Graetz GammaTwin User manual

Operaon Manual
for the
GRAETZ Dose Rate Meter
GammaTwin
Release: October 2017, Soware Version 1.4/1.5 Part Number 5641 9877
23.51
10.02

Soware Version 1.4/1.5
GammaTwinOperaon Manual
2

GammaTwin
Soware Version 1.4/1.5
Operaon Manual
3
1. Inhalt
1. Preface .................................................................... 5
2. Introducon ............................................................ 6
2.1 Possibilies of use ............................................... 6
2.2 Control and display ele men ts ............................. 8
3. Pung into operaon ............................................. 9
3.1 Before use ........................................................... 9
3.2 Switching ON/OFF ............................................... 9
3.3 Indicaon .......................................................... 10
3.3.1 „Twin“-indicaon .................................... 11
3.3.2 Dose indicaon ........................................ 12
3.3.3 Dose rate indicaon ................................ 13
3.4 Single-pulse indicaon and funconal control
signal ............................................................... 14
3.5 Display illuminaon .......................................... 14
3.6 Acknowledgement of acousc alarms .............. 15
4. Alarm messages .................................................... 16
4.1 Dose and dose rate alarm ................................. 16
4.2 Baery alarm .................................................... 17
5. The menu .............................................................. 17
5.1 Acvate the menu ............................................. 17
5.1.1 Menu operaon ....................................... 17
5.1.2 Dose and dose rate alarm thresholds ...... 18
5.1.3 Acouscs .................................................. 18
5.1.4 Dose ......................................................... 19

Soware Version 1.4/1.5
GammaTwinOperaon Manual
4
5.1.5 Seng basic display ................................. 19
6. Possible defects .................................................... 20
6.1 Error messages .................................................. 20
6.2 Baeries discharged ......................................... 20
6.3 Other error messages ....................................... 21
6.4 Switching on ...................................................... 21
7. Disposal of disused devices .................................. 24
8. Technical data ....................................................... 25
9. Repair and maintenance ....................................... 31

GammaTwin
Soware Version 1.4/1.5
Operaon Manual
5
1. Preface
To guarantee a faultless handling of the dose
rate meter GRAETZ GammaTwin, please read
the following operaon manual carefully.
Aer having inserted the baeries and swit-
ched on the instrument, the user can immedi-
ately start measurement of ionizing radiaon.
Please note: When applied in pulsed radiaon
elds the measuring results of the GammaTwin
are not reliable as measurements are based on
Geiger-Müller-tube technology. Thus, it is not
suitable for pulsed X-rays.

Soware Version 1.4/1.5
GammaTwinOperaon Manual
6
2. Introduction
2.1 Possibilies of use
The GammaTwin is a very robust, reliable
and easy to operate PTB-approved dose rate
meter. It is used for the safe determinaon of
the current local dose rate as well as the accu-
mulated dose of photon radiaon. The measu-
rands are the ambient dose equivalent rate
Ḣ*(10) as well as the ambient dose equivalent
H*(10).
The ambient dose equivalent rate Ḣ*(10) as
well as the ambient dose equivalent H*(10) are
PTB-approved.

GammaTwin
Soware Version 1.4/1.5
Operaon Manual
7
The GammaTwin oers the following features:
9acvatable acousc single-pulse indica-
on
9acvatable acousc funcon control
94 dose and dose rate alarm thresholds
each
9alternavely either simultaneous or
separate indicaon of dose or dose rate
9automac and connuous storage of
the dose
9simple menu navigaon
9illuminable display
9measured peak value of dose rate since
switching on the GammaTwin

3
6
4
5
Soware Version 1.4/1.5
GammaTwinOperaon Manual
8
2.2 Control and display elements
1 privileged direcon of irradiaon
and posion of reference point in
12 mm depth
2 LC- display
3 buon for ON/OFF and background
illuminaon
4 buon for menu and acousc
single-pulse indicaon
5 screw for baery compartment
6 sound source
1
2

GammaTwin
Soware Version 1.4/1.5
Operaon Manual
9
3. Putting into operation
3.1 Before use
Before starng the instrument the user has to
ensure that the GammaTwin is not damaged
and is correctly equipped with two baeries of
the type AAA 1,5 V (polarity acc. to the corres-
ponding label). For perfect contacng, please
ghten the screw of the baery compartment
by means of a coin. Remove the protecve foil
from the display.
3.2 Switching ON/OFF
The GammaTwin is turned on by briey press-
ing the -buon. The instrument passes a self-
test, in which the user may control the alarm
signal and all segments of the display. Further-
more, the serial number, the soware version
as well as the check sum of the soware are
displayed. The PTB-approved version of the
rmware and the corresponding check sum are
indicated in chapter 8 (technical data). If there
is no error, the GammaTwin is directly starng
operaon.

Soware Version 1.4/1.5
GammaTwinOperaon Manual
10
When pressing the buon for more than 2
seconds during operaon, the instrument is
switched o and the following safety check will
appear:
All acve sengs and the acousc single pulse
indicaon will be stored. An acvated acousc
alarm or acousc error messages have to be
acknowledged before switching o the Gam-
maTwin.
3.3 Indicaon
The user can either select the separate indica-
on of dose and dose rate or the simultaneous
indicaon („Twin“-indicaon). The dual „Twin“-
indicaon is preset by GRAETZ.

GammaTwin
Soware Version 1.4/1.5
Operaon Manual
11
3.3.1 „Twin“-indicaon
Displayed elements:
• bargraph (analog/logarithmic
indicaon of the dose rate)
• dose rate
• dose
• acvated dose rate alarm threshold
(DRA)
• acvated dose alarm threshold (DA)
• funcon control
(ashing once per second)
• loudspeaker symbol (when acousc
single-pulse indicaon is acvated)
• baery symbol

Soware Version 1.4/1.5
GammaTwinOperaon Manual
12
3.3.2 Dose indicaon
Displayed elements:
• bargraph (shows the dose in per-
centages of the acvated alarm
threshold)
• dose
• acve alarm thresholds (DRA/DA – as
shown above)
• funcon control (ashing once per
second)
• loudspeaker symbol (when acousc
single-pulse indicaon is acvated)
• baery symbol

GammaTwin
Soware Version 1.4/1.5
Operaon Manual
13
3.3.3 Dose rate indicaon
Displayed elements:
• bargraph (analog/logarithmic
indicaon of the dose rate)
• dose rate
• measured peak value since switching
on the GammaTwin
• acve alarm thresholds (DRA/DA – as
shown above)
• funcon control (ashing once per
second)
• loudspeaker symbol (when acousc
single-pulse indicaon is acvated)
• baery symbol

Soware Version 1.4/1.5
GammaTwinOperaon Manual
14
3.4 Single-pulse indicaon and funconal
control signal
By briey pressing the buon the acousc
single-pulse indicaon is acvated respecvely
deacvated. When acvated, the loudspeaker
symbol is displayed and each GM-tube pulse
triggers a brief sound.
This seng can be changed in the menu
„Acouscs“, where it is possible to reduce the
pulse rate of the background radiaon in such
a way that only the tweneth GM-tube pulse
will trigger a sound. Addionally, a funconal
control signal sounds approx. every two min-
utes, indicang the GammaTwin is correctly
working. This signal can be combined with both
acousc single-pulse indicaons.
3.5 Display illuminaon
During operaon the display illuminaon can
be turned on by pressing the buon .
By briey pressing the key once again, the illu-
minaon is turned o. For saving the baeries
energy, the illuminaon automacally turns o
aer approx. 10 seconds.

GammaTwin
Soware Version 1.4/1.5
Operaon Manual
15
During menu operaon it is not possible to
switch illuminaon on or o, because then the
corresponding buon has another funcon.
Thus, the illuminaon has to be acvated be-
fore menu operaon. Illuminaon will switch
o when leaving the menu.
3.6 Acknowledgement of acousc alarms
Triggered alarms or error messages can be ac-
knowledged by briey pressing the buon .
Then, the acousc alarm signal is switched o,
but the opcal warning remains sll acve.
In case of serious errors, the GammaTwin auto-
macally switches o aer acknowledgement
(see chapter 6).

Soware Version 1.4/1.5
GammaTwinOperaon Manual
16
4. Alarm messages
4.1 Dose and dose rate alarm
When an acvated alarm threshold is exceed-
ed, the display is blinking.
In case of an exceeded dose alarm threshold,
the display switches to the dose indicaon
and in case of an exceeded dose rate alarm
threshold to the dose rate indicaon. If both,
the dose and the dose rate alarm threshold are
exceeded, the „Twin“-indicaon is displayed.
In case of a dose rate alarm short constant
acousc warning signals will be triggered. If the
acousc alarm has been acvated in the menu,
the warning signal sounds every second.
Such an alarm can be acknowledged by means
of the -key. Thus, the acousc alarm is swit-
ched o and the menu returns to the selec-
ted standard display. But, the triggered alarm
threshold is sll blinking.
The acousc alarm will be re-triggered as soon
as the measured value falls below an already
acknowledged dose rate alarm threshold and is
then exceeded again.

GammaTwin
Soware Version 1.4/1.5
Operaon Manual
17
4.2 Baery alarm
The GammaTwin regularly checks the charge of
the baeries. The displayed baery symbol in-
dicates the residual capacity. When this signal
starts blinking, the baeries have to be repla-
ced as soon as possible (please also see chapter
6.2).
5. The menu
5.1 Acvate the menu
The GammaTwin’s menu is acvated by press-
ing the buon for at least 2 seconds. If for 30
seconds no key is pressed during menu opera-
on, the instrument automacally returns to
the preset standard display. Sengs changed
within a menu point will be taken over and
stored.
5.1.1 Menu operaon
There are ve dierent menu points the user
can scroll through by means of the -buon.

Soware Version 1.4/1.5
GammaTwinOperaon Manual
18
5.1.2 Dose and dose rate alarm thresholds
In these both menus the acvated alarm
threshold is marked by a hook. By using the
buon it is possible to scroll through the single
alarm thresholds. With the -buon the inver-
se shown alarm threshold is chosen (black line,
white gures) and acvated. By pressing this
-buon again, the next menu point is selected.
5.1.3 Acouscs
By using the -buon it is possible to scroll
through the lines. With the -buon the inver-
se shown opon can be selected. This selecon
is marked by a hook. By pushing the buon

GammaTwin
Soware Version 1.4/1.5
Operaon Manual
19
in the line „accept“, the seng is taken over
and the GammaTwin changes to the next menu
point.
5.1.4 Dose
In this menu the user can irrevocably clear the
accumulated dose by pressing the buon . By
using the buon it is possible to scroll to the
next menu point without clearing the dose.
5.1.5 Seng basic display
Here, the user sets the indicaon which will ap-
pear when starng the GammaTwin and when
leaving the menu.

Soware Version 1.4/1.5
GammaTwinOperaon Manual
20
The acvated basic seng is marked by a hook.
By means of the -buon, it is possible to
scroll through the display indicaons. With the
buon the inverse shown line can be selec-
ted and taken over as standard display. For lea-
ving the menu, the buon has to be pressed
again.
6. Possible defects
6.1 Error messages
Each me being switched on, the GammaTwin
passes a self-test. Addionally, all important
funcons are connuously controlled during
operaon. In case of an error, the user will be
informed by a corresponding error message.
6.2 Baeries discharged
If baery voltage falls below 1.8 volt, opera-
on of the GammaTwin cannot be connued.
By pressing the buon the instrument swit-
ches o. Remove and dispose the baeries im-
mediately, even if there are no new baeries
available.
Table of contents
Popular Measuring Instrument manuals by other brands

Hach
Hach DR/2010 instruction sheet

EUTECH INSTRUMENTS
EUTECH INSTRUMENTS CON 6 instruction manual

Chauvin Arnoux
Chauvin Arnoux C.A 1226 user manual

YOKOGAWA
YOKOGAWA 96035 user manual

Dwyer Instruments
Dwyer Instruments DPMA-5 Series Installation and operating instructions

ELOKON
ELOKON ELOSHIELD instruction manual