manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Gems Sensors
  6. •
  7. Switch
  8. •
  9. Gems Sensors ELS-950 Series Training manual

Gems Sensors ELS-950 Series Training manual

ELS-950 Series
Electro-OpticLevelSwitch
Instruction Bulletin No. 225480
Installation
1. For NPT threaded units:
a. Use Teflon (TFE) thread tape or Permatex #80725 plastic pipe sealant to seal thread. Caution: Pipe sealant must not come in
contact with prism surface.
b. Thread sensor into tank wall and tighten by hand. Further tighten an additional one to two threads past hand-tightness. (Avoid
overtightening, as this may damage threads.)
2. For 1/2" and M12 straight threaded units:
a. Thread sensor into tank wall by hand until sensor housing contacts tank wall surface. ( See below for mating o-ring gland design
information.)
b. The sensor shall be torque tightened to 12 to 24 in-lb ( 1.4 to 2.7 N-m). This range is slightly more than “hand-tight” to insure sealing,
and less than that which would cause housing or thread damage.
c. CAUTION: Under no circumstances shall the sensor be tightened more than 1/6 turn beyond that required to seat the bottom of the
sensor housing with the outside wall of the tank.
3. Sensor may be installed in horizontal or up to 45° from horizontal plane for best operation. Vertical mounting is possible, may be droplet sensitive,
depending on viscosity of media.
4. Caution: Do not install sensor close to infrared sources or incandescent light. Reflective surfaces may interfere with proper operation of the optical
sensor. (Consult Gems Sensors if prism is to be mounted less than 2” from any reflective surface.)
5. Connect voltage supply (Vdc + 10%) Between red (+) and black (-) of the unit.
Caution: Do not connect output (blue wire) to Vdc power without a load.
See wiring diagrams for output configuration.
Materials
Housing and Prism
Operating Pressure
Operating Temperature
Input Voltage
Output Configuration
Current Consumption
(No Load)
Repeatability
AmbientLightImmunity
Approval
Specifications
Polysulfone
0 to 250 psig, (0 to 17 bar) Max.
-40°F to 230°F (-40°C to +110°C)
See Part Number Chart
Open drain may sink 40mA max.
~ 4mA @ 5Vdc
~ 10mA @ 12 Vdc
±1 mm
See Ambient Light Test
U.L. Pending
Specifications
Note: Not for use in freezing liquids
The test depicted below was performed to determine the sensor
immunity level to ambient light. The recorded ambient is the maximum
level at which the sensor performs/detects normally.
Note: 1 Ft-Cd = 10.7 Lux
Above testing is based on minimum readings of
at least two (2) samples.
Variac
Test
Sensors Light Meter
GE Model #214
Test Container(s)
Ú
SpecialcautionshouldbetakentomeetEN61000-4-5SurgeImmunityifanyofthefollowingconditionsapplytotheinstallation:Theproduct
isinstalledoutside;thecable is greaterthan30metersinlength.Inorder to meettheSurgeImmunityrequirements,thefollowingconditions
must be followed during installation:
1. Shielded cable must be used, and the shield must be tied to earth ground (not power supply ground) on at least one end of the cable
shield/drain wire. The shield must be maintained all the way from sensor to the power supply.
2. If unshielded cable is used, an earth grounded metal conduit can be used to replace the shielded cable.
3. For the sensor with metal body or enclosure the body/enclosure must be grounded to earth. If a protective metal housing is used, the
metal housing should be grounded to earth.
4. If a protective plastic housing is used, the housing must be able to withstand at least 2 kV from the housing to earth ground.
MaxAmbient Light
Black Container
390 FT-CD
Opaque Container
30 FT-CD
tupnI rewoP noitautcA noitidnoC
epyTgnitnuoM
TPN"4/1B2-FNU02-"2/18-1x21M
CDV5 %01±
teW 405422105422805422
yrD 5054222
05422905422
21 CDV %01±
teW 605422305422015422
yrD 705422526322115422
Gems Sensors Inc.
One Cowles Road
Plainville, CT 06062-1198
Toll-Free: 1-800-378-1600
• Gems products must be maintained and installed in strict accordance with the National
Electrical Code and the applicable Gems Product Instruction Bulletin that covers
installation, operation and proper maintenance. Failure to observe this information may
result in serious injury or damages.
•Please adhere to the pressure and temperature limitations shown throughout this catalog
for our sensors. These limitations must not be exceeded. These pressures
and temperatures take into consideration possible system surge pressures, temperatures
and their frequencies.
•Selection of materials for compatibility with the media is critical to the life and operation of
Gems products. Take care in the proper selection of materials of construction, testing is
required.
•Our sensors have been designed to resist shock and vibration. However, shock and
vibration should be minimized.
•Filter liquid media containing particulate and/or debris to ensure the proper operation of
our products.
•Electrical entries and mounting points in an enclosed tank may require liquid/vaporsealing.
•Our sensors must not be field-repaired.
•Physical damage sustained by product may render it unserviceable.
Important Points:
Mating O-Ring Gland Design
1/2" Straight Thread
M12 X 1 Straight Thread
SAE #5 TUBE FITTING 12° .555
.550 .108
.093
1/2-20 UNF 2B
.020
[.50mm]
∅ .570 ± .002
[14.5±.05 mm]
M12 X 1 - 8G
45°
.150
[3.80mm]
∅(AT TOP)
Dimensions
11/16” NPT Mounting M12 x 1 1/2”-20 UNF
Electrical Termination: Lead Wires, 22 AWG, TPE Jacketed, 6” to 8” Extended
1 7/16”
(35.36 mm)
11/16” HEX
(17.5 mm)
1/4”
NPT
1 7/16”
(35.36 mm)
11/16” HEX
(17.5 mm)
M12 x 1 O-Ring
1 7/16”
(35.36 mm)
11/16” HEX
(17.5 mm)
1/2” - 20 UNF O-Ring
Typical Wiring Diagrams
External Load
Note: May sink up to 40mA max. Use diode
suppression for inductive loads.
L
o
a
d+
-Sensor
Red
Black
Blue
5 or
12Vdc
CMOS / TTL Output
Max. current sink = 40mA
R = Pull-up Resistor (4.7K 1/4W)
+
-Sensor
Red
Black
Blue
5 Vdc
R
Output
Relay Output
+
-Sensor
Red
Black
Blue
5 or
12Vdc
V
*
Note: * For inductive loads, use diode suppression
+
-
Return Policy
Returns are accepted on stock items up to 30 days from date
of order. You must contact our Returns Department for a
Return Authorization (RA) number. Return the goods - freight
prepaid - in the original container and include original packing
slip.
C. O. D. returns are not accepted. Gems reserves the right
to apply restocking charges.
Tel: 860-793-4357
Fax: 860-793-4563
Maintenance
Sensor may require a periodic cleaning of prism surface. Chlorinated hydrocarbons must not be used for cleaning.
A mild detergent may be used to clean prism surface.

Other Gems Sensors Switch manuals

Gems Sensors FS-380 Series Training manual

Gems Sensors

Gems Sensors FS-380 Series Training manual

Gems Sensors S-925 Series Training manual

Gems Sensors

Gems Sensors S-925 Series Training manual

Gems Sensors FS-500 Series Training manual

Gems Sensors

Gems Sensors FS-500 Series Training manual

Popular Switch manuals by other brands

Omron ER1022 datasheet

Omron

Omron ER1022 datasheet

Crystal Image Technologies KVM-IP8 user manual

Crystal Image Technologies

Crystal Image Technologies KVM-IP8 user manual

LevelOne GTL-2872 Quick installation guide

LevelOne

LevelOne GTL-2872 Quick installation guide

EGS Appleton MD2SR Series instruction sheet

EGS

EGS Appleton MD2SR Series instruction sheet

Brocade Communications Systems 6505 quick start guide

Brocade Communications Systems

Brocade Communications Systems 6505 quick start guide

Alfa Network AFS316 Specifications

Alfa Network

Alfa Network AFS316 Specifications

D-Link DGS-3100-24 Specifications

D-Link

D-Link DGS-3100-24 Specifications

Cisco Linksys EZXS88W user guide

Cisco

Cisco Linksys EZXS88W user guide

Seojin Instech SMC instruction manual

Seojin Instech

Seojin Instech SMC instruction manual

Mellanox Technologies Spectrum SN2700 user manual

Mellanox Technologies

Mellanox Technologies Spectrum SN2700 user manual

Avea Kramer VP Series Connection guide

Avea

Avea Kramer VP Series Connection guide

Dell PowerEdge 1855 quick start guide

Dell

Dell PowerEdge 1855 quick start guide

Philips SWV2052W Specifications

Philips

Philips SWV2052W Specifications

GRAS RA0056 instruction manual

GRAS

GRAS RA0056 instruction manual

Fantec UMP-4U31-C user manual

Fantec

Fantec UMP-4U31-C user manual

Jensen RAINSWITCH Upgrade instructions

Jensen

Jensen RAINSWITCH Upgrade instructions

Lutron Electronics MAESTRO MS-OPS2 Installation instruction

Lutron Electronics

Lutron Electronics MAESTRO MS-OPS2 Installation instruction

Allied Telesis WebSmart AT-FS750/24POE datasheet

Allied Telesis

Allied Telesis WebSmart AT-FS750/24POE datasheet

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.