OperationalMatrixfortheModelDO600
Function/
Resulting
Action
Power
Status
Mode
Setting
RequiredButtonPressSequenceomments
On/OffAnyAnyMomentarypressoftheON/OFFbutton
WaterSaturated
AirCalibration
OnAnyPlaceElectrodeinCalibrationCap
Press&holdCAL/ ECALLbuttonfor2seconds
ZeroCalibrationOnAny
PlaceElectrodeinZeroSolution,waitforstabilityPress
&holdCAL/ ECALLbuttonfor2seconds
Alsoworkswithprobe
removed(Calatzerocurrent)
StoreeadingOnAnyMomentarypressoftheMODE/HOLDbuttonStoresandholdsreading
“HOLD”displayed
Holdelease
On
WhileIn
Hold
Mode
MomentarypressoftheMODE/HOLDbutton
EnterMemory
etrievalOnAny
MomentarypressoftheCalbuttonfollowedby
amomentarypressoftheMODE/HOLDbutton
(w/in4seconds)
Ifnodataisstoredin
memory,“End”isdisplayed
briefly,andthenreturnsto
lastmode.
ScrollStored
eadingsOnMemory
ecallMomentarypressoftheMODE/HOLDbutton
ExitMemory
etrievalOnMemory
ecallMomentarypressoftheCAL/ ECALL
ClearStored
MemoryOn
Any
Measure‐
mentMode
PressandholdtheON/OFFbuttonfor4seconds“clr”isdisplayed.
Change
Measurement
Mode
OnAny
PressandholdtheMODE/HOLDbuttonforat
leasttwoseconds(themodeswillscrolluntil
thebuttonisreleased)
EnterSalinity
CompensationOnAnyPressandreleasetheCAL/ ECALLbuttontwice
(momentary)insuccession(DisplaysSAL)
Changing
Salinity
CompensationOnSAL
MomentarypressoftheMODE/HOLDbutton
(eachbuttonpressincreasestheratioby1ppt
(partperthousand),thevaluecyclesfrom0to
50)
ExitSalinity
CompensationOnSAL
PressandreleaseCAL/ ECALLbuttonfor2
secondstoenterAltitudeCompensation,Or
PressandreleaseCalButtononemoretimeto
entermeasurementmode
Itisnecessarytopressthe
CAL/ ECALLbuttontosave
changes.Iftheunittimesout,
nochangesaresaved.
EnterAltitude
CompensationOnAnyor
SAL
PressCAL/ ECALLtwice(momentary).Unit
entersSalinityMode.PressCAL/ ECALLfor2
secondstoenterAltitudeCompensationMode.
(DisplaysAld)
Unittimesoutin5
secondsifnobuttonis
pressed,revertsto
previousmode.
Changing
Altitude
Compensation
OnAld
MomentarypressoftheMODE/HOLD
button(eachbutton‐pressincreasesaltitudeby
1,000ft,thevaluecyclesfrom0to20)
ExitAltitude
CompensationOnAldMomentarypressCAL/ ECALLbuttontoexit
andsavechanges.
Itisnecessarytopressthe
CAL/ ECALLbuttontosave
changes.Iftheunittimesout,
nochangesaresaved.
Change
Temperature
UnitsOffn/a(off
mode)
PressandholdtheCAL/ ECALLbuttonthen
momentarilypresstheON/OFFbutton. elease
theCAL/ ECALLbuttonafterthe"SELFCAL"
lights
Override
AutopowerOffOnAny
PressCAL/ ECALLbutton(momentary)then
simultaneouslypressandholdtheMODE/HOLD
andON/OFFbuttonsforfor2seconds.
Defaulteset
Off
n/a(off
mode)
SimultaneouslypressON/OFF,CAL/ ECALLand
MODE/HOLDmomentarily.“dFLt”willbe
displayed.