6HD780‐en‐GB_V1.09/16
ViewingMAX‐MIN‐AVG
Foragivenmeasurementsession,thismetercanrecordthehighest(MAX)andthelowest(MIN)
readingsforlaterrecall.
1. PresstheMAX‐MINbuttontoviewthecurrentMAXreadings(MAXiconappears).The
readingsonthedisplayarenowthehighestreadingslogged.
2. PresstheMAX‐MINbuttonagaintoviewthecurrentMINreadings(MINiconappears).The
readingsonthedisplayarenowthelowestreadingslogged.
3. PressMAX‐MINbuttononcemoretoviewthecurrentAVGreadings(AVGiconappears).
Thereadingsonthedisplayarenowthelowestreadingslogged.
ToexittheMAX‐MINmode,pressandholdtheMAX‐MINbuttonfor3seconds.TheMAX‐MIN‐
AVGiconswillswitchoffandthemeterwillreturntothenormaloperatingmode.
ZeroAdjust(Pressure)
TheunitmustbekeptintheuprightpositionwhenZEROINGorwhentakingmeasurements.To
ZEROthemeter:Beforeuseandwithoutfittingsattachedtothemeter,pressandholdtheH
buttonfor3secondsoruntilthedisplayshowsallzeroes.
SetupMode
Pressandholdthe/SETbuttonforthreesecondstoentertheSetupMode.Momentarily
pressingtheorbuttonswillcyclethroughoutthefunctions.Press toselectthedesired
function.SetupisdisabledinMIN/MAXmode.Theyare(inorder):
OFFSET………StoreoffsetsforT1andT2
SLP……Sleepmode
Note:PressandholdtheSETbuttonforthreesecondswhenintheSetModetoexit.
Offset
Usetheoffsetoptiontocompensateforaspecificthermocoupletemperatureerroror
mismatch.Theallowableadjustmentrangeis±5.0°C(9.0°F).
1. PlugthethermocoupleintotheT1orT2inputconnectorlocatedoneithersideofthemeter.
2. Placethethermocoupleinaknown,stableenvironment.
3. Allowthereadingstostabilize.Theprimarydisplayshowsthetemperature,whilethe
bottomdisplayshowstheoffsetvalue.
4. Usethe▲and▼buttonstoadjusttheoffsetuntiltheprimaryreadingmatchesthe
calibrationtemperature.Pressbuttontostorethevalueandmovetothenextsetup
modeoption.
SleepMode(AutoPowerOFF)
Withsleepmodeactive,themeterautomaticallypowersoffafter20minutesofinactivity.To
defeatthisfeature,whileinSLPmode,presstoindicateON(sleepmodeON)orOFF(sleep
modeOFF).Pressorbuttonstomakeaselection.Press tostorethenewsettingin
memory.