manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. E-one
  6. •
  7. Multimeter
  8. •
  9. E-one e+erne VC60B User manual

E-one e+erne VC60B User manual

VC60B/VC60D
3 1/2 digits digital display
AC Voltage/Insulation tester
User manual
Design and Manufacture by
Zhuhai E-one science and technology Co., Ltd
1
2
Digital intelligent insulation meter
Measuring range Resolution
Medium
resistance
Accuracy
5M~200M
50M~2G
1000M~20G
0.1M
1M
10M
1M
10M
20M
200M
100M
500M
50M
50M
20M
±(4%+5d)
±(10%+5d)
±(15%+5d)
Model
2-1 Insulation test
2-2 AC Voltage measuring:
B型
D型
Voltage Gear
500V
200M
1000V
2500V
1000V
2G
20G
2.Specification
Note:
750V 1V ±
(1.0% + 3d)
Range Resolution Accuracy
AC frequency response:40Hz~400Hz
Display: Sine-wave RMS (Average)
Ω
Ω
Ω
10M
20M
500V
500M~20G 10M
1M
20G
Ω
200M~20G
10M~2000M
10M
1M
2000M
20G
Ω
Ω
5M~200M
50M~2G
0.1M
1M
10M
20M
±(4%+8d)
250V
200M
2G
Ω
Ω
50M~2000M
2000M
Ω
100M~2000M
2000M
Ω
5M~200M
50M~2G
0.1M
1M
200M
2G
Ω
VC60B/VC60D AC voltage/Insulation tester adopts DC voltage
changer which is low consume, carried high voltage ratio and
inductance power stored. It transits the 9V voltage into
250V/1000V/2500V and process the insulation test with
analog/digital circuit. Widely use in solving problems of
electrical equipments and system.
1.Adopt 6*1.5V batteries, more durable.
2. Big screen design with big font display and
LED backlight, which makes the viewing more
direct and clear.
3.Voltage: 250V/500V/1000V(VC60) 500V/1000V/2500V
(VC60D)
4.Press lock function in measuring
5.Protection optimized, enable to bearing
surge of short-curcuit and residual voltage
Features:
Summarize:
Technique features:
1.General feature:
Display: 80X52mm LCD display, LED back light, Max.display“1999”
Over range indicate : LCD shows " 1" when it's over-range
Supply : 6*1.5V batteries or outer DC9V/1A power adapter.
(Caution: Batteries and adapter can't be use simultaneously)
Low voltage indicate: Symbol shows
static consume: About 10mA
Max.Current: About 280mA
Using condition: Temperature 0℃~40℃,Humidity30%RH~80%RH
Storage condition: -10℃~ +50℃, Relative humidity<85%RH
Dimension: 185*115*65mm
Weight: About 500g(Including batteries)
Accessory: 1*User manual,1*test pen, 1*test clamp
Medium resistance is a measuring lower value which indicates and ensures that
when operating the voltage won’t be lower than 90% of the tested voltage.
Create the 3th era
The waveform display multi-meter E-one
The professional manufacture
250V 500V 1000 V
POWER
VC60B
Insulation tester
M
G
illustration:
1
250V 500V 1000V
POWER
VC60B
Insulation tester
M
G
2
3
4
5
6
4
7
8
9
1011
Operation:
5-1 Insulation testing:
Maintenance:
5
6
Digital intelligent insulation meter
3
4
1. Outer DC power socket.
2. Gear selected key-press.
3. Measuring range key-press.
4. Hand-strap position.
5. Power key-press.
6. Test key-press, spin to lock.
7. Display screen.
8. Measurand ground terminal connected socket.
9. Voltage input terminal.
10.Ground terminal, connecting to protect the ring
electrode when testing for the cable.
11. Tested terminal socket.
Turn on the instrument by pressing the power key-press. Please
replace the batteries timely when the low power symbol displayed.
1. Connect the measurand ground terminal with the input wire
E, and connect the tested terminal to L.(Make the L terminal
connection hand in the air as much as possible. Please
connect to the COM terminal when testing for the cable.)
2. According to the detecting mission, select the proper gear, then
LCD will display the corresponding voltage; if the tested voltage is
unclear, try from the low gear, when it show as”1", change to the
higher gear to continue.
3. Select the proper measuring range, press the TEST key, then
the L and E terminal will have high voltage output, LCD display
symbol , the high voltage indicate light will light up.
4. Press down the TEST key to progress the measuring, readout
the value when it display stable. Release the TEST key when
finished, the LCD will display “1".
(If the TEST press-key down but the LCD still display “1", it’s
possibly cause by:
A: The tested resistance has over the selected range or over the
Max. measuring range.
B: The wire maybe not being well connected or disconnected.
5. Press the TEST key and spin clockwise to lock and
measuring continuously.
6.Caution: It’s prohibit from pushing/pulling the connecting
wire or touching the tested resistance when the TEST key is
being press down cause the L and E terminals will have high
voltage output.
If the readout bing unstable all the time, it’s possible because:
A: The connecting wire is too long.
B: The humidity of the testing environment may be a little high.
C: The outer layer of the tested resistance need to be
connected to the COM.
D: The tested resistance is unstable, lies in critical state.
7. When progressing the insulation test, please use the
conductor to protect around the measurand’s appearance
between the two tested terminals in order to eliminate the
measuring inaccuracy which generated from the current’s
leakage of the measurand’s appearance. Plus, the protected
conductor should be connected to the COM terminal with the
wire.
5-2 AC Voltage measuring
1. Press the gear of 750V~, connect the ACV terminal with
the red test pen, and the COM terminal with the black one.
2. Full contact two tested terminals and parallel to the
electric power with the testing probe. The readout on the
LCD screen is the measuring result
3. If the LCD displayed as “000", it indicates that the tested
voltage is zero or the connected wire may be
disconnected.
1. This instrument is a precision measuring instrument, please
avoid using under bad environment, it's not allow to change the
circuit, replace the component and calibrate and repair the meter
without obtaining the authorization of the Product Service Center
of our company.
2.If there is no display on the LCD even when there is power
left, then please check the connecting condition of the batteries.
3.The batteries should be replaced when the low power symbol
displayed. Make sure the measuring circuit is cut off. Loose the
screw of the battery box, take out the batteries and install the new
one, please pay attention to the polarity, after that, put back the
cover and fasten with screw.
4.If you don't use the meter for a long period, please take out
the batteries to prevent the electrolyte leakage which may affect
the usage of the meter or even damage the meter.
Please full read this user manual carefully before using. If you
have any further question during your usage, please feel free to
contact us.
Tel: 0756-8659101
Skype: e-one.caroline
QQ/Wechat: 373234353
Website: http://e.e-one.net.cn/
Further service:
Digital intelligent insulation meter

This manual suits for next models

1

Popular Multimeter manuals by other brands

Forch F72 manual

Forch

Forch F72 manual

Vichy VC97 Operation manual

Vichy

Vichy VC97 Operation manual

DPMS XPQ user manual

DPMS

DPMS XPQ user manual

HoldPeak 850E Operation manual

HoldPeak

HoldPeak 850E Operation manual

Eurotron Mini MMX II instruction manual

Eurotron

Eurotron Mini MMX II instruction manual

Format 55790010 instruction manual

Format

Format 55790010 instruction manual

Chauvin Arnoux C.A 6454 user manual

Chauvin Arnoux

Chauvin Arnoux C.A 6454 user manual

MAC TOOLS EM720T instruction manual

MAC TOOLS

MAC TOOLS EM720T instruction manual

GW Instek GDM-350B user manual

GW Instek

GW Instek GDM-350B user manual

Fluke LinkRunner Pro Quick reference guide

Fluke

Fluke LinkRunner Pro Quick reference guide

IDEAL 61-764 instruction manual

IDEAL

IDEAL 61-764 instruction manual

Tektronix DM 504A instruction manual

Tektronix

Tektronix DM 504A instruction manual

VOLTCRAFT VC611BT operating instructions

VOLTCRAFT

VOLTCRAFT VC611BT operating instructions

REED R5007 instruction manual

REED

REED R5007 instruction manual

PeakTech 3365 Operation manual

PeakTech

PeakTech 3365 Operation manual

Extech Instruments 407119A user guide

Extech Instruments

Extech Instruments 407119A user guide

PROSOL UK IEC-1010 Operator's instruction manual

PROSOL UK

PROSOL UK IEC-1010 Operator's instruction manual

TES TES-2620 instruction manual

TES

TES TES-2620 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.