Drive
Technical Operating IFU
• Connection of head: Insert the attachment
head [4] onto Gentlele-drive [5], joint
and push rmly to complete connection.
• Insertion of File: Insert the le into the
hole in the attachment head [4] by gently
pushing the le into the head. (grip by
friction power only)
•Gentlele drive operation:
• 1. Place the le in the canal and press the
on-o button [3] briey.
• 2. Verify that the light indicator [2] is on,
and not blinking. The Gentlele-drive [5]
will continue until the on-o button [3] is
pressed again briey.
• 3. Before extracting the le from the canal,
push the on-o button [3] again to stop
operation.
Charging & Maintenance
•Connect the charger’s plug to the unit’s
socket [1], located on the back lid of the
unit [2]. When the battery is full, charging
will stop automatically (Green light
indication on the charger will appear).
When charging is completed, unplug the
unit from the charger.
• Note: It is recommended to charge the
device prior to each procedure.
• Note: To prolong battery lifetime, it is
recommended to disconnect the device
from the charger after 4-5 hours.
• Alerts: When the battery charge is low, the
indicator light [2] on the back lid of the unit
will blink.
Control and Safety
•The Gentlele drive includes inherent
safety mechanisms.
1. Motor Power control - when resistance
is encountered, the motor’s rotation
speed declines, reaching a full stop in the
continue presence of high resistance.
2. Motor Load Control - When the resistance
in the canal rises, controller rmware
detects the increase in load and limits the
motor power to a safe predened value.
3. After stoppage of the le (should that
occur), the le may be released by
manually pulling it clock-wise.
Single use attachment head
GF-CAP-01
IFU
Gentlele - Drive
GF-DRV-02
Chahrging socket
➍
➊
➋
➌
➎