ECOTEST Gamma Sapiens User manual

UDKG-01A (UDKG-01APro)
“Gamma Sapiens”
INTELLECTUAL
GAMMA RADIATION
DETECTOR
Operating manual


1
Dear users,
You had chosen well if purchased a device of
“ECOTEST” trademark manufactured by “Sparing-Vist
Center”. Your device is easy to use and can be applied for
household purposes without special training of the user.
Should any questions arise, please contact our managers by
telephone (+38 032) 242-15-15, fax (+38 032) 242-20-15
We would greatly appreciate to receive your comments
and suggestions on its operation. The device is under
24-months (free of charge) warranty maintenance.
Best regards,
International Sales Department

2
Dear users,
We would like to draw your attention to the delivery kit of the
UDKG-01A (UDKG-01APro) “Gamma Sapiens” intellectual gamma
radiation detector that includes two AAA batteries. However, the
power for operation can also be supplied by two ААА nickel-metal-
hydride (NiMH) rechargeable batteries of no less than 800 mАh
capacity instead of batteries that come with the device. You can buy
these rechargeable batteries and their charger at any retail outlet.
Famous manufactures, such as Varta, arehighly recommended.
Choose an automaticbattery charger, preferably with a function to
control chargingof each rechargeablebattery.
Charge rechargeable batteries only when they are fully
discharged. Recharging partially discharged rechargeable batteries on
a repeated basis minimizes considerably their lifespan.
Do not use one rechargeable battery along with a non-
rechargeable one to power the gamma radiation detector. This will
lead to the rechargeable battery failure.

CONTENTS
1 DESCRIPTION AND OPERATION............................5
1.1 Purpose of use..................................................5
1.2 Technical specifications...................................6
1.3 Delivery kit ....................................................12
1.4 Design and principle of operation..................15
1.5 Labeling and sealing ......................................19
1.6 Packing of the detector...................................20
2 PURPOSE OF USE..................................................... 22
2.1 Operating limitations .....................................22
2.2 Preparation for operation and use ..................23
3 MAINTENANCE........................................................33
3.1 Technical maintenance of the detector...........33
3.2 Verification of the detector............................37
4 CERTIFICATE OF ACCEPTANCE ..........................52
5 PACKING CERTIFICATE......................................... 53
6 WARRANTY..............................................................54
7 REPAIR.......................................................................56

2
8 STORAGE...................................................................57
9 SHIPPING....................................................................58
10 DISPOSAL.................................................................59
WARRANTY CERTIFICATE .......................................60
APPENDIX А
TROUBLE RECORD DURING USЕ ............................61
APPENDIX B
PERIODICVERIFICATIONOFKEYSPECIFICATIONS..62
APPENDIX C
REPAIR...........................................................................63
APPENDIX D
STORAGE......................................................................64
APPENDIX E
VERIFICATION AND INSPECTION RESULTS.............65

3
This operating manual (hereinafter called the OM) is
intended to inform the user about the principles of
operation of the UDKG-01A (UDKG-01APro) “Gamma
Sapiens” intellectual gamma radiation detector and its
operation procedure. The OM contains all information
necessary for proper use of the detector and full
realization of its technical capabilities.
The UDKG-01A (UDKG-01APro) “Gamma Sapiens”
intellectual gamma radiation detector is easy to use
without the need of special training of the user.
The UDKG-01A (UDKG-01APro) “Gamma Sapiens”
intellectual gamma radiation detector is intended to work
with smartphones or tablets which run under the Android
operating system.

4
The OM contains the following abbreviations and
symbols:
DE - ambient dose equivalent of gamma
radiation;
DER - ambient dose equivalent rate of gamma
radiation;
QCD - Quality Control Department;
TS - technical specifications.

5
1 DESCRIPTION AND OPERATION
1.1 Purpose of use
The UDKG-01A (UDKG-01APro) “Gamma Sapiens”
intellectual gamma radiation detector (hereinafter called
the detector) is designed to measure ambient dose
equivalent (DE) and ambient dose equivalent rate (DER)
of gamma radiation and communicate measured data to a
smartphone or a tablet via the Bluetooth radio interface.
The detector is used to:
-monitor personal radiation safety;
-monitor apartments, buildings, and constructions,
household items, clothes, soil surface of grounds,
and vehicles exposed to radiation.

6
1.2 Technical specifications
1.2.1 Measurement range of gamma radiation DER
–from 0.1 to 9999.0 μSv/h for UDKG-01APro
version;
–from 0.1 to 5000.0 μSv/h for UDKG-01A version
1.2.2 Main relative permissible error limit in gamma
radiation DER measurement with 0.662 MeV energy
)10(
*
H
, %, with confidence probability of 0.95
(calibrated with 137Cs) is estimated by the formula:
- for UDKG-01APro version by the formula:
)10(
*
H
= 15+2/
)10(
*
H
, (1.1)
- for UDKG-01A version by the formula:
)10(
*
H
= 25+2/
)10(
*
H
, (1.2)
where
)10(
*
H
is a numeric value of measured gamma
radiation DER in μSv/h.
1.2.3 Measurement range of gamma radiation DE is
from 0.001 to 9999 mSv.

7
1.2.4 Main relative permissible error limit in gamma
radiation DE measurement with confidence probability of
0.95 is:
- 15 % for UDKG-01APro version;
- 25 % for UDKG-01A version.
1.2.5 Energy range of registered gamma radiation is
from 0.05 to 3.00 MeV.
1.2.6 Energy dependence of the measurement results
during gamma radiation DER and DE measurements in
the energy range from 0.05 to 1.25 MeV is not more than
± 25 %.
1.2.7 Complementary permissible error limit in
gamma radiation DER and DE measurements caused by
ambient air temperature variations from –18 to + 50 oC is
not more than 5 % per each 10 oC deviation from +20 oC.
1.2.8 Power for the detector is supplied by two AAA
batteries with a total voltage of 2.4 to 3.2 V. Operating
supply voltage is 3.0 V.

8
1.2.9 Complementary permissible error limit in
gamma radiation DER and DE measurements caused by
supply voltage variations of the detector within 2.4 - 3.2 V
is not more than 5 %.
1.2.10 Average current consumption of the detector is not
more than 20 mA.
1.2.11 Time of continuous operation of the detector
with a new rechargeable battery consisting of two AAA
cells of up to 1280 mAh capacity under natural radiation
background conditions is not less than 60 hours.
1.2.12 Setup time of operating mode and measurement
time of the detector is not more than 3 minutes.
1.2.13 Unstable readings of the detector during gamma
radiation DER measurement for an 8-hour continuous
operation is not more than 5 %.

9
1.2.14 The Bluetooth radio interface (hereinafter –
radio channel) serves as the interface for data
communications between the detector and the smartphone
or tablet personal computer (tablet PC).
1.2.15 The detector is able to perform data
communications via the interface within a distance of not
less than 5 m.
1.2.16 Mean time between failure is not less than
6000 hours.
1.2.17 Average service life of the detector is not less
than 6 years.
1.2.18 Mean time to repair of the detector is not more
than 12 hours.
1.2.19 Average shelf life of the detector is not less
than 1 year
1.2.20 Dimensions of the detector do not exceed
194095 mm.

10
1.2.21 Weight of the detector without batteries is not
more than 0.05 kg.
1.2.22 The detector meets the requirements of
technical specifications (TS) under the following
operating conditions:
- temperature from –18
to + 50
оС;
- relative humidity up to 95 % at + 35
оС, non-
condensing;
- atmospheric pressure from 84 to 106.7 kPa (from
645 to 800 mm Hg).
1.2.23 The detector is tolerant to sinusoidal vibration
effects of the group N1 according to GOST 12997-84 over a
frequency range of 10 –55 Hz, shift amplitude corresponds
to 0.15 mm.
1.2.24 The detector withstands exposure to shocks
according to GOST 12997-84 with the following parameters:
- shock pulse duration –9.5 ms;
- number of shocks –1000±10;
- maximum shock acceleration –100 m/s2.

11
1.2.25 The detector in its shipping container can
endure:
- ambient air temperature from –25 to +55 0С;
- relative air humidity up to (95±3) % аt +35 0С
temperature;
- shocks with 98 m/s2acceleration, shock pulse
duration –16 ms, number of shocks –1000 ± 10 in each
direction.
1.2.26 The detector is immune to constant or
alternating magnetic fields with (501) Hz of 400 A/m
intensity.
1.2.27 The detector tolerates exposure to gamma
radiation with DER of 1 Sv/h within 5 minutes.
1.2.28 Ingress protection rating of the detector’s
housing meets IP30 specifications according to DSTU EN
60529:2018.
1.2.29 The design of the detector supports
decontamination.

12
1.3 Delivery kit
1.3.1 The detector delivery kit consists of units and
maintenance documentation presented in:
- Table 1.1 for UDKG-01APro version;
- Table 1.2 for UDKG-01A version.

13
Тable 1.1 –Delivery kit of the UDKG–01AProdetector
Type
Item
Quan-
tity
Note
ВІСТ.412129.024-
02
UDKG–01APro
“Gamma
Sapiens”
intellectual
gamma
radiation
detector
1 pc.
ВІСТ.412129.022-
02 НЕ
Operating
manual
1 copy
ENERGIZER
AAA 1.5 V
battery
2 pcs.
Batteries
of other
manufac-
turers are
permitted
Case of model
491, Galvemo
Ltd
1 pc.
Package
1 set

14
Тable 1.2 –Delivery kit of the UDKG–01A detector
Type
Item
Quan-
tity
Note
ВІСТ.412129.024
—04-02
UDKG–01APro
“Gamma
Sapiens”
intellectual
gamma
radiation
detector
1 pc.
ВІСТ.412129.022-
02 НЕ
Operating
manual
1 copy
ENERGIZER
AAA 1.5 V
battery
2 pcs.
Batteries
of other
manufac-
turers are
permitted
Case of model
491, Galvemo
Ltd
1 pc.
Package
1 set

15
1.4 Design and principle of operation
1.4.1 General information
The detector is presented as a monoblock construction
with a built-in sensor of gamma radiation (the Geiger-
Muller counter), a printed-circuit board with electronic
components, and batteries.
The operation principle of the detector is based on
transformation of radiation by the Geiger-Muller counter
into the sequence of voltage pulses; the number of pulses
is proportional to the registered radiation intensity.
The power for operation is supplied by two AAA
batteries.

16
1.4.2 Design description
The detector is designed as a flat square plastic
housing with rounded corners.
The housing (Figure 1, 2) consists of the front (1) and
the back (2) covers.
On the right side of the front cover there is a disk-
shaped control and indication board (3) with a decorative
circle frame (4) around. The window (5) for two-color
LED indicator and the control button (6) are also
located in that area.
The battery compartment (7) for two AAA batteries is
in the middle of the back cover. The polarity signs are
indicated at the bottom of the battery compartment (7) for
proper insertion of batteries. The battery compartment is
closed with a lid (8).
A clip with an opening (9) fixed to the back cover is
used to attach the detector to a strap.
Other manuals for Gamma Sapiens
1
This manual suits for next models
2
Table of contents
Other ECOTEST Security Sensor manuals
Popular Security Sensor manuals by other brands

Vive
Vive LVA1074BED user manual

CAME
CAME SADRIN 205 installation manual

KUSSMAUL
KUSSMAUL 091-85 instruction manual

DIGMESA
DIGMESA FHK-LCD operating instructions

Infinite
Infinite L-com SRCN-C267-1 user guide

EGON HARIG
EGON HARIG flamtron FL (Ex) d 07 Series Assembly, Installation and Commissioning Instructions

Brica
Brica by-my-side 60004 instructions

Honeywell
Honeywell Lyric W1 installation instructions

STEINEL PROFESSIONAL
STEINEL PROFESSIONAL IR CM COM1-24 Information

Hoiax
Hoiax Alpha IP SMI 63001 instruction manual

schmersal
schmersal RSS 36-SD-ST-DU manual

System Sensor
System Sensor 885WP-B Installation and maintenance instructions