Fixturlaser SMC User manual

USER MANUAL Fixturlaser SMC


USER MANUAL FIXTURLASER SMC, edition 1.3 (November 2017)
CONTENT
General presentation
1.1
General settings
2.1
Supervisor module
3.1
Balancing module
4.1
Maintenance
5.1
Technical specification
6.1


1.1
1 GENERAL PRESENTATION
1.1 INTRODUCTION
We want to congratulate you on your choice and hope that you will be fully satisfied with it. For this
reason, we recommend that you read carefully the present user guide and more specifically the safety
instructions.
Designed to meet industrial requirements for enhancing productivity, FIXTURLASER SMC is a portable
multichannel instrument coming along with a large touchscreen, a 3-axis wireless sensor and numerous
accessories. This innovative all-in-one solution makes vibration analysis available to all users.
FIXTURLASER SMC is available in 2 versions. The FIXTURLASER SMC standard (red rubber) is for
safe area, and the FIXTURLASER SMC-EX (black rubber) for ATEX Zone II area. Always refer to the
product marking to know which version you are using. All functionalities described in this manual are
applicable to both versions.
1.2 SAFETY INSTRUCTIONS
The safety instructions delivered with the instrument (printed and on the USB stick) should be carefully
followed and the instrument should always be used within the limits specified here.
Instrument and operator safety is at risk when the instrument is used in conditions that are not
intended by ACOEM.
Dismantling the instrument for an internal operation is forbidden. The only parts for which dismantling is
allowed are the battery hatch, the battery and the hatch providing access to the connectors.
All the spare parts must be provided by ACOEM.

1.2
List of symbols and warning on the instrument
Warning: Whenever this symbol is present on the device, refer to the safety instructions and user
manuals.
The table below lists the warning signs and security present on the instrument.
Symbol
Signification
Position
Laser radiation
Refer to section "Laser Class 3R"
Rear side (bottom)
Laser radiation class 3R
Refer to section "Laser Class 3R"
Rear side (bottom)
Avoid exposure to beam
Refer to section "Laser Class 3R"
Rear side (bottom)
Light Hazard Group 2
Refer to section "High-power white LED
flash and stroboscope"
Rear side (bottom)
Refer to the safety instructions
documentation and the user manual
before connecting
Rear side (bottom)
Laser aperture
Refer to section "Laser Class 3R"
Rear side (top)
Battery replacement
Refer to section "Battery pack set-up"
Battery compartment
CE certification data. Refer to the copy of
the CE APT2069 type certificate
Rear side (center)
Do not dispose of this product as unsorted
household waste.
Refer to section "Dismantling/recycling"
Rear side (center)
FCC and IC number. Refer to section
"Radio"
Rear side (center)
External power connection: refer to
section "Main Power supply block"
Rear side (bottom)
List of symbols and warning on WLS
Warning: Whenever this symbol is present on the device, refer to the safety instructions and user
manual.
The table below lists the warning signs and security present on the WLS sensor.
Symbol
Signification
Warning: Refer to the safety instructions
and user manuals
Use a power source with limited power
(=> Mains Power supply bloc)
=> CE certification data
=> Dismantling/recycling

1.3
Laser
Laser maintenance: The laser does not need maintenance or adjustment excluding cleaning the glass
with a cotton swab. Always shut off completely the device FIXTURLASER SMC before this operation.
Caution: Use of controls or adjustments or performance of procedures other than those specified herein
may result in hazardous radiation exposure.
Electric connections
All external circuits connected to the instrument must be non-hazardous voltage sources and be energy
limited as explained in sections 6.3 and 9.4 of the IEC61010-1 standard.
Do not exceed maximum input voltage on the A/B/C/D connectors: Maximum input voltage ±24 V DC,
±24 V AC peak.
All external circuits connected to the instrument must carry non-dangerous voltage as defined in
Standard IEC61010-1 (Paragraph 6.6)
Do not exceed maximum input voltage and current intensity for the power supply delivered by the mains
block (see chapter “Mains power block” of the safety instructions).
Use the cables designed for the different types of measurement.
If possible, protect the unused connectors using the plastic caps provided.

1.4
1.3 FIRST POWER UP
The instrument turns on automatically a few seconds after being connected to the mains through the
power supply module. If the battery level is too low, the charge starts and goes on as long as the
instrument is connected to its power supply.
When using a new battery, leave the battery in charge for about 10 hours in order to achieve full charge.
Do not use the instrument prior to 2-3 hours of charge. Usual charging time is about 6 hours when
instrument is switched off.
On-Off
•Power on: Press the On-Off key
•Power off: Press the On-Off key, then message “Shutdown”
Battery charge
•Power off: Press the On-Off key, then message “Shutdown”
•Connect the charger on the main. The instrument is automatically powered on. During the charge,
you can continue to use it. For a faster charge, it is recommended to switch-off the instrument. A full
charge requires about 6 hours.
It is recommended to disconnect the charger from the main when you are not using it.

1.5
1.4 USER INTERFACE
FIXTURLASER SMC starts on its “Home screen”. It is a touch screen and a simple pressure on it gives
access to the application modules installed on the instrument and to 2 side panels:
For more details:
•Supervisor module: See CHAPTER 3
•Balancing module: See CHAPTER 4
•Status panel: See § 1.9
•Shortcuts panel: See § 1.10
The 2 side panels are accessible from any screen.
Access to
application
modules
Access to
status side
panel
Access to
shortcuts side
panel

1.6
1.5 CONNECTIONS
Connectors A to D on the top of the instrument
Those connectors can be used in industrial environment. They are IP65
Connector A
Channels 1. Use this connector when the instrument is set in single-channel mode.
•Contact A:
•Contact B: Common
•Contact C: Channel 1 signal input
•Contact D:
•Contact E: Channel 3 signal input
•Contact F: -24 Vdc output
Connector B
Channels 2, 3 and 4. Use this connector with wired triaxial sensor when the instrument is set in multi-
channel mode.
•Contact A:
•Contact B: Common
•Contact C: Channel 2 signal input
•Contact D: Channel 3 signal input
•Contact E: Channel 4 signal input
•Contact F: -24 Vdc output
Notes
For 2-channel measurement, use connectors A and B
If inputs are set as IEPE, the sensor is powered by a 4 mA constant current. The integrity of the sensor
is monitored and a warning is issued if peak voltage is out of about 1 to 22 V.
A
B
C
D
Antenna for WLS
sensor and Wi-Fi

1.7
Connector C
Tachometer input, Stroboscope output, Power supply input. It it is marked in yellow as well as all cables
used on this connector
•Contact A: Common
•Contact B: Instrument power input
•Contact C: 5 Vdc output
•Contact D: -24 Vdc output
•Contact E: Tacho signal input
•Contact F: Strobe signal output
Connector D
Ethernet, Microphone input, Audio output.
•Contact A: Common audio
•Contact B: Audio output
•Contact C: Microphone input
•Contact D: Lan_Tx+
•Contact E: Lan_Tx-
•Contact F: Lan_Rx-
•Contact G: Lan_Rx+
Warning
•Do not connect channels 1, 2, 3, 4 and tachometer input on a not buffered output or in parallel to
other instrument as their impedance is not maintained when changing configuration or when the
instrument is switched off.
•Respect the maximum input voltage to the A/B/C/D connectors: maximum input voltage ±24 VDC,
±24 volts peak AC.

1.8
Connectors E to I behind the trapdoor
Those connectors are behind a trapdoor. It must remain close in industrial environment to preserve the
IP65 protection. Use those connectors only in office environment.
Connector E:
Power supply input
Connector F:
Ethernet RJ45
Connector G:
USB 2 type A host (for USB memory stick)
Connector H:
USB 2 type B device (direct connection with PC)
Connector I:
Serial interface for maintenance purposes only
E
F
G
H

1.9
1.6 BUILT-IN SENSORS
The back of the instrument gives access to the built-in sensors.
For the stroboscope and the pyrometer, please see the safety instructions delivered with the instrument
(printed and on the USB STICK)
High-power LED for
stroboscope and
camera flash
Laser beam source
for pyrometer sighting
Pyrometer window
Camera window

1.10
1.7 WIRELESS SENSOR
FIXTURLASER SMC can be used with a wireless triaxial accelerometer (WLS
Sensor). This chapter described how to use FIXTURLASER SMC with this sensor.
WLS sensor battery
Use the USB cable and charger supplied with the sensor. The connector is protected
by the rubber cap on the top of the sensor. It can also be charged from any USB
interface with power supply (! measurement is not possible if the WLS is connected to
a USB PC port). Charging time is 8 hours with the standard 500 mA charge current.
Notes
•After recharging, the cap must be correctly and fully inserted.
•The sensor should not be used without its cap in wet or polluted environment.
•Opening of the cap should only be done in clean and dry environment.
•If the sensor has been in polluted environment, the cap opening must be done after cleaning and
drying, cap facing down to prevent the intrusion of foreign matter.
•WLS sensor battery replacement must be done by qualified personnel only.
WLS sensor first connection
1. Switch on WLS: press on/off for 2s, until the red LED is on, and wait until the red LED is off and the
blue LED is continuously on. See details below concerning LED indication.
2. On FIXTURLASER SMC from the home screen proceed as follows:
oCheck first if Wi-Fi is enabled in:
Shortcuts > Setting > > Wi-Fi = enabled
oSet FIXTURLASER SMC to work with a WLS sensor:
Shortcuts > Setting > > Accelerometer link = Wireless
oShortcuts > Setting > Wireless sensor:
▪Input WLS serial number (e.g., 10015)
▪Save the settings
▪Wait for about 30 s until the blue LED is flashing. Then 20 s later, the connection
must be set up. It can be seen when there is in the status summary or from the
status panel where you can see the sensor identification (e.g., WLS_10015). It can
also be checked from Wireless sensor setting with the Test function.
If Wi-Fi is enabled further connections are automatic after switch on of the instrument and the WLS
sensor.
Note: It is not possible to connect an instrument with a WLS sensor already connected to another
instrument.

1.11
WLS sensor switch off
Press on/off for 6s until the red LED flashes on. The sensor is also automatically switched off if there is
no connectivity during 10 min.
WLS LED indication
Red LED
Blue LED
Significance
WLS connected to the
charger
———
Charge in progress (LED brightness =
50%)
━━━━
Charge completed (LED brightness =
100%)
- - - - -
Charge error
Switch on
(press 2s on/off)
——— (< 10 s)
Off
Start in progress
——— (> 15 s)
Off
Error
———
WLS ready and not connected to the
instrument
- - - - -
WLS has detected the instrument. The
connection is only effective when you
have in FIXTURLASER SMC status
summary
Switch off
(press 7 s on/off)
- - - - -
Stop in progress
Off
Off
WLS is switched off
Low battery
- - - - -
Notation:
━━━━ brightness = 100%
——— brightness = 50%
- - - - - flashing
Blank any status

1.12
1.8 CONNECTION WITH PC
FIXTURLASER SMC can exchange data with the PC when the Communication module is run from
the home screen. See below the different settings according to the type of communication. Once the
communication is established the instrument is seen as an external drive from the PC (see § 6.1).
Using USB (Connector H)
No settings are required, the PC automatically detects the instrument after its connection.
Using USB memory (Connector G)
It is possible to use a USB memory stick connected on port G to exchange the following data:
•Firmware update: see § 2.11
•Issue balancing reports: see § 4.13
•Issue diagnostic reports: see § 3.7
•Copy screenshots taken with the device: see § 1.10
Note: USB memory stick format must be FAT32, NTFS format is not accepted. It is always possible to
reformat it to FAT32.
Using Ethernet (Connector D or F)
•Direct connection PC- FIXTURLASER SMC setting:
oOn PC: Setup Network
▪DHCP = No
▪Set the IP address: Ex 192.168.1.10
▪Mask = 255.255.255.0
oOn FIXTURLASER SMC:
Shortcuts > Setting > Network >
▪DHCP = No
▪Set the IP address = e.g., 192.168.1.12
Note: the first 3 numbers must be the same (192.168.1) and the last one different
from that of the PC (10 ≠12)
▪Mask = 255.255.255.0
Once the setting is done, to access the data use FIXTURLASER SMC IP address, e.g., in
the Explorer type “\\192.168.1.10\Data”.
•PC- FIXTURLASER SMC connection through LAN Network:
oOn PC: Setup Network
▪DHCP = Yes
oOn FIXTURLASER SMC:
Shortcuts > Setting > Network >
▪DHCP = Yes

1.13
Connect FIXTURLASER SMC to the LAN, set it in communication mode and wait
until it gets its IP address.
The IP address can be read in Setting > About information page.
If there is a DNS, it is also possible to access the instrument with its name. The name is
FIXTURLASER SMC_serial_number (e.g., FIXTURLASER SMC_10015).
Once the setting is done, to access the data use FIXTURLASER SMC IP address, e.g., in
the Explorer type “\\FIXTURLASER SMC_10015\Data”.
Using Wi-Fi
Check first if Wi-Fi is enabled in:
Shortcuts > Setting > > Wi-Fi = enabled
•Direct Wi-Fi connection PC- FIXTURLASER SMC setting without WLS sensor:
oOn FIXTURLASER SMC:
Setting > Network > Wi-Fi part
▪Enabled = Yes
▪Adhoc = Yes
▪SSID = My_FIXTURLASER SMC or other
▪Canal = 6
▪Authentication = none
▪DHCP = No
▪Set the IP address: Ex 192.168.1.16
Note: the first 3 numbers must be the same (192.168.1) and the last one different
from that of the PC (14 ≠16)
▪Mask = 255.255.255.0
oOn PC: Setup WIFI network
▪DHCP = No
▪Set the IP address, e.g., 192.168.1.14
▪Mask: 255.255.255.0
▪Scan Wi-Fi networks and select FIXTURLASER SMC SSID.
Once the setting is done, to access the data use FIXTURLASER SMC IP address, e.g., in
the Explorer type “\\192.168.1.16\Data”.
•Direct Wi-Fi connection PC- FIXTURLASER SMC-WLS setting:
oOn FIXTURLASER SMC:
▪First set the connection FIXTURLASER SMC-WLS: see § 1.7
▪Wait for sensor connection ( in status summary)
oOn PC: Setup Wi-Fi network
▪DHCP = Yes
▪Scan Wi-Fi networks and select FIXTURLASER SMC SSID (WLS_10015).
▪Input the security key: WLS serial number (Ex: 10015)
Once the setting is done, to access the data use FIXTURLASER SMC name, e.g., in the
Explorer type “\\FIXTURLASER SMC_10015\Data”.

1.14
•Wi-Fi LAN connection PC- FIXTURLASER SMC setting without WLS sensor:
Note: WLS sensor cannot be used simultaneously with this mode.
oOn FIXTURLASER SMC:
Setting > Network > Wi-Fi part
▪Enabled: Yes
▪Adhoc: No
▪Save the setting
Setting > Network > Wi-Fi part
▪Scan network until you detect the right SSID
▪Set Authentication, Encryption and Key according to selected SSID
▪DHCP: Yes
The IP address can be read in Setting > About information page.
If there is a DNS, it is also possible to access the instrument with its name. The name is:
FIXTURLASER SMC_serial_number (e.g., FIXTURLASER SMC_10015).
oOn PC: the PC must also be connected to the same Wi-Fi LAN

1.15
1.9 STATUS INDICATIONS
Status summary
The status is indicated at the bottom of the right-hand side of the screen
•General status
oDate and time
oBattery level of the instrument
oIf Wireless measurement mode is selected, WLS connection status:
▪connected
▪not connected
•During analogic measurement
oInput overload indication in percentage of time.
oSensor integrity indicator
•During wireless measurement
oSensor overload indication in percentage of time.
oBattery level of the sensor
Status panel
If you click on the status summary, the status panel appears on the right-hand side of the screen. It
shows the following information:
•Wi-Fi:
oOff: Wi-Fi is disabled in Shortcuts > Setting >
oNot connected: The instrument is not connected to a WLS sensor or another Wi-Fi
network
oWLS_NNNNN: The instrument is connected to a WLS sensor, NNNNN is the serial
number of the sensor.
oNetwork: Wi-Fi is connected to a network.
•Sensor battery: This information shows the battery level of the WLS sensor. It is not displayed if the
instrument is not connected to a WLS sensor.
•System battery: This information shows the battery level of the instrument.
•Next calibration date: Display the recommended date for the next calibration.
•The amount of available memory.

1.16
1.10 SHORTCUTS PANEL
From any screen, button opens the Shortcuts panel. It gives direct access to a group of functions.
The list of accessible functions depends of the current screen.
Photo
•From the Supervisor module:
oMeasurement list screen: Take an inspection picture
oMap mode at location level: Take a location picture
oMap mode at machine level: Take a machine picture
•From Balancing module, any screen: Take picture for your report (balanced machine, sensor
installation, weight mounting)
Take the picture:
: Take the picture. You must not move during few seconds.
Brightness adjustment.
Flash on and off.
Stored picture management:
Add a new picture.
See previous picture.
See next picture.
Add a comment to the current picture.
Delete the current picture.
Text note
•From Supervisor module, measurement list screen: input text inspection note directly from the
keyboard or from a list of predefined notes. The list of predefined notes is only available if this list is
created in PC database (in XPR menu “Libraries/Predefined notes”)
•From Balancing module, any screen: Balancing comment for the report
Table of contents
Other Fixturlaser Measuring Instrument manuals