FujiFilm DR-ID 1200MP User manual

897N120970A
This Operation Manual describes details on how to operate the power supply unit
DR-ID 1200MP and cautions to be observed when operating it.
Please read the Operation Manual, System Operation Manual and other manuals
for the related products before actually operating the power supply unit.
After reading this manual, store it nearby the power supply unit so that you can see
it whenever necessary.
Power Supply Unit
DR-ID 1200MP
Operation Manual
2nd Edition : April 2016
For Safe Operation
System
Conguration
(Product Overview)
Basic Operation
Daily Inspection and
Maintenance
Appendix
Maintenance and
Inspection

ii DR-ID 1200MP Operation Manual 897N120970A

iii
DR-ID 1200MP Operation Manual 897N120970A
Introduction
The power supply unit DR-ID 1200MP is a device which supplies power to the at panel sensor and
connects between the at panel sensor, image processing unit and control cabinet.
The power supply unit DR-ID 1200MP is an optional item of DR-ID 1200.
This Operation Manual includes descriptions of matters necessary when using the DR-ID 1200MP,
such as the equipment overview, operation procedures and precautions to observe, as well as daily
inspections and maintenance.
Accompanying documents were originally drafted in the English language.
Installation may only be conducted by authorized service personal.
CAUTIONS
1. No part or all of this manual may be reproduced in any form without prior permission.
2. The information contained in this manual may be subject to change without prior notice.
3. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
installation, relocation, remodeling, maintenance, and repair performed by other than dealers
specied by FUJIFILM Corporation.
4. FUJIFILM Corporation shall not be liable for malfunctions and damages of FUJIFILM
Corporation products due to products of other manufacturers not supplied by FUJIFILM
Corporation.
5. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
remodeling, maintenance, and repair using repair parts other than those specied by
FUJIFILM Corporation.
6. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
negligence of precautions and operating methods contained in this manual.
7. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from use
under environment conditions outside the range of using conditions for this product such as
power supply, installation environment, etc. contained in this manual.
8. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from
natural disasters such as res, earthquakes, oods, lightning, etc.
This system is classied as a medical device under EC Directive 93/42/EEC.
Process waste correctly, as stipulated by local law or any regulations that apply.
Caution : Rx Only in the United States (Federal law restricts this device to sale by or on the order
of a physician.)

iv DR-ID 1200MP Operation Manual 897N120970A
Open-Source Software Contained in This Product
This product contains third party’s software that is made available as open source software or free
software.
This software is provided “as is” with no warranty of any kind as to its merchantability or tness for any
particular purpose.
For the information on open source software contained in this product, please see the attached DVD.
Source codes for certain type of open source software used in this product are available at delivery cost.
If you would like to receive such source codes, please contact FUJIFILM dealer or the service
representatives at the agency from which you purchased this product. (Please be noted that any
inquiries concerning the contents of source codes should be directed to original licensors of open source
software.)
Note : FUJIFILM has successfully performed verication and validation testing on all third party
software and has conrmed its suitability to be used in this system.
Trademarks
All company names and product names described in this manual are the trademarks or registered
trademarks of FUJIFILM Corporation or their respective holders.
Windows is the registered trademark of US Microsoft Corporation in the U.S.A. and other countries.
Copyright © 2015-2016 FUJIFILM Corporation. All rights reserved.

v
DR-ID 1200MP Operation Manual 897N120970A
Contents
Introduction...........................................................................................................................iii
Chapter 1 For Safe Operation
1.1 Safety ...................................................................................................................... 1-1
Chapter 2
System Conguration (Product Overview)
2.1 DR-ID 1200MP........................................................................................................ 2-1
2.1.1 System Conguration ..............................................................................................2-1
2.2 Unit Names and the Functions................................................................................ 2-2
2.3 Locations of Labels and Signs ................................................................................ 2-2
2.3.1 Locations of Labels..................................................................................................2-2
2.3.2 Safety and Other Symbols.......................................................................................2-3
2.3.3 Symboles de sécurité et autres ...............................................................................2-4
Chapter 3 Basic Operation
3.1 Preparing the Flat Panel Sensor ............................................................................. 3-1
3.1.1 Number of the Connectable Flat Panel Sensors .....................................................3-1
3.1.2 Connecting/Disconnecting the Flat Panel Sensor Connector..................................3-1
3.1.3
Changing the Direction of the Flat Panel Sensor Connector ........................................ 3-2
3.1.4 Checking the Connection between the Flat Panel Sensor and the Connector........3-2
3.2 Starting Up and Shutting Down the DR-ID 1200MP................................................ 3-2
3.2.1 Starting Up the DR-ID 1200MP ...............................................................................3-2
3.2.2 Performing an Exposure .........................................................................................3-2
3.2.3 Shutting Down the DR-ID 1200MP..........................................................................3-2
Chapter 4 Daily Inspection and Maintenance
4.1 Daily User Inspection and Maintenance......................................................................4-1
4.1.1 Periodical Inspection................................................................................................4-1
Appendix A Specications
A.1 Specications ..........................................................................................................A-1
A.1.1 Power Supply Conditions........................................................................................ A-1
A.1.2 Environmental Conditions....................................................................................... A-1
A.2 External View and Weight .......................................................................................A-2
A.2.1 DR-ID 1200MP ....................................................................................................... A-2
Appendix O Use of Optional Items
O.1 Optional Items ....................................................................................................... O-1
O.1.1 SE Cable ....................................................................................................................O-1
O.1.2 Using the Retaining Bracket for MP ......................................................................... O-1
O.1.3 Relay unit for AC bucky ..................................................................................................
Maintenance and Inspection

vi DR-ID 1200MP Operation Manual 897N120970A

1-1
For Safe Operation
1
DR-ID 1200MP Operation Manual 897N120970A
Chapter 1 For Safe Operation
1.1 Safety
Before using the DR-ID 1200MP, read this section thoroughly to ensure that you use the product
properly.
Electric Shock Warnings and Cautions
WARNING
The power supply to this device is AC100 to 240V.
To avoid electric shocks, users should always take the following precautions:
● Do not open any covers when it is not necessary.
● Install the equipment in a location where it will not be exposed to water.
● Check that the equipment is securely earthed.
● Check that all of the cables are completely and securely connected.
WARNING
Do not touch the patient’s body while touching the control cabinet and the image processing
unit. Otherwise, the patient may receive an electric shock.
WARNING
Do not use a multiple tap connector or extension cable for powering the devices constituting
the system. Otherwise, re or electric shock may occur due to the electrical load exceeding the
allowable limit.
WARNING
Observe the following precautions when using the cables.
● Do not touch the plug and connector with wet hands. Otherwise, electric shock may result,
causing death or severe injury.
● Hold the plug or connector when removing the cable.
Pulling the cable or carrying by holding it may damage the cable, causing re or electric shock.
● Do not damage or remodel the cable.
Do not place a heavy object on the cable or lay it under the at panel sensor. Do not step on,
pull, forcibly bend, or bundle the cable. Otherwise, re or electric shock may result.
● Do not use the at panel sensor for the radiographic examination stand if its cable becomes
overloaded. Otherwise, the cable may be damaged, causing re or electric shock.
WARNING
Do not turn on the system with dew condensation on the at panel sensor. Otherwise, re or
electric shock may result.
WARNING
Do not use the equipment in a location where metal particles could come into the equipment.
This may cause an electric shock.

1-2
For Safe Operation
1
DR-ID 1200MP Operation Manual 897N120970A
WARNING
Do not disassemble or remodel the equipment. Otherwise, re or electric shock may result.
Keep away from the parts inside the product, which may cause electric shock. If you touch
them accidentally, death or severe injury may result.
WARNING
Do not hit or drop the equipment or subject it to severe shock. Otherwise, the equipment may
be damaged. If the damaged equipment is used, re or electric shock may result.
WARNING
Make sure to use the optional parts recommended by us. Failure to use the optional parts
recommended by us may result in damage to the equipment and/or electric shock and injury.
CAUTIONS
As the cables of the equipment are long, be careful not to entangle the cables during use.
Also, be careful not to trip over the cables. Falls could result in injury.
CAUTIONS
Follow the specied procedure when turning off the equipment. Otherwise, the at panel
sensor could be damaged by thermal shock.
CAUTIONS
Do not store magnetic media near the DR system and control cabinet. Otherwise, magnetism
generated by the equipment may cause the data to be lost.
CAUTIONS
Keep the equipment away from patient’s body uids, chemicals, water, etc.
Otherwise, it may become damaged, causing re or electric shock.
Explosion Warnings
WARNING
Because this equipment is not explosion-proof, do not use combustible and explosive gases
near the equipment.
WARNING
Flammable gasses may stay in the room after disinfection. If you turn the system on just after
disinfection, ensure that the room is well ventilated before powering on the system.

1-3
For Safe Operation
1
DR-ID 1200MP Operation Manual 897N120970A
Warnings for Abnormalities
WARNING
If any of the following occurs, immediately turn off the power of each unit, unplug the power
cable from the outlet, and then contact our ofcial dealer or FUJIFILM Representative.
● When smoke, strange odor, or abnormal sound is present.
● When a foreign object (such as a metal object) or liquid enters the product.
● When the equipment is dropped or hit and is damaged.
Avertissements relatifs aux anomalies
AVERTISSEMENT
Si l’une des conditions répertoriées ci-après se produit, mettez immédiatement chaque unité
hors tension, débranchez le cordon d’alimentation de la prise secteur, puis contactez notre
revendeur agréé ou notre représentant FUJIFILM.
● En cas de présence de fumée, d’une odeur étrange ou d’un bruit anormal.
● En cas de pénétration d’un corps étranger (comme un objet métallique) ou d’un liquide dans
le produit.
● En cas d’endommagement de l’équipement suite à une chute ou à un impact.
Installation Precautions
CAUTIONS
Do not install the equipment in a location with the following conditions.
● Where the temperature changes sharply.
● Close to heat sources such as a heater.
● Where the equipment may be exposed to water due to water leakage or ingress.
● Where corrosive gas may be generated.
● Where there is excessive dust.
● Where the equipment is subject to frequent or excessive vibration/shock.
● Where the equipment is exposed to direct sunlight.
● Where there is no ventilator.
CAUTIONS
When the power supply unit is installed, be sure to use the Retaining bracket for MP or xing
metal ttings to prevent its overturning.
CAUTIONS
Use the equipment on a at place. If the equipment falls, it may cause damage to the equipment
or personal injury.
CAUTIONS
For veterinary or mobile applications, contact our ofcial dealer or FUJIFILM Representative.
CAUTIONS/ATTENTION
Do not place any object in a place where removal of the power cable is prevented.
Ne placez aucun objet à un emplacement gênant le débranchement du câble d’alimentation.

1-4
For Safe Operation
1
DR-ID 1200MP Operation Manual 897N120970A
CAUTIONS
To ensure optimal image quality, it is recommended that you do not use the at panel sensor
near devices (motor, transformer, switching supply, etc.) that generate electromagnetic noise.
CAUTIONS
Do not install the power supply unit in a place where it may be contacted inadvertently. In
addition, take care not to make contact with the power supply unit except when operating the
main switch. If the fan inside the power supply unit malfunctions, the power supply unit may
become hot, causing injury.
Connection Instructions
WARNING
Make sure that the devices to be connected to the equipment are authorized for connection.
Precautions on External Network Connection
CAUTIONS
When a setting of the network to which the equipment is connected has been changed,
check that the change does not affect the system operation and take measures if necessary.
The setting change may include the following:
● Change of connection destination
● Addition of devices
● Removal of devices
● Update of devices
● Upgrade of devices
Warnings and Cautions on Network
WARNING
Make sure to use the optional parts and networks recommended by us. Failure to use the
optional parts and networks recommended by us may result in damage to the equipment and/
or electric shock and injury.
CAUTIONS
Connect to the Ethernet Network of 100BASE-TX or 10BASE-T prescribed in the IEEE standard 802.3.
Do not connect telephone lines to LAN connector. Only UTP-type straight LAN cables of 4-pair
Category 5 cable (CAT 5E) or higher are appropriate for connection to this connector.
CAUTIONS
After connecting this system to the network with other systems,conrm that the other systems
are not affected. If they are affected, take countermeasures such as network separation.

1-5
For Safe Operation
1
DR-ID 1200MP Operation Manual 897N120970A
System Isolation Instructions
WARNING
To ensure complete system isolation, never install any unauthorized items.
When it is necessary to install authorized optional items, contact our ofcial dealer or FUJIFILM
Representative.
WARNING
Keep the products other than those used for patients out of their reach to ensure appropriate
system isolation.
For details on the products that can be installed in the patient environment, see “2.1.1 System
Conguration”.
WARNING
In normal use, have a patient take a proper positioning for exposure. The operator should operate the
system in a place where safety from radiation is ensured. The operator should also make sure before
exposure that no one but the patient is in the exposure area and the operating area of the system.
Software Precautions
CAUTIONS
Do not install additional software to the system. Do not uninstall any of the software
preinstalled in the system.
The system is preinstalled with the appropriate software. If other software is installed or if the
existing software is uninstalled, various operational errors may result.
Disinfection Instructions
WARNING
Conrm that the respiratory density of disinfectant including solvent is under legal regulation.
Certain disinfectants may damage health. When using a disinfectant, follow instructions
supplied by the manufacturers.
WARNING
Do not use the following disinfectants or sterilizers at the time of disinfection. Quality,
performance and safety of the equipment cannot be assured.
● Chloric disinfectant which is strongly corrosive to metals and rubber parts.
● Disinfectant whose uses on metals, plastics, and coating are forbidden according to the
instructions supplied with the disinfectant.
● Formalin gas and disinfectant sprays that may get inside the equipment.
● Ultraviolet sterilizers
Disinfectant ethanol is recommended for disinfection. Carefully read the instructions and
cautions supplied with the disinfectant before use.
For details on the disinfectant, contact a FUJIFILM dealer or the service representatives at the
agency from which you purchased the disinfectant.

1-6
For Safe Operation
1
DR-ID 1200MP Operation Manual 897N120970A
Other Warnings and Cautions
WARNING
No modication of this equipment is allowed.
CAUTIONS
Install the system in accordance with what is provided by IEC 60601-1:2005 + A1:2012 Chapter
16. Contact our ofcial dealer or FUJIFILM Representative for installation (except the at panel
sensor) of the system.
CAUTIONS
Do not hit or drop the equipment. Otherwise, injury or damage to images, etc. may result.
CAUTIONS
Be sure to inspect the system periodically.
To assure optimum performance of the equipment, it is necessary to systematically perform
maintenance and inspection. For information on maintenance and inspection, contact our
ofcial dealer or FUJIFILM Representative.
CAUTIONS
Do not perform maintenance and inspection while the equipment is used for a patient.
CAUTIONS
The institution where the equipment is installed is responsible for its use and maintenance.
In addition, this equipment should not be used by persons other than doctors or suitably
trained staff.
Contraindications and Prohibitions
No contraindications present.
Classication
● According to the type of protection against electrical shock
Class 1 equipment
● According to the degree of protection against electrical shock
No applied part
● According to the degree of protection against harmful ingress of water
IP00
● According to the degree of safety of application in the presence of a ammable anesthetics
mixture with air or with oxygen or nitrous oxide.
Equipment not suitable for use in the presence of a ammable anesthetics mixture with air or with
oxygen or nitrous oxide.
● According to the mode of operation
CONTINUOUS OPERATION

2-1
System Configuration (Product Overview)
2
DR-ID 1200MP Operation Manual 897N120970A
Chapter 2 System Configuration
(Product Overview)
2.1 DR-ID 1200MP
2.1.1 System Conguration
Flat panel
sensor
Control cabinet DR-ID 1200MC
Image processing unit DR-ID 300CL*1
Power supply unit DR-ID 1200MP
Flat panel sensor unit
DIGITAL RADIOGRAPHY DR-ID 1200
• The products in can be installed in patient environment.
• Up to two at panel sensors can be connected to one power supply unit in wired communication
mode. Up to two power supply units can be used.
• When the at panel sensors are used with three to four different techniques, two power supply
units are required.
*1 F
or detail specication of image processing unit, please refer to “DR-ID 300CL Operation Manual”.

2-2
System Configuration (Product Overview)
2
DR-ID 1200MP Operation Manual 897N120970A
2.2 Unit Names and the Functions
Unit names and the functions of the DR-ID 1200MP are described below.
Power supply unit (DR-ID 1200MP)
Power status LED
Main switch
Name Description
Main switch Supplies the power to the at panel sensor and the inside of the power supply unit.
Power status LED Displays ON/OFF of the power supply unit.
2.3 Locations of Labels and Signs
Locations of labels and signs afxed to the DR-ID 1200MP, and the relevant safety signs are shown below.
2.3.1 Locations of Labels
Power supply unit (DR-ID 1200MP)
DR-ID 1200MP
System Label
Do Not Step On
This Surface label
DR-ID 1200MP
Rating Label
Bracket for MP
Identification Label
DR-ID 1200MP
Caution Label 1
DR-ID 1200MP
Caution Label 2
For the types of connectable
cables, consult our ofcial
dealer or FUJIFILM
Representative
Pour en savoir plus sur les
types de câbles connectables,
contactez notre revendeur agréé
ou notre représentant FUJIFILM.
DR-ID 1200MP Caution Label 2
DR-ID 1200MP System Label
Do Not Step On This Surface label
DR-ID 1200MP Rating Label
Bracket for MP Identification Label
DR-ID 1200MP Caution Label 1

2-3
System Configuration (Product Overview)
2
DR-ID 1200MP Operation Manual 897N120970A
2.3.2 Safety and Other Symbols
The following safety symbols are used in the labels or on its body.
Symbol Description
CE marking
Caution (See “2.3.1 Locations of Labels” (page 2-2).)
OFF (To indicate disconnection from the mains, at least for mains switches or their
positions, and all those cases where safety is involved.)
ON (To indicate connection to the mains, at least for mains switches or their positions, and
all those cases where safety is involved.)
Protective earth (ground)
Alternating current
This symbol indicates that this product is not to be disposed of with your household waste,
according to the WEEE Directive (2002/96/EC) and your national law. This product should be
handed over to a designated collection point.
Improper handling of this type of waste could have a possible negative impact on the
environment and human health due to potentially hazardous substances that are generally
associated with EEE.
At the same time, your cooperation in the correct disposal of this product will contribute to the
effective usage of natural resources.
For more information about waste, please contact our ofcial dealer or FUJIFILM Representative.
Year of manufacture
Refer to Instruction Manual/Booklet
No stepping on surface

2-4
System Configuration (Product Overview)
2
DR-ID 1200MP Operation Manual 897N120970A
2.3.3 Symboles de sécurité et autres
Les symboles de sécurité suivants sont utilisés sur les étiquettes ou sur le corps de l’équipement.
Symbole Description
Marquage CE
Attention (Voir « 2.3.1 Emplacement des étiquettes » (page 2-2).)
HORS TENSION (Pour indiquer une déconnexion de l’alimentation secteur, au moins au niveau
des interrupteurs secteurs ou leur position, et tous les cas dans lesquels la sécurité est en jeu.)
SOUS TENSION (Pour indiquer une connexion à l’alimentation secteur, au moins au niveau
des interrupteurs secteurs ou leur position, et tous les cas dans lesquels la sécurité est en jeu.)
Protection via mise à la terre (masse)
Courant alternatif
Ce symbole indique que ce produit ne doit pas être mis au rebut avec les déchets ménagers,
conformément à la directive DEEE (2002/96/CE) et à la législation nationale en vigueur. Ce produit doit
être remis à un centre de collecte approprié.
Une manipulation incorrecte de ce type de déchet peut avoir un impact négatif sur l’environnement et
sur la santé humaine, en raison des substances potentiellement dangereuses généralement associées
aux EEE.
Votre coopération pour la mise au rebut correcte de ce produit contribuera en outre à une utilisation
efcace des ressources naturelles.
Pour en savoir plus sur les déchets, contactez notre revendeur agréé ou notre représentant FUJIFILM.
Année de fabrication
Consultez le mode d’emploi
Ne montez pas sur la surface

3-1
Basic Operation
3
DR-ID 1200MP Operation Manual 897N120970A
Chapter 3 Basic Operation
3.1 Preparing the Flat Panel Sensor
This section describes how to prepare the at panel sensor.
3.1.1 Number of the Connectable Flat Panel Sensors
Up to two at panel sensors can be connected to
one power supply unit in wired communication mode. Up to two power supply units can be used.
The types of connectable at panel sensors are DR-ID 1201SE, DR-ID 1202SE, DR-ID 1211SE,
DR-ID 1212SE and DR-ID 1213SE.
3.1.2 Connecting/Disconnecting the Flat Panel Sensor
Connector
When multiple at panel sensors are used, make sure that the READY lamp among the status lamps of the at
panel sensor to be used for an exposure is lit.
Connect/Disconnect the connector straight to the at panel sensor. If connected/disconnected at an angle, the
connector may be damaged.
1Disconnect the connector.
Press the latches on both sides of the connector.
2 Connect the connector.
Press the connector into the insertion section.
Make sure that the latches on both sides are
properly engaged when connecting the connector.
If the connector is inserted incompletely, the power
may turn off.

3-2
Basic Operation
3
DR-ID 1200MP Operation Manual 897N120970A
3.1.3
Changing the Direction of the Flat Panel Sensor Connector
The direction of the connector of the at panel sensor can be changed, depending on how it is
inserted into the radiographic examination stand.
To change the direction, contact our ofcial dealer or FUJIFILM Representative.
When shipped After changing the direction
3.1.4 Checking the Connection between the Flat Panel Sensor
and the Connector
After connecting the connector of the SE cable to the at pane sensor, check the power ON/OFF
state and communication state with the status LEDs.
POWER (Blue)
(The power on/off state of the
at panel sensor is displayed.)
On Power ON
Off Power OFF
LINK (White)
On Connected
Off Communication not possible.
3.2 Starting Up and Shutting Down the
DR-ID 1200MP
This section explains how to start up and shut down the DR-ID 1200MP.
3.2.1 Starting Up the DR-ID 1200MP
1Press the ON side of the main switch of the power supply unit.
Make sure that the power status LED is lit in blue.
Start up the at panel sensor with the initial settings properly made by our ofcial dealer or FUJIFILM Representative.
3.2.2 Performing an Exposure
1
When performing an exposure, make sure that the READY status lamp on the at panel sensor is lit.
3.2.3 Shutting Down the DR-ID 1200MP
1Press the OFF side of the main switch of the power supply unit.
Make sure that the power status LED is off.

4-1
Daily Inspection and Maintenance
4
DR-ID 1200MP Operation Manual 897N120970A
Chapter 4 Daily Inspection and
Maintenance
4.1 Daily User Inspection and Maintenance
During maintenance and inspection, strictly observe precautions contained in “Chapter 1 For Safe
Operation” in this manual for you to use the DR-ID 1200MP under best conditions.
4.1.1 Periodical Inspection
Inspection Every Three Months
Once every three months, remove any dirt or dust accumulated in each part of the equipment using
a vacuum cleaner or air duster, clean each part with a slightly moistened soft cloth and then wipe off
any moisture with a dry cloth.
See “2.2 Unit Names and the Functions” (page 2-2).
CAUTIONS
Be sure to turn off the power before cleaning each part of the device.
No. Unit No. Unit
1Power supply unit 2Power supply unit
Air lter (1)
CAUTIONS
Ensure sufcient space when cleaning the equipment on a table, etc.
Air lter
Clean the air lter on the rear of the power
supply unit with a vacuum cleaner.
Remove the louver while pressing the lever
on the right side, and then clean the air lter
with a vacuum cleaner after detaching it from
the louver.
Air filter

4-2
Daily Inspection and Maintenance
4
DR-ID 1200MP Operation Manual 897N120970A
Table of contents
Other FujiFilm Power Supply manuals
Popular Power Supply manuals by other brands

Multiform
Multiform PS020A user manual

Whelen Engineering Company
Whelen Engineering Company ISP188 installation guide

Monacor
Monacor PSS-510 DC operating instructions

Thermal Dynamics
Thermal Dynamics PakMaster XLPLUS 50 Service manual

Deutronic
Deutronic D-IPS 350/3C manual

3B SCIENTIFIC PHYSICS
3B SCIENTIFIC PHYSICS 1008690 instruction sheet