
Page 2 Version Error! Reference source not found.
General Information
On the basis of this calibration guide you are able to recalibrate the measuring instruments humimeter RH1, RH2, RH5 and
RH6 by using humidity standards and the calibration device. If the accuracy check performed before the recalibration
shows that the accuracy is still high enough, you do not need to recalibrate your device. Depending on your accuracy you
have two options to recalibrate or adjust your RHx. To reach the specified tolerance of ±1,5% in the range from 0% to
90%rh you should calibrate your instrument with the two point calibration. The second option is a one point recalibration.
An overview of proceeding the two point calibration.
Assemble the sword sensor and the calibration device with the 35% humidity standard
The time for conditioning is two hours.
Switch on the device and annotate the shown measuring value.
If the difference between the shown value and the calibration value is bigger than 2,0 % rh, we advice you to
recalibrate your RHx.
oNavigate to the menu Calibrate – Ic => V assume the active IC-Calibration value of the 35% and annotate this.
oDon´t push any key until the device switches off automatically after 4 minutes.
Assemble the sword sensor and the calibration device with the 80% humidity standard
The time for conditioning is two hours.
Switch on the device and annotate the shown measurement value.
If the difference between the shown value and the calibration value is bigger than 2,0%rh, we advice you to recalibrate
your RHx.
Navigate to the menu Calibrate – Ic => V assume the active IC-Calibration value of the 80% and annotate this.
Don´t push any key until the device switches off automatically after 4 minutes.
From the menu Calibrate – Ic => V enter the original IC- Values of the device to the MS-Excel-Calculating sheet at
the associated Voltages (V). Also fill in the values which are taken and annotated at 35% and 80% humidity.
At the menu Calibrate – Ic => V you have to enter the calculated new IC-values to the associated voltages
After entering the new values, leave the menu Ic => V by pressing Escape (Keys and ) to save the new
calibration.
The recalibration is finished now.
Details of proceeding the calibration
Prepare the calibration
To reach the best results, the calibration device, the humidity standard and the measuring instrument must have the same
temperature. This Temperature should be between 20°C and 25°C. The best way to manage this is, when you store all
three components together in a room without a thermal fluctuation over night.
Assemble the sensor-sword and the calibration device:
In this picture you see the single parts of the calibration device and a vial with
the humidity standard next to an RH5 and an RH1.