the value
for the max capacity and enter it into the 9840 using the same buttons: + , — ,◄, and ►.
For this example, the max capacity of the 1210AF is 10K,
so we enter “10000” and then press ENTER.
*Or you can enter the appropriate SEB Output Value as your full-scale output value.
(14)
(13)
Enter the voltage excitation
using buttons +, —, ◄, and ►.
T hen press ENTER.
(14)
7418 East Helm Drive • Scosdale, Arizona 85260 • 480.948.5555 • www.interfaceforce.com
Setup Guide 9840 v1.1 09-18-2018
Page 5of 7
(16)
(15)
(17)
Check that there are no masses (no load)
calibration begins with a zero balance.
When ready, press ENTER.
While the screen is displaying “Reading...”,
the 9840 is calibrating. Wait for just a few
With that information from the calibration
buttons +, —, ◄, and ►.
Then press ENTER.
For the next display, you will be
required to enter the mV/V for the
full-scale output of the direction that
you are going to use. In this example,
we will pick “Tension”.
We see that the value of 4.07906 is the
full-scale mV/V, correlating with the max
capacity of 10K that we entered earlier.
7418 East Helm Drive • Scosdale, Arizona 85260 • 480.948.5555 • www.interfaceforce.com
Setup Guide 9840 v1.1 09-18-2018
Page 6of 7
(18a) (18b)
Calibration setup is now complete. To return to the main menu, press ESC.
7418 East Helm Drive • Scosdale, Arizona 85260 • 480.948.5555 • www.interfaceforce.com
Setup Guide 9840 v1.1 09-18-2018
Page 7of 7