Speculum insertion and removal: Position the selected speculum on the
otoscope socket (speculum size is marked on its outside). Turn the specu-
lum clockwise until you feel a resistance. To remove the speculum, turn it
counter-clockwise until it is free of the otoscope socket.
mAiNteNANce
Replacing the bulb: Unscrew the instrument head to remove it from the bat-
tery handle. The bulb is in the bottom section of the instrument head. Re-
move the bulb from the instrument head with your thumb and forenger or a
suitable tool. Insert the new bulb rmly.
Battery use and care: Use two AA alkaline batteries only. If the otoscope will
not be used for an extended period of time, remove the batteries. Dispose of
used batteries in accordance with all applicable laws and regulations.
Cleaning and sterilization: Dampen a clean cloth with a mild detergent and
water solution, a solution of 70% isopropyl alcohol and water solution, or a
10% bleach and water solution. Wring out the cloth to remove excess uid. Use the cloth to wipe
all external surfaces of the otoscope. Ensure that no fluid enters the otoscope. Clean the instru-
ment glass with a cotton swab dipped in isopropyl alcohol. The reusable ear specula may be steril-
ized in 134°C saturation steam for a dwell time of ten minutes.
SpeciFicAtiONS
Optical specications Illumination form Fiber optics transmit light
Lens magnication 2.5x
Otoscope port View tympanic membrane using the FiberOptic Otoscope bulb
Ear specula specications Ear specula diameter 2.5mm, 4mm
Light port specications Light port Yes. An otoscope bulb can be tted to the port.
Electrical specications Bulb model T4954 2.5V, 1.6W
Rated voltage Two AA alkaline batteries (included)
Operation mode On time Intermittent; ≤ one minute
Off time ≥ ve minutes
Electrical safety standard IEC60601-1, IEC60601-1-2
Environmental requirements Use Temperature 50°F — 95°F (10°C — 35°C)
Relative humidity 30% — 75%
Atmospheric pressure 10.15 psi — 15.37 psi (70 kPa — 106 kPa)
Shipping and storage Temperature -40°F — 131°F (-40°C — 55°C)
Relative humidity 10% — 90%
Atmospheric pressure 7.25 psi — 15.37 psi (50 kPa — 106 kPa)
Replacement parts Specula size 2.5mm (8617-1); Specula size 4.0mm (8617-2); Replacement bulb (8617-3)
WArrANty
GF Health Products, Inc. warrants the Grafco 8617 FiberOptic Otoscope (not including specula) for a
period of one year for defects in workmanship and materials. During the warranty period, defec-
tive items will be repaired or replaced at manufacturer's option at no charge.
bulb
battery
handle
socket
instrument
head
speculum
8617
FiberOptic
Otoscope
GF, Graham-Field, and Grafco are registered trademarks of GF Health Products, Inc.