Gaumard S415.100 User manual

User Guide 12.7.1
© Gaumard Scientic Company, 2012
All Rights Reserved
www.Gaumard.com
Blood Pressure Training
System with Speakers
S415.100
®
Gaumard
Simulators for Health Care Education
The Blood Pressure Training System with Speakers is an interactive
educational system developed to assist a certied instructor. It is not
a substitute for a comprehensive understanding of the subject matter
and not intended for clinical decision making.

2 | User Guide | Blood Pressure Training System with Speakers |

| Blood Pressure Training System with Speakers | User Guide | 3
Care and Cautions
Overall Warnings .................................................................................................. 5
General ................................................................................................................... 5
Latex warning ......................................................................................................... 5
Cleaning ................................................................................................................. 5
Operating Conditions .............................................................................................. 5
Storage Conditions ................................................................................................. 5
BP Arm Set Up
Control Panel ........................................................................................................ 7
Power Supply........................................................................................................ 7
Speaker.................................................................................................................. 7
Omni™ Setup........................................................................................................ 7
Blood Pressure Arm Setup.................................................................................. 8
BP Arm Features
Overview.............................................................................................................. 10
Features............................................................................................................... 10
Palpable Pulse...................................................................................................... 10
Korotkoff Sounds .................................................................................................. 10
Auscultation ......................................................................................................... 10
Using the Omni™Blood Pressure Controller
Omni™.................................................................................................................. 12
Main Screen ......................................................................................................... 12
Omni Menus ......................................................................................................... 12
Pulse Menu........................................................................................................... 13
Au. Gap menu ...................................................................................................... 13
Volume menu........................................................................................................ 13
Calib menu ........................................................................................................... 13
Help ...................................................................................................................... 14

4 | User Guide | Care and Cautions |
Care and Cautions

| Care and Cautions | User Guide | 5
Overall Warnings
Damage caused by misuse may void the
manufacturer’s warranty. Failure to comply with
the following guidelines may result in injury or
damage to the equipment. Additional warnings
are found throughout the document.
GENERAL
• This simulator is constructed of material
that approximates skin texture. Therefore,
in handling the model, use the same gentle
techniques as you would in working with a
patient.
• Do not wrap this or any other Gaumard product
in newsprint.
• Indelible marks made with ballpoint pens, ink or
marker cannot be removed.
• Do not use alcohol, acetone, Betadine® or any
other antiseptic which contains iodine in this
or any Gaumard® simulator. These products
could damage or stain the skin of the simulator.
LATEX WARNING
• Vein tubing contains latex which may cause
allergic reactions. Users allergic or sensitive to
latex should avoid contact. Discontinue use of
this product and seek medical attention if an
allergic reaction occurs.
CLEANING
• Clean the skin of the simulator after every
training session. The skin should be cleaned
with a cloth, dampened with diluted liquid dish
washing soap.
• Do not clean with harsh abrasives.
• Do not use povidone iodine on the simulator.
• The simulator is “splash-proof” but not water-
proof. Do not submerge or allow water to enter
the interior of the simulator.
OPERATING CONDITIONS
Operating the Blood Pressure Training System
outside these ranges may affect performance:
• Operating temperature: 50°- 95° F (10°- 35° C).
• Humidity: 5%-95% (non-condensing).
STORAGE CONDITIONS
• Improper storage may damage the simulator.
• Keep it stored in the box and/or bag provided.
• Do not stack or store heavy materials on top of
the box and/or bag.
• Storage temperature: 32°- 113° F (0°- 45° C).
• Humidity: 40%-60% (non-condensing).

6 | User Guide | BP Arm Set Up |
BP Arm Set Up

| BP Arm Set Up | User Guide | 7
Speaker
Connect the speakers to the audio jack on the
control panel to listen to the Korotkoff sounds
while measuring the blood pressure.
Omni™ Setup
Omni controls the Blood Pressure Training
System with the touch of button.
1. Connect the communication cable to the port
labeled Omni on the control panel.
2. Connect the other end of the communication
cable to Omni.
Control Panel
A panel, near the shoulder, encloses all of the
blood pressure arm connectors. Each connector
includes a label for quick reference.
The control panel, located near the shoulder,
provides quick access to all the blood pressure
arm connectors. Each connector is labeled for
quick reference.
Power Supply
1. Unpack the system and place the blood
pressure arm on a at clean surface. Access
to electrical power is required.
2. Connect the power supply to a wall outlet.
3. Connect the power supply cable to the power
input in the control panel.

8 | User Guide | BP Arm Set Up |
A startup screen is shown while Omni is
detecting the simulator features.
After the start up screen, Omni will automatically
proceed to the main screen.
WARNING
Do not connect the Blood Pressure Arm
or Omni to a computer, LAN network
or unauthorized diagnostic equipment
using the communication cable
(Ethernet cable). Doing so will cause
serious damage to the equipment.
Blood Pressure
Arm Setup
1. Wrap the blood pressure cuff around the
arm, midway between the deltoid and elbow.
Ensure the arrow on the cuff points
toward the elbow following the direction
of blood ow in the artery.
2. Connect the blood pressure cuff extension
to the connector labeled BP on the control
panel.
Reverse the setup steps to disassemble the
simulator and turn off Omni.

| BP Arm Features | User Guide | 9
BP Arm Features

10 | User Guide | BP Arm Features |
4. Slowly release the cuff and listen for Korotkoff
sounds.
5. Ask the provider to record the systolic
pressure, the diastolic pressure, the limits of
the auscultation gap (if any) and the pulse
rate.
6. Compare the values recorded by the provider
with those originally programmed in Omni.
Overview
The Blood Pressure Training System consists of
a full-size adult left arm. This is a versatile training
tool developed to assist health professionals
teach the processes and skills required to
perform blood pressure auscultation procedures
and techniques.
Features
PALPABLE PULSE
Programmable, palpable radial pulse is present
when the cuff pressure is less than the selected
systolic blood pressure.
KOROTKOFF SOUNDS
Korotkoff sounds, K1 through K4 (K5 is silence)
are audible between systolic and diastolic
pressures. The sounds are automatically
silenced if the auscultation gap is enabled, and
adjusted depending on the selected heart rate
and the rate of cuff deation.
AUSCULTATION
The simulator operates with a modied blood
pressure cuff. Connect the BP cuff extension
to the arm’s control panel before use. Use a
conventional stethoscope to auscultate Korotkoff
sounds in the antecubital area.
INSTRUCTIONS FOR USE
1. Follow the instructions listed in the Set Up
section to assemble the simulator.
2. Place the bell of the stethoscope on the
antecubital region of the arm.
3. Pump up the cuff until the radial pulse is
absent.

| Using the Omni™ Blood Pressure Controller | User Guide | 11
Using the Omni™
Blood Pressure
Controller

12 | User Guide | Using the Omni™ Blood Pressure Controller |
NAVIGATION MENUS
Edit Blood Pressure settings using the navigation
menus. Select a menu item by pressing the
soft key buttons located on the right side of the
controller. Each menu has additional submenus.
Use the submenus to decrease, increase,
accept, or cancel values.
OMNI MENUS
SET BP MENU
Use this menu to set the blood pressure in the
simulator.
Select systolic or diastolic and adjust the pressure
using the + and - buttons. Press OK to save the
changes and return to previous screen. The
default value for the blood pressure is 120/80
mmHg.
Omni™
Omni was designed to assist health professionals
teach the processes and skills required to
perform blood pressure auscultation procedures
and techniques.
MAIN SCREEN
The Omni main screen is divided into three
sections which are, Current Settings, Feedback
Graphics and Navigation Menus.
CURRENT SETTINGS
Current settings information is listed in the upper
center of the screen.
FEEDBACK GRAPHICS
The feedback graph located in the lower center
of the screen displays, in real time, the BP cuff
pressure.

| Using the Omni™ Blood Pressure Controller | User Guide | 13
PULSE MENU
Use this menu to set the heart rate. Adjust the
beats per minute using the + and - buttons. Press
OK to save the changes and return to previous
screen.
AU. GAP MENU
Enable and set an interval of pressure where
Korotkoff sounds dissipate. The auscultatory gap
is disabled by default.
Pressing the ENABLE button activates the
UPPER and LOWER menus. Use these menus
to set the limits of the auscultatory gap.
Adjust the limits of the auscultatory gap using the
+ and - buttons. Press OK to save the changes
and return to previous screen. The default upper
and lower limits for the auscultatory gap are 100
mmHg and 90 mmHg, respectively.
VOLUME MENU
Increase or lower the volume of Korotkoff sounds
using the VOLUME menu. Press OK to save
the changes and return to previous screen. The
default value for the Korotkoff sounds volume is
5.
CALIB MENU
Calibrate the sensors inside the simulator
using this option. Set the standard against
which auscultation will be evaluated during the
exercise. Blood pressure cuff calibration should
be performed only when the Korotkoff sounds do
not match the systolic and diastolic values set
from Omni.
It is not necessary to re-calibrate Omni
for each training session. Omni will
remember previous calibration settings.
CALIBRATION INSTRUCTIONS
1. Assemble the system and turn Omni on.
2. Place the blood pressure cuff on the arm.
3. Connect the cuff’s extension to connector on
the control panel.

14 | User Guide | Using the Omni™ Blood Pressure Controller |
After pressing the DEFAULT button, a message
on the screen will notify the user about the status
of the reset.
HELP
The help window provides access to global
settings such as backlight time and Omni/
Simulator serial number. This menu is available
in the main screen and throughout the control
menus for convenience.
BACKLIGHT
The default value for Omni’s backlight timer is 10
minutes. After 10 minutes the backlight will turn
off.
To alter the backlight duration, adjust the
backlight timer with the plus or minus sign. Press
OK to accept the changes.
SERIAL NUMBER
View Omni and simulator serial number. Press
OK to return to the HELP menu.
4. Select the CALIB menu from the main screen.
5. Press NEXT to begin the calibration
procedure.
6. You will be prompted to set the BP cuff to a
certain pressure. Hold that pressure constant
and press the NEXT button.
7. Omni will now prompt you to set the BP cuff
to 20 mmHg. Hold that pressure constant and
press the NEXT button.
Follow the text cue on the screen and repeat the
procedure for each pressure level. The facilitator
should follow the text cue on the screen to inate
the cuff just one value at a time, until nished.
After calibrating the sensor, Omni will return to
the main screen and display the reading shown
on the sphygmomanometer.
FACTORY SETTING
The DEFAULT option is used to restore sensors
back to factory settings, erasing any calibrations
performed by users. When restoring the sensors
to factory settings, do not practice auscultation.
This action will interfere with the reset.

| Appendix | User Guide | 15
Appendix

16 | User Guide | Appendix |
Consumables and
Replacements
SELECTED PARTS LIST
Contact Gaumard Scientic for a complete list
of consumable and replacement parts and their
prices.
Item ID Description
S415.100.002 Blood Pressure Cuff
Kit
S415.100.003 Soft Carrying Bag r
Arm
S415.100.008 2 Arm Skins
S415.100.009 Synthetic Blood
Concentrate
S415.100.010 Speakers
S415.100.180 Omni Controller
S415.100.005 Power Supply 100-
240 VAC
S415.100.180.1 Communication Cable
S415.100.180.2 USB Cable
Troubleshooting
GENERAL TROUBLESHOOTING GUIDE
Use the following table to nd causes and
solutions to a number of possible problems.
Symptom Possible
Cause Solution
Communication
never gets
established or
is lost
Data cable
is not
connected
Ensure the
Ethernet
cable is
plugged into
Omni and
the simulator
Ethernet
cable is
damaged
Verify the
power cable
is connected
to the
simulator
No radial pulse
Pulse rate
was set to
0 BPM
Increase the
pulse rate
using the
pulse menu
The readings
from the
simulator do not
match Omni
Improper
calibration
Calibrate
simulator
following the
calibration
instructions
listed in the
Using the
Omni section
Korotkoff
sounds are
absent or
not heard at
desired volume
level
Volume
level is set
to low
Adjust the
volume
control to get
it to desired
level

| Appendix | User Guide | 17
Warranty
EXCLUSIVE ONE-YEAR LIMITED WARRANTY
Gaumard warrants that if the accompanying Gaumard product proves to be defective in material or
workmanship within one year from the date on which the product is shipped from Gaumard to the
customer, Gaumard will, at Gaumard’s option, repair or replace the Gaumard product.
This limited warranty covers all defects in material and workmanship in the Gaumard product, except:
• Damage resulting from accident, misuse, abuse, neglect, or unintended use of the Gaumard product;
• Damage resulting from failure to properly maintain the Gaumard product in accordance with Gaumard
product instructions, including failure to property clean the Gaumard product; and
• Damage resulting from a repair or attempted repair of the Gaumard product by anyone other than
Gaumard or a Gaumard representative.
This one-year limited warranty is the sole and exclusive warranty provided by Gaumard for
the accompanying Gaumard product, and Gaumard hereby explicitly disclaims the implied
warranties of merchantability, satisfactory quality, and tness for a particular purpose. Except
for the limited obligations specically set forth in this one-year limited warranty, Gaumard will not be
liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract,
tort, or any other legal theory regardless of whether Gaumard has been advised of the possibilities of
such damages. Some jurisdictions do not allow disclaimers of implied warranties or the exclusion or
limitation of consequential damages, so the above disclaimers and exclusions may not apply and the
rst purchaser may have other legal rights.
This limited warranty applies only to the rst purchaser of the product and is not transferable.
Any subsequent purchasers or users of the product acquire the product “as is” and this limited warranty
does not apply.
This limited warranty applies only to the products manufactured and produced by Gaumard.
This limited warranty does not apply to any products provided along with the Gaumard product that are
manufactured by third parties. For example, third-party products such as computers (desktop, laptop,
tablet, or handheld) and monitors (standard or touch-screen) are not covered by this limited warranty.
Gaumard does not provide any warranty, express or implied, with respect to any third-party products.
Defects in third-party products are covered exclusively by the warranty, if any, provided by the third-party.
Any waiver or amendment of this warranty must be in writing and signed by an ofcer of Gaumard.
In the event of a perceived defect in material or workmanship of the Gaumard product, the rst purchaser
must:
1. Contact Gaumard and request authorization to return the Gaumard product. Do NOT return the
Gaumard product to Gaumard without prior authorization.
2. Upon receiving authorization from Gaumard, send the Gaumard product along with copies of (1) the
original bill of sale or receipt and (2) this limited warranty document to Gaumard at 14700 SW 136
Street, Miami, FL, 33196-5691 USA.
3. If the necessary repairs to the Gaumard product are covered by this limited warranty, then the rst
purchaser will pay only the incidental expenses associated with the repair, including any shipping,
handling, and related costs for sending the product to Gaumard and for sending the product back
to the rst purchaser. However, if the repairs are not covered by this limited warranty, then the rst
purchaser will be liable for all repair costs in addition to costs of shipping and handling.
Extended Warranty In addition to the standard one year of coverage, the following support plans are
available:
• Two-Year Extension (covers second and third years)
• Call for pricing (USA only)

Contact Us
Before contacting Tech Support you must:
1. Have the simulator’s Serial Number
2. Be next to the simulator if troubleshooting is needed.
Phone: Toll-free in the USA: (800) 882-6655
Worldwide: 01 (305) 971-3790
Fax: (305) 667-6085
Post: Gaumard Scientic
14700 SW 136 Street
Miami, FL 33196-5691
USA
Ofce hours: Monday-Friday, 8:30am - 4:30pm EST (GMT-5, -4 Summer Time)
General Information
Gaumard®, ZOE®, Michelle®, Mike®, PEDI®, Susie Simon®, Susie®, Simon® Code Blue®, SIMA Models®,
SIMA GYN/AID®, Virtual Instruments®, Codemaker®, Code Blue®, NOELLE®, Simulation Made Easy™,
HAL®, CPRLink™, Zack™, RITA™, Chloe™, Seatbelt Susie™, Krash Kids™, Premie™, GIGA®, Omni™,
SmartSkin™ are Trademarks of
© Gaumard Scientic Company, 2012
All rights reserved.
Gaumard®, NOELLE®, HAL® are trademarks of Gaumard Scientic Company, Inc.
Patented; Other Patents Pending