Non-contact voltage
With the red NCV tab on the tip of the meter close
to an AC voltage, press and hold the NCV button.
The NCV LED will light and the beeper will beep. The
closer you get to AC voltage, the louder the beep.
The NCV function is sensitive enough to detect
24VAC on thermostats.
Hi voltage indicator
In any VAC/VDC range, when you touch a voltage
greater than 30V, the beeper will beep and the red Hi-
V LED will blink. BE CAREFUL!
Microamps
For measuring the flame diode current in a heater
control.
Capacitance
For motor-start and motor-run capacitors.
Disconnect the capacitor from power first. Short the
terminals to discharge the capacitors. Disconnect
any resistors that might be between the terminals of
the capacitor.
MIN/MAX
Press MIN/MAX once to begin recording MIN and
MAX. Press MIN/MAX to select current reading’s MIN
or MAX. Hold down for 2 seconds to exit MIN/MAX
function.
Temperature
Plug any K-type thermocouple directly into the
meter to measure temperature. Temperature meas-
urement will be accurate even in fast changing envi-
ronments because of excellent temperature compen-
sation. One thermocouple is included. No adapter is
required.
Backlight (HS36)
Press the button to activate the backlight for
approximately 60 seconds.
True RMS (HS36)
Digital multimeters use two different types of AC
sensing. The most common is average sensing, nor-
malized to a true RMS value of a sine wave. The
other is true RMS sensing. The actual true RMS
value is sensed for a wave form within the limits of
the crest factor. Either sensing method will give the
same results on a clean sine wave but they may dif-
fer on a non-sinusoidal waveform.
Field
q
F calibration
For accuracies of r1qF, calibrate to a known tem-
perature. A glass of stabilized ice water is very close
to 32qF (0°C) and is usually very convenient but any
known temperature can be used.
1. Select the 400qF range.
2. Remove back case and hold the battery in place
with a rubber band so terminals are touching.
3. Stabilize a large cup of ice water.
4. Immerse the thermocouple probe and let it stabi-
lize.
5. Adjust VR3 (lower right corner of PCB) to get
close to 32qF (0°C) then adjust VR2 (left of VR3)
to get within 0.1qF (0.05°C) of 32qF (0°C).
6. To calibrate in qC, close the jumper that is to the
left of VR3.
Disable auto off
Set to OFF position, press and hold RNG (HS35)
or MIN/MAX (HS36) button while turning rotary dial
to desired range position. Release the button when
LCD displays normally. Note: “APO” annunciator will
be missing from the display. The Auto Power Off
mode is on when “APO” indicated on the display.
Attach to Fieldpiece accessory head
Connect your Fieldpiece accessory head directly
to the top of HS series and switch to range indicated
by head. Visit www.fieldpiece.com for more info.
Symbols used:
Caution, risk of electric shock
Caution, refer to manual.
Ground
Double insulation
Using & storing test leads
Because the wire insulation is silicone the leads
will stay flexible in cold weather and will not melt if
bumped by a soldering iron.
Disconnect top half of test lead and plug tip direct-
ly into meter to make voltage testing easy. Use with
included alligator clip (ASA2) as shown for even eas-
ier operation.
For convenient lead storage, wrap the leads as
shown. Pull leads around front between overhanging
tips, twist, and pull over one of the lead plugs.
OPERATOR’S MANUAL
For your safety...
General: Disconnect the test leads before opening
the case. Inspect the test leads for damage to the
insulation or exposed metal. Replace if suspect.
Never ground yourself when taking electrical meas-
urements. Do not touch exposed metal pipes, outlets,
fixtures, etc., which might be at ground potential.
Keep your body isolated from ground by using dry
clothing, rubber shoes, rubber mats, or any approved
insulating material. When disconnecting from a cir-
cuit, disconnect the "RED" lead first, then the com-
mon lead. Work with others. Use one hand for test-
ing. Turn off power to the circuit under test before cut-
ting, unsoldering, or breaking the circuit. Keep your
fingers behind the finger guards on the probes. Do
not measure resistance when circuit is powered. Do
not apply more than rated voltage between input and
ground.
All voltage tests: All voltage ranges will withstand
up to 600V. Do not apply more than 600VDC or
600VAC.
AC tests: Disconnect the meter from the circuit
before turning any inductor off, including motors,
transformers, and solenoids. High voltage transients
can damage the meter beyond repair. Do not use
during electrical storms.
Maintenance
Clean the exterior with clean dry cloth. Do not use
liquid.
Battery replacement: When the multimeter dis-
plays " " the battery must be replaced. Disconnect
and unplug leads, turn meter off, and remove the bat-
tery cover. Replace the battery with a NEDA type
1604 9V battery.
Limited warranty
This meter is warranted against defects in materi-
al or workmanship for one year from date of pur-
chase. Fieldpiece will replace or repair the defective
unit, at its option, subject to verification of the defect.
This warranty does not apply to defects resulting
from abuse, neglect, accident, unauthorized repair,
alteration, or unreasonable use of the instrument.
Any implied warranties arising from the sale of a
Fieldpiece product, including but not limited to
implied warranties of merchantability and fitness for a
particular purpose, are limited to the above.
Fieldpiece shall not be liable for loss of use of the
instrument or other incidental or consequential dam-
ages, expenses, or economic loss, or for any claim of
such damage, expenses, or economic loss.
State laws vary. The above limitations or exclu-
sions may not apply to you.
Service
Return any defective HS35/36 to Fieldpiece for
warranty service along with proof of purchase.
Contact Fieldpiece for out of warranty repair charges.
www.fieldpiece.com
!
AUTO-RANGING
DIGITAL STICK
METERS:
HS35
HS36
Function Legend
Continuity
VDC Volts DC
MFD Capacitance
°C/°F Temperature
ȝADC Microamps DC
:
:
Resistance
Hz Frequency
VAC Volts AC
v08
SPECIFICATIONS
Display: 3¾ digit liquid crystal display (LCD) with a
maximum reading of 3999.
Analog bar graph: 41 segments with measure-
ments 20 times per second.
Range (RNG): Manual ranging mode (HS35)
Overrange: "OL" mark indication.
Auto power off: 30 minutes.
Operating environment: 32 to 122qF (0 to 50°C) at
<70% R.H.
Storage temperature: -4 to 140qF (-20 to 60°C), 0
to 80% R.H. with battery removed.
Accuracy: Specifications good in ambient condi-
tions of 73qF r9qF (23°C r5°C), <75% relative
humidity.
Temperature coefficient: 0.1u(specified accuracy)
per qF/°C. (32 to 64qF (0 to 18°C), 82 to 122qF(28
to 50°C)).
Power: Single standard 9-volt battery, NEDA 1604,
JIS 006P, IEC 6F22.
Battery life: 300 hours typical with alkaline.
Accessories: One pair test leads, one pair alligator
clips, k-type thermocouple, 9V battery (installed),
and operating instructions.
Safety: UL,CE,Cat III600V,UL61010-1,IEC/EN61010-1.
Temperature
Range: -30 to 1000qF (-34 to 538°C)
Resolution: 0.1qF/°C
Accuracy: r1qF, 32 to 120qF (0 to 49°C),
r1% + 1.5qF, -4 to 750qF (-20 to 399°C),
r2% + 4qF, -30 to -4qF (-34 to -20°C),
r2% + 4qF, 750 to 1000qF (399 to 538°C)
Sensor type: K-type thermocouple
Overload protection: 60 VDC or 30 VAC rms
Continuity
Audible indication: Less than 40:
Response time: 500ms
Green LED will be on continuously.
Overload protection: 500VDC or AC rms
DC volts
Ranges: 400mV, 4000mV, 40V, 400V, 600V
Resolution: 0.1mV
Accuracy: r(0.5% rdg + 2 dgt)
Input impedance: 3M:on V inputs, 10M:on mV
input
Overload protection: 600VDC or AC rms
Transient protection: 6kV for 10Psec
AC volts (50Hz - 500Hz)
Ranges: 400mV, 4V, 40V, 400V, 600V
Conversion: True RMS (HS36), average (HS35)
Crest factor: Less than or equal to 3 (HS36)
Resolution: 0.1mV
Accuracy:
r(1.2% rdg + 5 dgts) 50Hz ~ 60Hz on 400mV range
r(1.5% rdg + 5 dgts)
r(2% rdg + 5 dgts) on 600V range
Input impedance: 3M:
Overload protection: 600V DC or AC rms
Transient protection: 6kV for 10ȝsec
DC current
Ranges: 400ȝA, 4000ȝA
Resolution: 0.1ȝA
Accuracy: r(1.0% rdg + 2 dgt)
Voltage burden: 1V, (8V on 4000ȝA range)
Overload protection: 500V DC or AC rms
Resistance
Ranges: 400:, 4k:, 40k:, 400k:, 4M:, 40M:
Resolution: 0.1:
Accuracy:
r(1.0% rdg + 4 dgts) on 400:to 400k:ranges
r(1.5% rdg + 4 dgts) on 4M:range
r(3.0% rdg + 5 dgts) on 40M:range
Open circuit volts: -0.45VDC typical, (-1.2VDC on
400:range)
Overload protection: 500VDC or AC rms
Capacitance
Range: 400ȝF
Resolution: 0.1ȝF
Accuracy: r(3% rdg + 5 dgts)
Overload protection: 500V DC or AC rms
Frequency
Ranges: 4k, 40k, 400k, 4M, 40MHz
Resolution: 1Hz
Accuracy: r(0.1% rdg + 3 dgts)
Sensitivity: 10Hz ~ 4MHz: > 1Vrms, 4MHz ~
40MHz: > 2Vrms, <5Vrms
Minimum pulse width: > 25ns
Duty cycle limits: > 30% and < 70%
Overload protection: 500VDC or AC rms
Voltage
Frequency
DC Current < 4000
P
ADC
Selecting Ranges
For DC voltage, set the meter to the VDC parame-
ter instead of VAC as shown above.
For all ranges choose a range just above the value
you expect. If display reads “OL” (overload), select a
higher range. If display shows less than three num-
bers, select a lower range for better resolution.
Resistance
Capacitance
Temperature
Works with Fieldpiece accessory
heads!
Connect to Fieldpiece accessory heads by simply
attaching them to the top of meter (1) or attach
remotely through leads (2). For most heads, move
dial to range shown (1). For the AAC clamp (ACH4),
move dial to VAC range (2).
Disconnect test
leads from volt-
age before plug-
ging in thermo-
couple!
Insure the tem-
perature being
measured is stable.
Maintain good con-
tact between the
thermocouple and
what's being meas-
ured.
CAT.III
300V
400A
CLAMP
ACH4
AC Current
Clamp
1AAC / 1mVAC
400AAC MAX
!
(1) (2)
WARNINGS
DISCONNECT AND UNPLUG TEST LEADS
before opening case.
TEST NCV FUNCTION ON KNOWN LIVE
WIRE before using.
DO NOT APPLYVOLTAGE greater than
30VAC or 60VDC to the thermocouple or the
jacks when the rotary dial is on OF.
REMOVE THE THERMOCOUPLE when tak-
ing voltage measurements.
DISCONNECTTHE TEST LEADS when taking
temperature measurements.
DO NOT APPLY VOLTAGE TO THE JACKS
when the rotary dial is on microamps. Even low
voltages can cause a current overload and blow
the fuse. Replace blown fuse to regain function.
!
!

This manual suits for next models

1

Other Fieldpiece Measuring Instrument manuals

Fieldpiece SMAN SM480VINT User manual

Fieldpiece

Fieldpiece SMAN SM480VINT User manual

Fieldpiece SC680 User manual

Fieldpiece

Fieldpiece SC680 User manual

Fieldpiece SMAN SM380VINT User manual

Fieldpiece

Fieldpiece SMAN SM380VINT User manual

Fieldpiece ATA1 User manual

Fieldpiece

Fieldpiece ATA1 User manual

Fieldpiece CAT45K2 User manual

Fieldpiece

Fieldpiece CAT45K2 User manual

Fieldpiece SVG3 User manual

Fieldpiece

Fieldpiece SVG3 User manual

Fieldpiece HS32 User manual

Fieldpiece

Fieldpiece HS32 User manual

Fieldpiece SOX3 User manual

Fieldpiece

Fieldpiece SOX3 User manual

Fieldpiece ADMN2 User manual

Fieldpiece

Fieldpiece ADMN2 User manual

Fieldpiece SOX3 User manual

Fieldpiece

Fieldpiece SOX3 User manual

Fieldpiece SC55 User manual

Fieldpiece

Fieldpiece SC55 User manual

Fieldpiece SMAN4 User manual

Fieldpiece

Fieldpiece SMAN4 User manual

Fieldpiece SCM3 User manual

Fieldpiece

Fieldpiece SCM3 User manual

Fieldpiece SMG5 User manual

Fieldpiece

Fieldpiece SMG5 User manual

Fieldpiece JL3MN User manual

Fieldpiece

Fieldpiece JL3MN User manual

Fieldpiece FPSC680 User manual

Fieldpiece

Fieldpiece FPSC680 User manual

Fieldpiece STA2 User manual

Fieldpiece

Fieldpiece STA2 User manual

Fieldpiece SC54 User manual

Fieldpiece

Fieldpiece SC54 User manual

Fieldpiece SVG3 User manual

Fieldpiece

Fieldpiece SVG3 User manual

Fieldpiece JL3PR User manual

Fieldpiece

Fieldpiece JL3PR User manual

Fieldpiece ACH User manual

Fieldpiece

Fieldpiece ACH User manual

Fieldpiece SRL8 User manual

Fieldpiece

Fieldpiece SRL8 User manual

Fieldpiece SC680 User manual

Fieldpiece

Fieldpiece SC680 User manual

Fieldpiece SC57 User manual

Fieldpiece

Fieldpiece SC57 User manual

Popular Measuring Instrument manuals by other brands

Hanna Instruments HI 718 quick start guide

Hanna Instruments

Hanna Instruments HI 718 quick start guide

PCB Piezotronics IMI Sensors 621C40 Installation and operating manual

PCB Piezotronics

PCB Piezotronics IMI Sensors 621C40 Installation and operating manual

ICS Schneider Messtechnik HYDROTECHNIK MultiHandy 2025 operating instructions

ICS Schneider Messtechnik

ICS Schneider Messtechnik HYDROTECHNIK MultiHandy 2025 operating instructions

Moeller ps4-201-mm1 installation instructions

Moeller

Moeller ps4-201-mm1 installation instructions

MARTINDALE TEK100 instruction manual

MARTINDALE

MARTINDALE TEK100 instruction manual

Inficon 3000 Micro GC operating manual

Inficon

Inficon 3000 Micro GC operating manual

Nikon 50i instruction manual

Nikon

Nikon 50i instruction manual

SMA SUNNY SENSORBOX installation guide

SMA

SMA SUNNY SENSORBOX installation guide

CT Lab Vecto III user manual

CT Lab

CT Lab Vecto III user manual

Keysight PNA-X N5244A Service guide

Keysight

Keysight PNA-X N5244A Service guide

Hach PHC101 user manual

Hach

Hach PHC101 user manual

Satec PM130 series reference guide

Satec

Satec PM130 series reference guide

Young 27106T instructions

Young

Young 27106T instructions

CET PMC-680i user manual

CET

CET PMC-680i user manual

JPT YDFLP-E-80&100-M7-M-R user manual

JPT

JPT YDFLP-E-80&100-M7-M-R user manual

Kett PM-650 operating manual

Kett

Kett PM-650 operating manual

Knick Stratos Pro A4..MSPH series instruction manual

Knick

Knick Stratos Pro A4..MSPH series instruction manual

Test Products International Vib Meter 9085 user guide

Test Products International

Test Products International Vib Meter 9085 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.