
D:\Oude_D_schijf\Uniflo 1200\Uniflo_1200_Calibration_Manual_011015UK.docx
Page 4 of 17
Uniflo_1200_Calibration_Manual_011015UK.docx
1.0 General
This manual describes how to make corrections on the calibration of the pressure and temperature inputs of
the Uniflo 1x00.
Warning: Please note that Flonidan performs an accurate calibration in the factory as
part of the MID accreditation where the highest accuracy possible will be
insured. Any modification to the factory calibration will modify the initial
calibration and the metrological behaviour will be affected.
Following the procedure described in this manual will help in creating an accurate calibration of the
converter.
The calibration can be executed with use of the Configuration Program providing the user has the
appropriate user rights.
2.0 Equipment
To be able to apply a calibration to the Uniflo 1x00 following equipment is required:
1. Pressure reference (error ≤0.1 %)
2. Inert gas bottle with pressure controller or air pump
3. Temperature reference (error ≤0.1 °C)
4. Temperature controlled bath or at least 2 insulated cans, one with melting ice and one with pre-
heated water at approximately 40°C
5. PC with Uniflo 1x00 Config program
6. Optical reading head for Uniflo 1x00
2.1 Location
The calibration should be done in a temperature controlled room where the reference temperature is
approximately 21°C and should not vary more than ± 0.5°C during the calibration procedure.
Alternatively the calibration can be done on site but a higher uncertainty has to be excepted. It is important to
verify the operating conditions of the reference equipment.
Warning: Using reference equipment on site may cause safety conflicts since frequently
reference equipment cannot be used inside hazardous locations. Please insure
the safe use of the reference equipment inside the hazardous location.
2.2 Preparing the calibration
Before the calibration procedure can start a number of preparations should be handled.
1. Make sure that the Uniflo 1x00 is temporized for 24 hours in the calibration location
2. Check the pressure range of the Unilfo 1x00
a. Range 1: 0,6 to 2 bara
b. Range 2: 0,9 to 6 bara
c. Range 3: 2.5 to 14 bara
d. Range 4: 5 to 30 bara
e. Range 5: 13 to 80 bara
Use a reference for pressure matching the Uniflo 1x00 pressure range
3. Use 1 or 2 temperature controlled bath for setting the required reference temperatures and use a
calibrated temperature reference to measure the liquid temperature. For temperatures below 0°C a
non-freezing liquid has to be used. Alternatively 2 insulated cans can be used where one is filled with
melting ice and the other one with a pre-heated liquid at about 40°C. Please note that the
temperature is falling during the calibration and therefore the reference reading and the
measurement has to be done within a short time to avoid unwanted errors.
4. A PC with installed version of Uniflo 1x00 Configuration Program and an optical infra-red head for
reading data from the Uniflo 1x00 is required.