manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Hanna Instruments
  6. •
  7. Measuring Instrument
  8. •
  9. Hanna Instruments HI 96735 User manual

Hanna Instruments HI 96735 User manual

1. RANGE/GLP/ key: press to change the range or press and hold
for three seconds to enter
GLP mode
. In
calibration mode
press to edit
the date and time.
2. CAL CHECK key: press to perform the validation of the meter, or
press and hold for three seconds to enter
calibration mode
.
3. ZERO/CFM key: press to zero the meter prior to measurement, to
confirm edited values or to confirm factory calibration restore. Press
and hold for three seconds to start a preprogramed countdown prior
to measurement.
4. READ/ /UNIT key: In
measurement mode
, press to make a
measurement, or press and hold for three seconds to change the
measurement unit. In
GLP mode
press to view the next screen.
5. ON/OFF key: to turn the meter on and off.
6. Liquid Crystal Display (LCD)
7. Cuvette alignment indicator
8. Cuvette holder
DISPLAY ELEMENTS DESCRIPTION:
INSTRUCTION MANUAL
HI 96735HI 96735
HI 96735HI 96735
HI 96735
www.hannainst.com
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is
not damaged. If any damage occured during shipment, please notify
your Dealer.
Each HI 96735 Ion Selective Meter is supplied complete with:
• Two Sample Cuvettes and Caps
• 9V Battery
• Instruction Manual
Note:Save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.
Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using the meter.
This manual will provide you with the necessary information for correct use
of the instrument. If you need additional technical information, do not
hesitate to e-mail us at [email protected] or view our worldwide
contact list at www.hannainst.com.
Functional description:
Light High: There is too much light to perform
a measurement. Please check the preparation of
the zero cuvette.
Light Low: There is not enough light to perform
a measurement. Please check the preparation of
the zero cuvette.
No Light: The instrument cannot adjust the
light level. Please check that the sample does
not contain any debris.
Inverted Cuvettes: The sample and the zero
cuvette are inverted.
Zero: A zero reading was not taken. Follow the
instructions of the measurement procedure for
zeroing the meter.
Under Range: A blinking “200” indicates that
the sample absorbs less light than the zero
reference. Check the procedure and make sure
you use the same cuvette for reference (zero)
and measurement.
Over Range: A flashing value of the maximum
concentration indicates an over range condition.
The concentration of the sample is beyond the
programmed range: dilute the sample and
re-run the test.
Standard Low: The standard reading is less
than expected.
Standard High: The standard reading is higher
than expected.
Errors and warnings:
Measurement procedure:
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the LCD
displays dashes and “P1” (Low range), “P2”
(Medium range) or “P3” (High range), the
meter is ready. The code that appears on
the secondary display is the one of the
last selected range. If necessary, press
RANGE/GLP/ to change range. The
blinking “ZERO” indicates that the
instrument needs to be zeroed first.
3• Add 0.5 mL of unreacted sample to the
cuvette. Add 0.5 mL of HI 93735IND-0
indicator reagent. With the plastic dropper
fill the cuvette up to the 10 mL mark
adding HI 93735A-XR (X=L, M, H)
reagent appropriate to the selected range.
4• Add two drops of HI 93735B-0 buffer
reagent. Replace the cap and invert 5 times
to mix.
5• Place the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
6• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the display,
depending on the measurement phase.
7• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and ready
for measurement. Remove the cuvette.
8• Add the content of 1 packet of HI 93735C-0
fixing reagent. Replace the cap and shake
gently to mix 20 seconds.
9• Replace the cuvette into the holder and
ensure that the notch on the cap is positioned
securely into the groove.
10•Press READ/ /UNIT. In all cases the lamp,
cuvette and detector icons will appear on the
display, depending on the measurement phase.
Measurement
1. The measuring scheme (lamp, cuvette, detector), appears during
different phases of zero or reading measurement
2. Error messages and warnings
3. The battery icon indicates the charge state of the battery
4. The hourglass appears when an internal check is in progress
5. Status messages
6. The chronometer appears when the reaction timer is running
7. The month, day and date icons appear when a date is displayed
8. Four digit main display
9. Measuring units
10. Four digit secondary display
ON ZERO READING:
ON SAMPLE READING:
DURING CALIBRATION PROCEDURE:
OTHER ERRORS AND WARNINGS:
Battery Low: The battery must be replaced
soon.
Dead Battery: This indicates that the battery
is dead and must be replaced. Once this
indication is displayed, normal operation of the
instrument will be interrupted. Change the
battery and restart the meter.
Cap Error: Appears when external light enters
in the analysis cell. Assure that the cuvette cap
is present.
Cooling lamp: The instrument waits for the
lamp to cool down.
Hardness ISM
3
6
9
10
10 mL
4
5
8
Technical specifications:
Range LR 0 to 250 mg/L
MR 200 to 500 mg/L
HR 400 to 750 mg/L
Resolution 1 mg/L from 0 to 100 mg/L
5 mg/L from 100 to 750 mg/L
Accuracy LR ±5mg/L ±4% of reading@ 25°C
MR ±7 mg/L ±3% of reading@ 25°C
HR ±10 mg/L ±2% of reading@ 25°C
Light Source Light Emitting Diode
Light Detector Silicon Photocell with narow band interference
filter @ 466 nm
Method Adaptation of the EPA recommended method
130.1. The reaction between calcium,
magnezium and the reagents causes a
red-violet tint in the sample.
Environment 0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
Battery Type 1 x 9 volt
Auto-Shut off After 10' of non-use in measurement mode;
after 1 hour of non-use in calibration mode;
with last reading reminder.
Dimensions 192 x 104 x 69 mm (7.6 x 4.1 x 2.7”)
Weight 320 g (11.3 oz.).
2
For more details about spare parts and accessories
see “Accessories”.
11•Then the date of last calibration (e.g.:
“01.08.2009”) appears on the display, or
“01.01.2009” if the factory calibration
was selected before. In both cases the year
number is blinking, ready for date input.
12•Press RANGE/GLP/ to edit the desired
year (2009-2099). If the key is kept
pressed, the year number is automatically
increased.
13•When the correct year has been set, press
ZERO/CFM or READ/ /UNIT to confirm.
Now the display will show the month
blinking.
14•Press RANGE/GLP/ to edit the desired
month (01-12). If the key is kept pressed,
the month number is automatically
increased.
15•When the correct month has been set, press
ZERO/CFM or READ/ /UNIT to confirm.
Now the display will show the day blinking.
16•Press RANGE/GLP/ to edit the desired
day (01-31). If the key is kept pressed,
the day number is automatically increased.
Note:It is possible to change the editing from
day to year and to month by pressing
READ/ /UNIT.
17•Press ZERO/CFM to save the calibration
date.
18•The instrument displays “Stor” for one
second and the calibration is saved.
19•The instrument will return automatically
to
measurement mode
by displaying dashes
on the LCD.
REAGENT SETS
HI 93735-00 Reagents for 100 tests (LR, 0 to 250 mg/L)
HI 93735-01 Reagents for 100 tests (MR, 200 to 500 mg/L)
HI 93735-02 Reagents for 100 tests (HR, 400 to 750 mg/L)
HI 93735-0 Reagents for 300 tests (LR - 100 tests, MR - 100 tests,
HR - 100 tests)
OTHER ACCESSORIES
HI 96735-11 CAL CHECK™ Standard Cuvettes (1 set)
HI 740029P 9V battery (10 pcs.)
HI 731318 Cloth for wiping cuvettes (4 pcs.)
HI 731331 Glass cuvettes (4 pcs.)
HI 731335 Caps for cuvettes (4 pcs.)
HI 93703-50 Cuvette cleaning solution (230 mL).
Warranty
HI 96735 is warranted for two years against defects in workmanship and
materials when used for its intended purpose and maintained according
to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact your dealer. If under warranty, report the
model number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified
of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service Department
and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
Recommendations for Users
Before using these products, make sure that they are entirely suitable for your specific application and for
the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other electronic equipment, this
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument
safety do not use or store the instrument in hazardous environments.
Hanna Instruments reserves the right to modify the
design, construction, or appearance of its products
without advance notice.
For additional information, contact your
dealer or the nearest
Hanna Customer Service Center.
To find the Hanna Office in your area,
visit our web site
www.hannainst.com
IST96735 11/15
Accessories:
3
Warning:do not validate or calibrate the instrument with standard
solutions other than the Hanna CAL CHECK™ Standards, otherwise
erroneous results will be obtained.
For accurate validation and calibration results, please perform tests at
room temperature (18 to 25°C; 64.5 to 77.0°F).
Validation and Calibration procedures
VALIDATION
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
3• Place the CAL CHECK™ Standard
HI 96735-11 Cuvette A into the holder
and ensure that the notch on the cap
is positioned securely into the groove.
4• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the display,
depending on the measurement phase.
5• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and
ready for validation.
6• Remove the cuvette.
7• Place the CAL CHECK™ Standard
HI 96735-11 Cuvette B into the holder
and ensure that the notch on the cap is
positioned securely into the groove.
Validation
4-5
7
8
9
Calibration
3
4
5-6
8
9-10
11-13
or
14-15
or
16
17
LAST CALIBRATION DATE
1• Press and hold for three seconds
RANGE/GLP/ to enter
GLP mode
.
The calibration month and day will appear
on the main display and the year on the
secondary display.
2• If no calibration was performed, the factory
calibration message, “F.CAL” will appear
on the main display and the instrument
returns to
measurement mode
after three
seconds.
GLP
Last
Calibration
Date
In
GLP mode
, the last calibration date can be verified and the factory
calibration can be restored.
Factory
Calibration
Restore
3-4
2
3-4
Battery management
To save the battery, the instrument shuts down after 10 minutes of non-
use in
measurement mode
and after 1 hour of non-use in
calibration
mode.
If a valid measurement was displayed before
auto-shut off, the value is displayed when the
instrument is switched on. The blinking “ZERO”
means that a new zero has to be performed.
One fresh battery lasts for arround 750 measurements, depending on the
light level.
The remaining battery capacity is evaluated at the instrument startup
and after each measurement.
The instrument displays a battery indicator with three levels as follows:
•3 lines for 100 % capacity
•2 lines for 66 % capacity
•1 line for 33 % capacity
•Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken any
more, the instrument shows “dEAd bAtt” and turns off.
To restart the instrument, the battery must be replaced with a fresh one.
To replace the instrument’s battery, follow the steps:
•Turn the instrument off by pressing ON/OFF.
•Turn the instrument upside down and remove the battery cover by
turning it counterclockwise.
•Extract the battery from its location and replace it with a fresh one.
•Insert back the battery cover and turn it clockwise to close.
8• Press CAL CHECK key and the lamp,
cuvette and detector icons together with
“CAL CHECK” will appear on the display,
depending on the measurement phase.
9• At the end of the measurement the display
will show the validation standard value.
The reading should be within specifications as
reported on the CAL CHECK™ Standard
Certificate. If the value is found out of
specifications, please check that the cuvettes
are free of fingerprints, oil or dirt and repeat
validation. If results are still found out of
specifications then recalibrate the instrument.
CALIBRATION
Note:It is possible to interrupt the calibration
procedure at any time by pressing CAL
CHECK or ON/OFF keys.
1• Turn the meter on by pressing ON/OFF.
2• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
3• Press and hold CAL CHECK for three seconds
to enter
calibration mode.
The display will
show “CAL” during calibration procedure. The
blinking “ZERO” asks for instrument zeroing.
4• Place the CAL CHECK™ Standard
HI 96735-11 Cuvette A into the cuvette
holder and ensure that the notch on the cap
is positioned securely into the groove.
5•Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the display,
depending on the measurement phase.
6• After a few seconds the display will show
“-0.0-”. The meter is now zeroed and
ready for calibration. The blinking “READ”
asks for reading calibration standard.
7• Remove the cuvette.
8• Place the CAL CHECK™ Standard
HI 96735-11 Cuvette B into the holder
and ensure that the notch on the cap is
positioned securely into the groove.
9•Press READ/ /UNIT and the lamp, cuvette
and detector icons will appear on the display,
depending on the measurement phase.
10•The instrument will show for three seconds
the CAL CHECK™ standard value.
Note:If the display shows “STD HIGH”, the
standard value was too high. If the
display shows “STD LOW”, the standard
value was too low. Verify that both
CAL CHECK™ Standard HI 96735-11
Cuvettes, A and B are free from
fingerprints or dirt and that they
are inserted correctly.
11•At the end of measurement, the instrument
directly displays the hardness in the
last unit selected on the LCD. Press the
READ/ /UNIT key repeatedly to change
the reading unit: mg/L, °f, °D and °E
respectively. The conversion factors are as
follows:
1 mg/L = 0.1 °f = 0.0556 °D = 0.07 °E
INTERFERENCES:
Interference may be caused by excessive amounts
of heavy metals.
FACTORY CALIBRATION RESTORE
It is possible to delete the calibration and restore
factory calibration.
1• Press RANGE/GLP/ to enter
GLP mode
.
2• Press READ/ /UNIT to enter in the factory
calibration restore screen. The instrument
asks for confirmation of user calibration
delete.
3• Press ZERO/CFM to restore the factory
calibration or press RANGE/GLP/ again
to abort factory calibration restore.
4• The instrument briefly indicates “donE”
upon restoration of factory calibration prior
returning to
measurement mode
.
2
1
18
11
Note:If the sample is very acidic, some extra
drops of HI 93735B-0 buffer reagent
may be added.
Use the Hanna CAL CHECK™ cuvettes (see
“Accessories”) to validate or calibrate instruments.

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 96700 User manual

Hanna Instruments

Hanna Instruments HI 96700 User manual

Hanna Instruments HI 9023C User manual

Hanna Instruments

Hanna Instruments HI 9023C User manual

Hanna Instruments Total Chlorine HI 711 User manual

Hanna Instruments

Hanna Instruments Total Chlorine HI 711 User manual

Hanna Instruments edge pH User manual

Hanna Instruments

Hanna Instruments edge pH User manual

Hanna Instruments HI 93703-11 User manual

Hanna Instruments

Hanna Instruments HI 93703-11 User manual

Hanna Instruments HI9813-51 User manual

Hanna Instruments

Hanna Instruments HI9813-51 User manual

Hanna Instruments HI 96821 User manual

Hanna Instruments

Hanna Instruments HI 96821 User manual

Hanna Instruments HALO2 HI9810442 User manual

Hanna Instruments

Hanna Instruments HALO2 HI9810442 User manual

Hanna Instruments HI5321 User manual

Hanna Instruments

Hanna Instruments HI5321 User manual

Hanna Instruments HI99165 User manual

Hanna Instruments

Hanna Instruments HI99165 User manual

Hanna Instruments HI 9033 User manual

Hanna Instruments

Hanna Instruments HI 9033 User manual

Hanna Instruments HI 5315 User manual

Hanna Instruments

Hanna Instruments HI 5315 User manual

Hanna Instruments HI 8564 User manual

Hanna Instruments

Hanna Instruments HI 8564 User manual

Hanna Instruments HI97750 User manual

Hanna Instruments

Hanna Instruments HI97750 User manual

Hanna Instruments HI96738 User manual

Hanna Instruments

Hanna Instruments HI96738 User manual

Hanna Instruments Checker HI707 User manual

Hanna Instruments

Hanna Instruments Checker HI707 User manual

Hanna Instruments HI 93712 User manual

Hanna Instruments

Hanna Instruments HI 93712 User manual

Hanna Instruments HALO 2 HI9810352 User manual

Hanna Instruments

Hanna Instruments HALO 2 HI9810352 User manual

Hanna Instruments HI 98186 User manual

Hanna Instruments

Hanna Instruments HI 98186 User manual

Hanna Instruments HI 3221 User manual

Hanna Instruments

Hanna Instruments HI 3221 User manual

Hanna Instruments HI762 User manual

Hanna Instruments

Hanna Instruments HI762 User manual

Hanna Instruments HI 98143 User manual

Hanna Instruments

Hanna Instruments HI 98143 User manual

Hanna Instruments HI 84429 User manual

Hanna Instruments

Hanna Instruments HI 84429 User manual

Hanna Instruments HI 84429 User manual

Hanna Instruments

Hanna Instruments HI 84429 User manual

Popular Measuring Instrument manuals by other brands

QUARTA RADEX ONE user manual

QUARTA

QUARTA RADEX ONE user manual

Katronic KATflow 170 Operating and safety instructions

Katronic

Katronic KATflow 170 Operating and safety instructions

Lumel N30P user manual

Lumel

Lumel N30P user manual

Easy-Laser E980 manual

Easy-Laser

Easy-Laser E980 manual

Elcontrol STAR 3 user manual

Elcontrol

Elcontrol STAR 3 user manual

Agilent Technologies NFA Series Setup guide

Agilent Technologies

Agilent Technologies NFA Series Setup guide

Endress+Hauser nivotester FTC 420 Quick installation guide

Endress+Hauser

Endress+Hauser nivotester FTC 420 Quick installation guide

Flybox Omnia57-80 Series manual

Flybox

Flybox Omnia57-80 Series manual

bmcm AMS42-LAN16fx manual

bmcm

bmcm AMS42-LAN16fx manual

Schäfer RMS 442 quick start guide

Schäfer

Schäfer RMS 442 quick start guide

PCB Piezotronics 352A21/NC Installation and operating manual

PCB Piezotronics

PCB Piezotronics 352A21/NC Installation and operating manual

Schaller humimeter BMA user manual

Schaller

Schaller humimeter BMA user manual

THEIS TKL-7 operating instructions

THEIS

THEIS TKL-7 operating instructions

HYDROMETER IZAR CENTER user manual

HYDROMETER

HYDROMETER IZAR CENTER user manual

Omega HHF143 user guide

Omega

Omega HHF143 user guide

amc VIBRO AVM 1002D user manual

amc VIBRO

amc VIBRO AVM 1002D user manual

Mollet LF20 Series Instruction

Mollet

Mollet LF20 Series Instruction

DayTronic 3000PLUS instruction manual

DayTronic

DayTronic 3000PLUS instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.