ENDA EI141 Series User manual

ENDA EI141 PROGRAMMABLE INDICATOR
Thank you for choosing ENDA EI141 Programmable Indicator.
SET
EI141
ENDA
PROGRAMMABLE INDICATOR
EI141-E-09-160217
Order Code : EI141 -
1 - Supply Voltage
230VAC...230V AC
24VAC.....24V AC
12VAC.....12V AC
SM...........9-30V DC /7-24V AC
1/3
SİSEL MÜHENDİSLİK ELEKTRONİK SAN. VE TİC. A.Ş.
Şerifali Mah. Y.Dudullu 34775
ÜMRANİYE/İSTANBUL-TURKEY
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr
Barbaros Cad. No:18 ENDA
TM
Please read this document carefully before using this product. The guarantee will be invalidated if the device is damaged by not
following instructions detailed in the manual. The company shall not be responsible for any damage or losses however caused,
which may be experienced as a result of the installation or use of this product.
- 35x77mm sized.
- 4 digits display.
- Easy to use by front panel keypad.
- Display scale can be adjusted between -1999 and 4000.
- Decimal point can be adjusted between 1. and 3. digits.
- Measurement unit can be displayed.
- Selectable four different standard input types (0-20mA, 4-20mA, 0-1V, 0-10V)
- User can calibrate the device according to specified input type.
- Sampling time can be adjusted in four steps.
- Stores maximum and minimum measurement values.
- The maximum or the minimum values can be hold on the display.
- Current and voltage calibration can be performed.
- Parameter access protection on 3 levels.
- Easy connection by removable screw terminal.
- CE marked according to European Norms.
TECHNICAL SPECIFICATIONS
ENVIRONMENTAL CONDITIONS
Height Max. 2000m
Ambient / Storage Temperature
Max. Relative Humidity
0 ... +50 °C/-25 ... +70°C (With no icing)
Rated Pollution Degree
80% Relative humidity for temperatures up to 31°C, decreasing linearly to 50% at 40°C.
According to EN 60529 Front panel : IP65
Rear panel : IP20
Do not use the device in locations subject to corrosive and flammable gases.
1
Supply
Power Consumption Max. 7VA
2.5mm² screw-terminal connections
Wiring
Date Retention
EMC
Safety Requirements
EN 61326-1: 2013
EEPROM (Min. 10 years)
EN 61010-1: 2010 P ( ollution degree 2, overvoltage category II, measurement category I)
ELECTRICAL CHARACTERISTICS
230V AC +10 -20 or 12/24V AC ±10 , 50/60Hz or optional .% % % 9-30V DC / 7-24V AC ±10% SMPS
EI141 Can not be used if measurement category II, III or IV is required.
While the current measuring mode, input impedance becomes 5Ω . Therefore, in current mode, the device must not be connected any
voltage input. Otherwise, the device is broken. While the device is running in the voltage measurement mode and if required to change
to current measurement mode, then firstly the voltage inputs must be removed and after that, input type must be changed to one of the
current measurement modes.
Input Type
0-1V DC voltage
0-10V DC voltage
0-20mA DC current
4-20mA DC current
Input Empedance
Approx. 11kΩ (terminal voltage limits: min. = -2V, max. = 30V)
Approx. Ω(terminal voltage limits: min. = -2V, max. = 30V) 11k
Approx. Ω 5 (applicable terminal voltage is max. 50mA.)
Approx. Ω(applicable terminal voltage is max. 50mA.) 5
0V
±0,5% (of full scale)
0V ±0,5% (of full scale)
0mA ±0,5% (of full scale)
0mA ±0,5% (of full scale)
1.1V
14V
25mA
25mA
Min. Max.
Measurement Range Measurement Accuracy
HOUSING
Housing Type Suitable for flush-panel mounting according to DIN 43 700.
Dimensions W77xH35xD71mm
Weight Approx. 250g (after packing)
Enclosure Material Self extinguishing plastics
While cleaning the device, solvents (thinner, benzine, acid etc.) or corrosive materials must not be used.

SET
EI141
ENDA
PROGRAMMABLE INDICATOR
INPUT
+
GND
1 2 3 4 5 6 78910 11 12
EI141-230VAC
PROGRAMMABLE INDICATOR
ENDA INDUSTRIAL ELECTRONICS
230V AC +10% - 20%
50/60Hz 7VA
SN: XXXXXXXXX
8 680407 703340
Made In Turkey
INPUT
+
GND
1 2 3 4 5 6 78910 11 12
EI141-24VAC
PROGRAMMABLE INDICATOR
ENDA INDUSTRIAL ELECTRONICS
24V AC ±10%
50/60Hz 7VA
SN: XXXXXXXXX
8 680407 703364
Made In Turkey
INPUT
+
GND
1234567 8 9 10 11 12
EI141-SM
PROGRAMMABLE INDICATOR
ENDA INDUSTRIAL ELECTRONICS
SN: XXXXXXXXX
8 680407 703388
Made In Turkey
9-30V DC / 7-24V AC
±10% 7VA
EI141-E-09-160217
2/3
SİSEL MÜHENDİSLİK ELEKTRONİK SAN. VE TİC. A.Ş.
Şerifali Mah. Y.Dudullu 34775
ÜMRANİYE/İSTANBUL-TURKEY
Tel : +90 216 499 46 64 Pbx. Fax : +90 216 365 74 01
url : www.enda.com.tr
Barbaros Cad. No:18 ENDA
TM
TERMS
1) Measurement value, measurement unit, the minimum or the maximum measured values are
displayed in the run mode.
Parameter name, parameter value or a user defined unit is displayed in the programming mode.
3) De
Used for making the minimum and the maximum measured values equal in the run mode.
crement or parameter selection key in the programming mode.
2) Increment or parameter selection key in the programming mode.
Used for displaying measurement unit or the max. measured value in the run mode.
4) Used for selecting run and programming modes, displaying
measurement unit or making the minimum and the maximum measured values equal.
adjusting parameters,
4 digits 7 segment yellow LED display
( 2 ),( 3 ),( 4 ),( 5 ) Keypad
( 1 ) Digital display
12.5mm
Micro switch
Character height
Panel cut-out
71,5mm
29,5mm
Note :
1) Panel thickness should be maximum 7 mm.
2) There must be at least 60mm free space
behind the device, otherwise it would be difficult to
remove it from the panel.
DIMENSIONS
Flush mounting
clamp
Panel
Depth
2
71mm 5mm
Flush mounting
clamp
1
Rubber
packing
SETSET
EI141
ENDA
PROGRAMMABLE INDICATOR
35mm
77mm
For removing mounting clamps ;
- Push flush mounting clamps in
direction 1 as shown in the figure below.
Then pull out the clamps in direction 2 .
INPUT
+
GND
1 2 3 4 5 6 7 8 9 10 11 12
EI141-230VAC
PROGRAMMABLE INDICATOR
ENDA INDUSTRIAL ELECTRONICS
230V AC +10% - 20%
50/60Hz 7VA
SN: XXXXXXXXX
8 68040 7 70334 0
Made In Turkey
Flush mounting
clamp
1
22
CONNECTION DIAGRAM
ENDA EI141 is intended for installation in control panels. Make sure that the device is used only for intended purpose.
grounded on the instrument side. During an installation, all of the cables that are connected to the device must be free of electrical power. The
device must be protected against inadmissible humidity, vibrations, severe soiling. Make sure that the operation temperature is not exceeded.
All input and output lines that are not connected to the supply network must be laid out as shielded and twisted cables. These cables should not
be close to the power cables or components. The installation and electrical connections must be carried out by a qualified staff and must be
according to the relevant locally applicable regulations.
The shielding must be
2
1
184-253V AC
50/60Hz 7VA
230V AC
Supply
Switch
Cable size: 1,5mm²
Fuse
F 100 mA
250V AC
Neutral
Line
SUPPLY :
NOTE :
Fuse should
be connected
1) Mains supply cords shall meet the requirements of IEC 60227 or IEC 60245.
2) In accordance with the safety regulations, the power supply switch shall bring the identification of the relevant instrument and it should
be easily accessible by the operator.
Note :
Holding screw
0.4-0.5Nm
Equipment is protected throughout
by DOUBLE INSULATION.

S.Cod.
d.C.sc.
Uo.sc.
D.Ca.s.
S.Cod.
a.CAL.
1.CA.
10.CA.
If key is pressed and held for 5
seconds programming mode is entered
U.opt. d.Cal. SECU.
d.pnt.
Cal.t.
l.sCl.
H.sCl.
l.inp.
H.inp.
CAL.t.=U.ınP.
Yes
No
d.Cnf.
i.typ.
ratE
Hold
dsp.C.
Unit
If first and then keys are
pressed together , measurement
unit appears. See Unit parameter
for programming.
If first and then keys are
pressed together, the maximum and the
minimum measurement values become
equal to the measured value at that time
and the message res. appears on
display.
Pressing this key the minimum
measured value up to that time
appears.
Pressing this key the maximum
measured value up to that time
appears
The minimum measured
value
The maximum
measured value
Measurement unit 870 1453
Bar.
&
Programming Mode
3/3
Running Mode
Measurement
value
1000
Res.
EI141-E-09-160217
SET
SET
SET
SET SET
SET
SET
CAL.t. = Calibration type.
Selectable as S.inP. or .
If S.inp. is selected, input type is one the
four standard input types. If U.inP is
selected, input types can be modified.
See NOTE 1 for modification.
U.inP
i.typ. = Input type.
Input type can be selected as 0-20mA,
4-20mA, 0-1V or 0-10V.
See NOTE 1 for programming.
rAte = Sampling rate.
Measurement is performed at each 200ms.
However,
for FASt rate, each measurement is displayed.
for SLo.1 rate, the average of 4 successive
measurements is displayed.
for SLo.2 rate, the average of 8 successive
measurements is displayed.
for SLo.3 rate, the average of 16 successive
measurements is displayed.
See NOTE 1 for programming.
Hold = Display holding parameter.
Selecting NonE , this parameter becomes
inactive.
Selecting Lo. , always the minimum
measured value is displayed.
Selecting Hi. , always the maximum
measured value is displayed.
dSP.C. = Display configuration.
Selectable as PrcS. or Pr.Un. . If PrcS. is
selected, process value appears. If Pr.Un.
is selected, process value and then
measurement unit are displayed 4 and 2
seconds successively.
See NOTE 1 for programming.
Unit = Measurement unit.
A constant, a message etc. to be displayed
can be entered. If a decimal point is desired,
it should be included before entering the
character.
For including decimal point first, then,
keys are pressed and held together. And
then, by using key decimal point can be
adjusted.
d.Pnt. = Decimal point.
Decimal point can be adjusted between
1. and 3. digits.
See NOTE 1 for programming.
L.SCL. = Lower limit for the scale.
It can be adjusted between -1999 and
(H.SCL. -100).
See NOTE 1 for programming.
H.SCL. = Upper limit for scale.
It can be adjusted between (L.SCL. +100)
and 4000.
See NOTE 1 for programming.
SET
At this state, the reference voltage or
current that corresponds to L.SCL. Parameter
is applied to the input.
To initialize calibration, first then
keys are pressed together and held until ‘CAL'
message appears. See NOTE 2.
SET
At this state, the reference voltage or
current that corresponds to H.SCL.
parameter is applied to the input.
To initialize calibration, first then
keys are pressed together and held until
‘CAL' message appears. See NOTE 2.
SET
SET
SET
s.Cod. = Access code for calibration menu.
This parameter should be 222.
See NOTE 1 for programming.
A.CAL. = Current calibration.
At this state, 20 mA current is applied to the
input of the device.
10.CA. = 10V
At this state, 10V is applied to the input of
the device.
input calibration.
1.CA. = 1V input calibration.
At this state, 1V is applied to the input of
the device.
For initializing calibration, first then
keys are pressed together and held until ‘CAL'
message appears. See NOTE 3.
To initialize calibration, first then
keys are pressed together and held until ‘CAL'
message appears. See NOTE 3.
To initialize calibration, first then
keys are pressed together and held until ‘CAL'
message appears. See NOTE 3.
s.Cod. = Access code for safety menu.
This parameter should be 333.
See NOTE 1 for programming.
d.C.Sc. = ConF. menu protection level
parameter.
nonE = No menu is seen.
P. no = Menu is seen but can not be programmed.
P.yES. = Menu is seen and programming is possible.
See NOTE 1 for programming.
Uo.Sc. = U.oPt. menu protection level
parameter.
nonE = No menu is seen.
P. no = Menu is seen but can not be programmed.
P.yES. = Menu is seen and programming is possible.
See NOTE 1 for programming.
d.Ca.S. = d.CAL. menu protection level
parameter.
nonE = No menu is seen.
P. no = Menu is seen but can not be programmed.
P.yES. = Menu is seen and programming is possible.
See NOTE 1 for programming.
CAL. CAL.
C.End C.End
s.Err.
S.Err.
C.Err.
C.Err.
The message on the left flashes approximately 5
seconds and calibration is completed.
The message on the left flashes approximately 5
seconds and calibration is completed.
If calibration is success, the C.End
message appears for 1 second.
If calibration is unsuccess, the S.Err
message appears for 1 second and
passed to the next step.
.
If the difference between the reference voltages or currents applied for the calibration of H.inP. and L.inP. is lower than one
half of the full scale, this error message appears on the display.
For example: Assume that the selected input type is 0-1V. In this case, if the difference between the reference voltages
applied for calibration of H.inP. and L.inP. is lower than 0.5V, this error message appears.
If the reference voltage or current applied to the input for calibration is too high or too low, this error message appears.
ERROR MESSAGES
CAL okey?CAL okey?
Yes Yes
No No
NOTE 2 NOTE 3
SET
For adjusting a selected parameter first press and hold key. Then, by using keys adjustment
can be made.
Parameter adjustment method
NOTE 1
If increment key is pressed and held 0.6 seconds, the value of the selected parameter changes rapidly. If
waited enough, the value increases 100 at each step. After 1 second following the release of the key, initial
condition is returned. The same procedure is valid for the decrement key.
Entering from programming mode to run mode:
If no key is pressed within 20 seconds during programming mode, data is stored automatically and the run mode is entered.
Alternatively the same function occurs by pressing key and holding for 5 seconds after pressing key and
entering main menu.
SET
If calibration is success, the C.End
message appears for 1 second.
If calibration is unsuccess, the .Err
message appears for 1 second and
passed to the next step.
C.
This manual suits for next models
6
Table of contents
Other ENDA Touch Panel manuals