RIA15
Endress+Hauser 9
Installation
Mounting location Panel housing
The device is designed for use in a panel.
Required panel cutout 45x92 mm (1.77x3.62 in)
Field housing
The field housing version is designed for use in the field. The unit is mounted directly on a wall, or
on a pipe with a diameter of up to 2 " with the aid of an optional mounting bracket. An optional
weather protection cover protects the device from the effects of weather conditions.
Orientation Panel housing
The orientation is horizontal.
Field housing
The device must be mounted in such a way that the cable entries point downwards.
Environment
Ambient temperature range –40 to 60 °C (–40 to 140 °F)
At temperatures below –25 °C (–13 °F) the readability of the display can no longer be
guaranteed.
Storage temperature –40 to 85 °C (–40 to 185 °F)
Climate class IEC 60654-1, Class B2
Altitude Up to 5 000 m (16 400 ft) above MSL in accordance with IEC61010-1
Degree of protection Panel housing
IP65 at front, IP20 at rear
Field housing
IP67, NEMA4x (aluminum housing)
Electromagnetic
compatibility
• Interference immunity:
As per IEC61326 (Industrial Environments) / NAMUR NE 21
Maximum measured error < 1 % of MR
• Interference emission:
As per IEC61326, Class B
Electrical safety Class III, overvoltage protection category II, pollution degree 2