
SKU 92573 For technical questions please call 1-800-444-3353 PAGE 9
Corresponds to one cycle of observation.
Corresponds to two cycles of observation.
Corresponds to three cycles of observation.
Corresponds to four or more cycles of observation.
FIGURE D
8. As the “EXECUTED CYCLES OF OBSERVATION” (14) averages, it analyzes the
current value taken and compares it to the values of the previous observations.
If the values exceed a defined value, the current result is displayed instead of the
average. For example, if the average of three previous observations is
0.20 uSv/h, and the current value of 0.80 uSv/h is registered, the result of the
fourth observation will not be averaged and the LCD Display (1) will read
0.80 uSv/h with the icon “ “. This allows the Radiation Detector to detect sharp
changes of dose rate.
To Turn Off The Radiation Detector:
1. To exit the menu, press the “OFF” Button (4). Then, press the “OFF” Button (4)
repeatedly until all messages on the LCD Display (1) disappear. (See Figure B.)
INSPECTION, MAINTENANCE, AND CLEANING
1. Before each use: Inspect the general condition of the Radiation Detector.
Check for broken or cracked parts, and any condition that may affect the proper
operation of the product. If a problem occurs, have the problem corrected before
further use. Do not use damaged equipment.
2. When changing Batteries: Refer to page 4for complete instructions.
CAUTION! Never burn batteries, as they can explode in a fire. Contact your
local solid waste authorities for instructions on correct disposal or recycling of
batteries.
3. When cleaning: Use a clean cloth with a mild detergent. Then, dry. Do not
immerse the Radiation Detector in any liquid. Do not use chemical substances
such as acids, alkali, solvents, etc. to clean.
4. When storing: Always remove the Batteries prior to storing the Radiation Detec-
tor for long periods of time. Make sure to store the unit in a safe, clean, dry
location out of reach of children and other non-authorized people.