
2 If the difference is ±2℃ or less, the instrument is correct.
Caution…Do not touch the sensor during this work.
We recommend to short-circuit the connect pins removing the sensor.
Ask HOZAN dealer for calibration and adjustment. We are able to issue test data,
calibration certificates and traceability charts on request.
You can perform calibration (judging if the unit is correct) and adjustment by yourself, but
it is not recommended for precise measuring and we are not able to guarantee its result.
Only calibrating procedures are written as follows. We send a detailed manual for
adjustment when requested.
Prepare a standard generator or direct current power supply and a voltmeter for
calibration and adjustment.
:23℃ or closer
:Leave the thermometer to be calibrated as
well as the instruments to use for calibration
for 30 minutes.
Recommended ambient temperature
Preparation
Calibration
1 First, check if the standard contact temperature given above is correct.
2 Pull out the sensor and wire like below figure. Turn the power of the instrument on.
Follow the instructions of the standard generator regarding its handling.
(The H-767 has a function of compensating ambient temperature automatically.
Use a proper standard generator with a thermo-probe.)
3 Apply voltage corresponding to 400℃ from the standard generator.
Calibration
In case of using a DC standard generator for testing thermoelectric thermometer
●
1 While installing the sensor, turn the power on with wiring nothing. After ten minutes
of warm-up, make a comparison between displayed temperature and the actual
ambient temperature.
Calibration of standard contact temperature setting
●
Sensor socket
Thermo-probe
+
-
Standard generator
4