GW Instek ASR-6000 Series User manual

Multi-phase Programmable AC/DC
Power Source
ASR-6000 Series
QUICK START GUIDE
ISO-9001 CERTIFIED MANUFACTURER

This manual contains proprietary information, which is protected by
copyright. All rights are reserved. No part of this manual may be
photocopied, reproduced or translated to another language without
prior written consent of Good Will company.
The information in this manual was correct at the time of printing.
However, Good Will continues to improve products and reserves the
rights to change specification, equipment, and maintenance
procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.

Table of Contents
3
Table of Contents
INTRODUCTION ............................................................. 4
ASR-6000 Series Overview .......................................4
Appearance .............................................................5
SET UP .......................................................................... 13
Input Terminal Connection .................................... 13
Copper Plate Introduction ..................................... 13
Single Phase Connection ....................................... 14
Delta Connection................................................... 15
Y Connection ......................................................... 16
Output Terminal Connection ................................. 17
1P2W Output Connection ...................................... 17
1P3W Output Connection ...................................... 19
3P4W Output Connection ...................................... 20
MISCELLANEOUS .......................................................... 21
Firmware Update ................................................... 21
Specifications ........................................................ 23
Information of Name Order................................... 25
ASR-6000 Dimensions ........................................... 26
Declaration of Conformity ..................................... 27

Table of Contents
4
INTRODUCTION
ASR-6000 Series Overview
Series lineup
The ASR-6000 series consists of 2 models, the ASR-6450 and ASR-
6600, differing in capacity. Note that throughout the user manual, the
term “ASR-6000” refers to any of the models, unless stated otherwise.
1P Output Condition
Model Name
Power Rating
Max. Output Current
Max. Output Voltage
ASR-6450
4500 VA
45 / 22.5 A
350 Vrms / 500 Vdc
ASR-6600
6000 VA
60 / 30 A
350 Vrms / 500 Vdc
1P3W Output Condition
Model Name
Power Rating
Max. Output Current
Max. Output Voltage
ASR-6450
3000 VA
15 / 7.5 A
700 Vrms / 1000 Vdc
ASR-6600
4000 VA
20 / 10 A
700 Vrms / 1000 Vdc
3P Output Condition (Pre phase)
Model Name
Power Rating
Max. Output Current
Max. Output Voltage
ASR-6450
1500 VA
15 / 7.5 A
350 Vrms / 500 Vdc
ASR-6600
2000 VA
20 / 10 A
350 Vrms / 500 Vdc

INTRODUCTION
5
Appearance
Front Panel
F 1
F 2
F 3
F 4
F 5
F 6
F 7
F 8
Programmable AC/DC Power Source
6450
-
ASR
0Lock
Unlock
1
Wave
23
456
78 9
V F I rms
Menu Test Preset
Enter
Cancel
Shift
Range
Off Phase ALM CLR
On Phase IPK CLR
V-Limit F-Limit IPK-Limit
Hardcopy
: Long Push
Output
/
Local
Mode
Phase
1 2 3 4 5 6 7 8
A
9
B
C
D
E
F
GHIJK
Item Index
Description
1
Power switch button
2
USB interface connector (A Type)
3
LCD screen
4
Function keys (blue zone)
5
Menu key
6
Test key
7
Preset key
8
Scroll wheel
9
Range key/Output mode key
A
Arrow keys
B
Output key
C
Shift key

ASR-6000 Series Quick Guide
6
D
Cancel key
E
Enter key
F
Irms/IPK-Limit button
G
Lock/Unlock button
H
F/F-Limit button
I
V/V-Limit button
J
Numerical Keypad with additional “Shift + key”
shortcut functions (green zone)
K
Air inlet
Item
Description
Power Switch
Turn on the mains power
USB A Port
The USB port is used for data transfers
and upgrading software. Also, it is
available for screenshot hardcopy.
It supports FAT32 format with maximum 32G storage.
LCD Screen
Displays the setting and measured
values or menu system
Function Keys
Assigned to the functions displayed on
the right side of the screen.

INTRODUCTION
7
Menu Key
Enters the Main menu or goes back to
one of the display modes.
Test Key
Puts the instrument into the Sequence
and Simulation control mode.
Preset Key
Puts the instrument into Preset mode.
Arrow Keys
The arrow keys are used to select the digit
power of a value that is being edited.
Range Key
Switches between the 100V, 200V and
AUTO ranges
Output Mode
+
Selects between the AC+DC-INT, AC-
INT, DC-INT, AC+DC-EXT, AC-EXT,
AC+DC-ADD, AC-ADD, AC+DC-Sync,
AC-Sync and AC-VCA modes.
Scroll Wheel
Used to navigate menu items or for
increment/decrement values one step at
a time.
Output Key
Turns the output on or off.
Shift Key
Turns on the shift state, which enables
shortcut operations with an icon
indicated on the top status bar. The shift
state, which allows continuous shortcut
operations, is kept until another press
on shift key again.
When performing shortcut operations, press
shift key followed by another shortcut function
key. Do Not press both shift key and shortcut
function key simultaneously.
Cancel Key
Used to cancel function setting menus
or dialogs.

ASR-6000 Series Quick Guide
8
Enter Key
Confirms selections and settings.
Irms
Used for setting the maximum output
current.
IPK-Limit
+
Used to set the peak output current limit
value.
Lock/Unlock Key
Used to lock or unlock the front panel
keys except output key. Simply press to
lock, whilst long press to unlock.
F
Used for setting the output frequency
(DC mode N/A).
F-Limit
+
Used for setting the output frequency
limit value (DC mode N/A).
V
Used for setting the output voltage.
V-Limit
+
Used for setting the output voltage limit
value.
Keypad
Used to input power of a value directly.
The key is used to input decimal /
plus or minus.

INTRODUCTION
9
On Phase
+
Sets the on phase for the output voltage.
Off Phase
+
Sets the off phase for the output voltage.
Output
Waveform
+
Selects between the Sine, Square,
Triangle and ARB 1~16 waveforms
(not available for DC-INT, AC+DC-EXT
and AC-EXT).
Local Mode
+
Switches operation back to local mode
from remote mode.
IPK CLR
+
Used to clear peak output current value.
ALM CLR
+
Clears alarms.
Hardcopy Key
+
Used to take a screenshot. Make sure an
USB flash disk in well inserted before
the action.
Output Phase
+
Used to prompt the output phase
window where 1P2W, 1P3W and 3P4W
modes are available for selection.

ASR-6000 Series Quick Guide
10
Rear Panel
GPIB
LAN RS232C EXT I/OOUTIN
S M
3P 1P3W
L
N
L
L
N
N
1P
INPUT
:L-N
( )
:L-L
( ) 200 240V 1 /3F F
8kVA MAX.
47 63Hz
OUTPUT
AC 0 350V 1 2kHz
DC 500V 500V
N2L2 L3 N3
N1L1 SENSING
L3
N
L2
L1
N
N
X
N
L2
L1
N
X
N2
L2
L3
N3
N1
L1
2
1346789 5
Item Index
Description
1
Output terminal
2
AC power input terminal
3
Remote sensing input terminal
4
External I/O connector
5
External IN/OUT connection in parallel function
6
RS232 connector
7
Ethernet (LAN) connector
8
USB interface connector (B Type)
9
Optional interface Slot
GPIB card (ASR-003)
DeviceNet card (ASR-004)
CAN BUS card (ASR-005)

INTRODUCTION
11
Item
Description
Output Terminal
Output terminal
(M4 screw type, 8 ~ 18 AWG)
(Screw torque value:18kgf-cm)
AC Power Input
Terminal
AC inlet
(M4 screw type, 8 ~ 18 AWG)
(Screw torque value:18kgf-cm)
Remote Sensing
Input Terminal
Remote sensing input terminal is
for compensation of load wire
voltage drop.
(M2.5 screw type, 12 ~ 30 AWG)
(Screw torque value: 0.5N*m)
(Strip length: 7 ~ 8mm)
External Control
I/O Connector
Used to control ASR-6000 externally
by using the logic signal and
monitor Sequence function status.
External IN/OUT
Connection in
Parallel Function
The IN (Slave) and OUT (Master)
ports are used for connection with
external unit in parallel function.

ASR-6000 Series Quick Guide
12
RS232C
Connector
The RS232C connector for
controlling the ASR-6000 remotely.
Ethernet LAN
Port
The Ethernet port is used for
remote control.
USB B-type Port
USB port for controlling the ASR-
6000 remotely.
Optional GPIB
Connector
The optional GPIB connector for
controlling the ASR-6000 remotely.
Optional CAN
BUS Connector
The optional CAN BUS connector
for controlling the ASR-6000
remotely.
Optional
DeviceNet
Connector
The optional DeviceNet connector
for controlling the ASR-6000
remotely.

SET UP
13
SET UP
Input Terminal Connection
Background
Basically, the input terminal, which is located in
the rear panel of ASR unit, can be connected
through 3 methods: Single Phase, Delta and Y
Connection. Depending on varied input
connections, use the corresponding cooper plates
and power cords for connection. Refer to the
following chapters for details of each connection.
Copper Plate Introduction
Voltage Range of
Input Connections
Input
Connection
Voltage
Range
Single
Single Phase 200 - 240V: L, N and G
Delta
Three Phase 200 - 240V: L1, L2, L3 and G
Y
Three Phase 200 - 240V: L1, L2, L3, N and G
Copper Plate
Description
Copper Plate
Description
62SR-6K0CP101
Copper plate for delta
connection input
62SR-6K0CP201
Copper plate for single phase
and Y connection input
62SR-6K0CP301
Copper plate for delta
connection input
Copper Plate
Quantity of Input
Connections
Input
Connection
Quantity of
Copper Plate
Single
62SR-6K0CP201*2pcs
Delta
62SR-6K0CP101*1pcs, 62SR-6K0CP301*2pcs
Y
62SR-6K0CP201*1pcs

ASR-6000 Series Quick Guide
14
Single Phase Connection
Steps
1. Assemble the two
copper plates specific
for Single phase input
connection with the
AC input terminals.
The first plate is for L1,
L2 and L3 terminals,
while the other plate is
for N1, N2 and N3
terminals.
2. Connect the AC power
cords to the AC input
terminals.
Red Line (L)
Black Neutral (N)
Note
Power input cords are not included in this product.
For the specific installation on protective lid of
input terminal, please refer to User Manual.

SET UP
15
Delta Connection
Steps
1. Assemble the three
copper plates specific
for Delta input
connection with the
AC input terminals.
The 1st plate is for N3
and L1 terminals. The
2nd second plate is for
L3 and N2 terminals,
while the 3rd plate is
for L2 and N1
terminals.
2. Connect the AC power
cords to the AC input
terminals.
Red Line (N2)
Green Neutral
(N1)
Yellow Neutral
(N3)
Note
Power input cords are not included in this product.
For the specific installation on protective lid of
input terminal, please refer to User Manual.

ASR-6000 Series Quick Guide
16
Y Connection
Steps
1. Assemble the copper
plate specific for Y
input connection with
the AC input
terminals. The copper
plate is for N1, N2 and
N3 terminals.
2. Connect the AC power
cords to the AC input
terminals.
Red L3
Green L2
Yellow L1
Blue Neutral
Note
Power input cords are not included in this product.
For the specific installation on protective lid of
input terminal, please refer to User Manual.

SET UP
17
Output Terminal Connection
Background
The output terminal can output power in three
mode: 1P2W, 1P3W and 3P4W. Select applicable
output mode, via panel configurations, in
accordance with varied applications.
WARNING
Be aware of dangerous voltages. Ensure that the
power to the instrument is disabled before
handling the power supply output terminals.
Failing to do so may lead to electric shock.
CAUTION
After configuring phase settings via the front panel,
please make sure the cords connection on the rear
panel is corresponding to the set configuration.
1P2W Output Connection
Steps
1. Disconnect the unit from the mains power
socket and turn the power switch off.
2. Assemble the two
copper plates specific
for 1P2W output
connection with the AC
output terminals. The
first plate is for N*3
terminals, while the
other plate is for L*3
terminals.

ASR-6000 Series Quick Guide
18
3. Connect the output
wires to the AC output
terminals as follows:
Red Line (L)
Black Neutral (N)
Note
The input & output terminals necessitate
connectivity through ring-type connectors.
For the specific installation on protective lid of
output terminal, please refer to User Manual.
Grounded Neutral Output for 1P2W output only:
ASR-6000 allows for a grounded return on the neutral
output. It is suit for the medical industry that required
between ground with neutral is 0 V essentially. And
possible to mitigate ground loops that is ideal for
reduce ground noise and isolate sensitive equipment
from the effects of ground loops.
WARNING
Because the neutral has been referenced to the
chassis ground, be careful electric shock by yourself.

SET UP
19
1P3W Output Connection
Steps
1. Disconnect the unit from the mains power
socket and turn the power switch off.
2. Connect the output
wires to the AC output
terminals as follows:
Yellow Line (L1)
Green Line (L2)
Blue Neutral (N)
Note
The input & output terminals necessitate
connectivity through ring-type connectors.
For the specific installation on protective lid of
output terminal, please refer to User Manual.

ASR-6000 Series Quick Guide
20
3P4W Output Connection
Steps
1. Disconnect the unit from the mains power
socket and turn the power switch off.
2. Connect the output
wires to the AC output
terminals as follows:
Yellow Line (L1)
Green Line (L2)
Red Line (L3)
Blue Neutral (N)
Note
The input & output terminals necessitate
connectivity through ring-type connectors.
For the specific installation on protective lid of
output terminal, please refer to User Manual.
Other manuals for ASR-6000 Series
1
This manual suits for next models
2
Table of contents
Other GW Instek AC Power Distribution manuals

GW Instek
GW Instek APS-7000 Series User manual

GW Instek
GW Instek APS-7100 Owner's manual

GW Instek
GW Instek APS-1102 User manual

GW Instek
GW Instek ASR-2000 Series Owner's manual

GW Instek
GW Instek ASR-6000 Series User manual

GW Instek
GW Instek APS-7100 Owner's manual

GW Instek
GW Instek APS-7051 User manual

GW Instek
GW Instek APS-7050E User manual

GW Instek
GW Instek APS-1102A User manual