1.0 INTRODUCTION
1. Place the empt provided cup (0.5 litre) on the scale
and turn the scale on. It shows 0.0 gram.
2. Fill up the cup with the sample material (+/- 1.0g). The
fill quantit is shown on the upper left corner on the
displa of the humimeter FS2.
3. Make sure that the measuring chamber is completel
empt . It is important that no material is left in the
measuring chamber when ou turn on the device.
4. Switch on the humimeter FS2 b pressing the power
button ( ) for 3 sec.
5. As the next step, please do the self calibration. The
word “calibrate” will show up on our displa . Accept
b pressing the button.
6. In case a menu point is shown, press the left button
( ) until ou reach the measuring window.
7. Select the right calibration curve for our material
under test using the buttons or .
Electromatic Equipment Co., Inc.
-11-
9.0 WARRANTY
Electromatic Equipment Co., Inc. (Electromatic) warrants to the original purchaser
that this product is of merchantable qualit and confirms in kind and qualit with the
descriptions and specifications thereof. Product failure or malfunction arising out of
an defect in workmanship or material in the product existing at the time of deliver
thereof which manifests itself within one ear from the sale of such product, shall be
remedied b repair or replacement of such product, at Electromatic’s option, except
where unauthorized repair, disassembl , tampering, abuse or misapplication has
taken place, as determined b Electromatic. All returns for warrant or non-warrant
repairs and/or replacement must be authorized b Electromatic, in advance, with all
repacking and shipping expenses to the address below to be borne b the
purchaser.
THE FOREGOING WARRANTY’S IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO, THE
WARRANTY OF MERCHANTABILITY AND FITNESS FOR ANY PARTICULAR
PURPOSE OR APPLICATION. ELECTROMATIC SHALL NOT BE RESPONSIBLE
NOR LIABLE FOR ANY CONSEQUENTIAL DAMAGE, OF ANY KIND OR
NATURE, RESULTING FROM THE USE OF SUPPLIED EQUIPMENT, WHETHER
SUCH DAMAGE OCCURS OR IS DISCOVERED BEFORE, UPON OR AFTER
REPLACEMENT OR REPAIR, AND WHETHER OR NOT SUCH DAMAGE IS
CAUSED BY MANUFACTURER’S OR SUPPLIER’S NEGLIGENCE WITHIN ONE
YEAR FROM INVOICE DATE.
Some State jurisdictions or States do not allow the exclusion or limitation of
incidental or consequential damages, so the above limitation ma not appl to ou.
The duration of an implied warrant , including, without limitation, fitness for an
particular purpose and merchantabilit with respect to this product, is limited to the
duration of the foregoing warrant . Some states do not allow limitations on how long
an implied warrant lasts but, not withstanding, this warrant , in the absence of such
limitations, shall extend for one ear from the date of invoice.
Electromatic Equipment Co., Inc.
600 Oakland Ave. Cedarhurst, NY 11516 - USA
Tel: 1-800-645-7330 / Tel: 516-295-4300 / Fax: 516-295-4399
Ever precaution has been taken in the preparation of this manual. Electromatic assumes no
responsibilit for errors or omissions. Neither is an liabilit assumed for damages resulting
from the use of information contained herein. An brand or product names mentioned herein
are used for identification purposes onl , and are trademarks or registered trademarks of their
respective holders.

Electromatic Equipment Co., Inc.
-10-
8.0 TECHNICAL DATA
Resolution o the display 0,1% water content
0,5°C temperature
Measuring range 5 up to 50 % depending on the material
Operation temperature 0°C up to 40°C
Storage temperature -20°C up to 60°C
Temperature compensation Automaticall
Power supply 4 pcs. 1.5 Volt AA Alkaline batteries
(1800 measurements)
Auto Switch OFF After app. 6 minutes
Current consumption 55 mA (with light)
Display 128 x 64 matrix displa (lighted)
Dimensions 260 x 70 x 250 mm
Weight app. 1.3 kg (with batteries)
Degree o protection IP 40
Scope o supply Plastic case
Digital scale 1000g
Measuring cup 0.5 litre
4 x 1.5Volt AA Alkaline Batteries
Electromatic Equipment Co., Inc.
-3-
8. Fill up the measuring device with the sample material.
The filling needs to be done slowl and constantl to
ensure reproducible results.
9. The displa shows the measuring result.
10. To save the results in the stor menu press the
( button). The storage was successful when the
number in front of the s mbol increased. To reach
the store menu please press ( ) until the appears.
11. To name the saved results press the button.
12. Empt the humimeter and ensure that no grain rests
are accumulated in the measuring chamber.

Electromatic Equipment Co., Inc.
-4-
2.0 CALIBRATION CURVES
Name o Calibration
curve
Material
under test
Fill quantity
300g corn stand. corn standard 300g
300g corn hand corn hand reaped 300g
300g r e r e 300g
300g triticale triticales 300g
300g whea/spe.pe wheat / spelt peeled 300g
300g barle barle 300g
190g oats oats 190g
300g rape rape 300g
230g pumpkin pumpkin seed 230g
310g peas peas 310g
300g so bean so bean 300g
277g beetle beans beetle bean 277g
295g horse bean horse bean 295g
180g sunflower sunflower seed 180g
reference -------- --------
Reference: To test the humimeter FS2. Do not use for measuring the moisture!
Electromatic Equipment Co., Inc.
-9-
6.0 CHANGING BATTERIES
Your new device is provided with batteries.
Change the batteries:
1. Press with our finger onto the arrow of the batter
cap and pull it back.
2. Remove the empt batteries.
3. Put four new batteries in the device. Make sure the
position of the batter poles is correct.
4. Press down the batteries and close the cap.
7.0 DETERMINATION OF THE MATERIAL MOISTURE
The principle is a comparison measurement with the dr ing out method according to
DIN 10350: 1967 09. Take the measured sample and weigh it. Dr it out in an oven
and weigh it again. Following formula can be taken to determine the absolute
moisture:
Note: The dried material ma not be used an more for measuring purposes!
M
n
: Mass with average moisture content
M
t
: Mass of the dried sample
%F : Calculated absolute moisture
Exemption rom liability
For miss-readings and wrong measurements and of this resulting damage we refuse
an liabilit . This is a device for quick determination of moisture. The moisture
depends on multiple conditions and multiple materials. Therefore we recommend a
plausibilit check of the measuring results. Each device includes a serial number
and the guarantee stamp. If those are broken, no claims for guarantee can be
made. In case of a fault device, please contact Electromatic.
%F = x 100
M
n
- M
t
M
n

Electromatic Equipment Co., Inc.
-8-
4.0 MOST COMMON REASONS FOR MISS READINGS
• Product temperature out o application range
Material below 0°C resp. above 40°C (32 to 104 °F) ma cause fault
measurements. The storage of cold material in a warm storage area usuall
creates condensed water which ma lead to major measuring errors.
• Not adjusted material
Let our humimeter FS2 adjust to the surrounding temperature of the material for
approximatel half an hour.
A ver high temperature difference has a negative effect on the stabilit of the
measurement result.
• Wrong calibration curve
Before ou measure our sample, double check the correct selection of the
calibration curve.
• Wrong ill quantity
Fill exactl the right weight (± 1.0 g) in the measuring chamber.
• Wet or mouldy material
• Frozen measuring material.
5.0 DEVICE MAINTENANCE INSTRUCTIONS
To provide a long lives of our device please do not expose it to strong mechanical
loads or heat e.g. dropping it or direct sunlight exposure.
Clean our device using a dry cloth. The measuring chamber needs to be cleaned
with a dry and so t brush.
An kind of wet cleaning damages the device. The instrument is not rainproof. Keep
it in dr areas.
We recommend an interval ISO-conforming device check with the dr ing method.
Electromatic will also provide for a fee a test with a calibration certificate.
Electromatic Equipment Co., Inc.
-5-
3.0 DESIGN OF THE DEVICE
Batteries slot
(flipside)
Measuring
chamber
Filler assistance
Name of
calibration
Moisture content
POWER button

Electromatic Equipment Co., Inc.
-6-
3.1 Menu
There are 3 menu levels.
The s mbols show the ke functions in the different menus.
1. T pe Selection :
2. Store Menu :
3. Main menu:
Edit Logs
Manual Logs
Clear Logs
Print Logs
Last Log
All Logs
Clear
Send Logs
Manual Logs
Clear Logs
Options
Date / Time
Language
Unlock
°C / °F
Password
Reset
Materialcalib.
Status
Electromatic Equipment Co., Inc.
-7-
3.2 Keypad symbols
Measuring window:
: Rolling Menu
: Power ON / OFF
: Switch upper
: Switch lower
: Save
: Hold
: Watch the measuring rows
: Suppliers data can be added
Menu:
: Enter
: Switch upper
: Switch lower
: Exit
: Enter numbers
: Enter letters
: Next or right
: Left
: Yes
: No
: Shift
: OK

FS2
GRAIN MOISTURE METER
OPERATING INSTRUCTIONS
Distributed by: ABQ Industrial LP USA
Tel: +1 (281) 516-9292 / (888) 275-5772 eFax: +1 (866) 234-0451
Web: https://www.abqindustrial.net E-mail: [email protected]
Other ELECTROMATIC Measuring Instrument manuals

ELECTROMATIC
ELECTROMATIC CHECK LINE DTMB-200 User manual

ELECTROMATIC
ELECTROMATIC TI-007DL User manual

ELECTROMATIC
ELECTROMATIC PosiTector 6000 User manual

ELECTROMATIC
ELECTROMATIC Check-line DT-205LR User manual

ELECTROMATIC
ELECTROMATIC Check-Line RH1 Use and care manual

ELECTROMATIC
ELECTROMATIC STM-50 User manual

ELECTROMATIC
ELECTROMATIC CHECK-LINE MTG User manual

ELECTROMATIC
ELECTROMATIC TI-25M-MMX User manual