TheHOBOMX2300seriesdataloggersrecordandtransmittemperatureand/orrelative
humidity(RH)inoutdoororindoorenvironments.TheseBluetooth®LowEnergy‐enabled
loggersaredesignedforwirelesscommunicationwithamobiledevice.UsingtheHOBOmobile®
orHOBOconnect®app,youcaneasilyconfigurethelogger,readitout,andviewdataonyour
phoneortablet,orexportthedataforfurtheranalysis.Theloggercancalculateminimum,
maximum,average,andstandarddeviationstatisticsandbeconfiguredtoindicatealarmsat
thresholdsyouspecify.Theloggeralsosupportsburstlogginginwhichdataisloggedatafaster
intervalwhensensorreadingsareaboveorbelowcertainlimits.TheTempandTemp/RH
modelshaveinternalsensorswhiletheExternalTemp/RH,ExternalTemp,and2xExternal
Tempincludebuilt‐inexternalsensors,offeringawiderangeofsolutionsformonitoring
temperatureandRHinnumerousapplications.
Specifications
TemperatureSensor
RangeMX2301AandMX2305internalsensors:‐40to70°C(‐40to158°F)
MX2302Aexternaltemperaturesensor:‐40to70°C(‐40to158°F)
MX2303andMX2304externalsensors:‐40to100°C(‐40to212°F),with
tipandcableimmersioninfreshwaterupto50°C(122°F)foroneyear
Accuracy±0.25°Cfrom‐40to0°C(±0.45from‐40to32°F)
±0.2°Cfrom0to70°C(±0.36from32to158°F)
±0.25°Cfrom70to100°C(±0.45from158to212°F),MX2303and
MX2304only
ResolutionMX2301AandMX2302A:0.02°C(0.036°F)
MX2303,MX2304,andMX2305:0.04°C(0.072°F)
Drift<0.01°C(0.018°F)peryear
RelativeHumiditySensor*(MX2301A,MX2302Aonly)
Range0to100%RH,‐40°to70°C(‐40°to158°F);exposuretoconditions
below‐20°C(‐4°F)orabove95%RHmaytemporarilyincreasethe
maximumRHsensorerrorbyanadditional1%
Accuracy±2.5%from10%to90%(typical)toamaximumof±3.5%including
hysteresisat25°C(77°F);below10%RHandabove90%RH±5%typical
Resolution0.01%
Drift<1%peryeartypical
ResponseTime(typical,to90%ofchange)
TemperatureWithoutSolar
RadiationShield
WithRS1/M‐RSA
SolarRadiationShield
WithRS3‐BSolar
RadiationShield
MX2301A
internalsensor
17minutesinair
moving1m/sec
24minutesinair
moving1m/sec
NA
MX2302A
externalsensor
3minutes,45seconds
inairmoving1m/sec
7minutes,45
secondsinair
moving1m/sec
6minutes,30
secondsinair
moving1m/sec
MX2303/MX2304
externalsensors
3minutesinair
moving1m/sec;20
secondsinstirred
water
7minutesinair
moving1m/sec
4minutesinair
moving1m/sec
MX2305
internalsensor
17minutesinair
moving1m/sec
24minutesinair
moving1m/sec
NA
RHWithoutSolar
RadiationShield
WithRS1/M‐RSA
SolarRadiationShield
WithRS3‐BSolar
RadiationShield
MX2301A
internalsensor
30secondsinair
moving1m/sec
40secondsinair
moving1m/sec
NA
MX2302A
externalsensor
15secondsinair
moving1m/sec
30secondsinair
moving1m/sec
30secondsinair
moving1m/sec
HOBOMX2300Series
DataLogger
Models:
MX2301A,temp/RH
MX2302A,exttemp/RH
MX2303,2exttemp
MX2304,exttemp
MX2305,temp
IncludedItems:
Screws
Cableties
RequiredItems:
Mobiledevicewith
Bluetooth
HOBOmobileappfor
deviceswithiOS
HOBOconnectappfor
deviceswithAndroid™
Accessories:
Solarradiationshield
(RS3‐Bforusewith
MX2302A,MX2303,and
MX2304models;RS1orM‐
RSAforusewithMX2301A
andMX2305models)
Mountingbracketforsolar
radiationshield(MX2300‐
RS‐BRACKET),forusewith
MX2301AandMX2305
models
Replacementbattery
(HRB‐2/3AA)

HOBOMX2300SeriesDataLoggerManual
1‐800‐LOGGERS2www.onsetcomp.com
Specifications(continued)
Logger
OperatingRange‐40°to70°C(‐40°to158°F)
RadioPower1mW(0dBm)
TransmissionRangeApproximately30.5m(100ft)line‐of‐sight
WirelessDataStandardBluetoothLowEnergy(BluetoothSmart)
LoggingRate1secondto18hours
LoggingModesFixedinterval(normal,statistics)orburst
MemoryModesWrapwhenfullorstopwhenfull
StartModesImmediate,pushbutton,date&time,ornextinterval
StopModesWhenmemoryfull,pushbutton,date&time,orafterasetlogging
period
TimeAccuracy±1minutepermonth0°to50°C(32°to122°F)
BatteryType2/3AA3.6Voltlithium,userreplaceable
BatteryLife2years,typicalwithloggingintervalof1minuteandBluetoothAlways
Onenabled;5years,typicalwithloggingintervalof1minuteand
BluetoothAlwaysOndisabled.Fasterloggingintervalsandstatistics
samplingintervals,burstlogging,remainingconnectedwiththeapp,
excessivedownloads,andpagingmayimpactbatterylife.
MemoryMX2301AandMX2302A:128KB(63,488measurements,maximum)
MX2303,MX2304,andMX2305:128KB(84,650measurements,
maximum)
FullMemoryDownloadTimeApproximately60seconds;maytakelongerthefurtherthedeviceis
fromthelogger
DimensionsLoggerhousing:10.8x5.08x2.24cm(4.25x2.0x0.88in.)
Externaltemperaturesensordiameter:0.53cm(0.21in.)
Externaltemperature/RHsensordiameter:1.17cm(0.46in.)
Externalsensorcablelength:2m(6.56ft)
Solarradiationshieldbracket:10.8x8.3cm(4.25X3.25in.)
WeightLogger:75.5g(2.66oz)
Solarradiationshieldbracket:20.4g(0.72oz)
MaterialsAcetal,siliconegasket,stainlesssteelscrews
EnvironmentalRatingNEMA6andIP67
TheCEMarkingidentifiesthisproductascomplyingwithallrelevant
directivesintheEuropeanUnion(EU).
Seelastpage
*PerRHsensormanufacturerdatasheet

HOBOMX2300SeriesDataLoggerManual
1‐800‐LOGGERS3www.onsetcomp.com
LoggerComponentsandOperation
MountingHoles:Usetheholesatthetopandbottomofthe
loggertomountit(seeDeployingandMountingtheLogger).
AlarmLED:ThisLEDblinksredevery4secondswhenanalarm
istripped(unlessShowLEDisdisabledasdescribedin
ConfiguringtheLogger).
StatusLED:ThisLEDblinksblueevery4secondswhenthe
loggerislogging(unlessShowLEDisdisabledasdescribedin
ConfiguringtheLogger).Iftheloggeriswaitingtostartlogging
becauseitwasconfiguredtostart“OnButtonPush”orwitha
delayedstart,itwillblinkevery8seconds.
StartButton:Pressthisbuttontowakeupthelogger;boththe
alarmandstatusLEDswillblink.Oncetheloggerisawake,
pressthisbuttontomoveittothetopofthedeviceslistinthe
app.Pressthisbuttonfor3secondstostartorstopthelogger
whenitisconfiguredtostartorstop“Onbuttonpush”(see
ConfiguringtheLogger).BothLEDswillblinkfourtimeswhen
youpressthebuttontostartorstoplogging.Pressthisbutton
for10secondstoresetapassword(seeSettingaPassword).
ExternalSensor:Thisistheexternalprobeattachedtothe
bottomoftheloggerthatmeasurestemperatureor
temperature/RH.TheMX2302Aloggerhasoneexternalsensor
thatmeasuresbothtemperatureandRHandtheMX2304
loggerhasonesensorthatmeasurestemperatureonly.The
MX2303logger(shownatleftabove)hastwoexternal
temperaturesensors;theleftsensorischannel1andtheright
sensorischannel2.
Vent:TheRHsensorislocatedbehindthevent(MX2301A
modelonly).
DownloadingtheAppand
ConnectingtoaLogger
Installtheapptoconnecttoandworkwiththelogger.
1. IfyouareusinganiPhone®oriPad®:Download
HOBOmobilefromtheAppStore®.
IfyouareusinganAndroidphoneortablet:Download
HOBOconnectfromGooglePlay™.
2. OpentheappandenableBluetoothinthedevicesettingsif
prompted.
3. Pressthebuttonontheloggertowakeitup.
4. TaptheDevicesintheapp.(InHOBOmobile,tapLoggersat
thetopofthescreen.)Taptheloggerintheapptoconnect
toit.
Iftheloggerdoesnotappearorifitishavingtrouble
connecting,followthesetips:
Makesuretheloggeris“awake”bypressingthestart
button.ThealarmandstatusLEDswillblinkoncewhen
theloggerwakesup.Youcanalsopressthebuttona
secondtimetobringittothetopofthelistifyouare
workingwithmultipleloggers.
Makesuretheloggeriswithinrangeofyourmobile
device.Therangeforsuccessfulwirelesscommunication
isapproximately30.5m(100ft)withfullline‐of‐sight.
Ifthereareseveralloggersinthearea,movethelogger
toalocationwithfewerloggers.Interferencecan
sometimesoccurwhennumerousloggersareinone
location.
Ifyourdevicecanconnecttotheloggerintermittentlyor
losesitsconnection,moveclosertothelogger,within
sightifpossible.
Iftheloggerappearsintheapp,butyoucannotconnect
toit,closetheappandpowercyclethemobiledevice.
ThisforcesthepreviousBluetoothconnectiontoclose.
Onceconnectedtotheloggeryoucanselectoneofthe
followingactions:
ConfigureinHOBOmobileorinHOBOconnect.
Selectloggersettingsandloadthemontotheloggerto
startlogging.SeeConfiguringtheLogger.
ReadoutinHOBOmobileorinHOBOconnect.
Downloadloggerdata.SeeReadingOuttheLogger.
FullStatusDetailsinHOBOmobileorin
HOBOconnect.Checkinformationaboutthelogger.In
HOBOconnect,accesssomeoftheactionsdescribedin
thissection.
StartLogginginHOBOmobileorandthenin
HOBOconnect.Selectthisoptiontobeginlogging(ifthe
loggerisconfiguredtostart“OnButtonPush”as
describedinConfiguringtheLogger).
StopLogginginHOBOmobileorandthenin
HOBOconnect.Stoptheloggerfromrecordingdata.This
overridesanyStopLoggingsettingsdescribedin
ConfiguringtheLogger.
PageLoggerLEDinHOBOmobileorandthenin
HOBOconnect.Pressandholdthisoptiontoilluminate
thealarmandstatusLEDsfor4seconds.
LoggerPasswordinHOBOmobileorandthenin
HOBOconnect.Selectthistocreateapasswordforthe
loggerthatwillberequiredifanothermobiledevice
attemptstoconnecttoit.Toresetapassword,connect
tothelogger,tapSetLoggerPasskey,andselectResetto
AlarmLED
Mounting
Holes
StartButton
ExternalSensor
StatusLED
MX2303modelshown
Vent
MX2301Amodelshown

HOBOMX2300SeriesDataLoggerManual
1‐800‐LOGGERS4www.onsetcomp.com
FactoryDefaultinHOBOmobileortapandtapReset
inHOBOconnect..Youcanalsopressthebuttononthe
loggerfor10secondstoresetapassword.
UpdateFirmwareinHOBOmobileorandthenin
HOBOconnect.Whennewloggerfirmwareisavailable,
thisactionappearsinthelist.Selectitandfollowthe
instructionsonthescreen.Aloggerreadoutwillbe
completedautomaticallyatthebeginningofthe
firmwareupdateprocess.Iftheconnectionislost
betweentheloggerandthemobiledeviceduringthe
firmwareupdate,aFirmwareUpdatePendingStatus
displaysfortheloggerintheDeviceslist.Connecttothe
loggerandupdatethefirmwareagain(orselectRestore
LoggerifavailableinHOBOmobile).
Important:Beforeupdatingthefirmwareonthe
logger,checktheremainingbatterylevelbyselecting
FullStatusDetailsandmakesureitisnolessthan30%.
Makesureyouhavethetimetocompletetheentire
updateprocess,whichrequiresthattheloggerremains
connectedtothedeviceduringtheupgrade.
ForceOffload(HOBOmobileonly).Thismayappearifan
errorwasencounteredwhenloadingconfiguration
settings.Selectthistooffloadallthedataonthelogger
beforereconfiguringit.
ConfiguringtheLogger
Usetheapptosetupthelogger,includingselectingthelogging
options,configuringalarms,andothersettings.Thesesteps
provideanoverviewofsettingupthelogger.Forcomplete
details,seetheappuser’sguide.
1. Pressthebuttonontheloggertowakeitup.
2. Intheapp,taptheDevicesicon(inHOBOmobile,tap
Loggersatthetopofthescreen).Taptheloggerintheapp
toconnecttoit.Ifyouareworkingwithmultipleloggers,
youcanpressthebuttonontheloggerasecondtimeto
bringittothetopofthelist.
3. TapConfigureinHOBOmobileortapinHOBOconnect.
4. TapNameandtypeanameforthelogger(optional).In
HOBOmobile,tapDone.Ifnonameisselected,thelogger
serialnumberisusedasthename.
5. TapGrouptoaddtheloggertoagroup(optional).TapDone
orSave.
6. TapLoggingIntervalandchoosehowfrequentlythelogger
willrecorddataunlessoperatinginburstloggingmode(see
BurstLogging).Note:Ifyouconfigureanalarm,thelogger
willusetheloggingintervalyouselectedastherateto
checkforalarmconditions(alarmsarenotavailableifburst
loggingisconfigured).SeeSettingupAlarmsformore
details.
7. TapStartLoggingandselectwhenloggingwillbegin:
Now.Loggingwillbeginimmediatelyafterconfiguration
settingsareloadedonthelogger.
OnNextLoggingInterval.Loggingwillbeginatthenext
evenintervalasdeterminedbytheselectedlogging
interval.
OnButtonPush.Loggingwillbeginonceyoupressthe
buttonontheloggerfor3seconds.
OnDate/Time.Loggingwillbeginonadateandtimeyou
specify.Selectthedateandtime.InHOBOmobile,tap
Done.
TapDoneorSave.
8. TapStopLoggingandselecttheoptionsforwhenlogging
willend.
a. Chooseoneoftwomemoryoptions:
WhenMemoryFills.Theloggerwillcontinuerecording
datauntilthememoryisfull.
Never(WrapWhenFull).Theloggerwillcontinue
recordingdataindefinitely,withnewestdata
overwritingtheoldest.Thisoptionisnotavailableif
theLoggingModeissettoBurst(seeBurstLogging).
b. SelectOnButtonPushifyouwanttobeabletostop
loggingbypushingthebuttonontheloggerfor3
seconds.ForHOBOmobile:Notethatifyoualsochoose
OnButtonPushfortheStartLoggingoption,thenyou
willnotbeabletostoplogginguntil30secondsafter
loggingbegins.
c. Selectoneofthefollowingtimeoptionsforwhentostop
logging:
Never.Selectthisifyoudonotwanttheloggertostop
atanypredeterminedtimeframe.
OnDate/Time.Selectthisifyouwanttheloggerto
stoploggingonaspecificdateandtime.Selectthe
dateandtime.InHOBOmobile,tapDone.
After.Selectthisifyouwanttocontrolhowlongthe
loggershouldcontinueloggingonceitstarts.Choose
theamountoftimeyouwanttheloggertologdata
andthentapDone.Forexample,select30daysifyou
wanttheloggertologdatafor30daysafterlogging
begins.
d. TapDoneorSave.
9. TapLoggingMode.Selecteitherfixedorburstlogging.With
fixedlogging,theloggerrecordsdataforallenabledsensors
and/orselectedstatisticsattheloggingintervalselected
(seeStatisticsLoggingfordetailsonchoosingstatistics
options).Inburstmode,loggingoccursatadifferent
intervalwhenaspecifiedconditionismet.SeeBurst
Loggingformoreinformation.TapDoneorSave.
10. EnableordisableShowLED.IfShowLEDisdisabled,the
alarmandstatusLEDsontheloggerwillnotbeilluminated
whilelogging(thealarmLEDwillnotblinkifanalarmtrips).
YoucantemporarilyturnonLEDswhenShowLEDis
disabledbypressingthebuttonontheloggerfor1second.
11. EnableordisableBluetoothAlwaysOn.IfBluetoothAlways
Onisenabled,theloggerwill“advertise”orregularlysend
outaBluetoothsignalforthephoneortablettofindviathe
appwhileitislogging,whichusesbatterypower.When
BluetoothAlwaysOnisdisabled,theloggerwillonly
advertiseduringloggingwhenyoupressthebuttononthe
loggertowakeitup,therebypreservingasmuchbattery
poweraspossible.
12. Selectthesensormeasurementtypesthatwillbelogged.
BoththetemperatureandRHsensorsarerequiredto
calculatedewpoint,whichisanadditionaldataseries

HOBOMX2300SeriesDataLoggerManual
1‐800‐LOGGERS5www.onsetcomp.com
availableforplottingafterreadingoutthelogger.Youcan
alsosetupalarmstotripwhenasensorreadingrisesabove
orfallsbelowaspecifiedvalue.SeeSettingupAlarmsfor
detailsonenablingsensoralarms.NoteforMX2303models
only:Thefirsttemperaturesensorlistedischannel1and
thesecondischannel2(and“‐1”and“‐2”isusedinthe
columnheadingsinthedatafiletodifferentiatethetwo
sensors).
13. TapStartintheupperrightcorneroftheConfigurescreen
inHOBOmobileortapinHOBOconnect.
Loggingwillbeginbasedonthesettingsyouselected.Press
thestartbuttonontheloggerifyousetituptostartlogging
withabuttonpush.SeeDeployingandMountingtheLogger
fordetailsonmountingandseeReadingOuttheLoggerfor
detailsondownloading.
SettingupAlarms
Youcansetupalarmsfortheloggersothatifasensorreading
risesaboveorfallsbelowaspecifiedvalue,theloggeralarm
LEDwillblinkandanalarmiconwillappearintheapp.Thiscan
alertyoutoproblemssoyoucantakecorrectiveaction.
Tosetanalarm:
1. TaptheDevicesicon(inHOBOmobile,tapLoggersatthe
topofthescreen).Taptheloggerintheapptoconnectto
it.IftheloggerwasconfiguredwithBluetoothAlwaysOn
disabled,pressthebuttonontheloggertowakeitup.
Whenworkingwithmultipleloggers,youcanalsopressthe
buttonontheloggertobringittothetopofthelist.
2. TapConfigureinHOBOmobileortapinHOBOconnect.
3. Tapasensor(taptheEnableLoggingtogglein
HOBOconnectifnecessary).
4. EnabletheHighAlarminHOBOmobileorselectHighin
HOBOconnectifyouwantanalarmtotripwhenthesensor
readingrisesabovethehighalarmvalue.Dragtheslideror
typeavaluetosetthehighalarmvalue.
5. EnabletheLowAlarminHOBOmobileorselectLowin
HOBOconnectifyouwantanalarmtotripwhenthesensor
readingfallsbelowthelowalarmvalue.Dragtheslideror
typeavaluetosetthelowalarmvalue.
6. FortheDuration,selecthowmuchtimeshouldelapse
beforethealarmtripsandselectoneofthefollowing:
Cumulative.Thealarmwilltriponcethesensorreadingis
outoftheacceptablerangefortheselecteddurationany
timeduringlogging.Forexample,ifthehighalarmisset
to85°Fandthedurationissetto30minutes,thenthe
alarmwilltriponcethesensorreadingshavebeenabove
85°Fforatotalof30minutessincetheloggerwas
configured.
Consecutive.Thealarmwilltriponcethesensorreading
isoutoftheacceptablerangecontinuouslyforthe
selectedduration.Forexample,thehighalarmissetto
85°Fandthedurationissetto30minutes,thenthe
alarmwillonlytripifallsensorreadingsare85°For
aboveforacontinuous30‐minuteperiod.
7. TapDoneorSaveandrepeatsteps3–6fortheothersensor
ifdesired.
8. Intheconfigurationsettings,selectoneofthefollowing
optionstodeterminehowthealarmindicationsarecleared:
LoggerReconfigured.Thealarmindicationwilldisplay
untilthenexttimetheloggerisreconfigured.
SensorinLimits.Thealarmiconindicationwilldisplay
untilthesensorreadingreturnstothenormalrange
betweenanyconfiguredhighandlowalarmlimits.
9. TapStartinHOBOmobileorinHOBOconnect.
Whenanalarmtrips,theloggeralarmLEDblinksevery4
seconds(unlessShowLEDisdisabled),analarmiconappearsin
theapp,andanAlarmTrippedeventislogged.Thealarmstate
willclearwhenthereadingsreturntonormalifyouselected
SensorinLimitsinstep8.Otherwise,thealarmstatewill
remaininplaceuntiltheloggerisreconfigured.
Notes:
Alarmlimitsarecheckedateverylogginginterval.For
example,iftheloggingintervalissetto5minutes,then
theloggerwillcheckthesensorreadingsagainstyour
configuredhighandlowalarmsettingevery5minutes.
Theactualvaluesforthehighandlowalarmlimitsareset
totheclosestvaluesupportedbythelogger.Inaddition,
alarmscantriporclearwhenthesensorreadingiswithin
theresolutionspecifications.
Whenyoureadoutthelogger,alarmeventscanbe
displayedontheplotorinthedatafile.SeeLogger
Events.
BurstLogging
Burstloggingisaloggingmodethatallowsyoutosetupmore
frequentloggingwhenaspecifiedconditionismet.For
example,aloggerisrecordingdataata5‐minutelogging
intervalandburstloggingisconfiguredtologevery30seconds
whenthetemperaturerisesabove85°F(thehighlimit)orfalls
below32°F(thelowlimit).Thismeanstheloggerwillrecord
dataevery5minutesaslongasthetemperatureremains
between85°Fand32°F.Oncethetemperaturerisesabove
85°F,theloggerwillswitchtothefasterloggingrateandrecord
dataevery30secondsuntilthetemperaturefallsbackto85°F.
Atthattime,loggingthenresumesevery5minutesatthe
normallogginginterval.Similarly,ifthetemperaturefallsbelow
32°F,thentheloggerwouldswitchtoburstloggingmodeagain
andrecorddataevery30seconds.Oncethetemperaturerises
backto32°F,theloggerwillthenreturntonormalmode,
loggingevery5minutes.Note:Sensoralarms,statistics,andthe
StopLoggingoption“WrapWhenFull”arenotavailablein
burstloggingmode.
Tosetupburstlogging:
1. TaptheDevicesicon(inHOBOmobile,tapLoggersatthe
topofthescreen).Taptheloggerintheapptoconnectto
it.IftheloggerwasconfiguredwithBluetoothAlwaysOn
disabled,pressthebuttonontheloggertowakeitup.
Whenworkingwithmultipleloggers,youcanalsopressthe
buttonontheloggertobringittothetopofthelist.
2. TapConfigureinHOBOmobileortapinHOBOconnect.
3. TapLoggingModeandthentapBurstLogging.In
HOBOmobile,tapasensorunderBurstSensorLimits.

HOBOMX2300SeriesDataLoggerManual
1‐800‐LOGGERS6www.onsetcomp.com
4. SelectLowand/orHighandeithertypeordragthesliderto
setthelowand/orhighvalues.TapDoneinHOBOmobile.
5. Repeatstep4fortheothersensorifdesired.
6. TapDoneandrepeatsteps4–7fortheothersensorif
desired.
7. Settheburstlogginginterval,whichmustbefasterthanthe
logginginterval.Keepinmindthatthefastertheburst
loggingrate,thegreatertheimpactonbatterylifeandthe
shortertheloggingduration.Becausemeasurementsare
beingtakenattheburstloggingintervalthroughoutthe
deployment,thebatteryusageissimilartowhatitwouldbe
ifyouhadselectedthisrateforthenormallogginginterval.
TapDoneinHOBOmobile.
8. TapDoneorSave.
9. TapStartinHOBOmobileorinHOBOconnect.
Notes:
Thehighandlowburstlimitsarecheckedattheburst
loggingintervalratewhethertheloggerisinnormalor
burstcondition.Forexample,iftheloggingintervalisset
to1hourandtheburstloggingintervalissetto10
minutes,theloggerwillalwayscheckforburstlimits
every10minutes.
Ifhighand/orlowlimitshavebeenconfiguredformore
thanonesensor,thenburstloggingwillbeginwhenany
highorlowconditiongoesoutofrange.Burstloggingwill
notenduntilallconditionsonallsensorsarebackwithin
normalrange.
Theactualvaluesfortheburstlogginglimitsaresetto
theclosestvaluesupportedbythelogger.
Burstloggingcanbeginorendwhenthesensorreadingis
withintheresolutionspecifications.Thismeansthevalue
thattriggersburstloggingmaydifferslightlythanthe
valueentered.
Oncethehighorlowconditionclears,thelogging
intervaltimewillbecalculatedusingthelastrecorded
datapointinburstloggingmode,notthelastdatapoint
recordedatthenormalloggingrate.Forexample,the
loggerhasa10‐minuteloggingintervalandloggedadata
pointat9:05.Then,thehighlimitwassurpassedand
burstloggingbeganat9:06.Burstloggingthencontinued
until9:12whenthesensorreadingfellbackbelowthe
highlimit.Nowbackinnormalmode,thenextlogging
intervalwillbe10minutesfromthelastburstlogging
point,or9:22inthiscase.Ifburstlogginghadnot
occurred,thenextdatapointwouldhavebeenat9:15.
ANewIntervaleventiscreatedeachtimethelogger
entersorexitsburstloggingmode.SeeLoggerEventsfor
detailsonplottingandviewingtheevent.Inaddition,if
theloggerisstoppedwithabuttonpushwhileinburst
loggingmode,thenaNewIntervaleventisautomatically
loggedandtheburstconditioniscleared,evenifthe
actualhighorlowconditionhasnotcleared.
StatisticsLogging
Duringfixedlogging,theloggerrecordsdataforenabled
sensorsand/orselectedstatisticsatthelogginginterval
selected.Statisticsarecalculatedatasamplingrateyouspecify
withtheresultsforthesamplingperiodrecordedateach
logginginterval.Thefollowingstatisticscanbeloggedforeach
sensor:
Themaximum,orhighest,sampledvalue,
Theminimum,orlowest,sampledvalue,
Anaverageofallsampledvalues,and
Thestandarddeviationfromtheaverageforallsampled
values.
Forexample,aloggerisconfiguredwithboththetemperature
andRHsensorsenabled,andtheloggingintervalsetto5
minutes.TheloggingmodeissettofixedloggingwithNormal
andallfourstatisticsenabledandwithastatisticssampling
intervalof30seconds.Onceloggingbegins,theloggerwill
measureandrecordtheactualtemperatureandRHsensor
valuesevery5minutes.Inaddition,theloggerwilltakea
temperatureandRHsampleevery30secondsandtemporarily
storetheminmemory.Theloggerwillthencalculatethe
maximum,minimum,average,andstandarddeviationusingthe
samplesgatheredovertheprevious5‐minuteperiodandlog
theresultingvalues.Whenreadingoutthelogger,thiswould
resultin10dataseries(notincludinganyderivedseries,such
asdewpoint):twosensorseries(withtemperatureandRH
dataloggedevery5minutes)pluseightmaximum,minimum,
average,andstandarddeviationseries(fourfortemperature
andfourforRHwithvaluescalculatedandloggedevery5
minutesbasedonthe30‐secondsampling).
Tologstatistics:
1. TaptheDevicesicon(inHOBOmobile,tapLoggersatthe
topofthescreen).Taptheloggerintheapptoconnectto
it.IftheloggerwasconfiguredwithBluetoothAlwaysOn
disabled,pressthebuttonontheloggertowakeitup.
Whenworkingwithmultipleloggers,youcanalsopressthe
buttonontheloggertobringittothetopofthelist.
2. TapConfigureinHOBOmobileortapinHOBOconnect.
3. TapLoggingModeandthenselectFixedIntervalLoggingin
HOBOmobileorFixedLogginginHOBOconnect.
4. SelectNormaltorecordthecurrentreadingforeach
enabledsensorattheloggingintervalshownatthetopof
thescreen.Donotselectthisifyouonlywanttolog
statistics.
5. Selectthestatisticsyouwanttheloggertorecordateach
logginginterval:Maximum,Minimum,Average,and
StandardDeviation(averageisautomaticallyenabledwhen
selectingStandardDeviation).Statisticswillbeloggedforall
enabledsensors.Inaddition,themorestatisticsyourecord,
theshortertheloggerdurationandthemorememoryis
required.
6. TapStatisticsSamplingIntervalandselecttheratetouse
forcalculatingstatistics.Theratemustbelessthan,anda
factorof,thelogginginterval.Forexample,ifthelogging
intervalis1minuteandyouselect5secondsforthe
samplingrate,thentheloggerwilltake12samplereadings
betweeneachlogginginterval(onesampleevery5seconds
foraminute)andusethe12samplestorecordtheresulting
statisticsateach1‐minutelogginginterval.Notethatthe
fasterthesamplingrate,thegreatertheimpactonbattery
life.Becausemeasurementsarebeingtakenatthestatistics
samplingintervalthroughoutthedeployment,thebattery
usageissimilartowhatitwouldbeifyouhadselectedthis

HOBOMX2300SeriesDataLoggerManual
1‐800‐LOGGERS7www.onsetcomp.com
rateforthenormallogginginterval.TapDonein
HOBOmobile.
7. TapDoneorSave.
8. TapStartinHOBOmobileorinHOBOconnect.
SettingaPassword
Youcancreateanencryptedpasswordfortheloggerthatwill
berequiredifanotherphoneortabletattemptstoconnectto
it.Thisisrecommendedtoensurethatadeployedloggerisnot
mistakenlystoppedorpurposelyalteredbyothers.This
passwordusesaproprietaryencryptionalgorithmthatchanges
witheveryconnection.
Tosetapassword:
1. TaptheDevicesicon(inHOBOmobile,tapLoggersatthe
topofthescreen).Taptheloggerintheapptoconnectto
it.
2. TapLoggerPasswordinHOBOmobileortapandthen
inHOBOconnect.
3. TypeapasswordandthentapSaveorSet.
Onlythephoneortabletusedtosetthepasswordcanthen
connecttotheloggerwithoutenteringapassword;allother
mobiledeviceswillberequiredtoenterthepassword.For
example,ifyousetthepasswordfortheloggerwithyourtablet
andthentrytoconnecttothedevicelaterwithyourphone,
youwillberequiredtoenterthepasswordonthephonebut
notwithyourtablet.Similarly,ifothersattempttoconnectto
theloggerwithdifferentdevices,thentheywouldalsobe
requiredtoenterthepassword.Toresetapassword,pressthe
buttonontheloggerfor10secondsorconnecttothelogger
andthentapSetLoggerPasskeyandselectResettoFactory
DefaultinHOBOmobileortap,then,andtapResetin
HOBOconnect.
ReadingOuttheLogger
Tooffloaddatafromthelogger:
1. TaptheDevicesiconinHOBOmobile,tapLoggersatthetop
ofthescreen).Taptheloggerintheapptoconnecttoit.If
theloggerwasconfiguredwithBluetoothAlwaysOn
disabled,pressthebuttonontheloggertowakeup
communications.Whenworkingwithmultipleloggers,you
canalsopressthebuttonontheloggertobringittothetop
ofthelist.
2. TapReadoutinHOBOmobileortapinHOBOconnect.
3. InHOBOmobile:TaptheDataFilesiconandthentapthe
mini‐graphtoviewalargerversionofthegraphorshare
thefile.
InHOBOconnect:TaptheHOBOFilesiconandselectthe
filetoviewit.Tapandthentoexportandsharethe
data.
DatacanalsobeuploadedautomaticallytoHOBOlink,Onset’s
web‐basedsoftware,viatheapportheMXgateway.For
details,seetheappuser’sguideandseetheHOBOlinkhelpfor
detailsonworkingwithdatainHOBOlink.
LoggerEvents
Theloggerrecordsthefollowinginternaleventstotracklogger
operationandstatus.Youcanvieweventsinexportedfilesor
ploteventsintheapp.
ToploteventsinHOBOmobile,tapamini‐graphandthen
tap.Selecttheeventsyouwanttoplotandthentap
again.
ToploteventsinHOBOconnect,taptheHOBOFilesicon
andselectafiletoopen.Tapandthentap.Select
theeventsyouwanttoplotandtapOK.
InternalEventNameDefinition
HostConnectedTheloggerwasconnectedtothemobile
device.
StartedTheloggerstartedlogging.
StoppedTheloggerstoppedlogging.
Alarm
Tripped/Cleared
Analarmhasoccurredbecausethereading
wasoutsidethealarmlimitsorbackwithin
range.Note:Althoughthereadingmayhave
returnedtoanormalrangeduringlogging,
analarmclearedeventwillnotbeloggedif
theloggerwassetuptomaintainalarms
untilreconfigured.
NewIntervalTheloggerhasswitchedtologgingatthe
burstloggingrateorbacktothenormalrate.
SafeShutdownThebatteryleveldroppedbelowasafe
operatingvoltageandtheloggerperformed
asafeshutdown.
DeployingandMountingtheLogger
Followtheseguidelineswhendeployingthelogger:
AsolarradiationshieldisrequirediftheMX2301Aor
MX2305loggerortheexternalsensorsfroman
MX2302A,MX2303,orMX2304loggerwillbeinsunlight
atanytime.
WhenusingasolarradiationshieldwithanMX2301Aor
MX2305model,theloggermustbemountedusingthe
solarradiationshieldbracket(MX2300‐RS‐BRACKET)to
theundersideofthemountingplateasshownonthe
nextpage.
Formoredetailsonthesolarradiationshield,refertothe
SolarRadiationShieldInstallationGuideat
www.onsetcomp.com/manuals/rs1.
WhendeployinganMX2302Alogger,itisrecommended
thatthesensormustbemountedvertically.Ifitmustbe
mountedhorizontally,thenmakesuretheventonthe

HOBOMX2300SeriesDataLoggerManual
1‐800‐LOGGERS8www.onsetcomp.com
sideofthesensorisverticalorfacingdown.Ifthesensor
isbeingdeployedinanRS3‐Bsolarradiationshield,
mountitverticallyasshownbelow.
Whendeployingaloggerwithexternalsensors
(MX2302A,MX2303,andMX2304),mounttheloggerso
thesensorcableisnotbeingpulled.Leaveabout5cm(2
in.)ofdriploopinthecablewhereitcomesoutofthe
loggertopreventwaterfromenteringthelogger
housing.
ForMX2301AandMX2305loggersthatarenotbeing
deployedwithasolarradiationshieldorforloggerswith
externalsensors(MX2302A,MX2303,andMX2304),you
caneitherusetheincludedlargescrewsorcabletiesto
mounttheloggerviathemountingholes.Usethescrews
toattachtheloggertoawallorflatsurface.Usethe
cabletiestoaffixtheloggertoaPVCpipeormast.The
MX2301Aloggermustalsobemountedverticallyorwith
thesensorventfacingdownwhennotusingthesolar
radiationshield.
ProtectingtheLogger
Note:Staticelectricitymaycausetheloggertostoplogging.
Theloggerhasbeentestedto8KV,butavoidelectrostatic
dischargebygroundingyourselftoprotectthelogger.Formore
information,searchfor“staticdischarge”ononsetcomp.com.
BatteryInformation
Theloggerrequiresoneuser‐replaceable2/3AA3.6Vlithium
battery(HRB‐2/3AA).Batterylifeis2year,typicalwithalogging
intervalof1minute,butmaybeextendedto5yearswhenthe
loggerisconfiguredwithBluetoothAlwaysOndisabled.
Expectedbatterylifevariesbasedontheambienttemperature
wheretheloggerisdeployed,theloggingorsamplinginterval,
frequencyofoffloadingandconnectingtothemobiledevice,
numberofchannelsthatareactive,anduseofburstmodeor
statisticslogging.Deploymentsinextremelycoldorhot
temperaturesorloggingintervalfasterthan1minutecan
impactbatterylife.Estimatesarenotguaranteeddueto
uncertaintiesininitialbatteryconditionsandoperating
environment.
Toinstallorreplacethebattery:
1. UseaPhillips‐headscrewdrivertounscrewthefourscrews
fromthebackofthelogger.
2. Carefullyseparatethetopandbottomofthelogger
enclosure.
3. Removetheoldbatteryandinsertthenewbattery
observingpolarity.Itisrecommendedthatyoureplacethe
desiccant(DESICCANT2)whenreplacingthebattery.
4. Makesuretherubbersealiscleanandfreeofanydebris
andthencarefullyreassembletheloggerenclosureand
screwinthefourscrews.
WARNING:Donotcutopen,incinerate,heatabove85°C
(185°F),orrechargethelithiumbattery.Thebatterymay
explodeiftheloggerisexposedtoextremeheatorconditions
thatcoulddamageordestroythebatterycase.Donotdispose
oftheloggerorbatteryinfire.Donotexposethecontentsof
thebatterytowater.Disposeofthebatteryaccordingtolocal
regulationsforlithiumbatteries.
Unscrewthesefour
screwstoreplace
thebattery
DripLoop

HOBOMX2300SeriesDataLoggerManual
1‐800‐LOGGERS(564‐4377)•508‐759‐9500
www.onsetcomp.com/support/contact
©2016–2020OnsetComputerCorporation.Allrightsreserved.Onset,HOBO,HOBOmobile,HOBOconnect,and
HOBOlinkareregisteredtrademarksofOnsetComputerCorporation.AppStore,iPhone,andiPadareservice
marksorregisteredtrademarksofAppleInc.AndroidandGooglePlayaretrademarksofGoogleLLC.Bluetooth
andBluetoothSmartisaregisteredtrademarkofBluetoothSIG,Inc.Allothertrademarksarethepropertyoftheir
respectivecompanies.
Patent#:8,860,56920923‐K
FederalCommunicationCommissionInterferenceStatement
ThisequipmenthasbeentestedandfoundtocomplywiththelimitsforaClassBdigitaldevice,pursuanttoPart15oftheFCCRules.Theselimitsaredesignedtoprovide
reasonableprotectionagainstharmfulinterferenceinaresidentialinstallation.Thisequipmentgeneratesusesandcanradiateradiofrequencyenergyand,ifnotinstalledand
usedinaccordancewiththeinstructions,maycauseharmfulinterferencetoradiocommunications.However,thereisnoguaranteethatinterferencewillnotoccurina
particularinstallation.Ifthisequipmentdoescauseharmfulinterferencetoradioortelevisionreception,whichcanbedeterminedbyturningtheequipmentoffandon,theuser
isencouragedtotrytocorrecttheinterferencebyoneofthefollowingmeasures:
Reorientorrelocatethereceivingantenna
Increasetheseparationbetweentheequipmentandreceiver
Connecttheequipmentintoanoutletonacircuitdifferentfromthattowhichthereceiverisconnected
Consultthedealeroranexperiencedradio/TVtechnicianforhelp
ThisdevicecomplieswithPart15oftheFCCRules.Operationissubjecttothefollowingtwoconditions:(1)Thisdevicemaynotcauseharmfulinterference,and(2)thisdevice
mustacceptanyinterferencereceived,includinginterferencethatmaycauseundesiredoperation.
FCCCaution:Anychangesormodificationsnotexpresslyapprovedbythepartyresponsibleforcompliancecouldvoidtheuser'sauthoritytooperatethisequipment.
IndustryCanadaStatements
ThisdevicecomplieswithIndustryCanadalicense‐exemptRSSstandard(s).Operationissubjecttothefollowingtwoconditions:(1)thisdevicemaynotcauseinterference,and
(2)thisdevicemustacceptanyinterference,includinginterferencethatmaycauseundesiredoperationofthedevice.
Avisdeconformitépourl’IndustrieCanada
LeprésentappareilestconformeauxCNRd'IndustrieCanadaapplicablesauxappareilsradioexemptsdelicence.L'exploitationestautoriséeauxdeuxconditionssuivantes:(1)
l'appareilnedoitpasproduiredebrouillage,et(2)l'appareildoitacceptertoutbrouillageradioélectriquesubi,mêmesilebrouillageestsusceptibled'encompromettrele
fonctionnement.
TocomplywithFCCandIndustryCanadaRFradiationexposurelimitsforgeneralpopulation,theloggermustbeinstalledtoprovideaseparationdistanceofatleast20cmfrom
allpersonsandmustnotbeco‐locatedoroperatinginconjunctionwithanyotherantennaortransmitter.
KCStatement
해당 무선설비는 전파혼신 가능성이 있으므로 인명안전과 관련된 서비스는 할 수 없음
Translation:
Theservicerelatedtohumansafetyisnotallowedbecausethisdevicemayhavethepossibilityoftheradiointerference.
This manual suits for next models
5
Table of contents