HBM MP85A User manual

Operating Manual
English
MP85A Program Help

Hottinger Baldwin Messtechnik GmbH
Im Tiefen See 45
64239 Darmstadt
Tel. +49 6151 803-0
Fax +49 6151 803-9100
www.hbm.com
DVS: a3387-1.1 HBM: public
12.2016
© Hottinger Baldwin Messtechnik GmbH
Subject to modifications.
All product descriptions are for general information only.
They are not to be understood as a guarantee of quality or
durability.

Contents
MP85A a3387-1.1 en HBM: public 3
Contents Page
1 Quick start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.1 The MP85A measuring system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.2 What needs to be done?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.3 Setting up the hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.4 Firmware update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2 SYSTEM settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.1 General settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.2 Using passcodes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.3 Interface parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3 Prepare MP85A for your application . . . . . . . . . . . . . . . . . . . . . . . . 23
3.1 Setup of the measuring chain (MP85A and transducer). . . . . . . . . . . 23
3.1.1 Specifying a transducer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.1.2 Adjusting the measuring chain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.1.3 Signal conditioning (zero balance, low-pass filter) . . . . . . . . . . . . . . . 40
3.2 What needs to be considered when zero balancing and how long
does it take? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
3.3 How are the sample rate and filter settings connected? . . . . . . . . . . . 43
3.3.1 Transducer test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
3.3.2 What is the procedure for a transducer test and what needs to be
considered? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3.4 Additional functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3.4.1 Limit values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3.4.2 Digital outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
3.4.3 Digital inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
4 Settings for monitoring using tolerance windows . . . . . . . . . . . . 55
4.1 Principle procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
4.2 The Configure evaluation parameter window . . . . . . . . . . . . . . . . . . . 58
4.3 Control settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
4.3.1 Maximum measuring time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
4.3.2 Start conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62

Contents
4a3387-1.1 en HBM: public MP85A
4.3.3 Stop conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
4.3.4 End conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
4.4 Alarm window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
4.5 Range window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
4.6 Test measurement (start measurement manually) . . . . . . . . . . . . . . . 74
4.7 Important information about tolerance windows . . . . . . . . . . . . . . . . . 76
4.8 Defining tolerance windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
4.8.1 Creating and modifying tolerance windows . . . . . . . . . . . . . . . . . . . . 80
4.8.2 Window active. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
4.8.3 Consider window result in total result . . . . . . . . . . . . . . . . . . . . . . . . . 82
4.8.4 Crossing window with min/max evaluation . . . . . . . . . . . . . . . . . . . . . 83
4.8.5 Crossing window with evaluation of the y mean . . . . . . . . . . . . . . . . . 83
4.8.6 Crossing window with evaluation of the x mean . . . . . . . . . . . . . . . . . 84
4.8.7 y limit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
4.8.8 x limit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
4.8.9 Block window (end window) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
4.8.10 Switch test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
4.8.11 Haptic test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
4.8.12 Evaluation in real time window option (online window). . . . . . . . . . . . 91
4.8.13 Consider window sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
4.8.14 Window option y coordinates relative . . . . . . . . . . . . . . . . . . . . . . . . . 99
4.8.15 Window option x coordinates relative to . . . . . . . . . . . . . . . . . . . . . . 101
4.8.16 Evaluation for entry and exit of window . . . . . . . . . . . . . . . . . . . . . . 104
4.8.17 Statistics and distribution (bin counting) . . . . . . . . . . . . . . . . . . . . . . 113
5 Monitoring with tolerance band: settings. . . . . . . . . . . . . . . . . . . 119
5.1 Principle procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
5.2 The Configure evaluation parameter window . . . . . . . . . . . . . . . . . . 122
5.3 Control settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
5.3.1 Maximum measuring time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
5.3.2 Start conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
5.3.3 Stop conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126
5.3.4 End conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
5.4 Alarm window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135
5.5 Range window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136

Contents
MP85A a3387-1.1 en HBM: public 5
5.6 Test measurement (start measurement manually) . . . . . . . . . . . . . . 138
5.7 Defining a tolerance band . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
6 Monitoring with envelope curve . . . . . . . . . . . . . . . . . . . . . . . . . . 143
7 Saving results, data storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
7.1 Storage medium MultiMediaCard/SD Card (target MMC/SD Card
internal). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
7.2 Storage medium external (storage via CAN bus or Ethernet). . . . . . 148
7.3 Storage methods. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
7.4 Files to save . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
7.5 File names and directories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
7.6 File formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
7.7 Save statistical data, statistical processing. . . . . . . . . . . . . . . . . . . . 157
8 SAVE/LOAD PARAMETERS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159
9 MEASURE + VISUALIZE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
9.1 Display current values (measurement display) . . . . . . . . . . . . . . . . . 165
9.1.1 Possible error messages and instructions for correction . . . . . . . . . 168
9.1.2 What limitations apply for (transducer) scaling? . . . . . . . . . . . . . . . . 171
9.2 Measurement curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 171
9.3 Results of the last measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . 176
9.4 Statistics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
9.4.1 3D graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 178
9.4.2 Distribution (bin counting) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179
9.5 Retrieve saved data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180
9.6 Files on MMC/SD Card (copy to/from PC) . . . . . . . . . . . . . . . . . . . . 183
10 Plus tools. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
11 Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
11.1 Change log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
11.2 Display connection loss detection. . . . . . . . . . . . . . . . . . . . . . . . . . . 187
12 FAQ: Questions and answers about MP85A and PME
Assistant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189
12.1 How do I adjust the interface on the MP85A?. . . . . . . . . . . . . . . . . . 189
12.2 How do I set the IP address on my PC? . . . . . . . . . . . . . . . . . . . . . . 192

Contents
6a3387-1.1 en HBM: public MP85A
12.3 How do I connect the PME Assistant to the MP85A? . . . . . . . . . . . . 195
12.4 How do I recognize the file system on my PC and which should
I use? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 196
12.5 How do the options in the start window of the PME Assistant
work? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
12.6 What happens when connecting additionally to an existing
Ethernet connection? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199
12.7 Requirements for MMC/SD Cards . . . . . . . . . . . . . . . . . . . . . . . . . . 200
12.8 How are the sample rate and filter settings connected? . . . . . . . . . . 201
12.9 What does the data reduction do and how do I set it up?. . . . . . . . . 201
12.10 What possibilities are there to start, stop and end the
measurement? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205
12.11 How do I work when setting up the evaluation criteria? . . . . . . . . . . 219
12.12 What needs to be considered when zero balancing and how
long does it take? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 220
12.13 What is the procedure for a transducer test and what needs to
be considered? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221
12.14 What limitations apply for (transducer) scaling? . . . . . . . . . . . . . . . . 222
12.15 What has to be considered with parameter set switching?. . . . . . . . 223
12.16 How to determine changes to the device settings? . . . . . . . . . . . . . 224
12.17 Which error messages exist for the measuring/the process
status and how can I correct errors?. . . . . . . . . . . . . . . . . . . . . . . . . 224
12.18 What do the error messages with tolerance window violation
mean? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 226
12.19 What do the LEDs on the MP85A mean? . . . . . . . . . . . . . . . . . . . . . 234
12.20 What should be observed when saving process data, curves or
process results? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236
12.21 How does the timing diagram for a process-optimized
measurement look like? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 238
12.22 How does the timing diagram for a measurement with no data
loss look like? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 239
12.23 How can a digital output be set?. . . . . . . . . . . . . . . . . . . . . . . . . . . . 239
12.24 How do I create a protocol printout of a process? . . . . . . . . . . . . . . 240
12.25 How to transfer all MP85A settings to another device (cloning)? . . . 241

Contents
MP85A a3387-1.1 en HBM: public 7
12.26 How do I carry out a firmware update and can the update be
prevented? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 242
12.26.1 Download firmware update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 242
12.26.2 How to update the firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 243
12.27 What does a tolerance band look like?. . . . . . . . . . . . . . . . . . . . . . . 243
12.28 What does an envelope curve look like? . . . . . . . . . . . . . . . . . . . . . 244
12.29 What is TEDS? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244
13 Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 247

Contents
8a3387-1.1 en HBM: public MP85A

Quick start
MP85A a3387-1.1 en HBM: public 9
1 Quick start
The help for the PME Assistant consists of three
parts:
•The general PME Assistant Online Help.
•The help for the devices MP01, MP30, MP55, MP60
and MP70 of the PME family.
•This document on the MP85A.
1.1 The MP85A measuring system
The MP85A measuring system comprises the PME Assis-
tant program and the PME MP85A or MP85A DP hard-
ware. The MP85A DP is provided with a PROFIBUS inter-
face in addition to the CAN bus and Ethernet interface.
The mode of operation is however the same. The device
parameters must be set either via the CAN bus or the
Ethernet interface for both systems. If you have an older
MP85, please use the PME Assistant provided for this
device (up to Version 2.2).
The MP85A can be used for two different monitoring oper-
ations:
1. Monitoring a process with verification of a certain pro-
cess curve using tolerance windows.
The measured values are checked at different points
by the MP85A to see whether the sequence lies within
a certain range (tolerance window).
2. Monitoring a process for adherence to a tolerance
band or envelope curve.

Quick start
10 a3387-1.1 en HBM: public MP85A
Starting with firmware version 2.22 the limit value results
can be included into the OK/NOK evaluation with all mon-
itoring methods.
There is also Version MP85A-S for switch and haptic test
with tolerance windows. However tolerance band moni-
toring is not possible with these devices.
You can find more information about MP85A, e.g. mea-
surement sequence, the data reduction function or the
relation between the filters used and sample rate, in sec-
tion FAQ: MP85A and PME Assistant.
1.2 What needs to be done?
The following steps must be performed in order to prepare
the MP85A measuring system for a measurement. The
steps can be divided into two sections.
1. Setting up the hardware: connect transducer (sensor),
set up bus system or interface, connect power supply
etc.
This part of the setup is described in detail in
the MP85A Operating Manual. This online
Help comprises only a summary.
2. Setting up the device via PME Assistant software.
Set up the device using this software and a PC which
is connected to the MP85A via CAN bus or Ethernet
interface as described in this Help.
Procedure
1. Start the PME Assistant, see also PME Assistant
Online Help: Start options.

Quick start
MP85A a3387-1.1 en HBM: public 11
2. Connect to the MP85A, see How do I connect the
PME Assistant to the MP85A?.
3. Define the MP85A system settings: Select the lan-
guage, set the synchronization if using more than one
device and, if required, set the passcode.
4. Setup of the measuring chain (MP85A and trans-
ducer).
5. Determine and enter the start and evaluation criteria
for the process: depending on the evaluation mode
either for Tolerance windows or Tolerance band.
See also EASYteach for envelope curves.
6. Specify the necessary outputs (process results).
7. Save all parameters.
The structure of the MP85A settings tree in the left-hand
side of the program window corresponds to the sequence
in which you should carry out the settings.
If these steps have already been performed, the data can
be loaded from a file into the device, see SAVE/LOAD
PARAMETERS.
See also The MP85A measuring system
1.3 Setting up the hardware
This section is described in detail in the MP85A Operating
Manual. To summarize:
•Connect the transducers.
•Connect the power supply.

Quick start
12 a3387-1.1 en HBM: public MP85A
•Connect the bus system and set parameters
(addresses and transmission speed, etc.) if neces-
sary. Because some of these settings are important
for the program function, you will also find the most
important settings described in Configuring the
MP85A interface(s).
•Check the (hardware) synchronization if several
devices are involved.
•Install the software on the PC, see also PME Assis-
tant Online Help: Operating requirements and PME
Assistant Online Help: Installation of software PME
Assistant.
The first channel is always displayed as the x channel
(horizontal axis) and the second as the y channel (vertical
axis) in the graphs. Please connect the transducers suit-
ably so that a correct display of y(x) is achieved.
1.4 Firmware update
You update the firmware using the program PME-Update.
With the program a new firmware version can also be
transferred to several devices simultaneously.
With the MP85A process controller, in order to avoid inter-
fering with the running of a process (no measurement or
evaluation is carried out during an update), from firmware
version 2.22 you can specify that the firmware is updated
only after manual confirmation on the device (F-update:
allowed!). If confirmation does not occur within 15 min-
utes, then no change to the firmware is made.
Refer also to General settings (basic settings).

Quick start
14 a3387-1.1 en HBM: public MP85A

SYSTEM settings
MP85A a3387-1.1 en HBM: public 15
2 SYSTEM settings
2.1 General settings
For passcodes see the separate section Using the pass-
code.
The general settings are identical for all parameter sets.
Some details in the SYSTEM General settings section
are for information only, e.g. hardware revision or device
serial number. Besides that, you can carry out the fol-
lowing settings:
Language
Next to Sprache (or Language or Langue), if English is
not set, select the desired language. Normally, the lan-
guage version of the PME Assistant determines the lan-
guage of the MP85A, i.e. when you start the German ver-
sion of the Assistant, all menus and texts are also in
German.
After choosing a different language, all the windows are
refreshed and formed anew. The language of the main
menu (at the top in the window) as well as of the mes-
sages output by the program are however dependent on
the language version of the program, PME Assistant, and
are not changed.
Device name
Use the device name to improve the identification of a
device in a group of several PMEs.

SYSTEM settings
16 a3387-1.1 en HBM: public MP85A
The device name does not affect the operation of the
device.
In the menu Data storage the option Keep device name
on parameter change specifies that when loading a
parameter set the device name is retained, i.e. it is not
overwritten by the name saved in the parameter set.
Date/time
The date and time are only used for saving the results.
The displayed time corresponds to that at the last window
update, e.g. on calling the dialog. Therefore an exclama-
tion point is displayed behind the time. The entered time is
sent to the PME, when, after making a change, you quit
the entry field by either clicking another field or pressing
.
Hardware synchronization
You have to carry out this setting if you are using more
than one device. In this case you set one device as the
Master and all others as Slave. This prevents the carrier
frequency measuring amplifiers of the individual devices
interfering with one another over transducer cables that
run parallel. However, only the measuring amplifiers are
synchronized, not, say, the CAN bus interfaces or similar
functions. You will find further information on this topic in
the MP85A operating manual.
Firmware update only possible after release on the
device
Here you define whether a firmware update may always
take place or only when it has been confirmed on the
device itself. This means that you can inhibit a firmware

SYSTEM settings
MP85A a3387-1.1 en HBM: public 17
update from starting via the network for all devices, e.g. in
a running process (during an update no measurement or
evaluation occurs). The update is then only executed after
manual confirmation on the device (F-Update: Per-
mitted!). If the confirmation does not occur within 15 min-
utes, then no change to the firmware is made.
2.2 Using passcodes
A passcode can be set, thus allowing you to block unau-
thorized access. This is a number. Letters are not allowed
therefore we use the word passcode, not password.
If you have forgotten your passcode, use the master pass-
code 1702, to reset the system and to allow you to enter a
new passcode.
Menu commands and dialogs can be permanently
blocked from the EASYsetup (FASTpress Suite) program
module. If you do not have one of the entries listed in this
Help, then the MP85A has been configured specially for
you.
Prepare passcode usage (activate)
The passcode must first be defined under SYSTEM
General settings and the field Specify passcode. Turn
on the function with Activate passcode?
Yes. When-
ever you then start the PME Assistant program, you will
only be able to access the menu command SYSTEM
Passcode entry and the MEASURE + VISUALIZE menu
items.

SYSTEM settings
18 a3387-1.1 en HBM: public MP85A
Use SYSTEM General settings and Reactivate pass-
code to hide all passcode protected menu items and dia-
logs. If those menus shall be accessed again, a new
passcode entry has to be made.
Passcode entry
If the MP85A is protected by a passcode, you can only
access this menu item and menu MEASURE + VISU-
ALIZE. The correct passcode must be entered before you
can access all other menus. If still some menus are
missing, your MP85A is customized with program module
Assistant configuration.
2.3 Interface parameters
The dialog displays the parameters for the interfaces
installed, e.g. also the MAC address of the Ethernet inter-
face. Most settings can only be changed on the device
itself, see Configuring the MP85A interface(s).
With CAN bus the output rate (in multiples of 0.1 ms) and
the format for PDO messages (cyclic PDOs) can be set in
the dialog. The cyclic PDOs must however be started with
an interface command.
With an MP85ADP (with PROFIBUS option) you can set
the PROFIBUS address.
When using Ethernet you may also specify here a
gateway address. Use IPGatew1 to IPGatew4 to type in
the four parts of the gateway IP address. Gateways are
used, for example, to be able to reach other IP segments
or to send data via other protocols (network types).

SYSTEM settings
MP85A a3387-1.1 en HBM: public 19
See also Options, Display connection loss detection.
You can find the interface commands in the MP85A Oper-
ating Manual “Interface Description CAN/Pro-
fibus/Ethernet”.
Configuring the MP85A interface(s)
Ethernet interface
Parameterization can only be carried out manually; in the
PME Assistant the parameters for the interface are only
displayed. The PME Assistant software must not be con-
nected to the MP85A over this interface during configura-
tion.
On the MP85A enter the Ethernet address (IP address)
and the subnet mask. Operation with dynamic addresses
(DHCP) is not possible. You should clarify with your net-
work administrator which IP address (Ethernet address)
and subnet mask you have to set. If you only want to
establish a direct link between a PC and the MP85A, you
can use any address, e.g. 192.168.169.xxx. The last
group of three figures (xxx) must be a number between 1
and 254 and different for the PC and the MP85A. Make
sure in this case that your PC can switch over the
Ethernet connection to a direct link or use a "crossover
cable" (not a normal Ethernet cable).
Setting the address and subnet mask on the MP85A:
1. Press the SET key for at least two seconds.
The display CAN BUS appears.
2. Press the + key until ETHERNET appears in the dis-
play.

SYSTEM settings
20 a3387-1.1 en HBM: public MP85A
3. Press the SET key.
The MAC address is displayed.
4. Press the + key so that the first group of the IP
address appears.
5. Press SET to be able to change the value with + or -.
Otherwise press + to proceed to the next group.
6. If you have changed a number, press the SET key.
7. After the display of the IP address you come to the
display of the subnet mask. Change this according to
your requirements.
8. To save the changes press the SET key for at least
two seconds.
The flashing display Save? appears.
9. Confirm it by pressing SET.
In the second line a flashing double arrow and the text
Yes appear.
10. Confirm this query also by pressing SET.
CAN bus interface
You can only change most parameters manually; the PME
Assistant software must not be connected to the MP85A
over this interface during configuration. Terminate the pro-
gram if necessary.
You do not need to terminate or close the online help
when you terminate the PME Assistant software.
Setting the baud rate on the MP85A
1. Press the SET key for at least two seconds.
The display CAN BUS appears.
Other manuals for MP85A
2
Table of contents
Other HBM Controllers manuals
Popular Controllers manuals by other brands

Honeywell
Honeywell HA40 Technical manual

Outback Power Systems
Outback Power Systems FLEXMAX 80 user guide

BleBox
BleBox shutterbox user manual

Nicotra Gebhardt
Nicotra Gebhardt EKE 10-0070-5E-IM operating manual

NETGEAR
NETGEAR WFS709TP - ProSafe Smart Wireless Controller quick start guide

Samson
Samson Type 5824 Mounting and operating instructions