
Equilibrate and Characterize
If your µPro pump is fitted with a pressure transducer and you wish to use pre-pressurization,
equilibrate and characterize your system once the pump has been primed.
Equilibration
The equilibration conditions you choose will depend on your application. Generally, you will
want to equilibrate under conditions identical to those at the beginning of the file you will be
running. You will want to run equilibration with the system connected to the downstream
devices you will be using during the running of the file. An aim of equilibration is to determine
the operating pressure of the system so that system characterization accuracy is enhanced.
To equilibrate the system select REAL TIME operation (if you have not changed the menu from
the last point of these procedures, you are already in REAL TIME operation) from the Main
Menu by pressing softkey A.
The cursor will appear on the first line of the display. The flow rate you enter depends on your
application. Since the goal of equilibration is to determine the operating pressure of the system
at the desired flow, the final flow rate run to equilibrate should be the flow you will run in your
application. However, because syringes will need to compress a volume due to the
compressibility of fluids, you may want (for a relatively short period of time) to enter a higher
flow rate to accomplish this compression quickly. You can quickly and easily change the
entered flow rate by positioning the cursor in the FLOW field, enter a new flow rate and press
ENTER.
When you have entered the appropriate flow rate and press ENTER, the cursor will appear on the
second line, where you will enter the appropriate %B (if multiple syringes are present and
gradient mode is chosen) and press ENTER.
If you wish to activate one of the six external events you may do so (0 is OFF, 1 is ON). Press
enter through these fields so the cursor is positioned in the HI LIM (high pressure limit) field.
Enter a high pressure limit and press ENTER. The cursor will move to the LO (low pressure limit)
field. Enter an appropriate low pressure limit and press ENTER.
Replace the front cover if you have not done so already. Pressing RUN will cause the µPro
system to run the syringes under the specified conditions.
You will want to allow the system to pump for a period of time sufficient for the pressure to
stabilize at the final flow rate. Once the pressure has stabilized, you may characterize the
system.
2.17
REAL TIME
EDIT
RUN
SETUP/CONFIGURATION
MAIN MENU
A
B
C
D
Select softkey A
FLOW:XXXXX.XX
%B:XXX
ON:0 0 0 0 0 0
HI LIM:XXXXX LO:XXXX
A
A
B
C
D
Enter desired values
TIP: If multiple syringes are plumbed to a common outlet, as in gradient mode, you will want to
program the additional syringes (or % B, etc.) at the rates they will operate at during the
beginning of your file. For example, if the program you intend to run begins at 5% B,
program 5%B during this “REAL TIME” equilibration.