manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hanna Instruments
  6. •
  7. Chemical Test
  8. •
  9. Hanna Instruments HI 38072 User manual

Hanna Instruments HI 38072 User manual

Other Hanna Instruments Chemical Test manuals

Hanna Instruments HI 3834 User manual

Hanna Instruments

Hanna Instruments HI 3834 User manual

Hanna Instruments HI 3855 User manual

Hanna Instruments

Hanna Instruments HI 3855 User manual

Hanna Instruments HI 38049 User manual

Hanna Instruments

Hanna Instruments HI 38049 User manual

Hanna Instruments HI 38061 User manual

Hanna Instruments

Hanna Instruments HI 38061 User manual

Hanna Instruments HI 3822 User manual

Hanna Instruments

Hanna Instruments HI 3822 User manual

Hanna Instruments HI 3873 User manual

Hanna Instruments

Hanna Instruments HI 3873 User manual

Hanna Instruments HI 3844 User manual

Hanna Instruments

Hanna Instruments HI 3844 User manual

Hanna Instruments HI 38016 User manual

Hanna Instruments

Hanna Instruments HI 38016 User manual

Hanna Instruments HI 38067 User manual

Hanna Instruments

Hanna Instruments HI 38067 User manual

Hanna Instruments HI 3880 User manual

Hanna Instruments

Hanna Instruments HI 3880 User manual

Hanna Instruments HI 3847 User manual

Hanna Instruments

Hanna Instruments HI 3847 User manual

Hanna Instruments HI 3859 User manual

Hanna Instruments

Hanna Instruments HI 3859 User manual

Hanna Instruments HI 38017 User manual

Hanna Instruments

Hanna Instruments HI 38017 User manual

Hanna Instruments HI 3854 User manual

Hanna Instruments

Hanna Instruments HI 3854 User manual

Hanna Instruments HI 3875 User manual

Hanna Instruments

Hanna Instruments HI 3875 User manual

Hanna Instruments HI 98103 User manual

Hanna Instruments

Hanna Instruments HI 98103 User manual

Hanna Instruments HI 3897 User manual

Hanna Instruments

Hanna Instruments HI 3897 User manual

Hanna Instruments HI 3838 User manual

Hanna Instruments

Hanna Instruments HI 3838 User manual

Hanna Instruments HI 3845 User manual

Hanna Instruments

Hanna Instruments HI 3845 User manual

Hanna Instruments HI 38001 User manual

Hanna Instruments

Hanna Instruments HI 38001 User manual

Hanna Instruments HI 755 User manual

Hanna Instruments

Hanna Instruments HI 755 User manual

Hanna Instruments HI 3849 User manual

Hanna Instruments

Hanna Instruments HI 3849 User manual

Hanna Instruments HI 3850 User manual

Hanna Instruments

Hanna Instruments HI 3850 User manual

Hanna Instruments HI 3857 User manual

Hanna Instruments

Hanna Instruments HI 3857 User manual

Popular Chemical Test manuals by other brands

JBL PRO AQUATEST pH 7.4-9.0 INFORMATION FOR USE

JBL

JBL PRO AQUATEST pH 7.4-9.0 INFORMATION FOR USE

INDICAID 2120100 quick start guide

INDICAID

INDICAID 2120100 quick start guide

Elitech ENVOY 500 HDL manual

Elitech

Elitech ENVOY 500 HDL manual

LaMotte WaterLink Spin Touch quick start guide

LaMotte

LaMotte WaterLink Spin Touch quick start guide

JBL PRO AQUATEST O2 INFORMATION FOR USE

JBL

JBL PRO AQUATEST O2 INFORMATION FOR USE

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Instruction Manual
www.hannainst.com
DearCustomer,
Thank you for choosing a Hanna Product.
Please read the instruction sheet carefully before using the
test kit. It will provide you with the necessary information
for correct use of the kit. If you need additional information,
do not hesitate to e-mail us at [email protected].
Remove the chemical test kit from the packing material and
examine it carefully to make sure that no damage has
occurred during shipping. If there is any noticeable dam-
age, notify your Dealer or the nearest Hanna office
immediately.
Each kit is supplied with:
•HI93709A-0CitrateBufferReagent,packets(100pcs);
•HI93709B-0SodiumPeriodate Reagent, packets (100
pcs);
 Deionized Water, 2 bottles (2x230 mL);
• 1 checker disc (containing a mirror, a mirror holder
and the 38072 disc);
• 2 glass vials with caps;
•1 plastic pipette (3 mL).
Note: Any damaged or defective item must be returned in
its original packing materials.
HI 38072
Manganese
Test Kit for
Irrigation Water
ISTR38072 02/00 PRINTED IN ITALY
Manganese is not occurring in natural waters but it is
found in various salts and minerals, frequently associated
withiron compounds.Manganesesalts areusedas fertilizer's
additives, in ferroalloys (in the steel manufacture), in non-
ferrous alloys as it improves their corrosion resistance and
hardness.
Manganese is not considered to be toxic to man and
aquatic life and it is ingested as a trace nutrient, both in
food and water. Nonetheless manganese has been limited
in drinking water since it causes tenacious stains to laundry
and because it produces objectionable tastes in beverages.
Manganese is commonly found in domestic wastewater,
industrial effluents and receiving streams.
Note: mg/L is equivalent to ppm (parts per million).
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONSSPECIFICATIONS
SPECIFICATIONS
SIGNIFICANCE AND USESIGNIFICANCE AND USE
SIGNIFICANCE AND USESIGNIFICANCE AND USE
SIGNIFICANCE AND USE
CHEMICAL REACTIONCHEMICAL REACTION
CHEMICAL REACTIONCHEMICAL REACTION
CHEMICAL REACTION
INSTRUCTIONSINSTRUCTIONS
INSTRUCTIONSINSTRUCTIONS
INSTRUCTIONS
The reaction between manganese and reagents causes a
violet tint in the sample which is proportional to the
manganese concentration.
READ THE ENTIRE INSTRUCTIONS BEFORE USING THE KIT
• Verify that the mirror, its holder and the disc are pre-
installed in the checker disc. If not, install them as
shown in the figure.
Range 0 to 10.0 mg/L (ppm) as Mn
SmallestIncrement 0.5 mg/L (ppm) Mn
AnalysisMethod Colorimetric
Sample Size 7.5 mL
Number of Tests 100
CaseDimensions 235x175x115 mm (9.2x6.9x4.5")
Shipping Weight 1100 g (38.8 oz.) 7.5 mL
The chemicals contained in this kit may be hazardous if
improperly handled. Read the relevant Health and Safety
Data Sheet before performing this test.
HEALTH AND SAFETYHEALTH AND SAFETY
HEALTH AND SAFETYHEALTH AND SAFETY
HEALTH AND SAFETY
REFERENCESREFERENCES
REFERENCESREFERENCES
REFERENCES
Adaptation of the Periodate Oxidation Method, Standard
Methods for the Examination of Water and Wastewater,
18th edition.
• Usingthe plasticpipette,fill each glass
vial with 7.5 mL of sample (up to the
mark).
• Insert one of them into the left hand
opening of the checker disc. This is the
blank.
• Add to the other vial
Deionized Water up
to the 10 mL mark.
Replace the cap and
shake to mix.
• Remove the cap
and add 1 packet
of HI 93709A-0
reagent. Replace
the cap and swirl
to mix.
• Remove the cap
and add 1 packet
ofHI93709B-0re-
agent. Replace the
cap and swirl to
mix.
• Wait for 1 minute and 30 sec-
onds to allow reaction to occur.
This is the reacted sample.
• Remove the cap and insert
the reacted sample into the
right hand opening of the
checkerdisc.
• Holdthechecker disc so that
alight sourceilluminatesthe
samples from the top.
• Keep the checker disc at a distance of 30-40 cm (12-
16") from the eyes to match the color. Rotate the disc
while looking at the color test windows and stop when
you find the color match. Read the value in the result
window directly in mg/L (or ppm) of Manganese.
For best results: Perform the reading three times and
take the average value (divide by 3 the sum of the
three numbers). Intensely colored samples will make
the color matching difficult and they should be
adequately treated before performing the test. Sus-
pended matter in large amounts should be removed
by prior filtration.
Caution: Ultraviolet radiation may cause fading of colors.
When not in use, keep the disc protected from light,
in a cool and dry place.
Interferences:Calcium above 700 ppm, magnesium above
100000 ppm, chloride above 7000 ppm, iron above
5 ppm.