DeltaTRAK DT-8806H User manual

Please read this manual before operating unit.
Important safety information inside.
User Manual
Model 15004
DT-8806H Non-Contact Forehead Infrared Thermometer

Non-Contact Forehead IR Thermometer User ManualNon-Contact Forehead IR Thermometer User Manual
Contents
1. General Description.......................................................................1
2. Safety Information..........................................................................1
3. Features......................................................................................... 2
4. Intended Use.................................................................................. 3
5. Conguration.................................................................................4
6. Indicator..........................................................................................4
7. Description of Symbols.................................................................. 5
8. Technical Specications................................................................ 6
9. Advice For the 1st Use....................................................................7
10. Use................................................................................................8
11. Measuring Operation................................................................... 9
12. Advice.........................................................................................11
13. Maintenance and Cleaning.........................................................11
14. Troubleshooting..........................................................................12
15. This Appliance Conforms to the Following Standards...............14

1
Non-Contact Forehead IR Thermometer User ManualNon-Contact Forehead IR Thermometer User Manual
1. General Description
Non-Contact Forehead IR Thermometer is specically designed to take
human body temperature, regardless of room temperature.
2. Safety Information
• This device must only be used for the purposes described in this
instruction manual.
• This device must only be used in an ambient temperature range
between 10°C to 40°C (50°F to 104°F).
• Do not expose this thermometer to electric shocks. Do not expose
this thermometer to extreme temperature conditions of > 50°C
(122°F) or < 0°C (32°F).
• Do not use the device in relative humidity higher than 85%.
• Do not use the device near large electromagnetic elds such as
found with cordless or cell phones.
• Keep the device away from water and heat, including direct sunlight.
• Do not drop or strike the device, and do not use if damaged.
• Accuracy of measurements may be affected when the forehead is
covered by hair, perspiration, or clothing. (See Part 10-4)
• Maintain a measuring distance of 5cm to 15cm (2in to 5.9in). (See
Part 10-4)
• Allow device to acclimate to ambient temperature for 15 to 20
minutes before using.
• Temperature may also be taken behind the ear lobe. (See Part 10-5)
• Clean the device surface with a cotton pad lightly moistened with
70% Isopropyl alcohol.

22
Non-Contact Forehead IR Thermometer User Manual Non-Contact Forehead IR Thermometer User Manual
Important Information:
• Before taking temperature remove hair and perspiration from forehead.
• Select “Body” mode to measure body temperature .
• Select “Surface” mode to measure the surface temperature.
• Use of this thermometer is not intended as a substitute for
consultation with a physician.
• Should a problem occur with device, please contact retailer.
• Do not attempt to repair the device.
• Follow EMC guidelines for maintaining medical electronic products.
3. Features:
• Precise non-contact measurements
• User selectable °C or °F
• Selectable Body and Surface mode
• Set Alarm value
• Memory holds 32 points of temperature data
• Automatic data hold
• Auto power off
• Automatic selection range and display resolution 0.1°C (0.1°F)
• Backlight LCD display

3
Non-Contact Forehead IR Thermometer User Manual Non-Contact Forehead IR Thermometer User Manual
4. Intended Use
Non-Contact Forehead IR Thermometer is designed for body surface
and forehead temperature measurement for infants and adults without
contact to human body.
Non-Contact Forehead IR Thermometer can also be used to measure
the temperature of a baby bottle, bath or room temperature (by using
the Surface Temp function).
Normal Temperatures According To Measurement Method
Measurement
Method
Normal Temp °C Normal Temp °F
Rectal 36.6 to 38 97.8 to 100.4
Oral 35.5 to 37.5 95.9 to 99.5
Axillary 34.7 to 37.3 94.4 to 99.1
Ear 35.8 to 38 96.4 to 100.4
The temperature of the human body varies throughout the day. It
can also be inuenced by the following factors: age, sex, type and
thickness of skin.
Normal Temperatures According To Age
Age Temp °C Temp °F
0-2 years 36.4 to 38.0 97.5 to 100.4
3-10 years 36.1 to 37.8 97.0 to 100.0
11-65 years 35.9 to 37.6 96.6 to 99.7
> 65 years 35.8 to 37.5 96.4 to 99.5

44
Non-Contact Forehead IR Thermometer User Manual Non-Contact Forehead IR Thermometer User Manual
5. Conguration
1. IR Sensor
2. LCD Display
3. Mode Selection
4. Down Button
5. Up Button
6. Mode Button
7. Measurement Trigger
8. Battery Cover
6. Indicators
1. Surface Mode Symbol
2. Body Mode Symbol
3. Digital Readout
4. Battery Symbol
5. Log Number
6. Stored Temperatures
7. Temperature °C/°F
8. Buzzer symbol

5
Non-Contact Forehead IR Thermometer User Manual Non-Contact Forehead IR Thermometer User Manual
7. Description of Symbols
0197
The device is in accordance with Medical Device
Directive 93/42/ EEC
The device is in accordance with FCC Part 15 Subpart
B: 2007 /Radio Frequency Devices IC Regulation
ICEC-00 3:
2004 Interference-causing Equipment Standard-
Digital Apparatus
3V DC power supply (2x 1.5v AA Batteries)
Type B equipment
In order to protect the environment, please recycle the
battery according to the local regulations
Body
Surface
Indication of Mode
Attention! Consult accompanying documents.

66
Non-Contact Forehead IR Thermometer User Manual Non-Contact Forehead IR Thermometer User Manual
8. Technical Specications
Normal Conditions of Use
Display Resolution 0.1°C (0.1°F)
Operating Temperature 10°C to 40°C (50°F to 104°F)
Storage Temperature 0°C to 50°C (32°F to 122°F)
Humidity Rate ≤ 85%
Power DC 3V (2x 1.5v AA Batteries)
Size 149 x 77x 43 mm / 5.9 x 3 x 1.7 in (L X W X H)
Weight Gross: 0.88lbs/400 g
Net: 0.38lbs/172 g
Measuring Range
In Body Mode 32.0° to 42 .5°C (90° to 108°F)
In Surface Temp Mode 0° to 60°C (32° to 140°F)
Accuracy ± 0.3°C (0.54°F)
Measuring Distance 5 cm - 15 cm (2 in - 5.9 in)
Response Time 0.5 seconds
Non-Contact Body Infrared Thermometer Precision
32° to 35.9°C / 93.2° to 96.6°F ± 0.3°C / 0.5°F
According to ASTM
Standard E1965-1998
(2003)
36° to 39°C / 96.8° to 102.2°F ± 0.2°C / 0.4°F
39° to 42.5°C / 102.2° to
108.5°F ± 0.3°C / 0.5°F

7
Non-Contact Forehead IR Thermometer User Manual Non-Contact Forehead IR Thermometer User Manual
9. Advice for the First Use
For stable and reliable results, it is essential to check the Non-Contact
Body Infrared Thermometer and change as needed, as follows:
1st step: Take the temperature of a person using a conventional
thermometer; you will get 37.5°C (99.5°F) for instance.
2nd step: Take the temperature of the same person using the Non-
Contact Body Infrared Thermometer keeping the 5 cm to 15
cm (2 in to 5.9 in) distance between the thermometer and the
forehead. Take care to remove any obstacle which could alter
the measurement, such as hair, perspiration, etc. If you get
37.5°C (99.5°F), the Non-Contact Body Infrared Thermometer
is properly set and ready for use.
If you get a lower temperature, 36.4°C (97 .4°F) for example, your
difference is 1.1°C (2.2°F). You should adjust the temperature on the
Non-Contact Body Infrared Thermometer and add the difference, i.e.
1.1°C (2.2°F). Press “MODE” button for 2 seconds, the screen displays
F1. Press “MODE” button again until you get F3. Press “UP” button in
order to add the difference. For example: 1.1°C - 2.2°F.
3rd step: To check, take the temperature again using the Non-Contact
Body Infrared Thermometer.

88
Non-Contact Forehead IR Thermometer User Manual Non-Contact Forehead IR Thermometer User Manual
10. Use
10-1. Install battery.
10-2. Prior to the rst use or when inserting new batteries wait at least
10 minutes for device to warm up.
10-3. lf use is suspended for a long period, once turned on the device
will test the room temperature and delay start for 1 to 2 seconds.
10-4. Aim device at the forehead (see the diagram below for the
positioning), from a distance of 5 cm (2 in), press the measuring
key; the temperature is displayed immediately. Ensure there is no
hair, perspiration, cosmetics or clothing covering the forehead.
10-5. When the room temperature is signicantly different, or there
is perspiration on the forehead, temperature may also be
taken behind the ear lobe. Ensure there is no hair, perspiration,
cosmetics or clothing covering the area.

9
Non-Contact Forehead IR Thermometer User Manual Non-Contact Forehead IR Thermometer User Manual
11. Measuring Operation
11-1. Choosing the Temperature Unit - F1 Function
Press “MODE” button for 2 seconds, the screen displays: F1. Select
“DOWN” for degrees Celsius, “UP” for degrees Fahrenheit.
11-2. Alarm Setup - F2 Menu
Press “MODE” button for 2 seconds. The screen displays: F1. Press
“MODE” button twice to get F2. Select “UP” to increase the threshold
by 0.1°C (0.1°F), or “DOWN” to reduce it by 0.1°C (0.1°F).
Note: The alarm threshold default value is 38°C (100.4°F).
11-3. Total Difference - F3 Menu
To adjust the total variation of the Non-Contact Body Infrared
Thermometer, press “MODE” button for 2 seconds. The screen
displays: F1. Press “MODE” button two times to get F3. Select “UP”
to increase the difference by 0.1°C (0.1 °F), or “DOWN” to reduce it
by 0.1°C (0.1 °F). ln the case of seasonal or environmental changes,
verication and adjustment should be completed.
Note: This function is only effective in “Body” mode.
11-4. Buzzer ON/OFF - F4(F3) Menu
Press “MODE” button for 2 seconds. The screen displays: F1. Press
the “MODE” button three times to get F4 (F3).
Select “UP” to open the buzzer (a sound icon “ ” is displayed on the
LCD screen). Press “DOWN” to stop it (the icon disappears).
11-5. Exiting the Setting Mode
Press “MODE” button until the screen turns off.

1010
Non-Contact Forehead IR Thermometer User Manual Non-Contact Forehead IR Thermometer User Manual
11-6. The Non-Contact Body Infrared Thermometer is designed to
take the body temperature of a human being. Measurement range for
“Body” mode is 32°C to 42.5°C (86°F to108°F). In “Surface” mode you
can also use the Non-Contact Body Infrared Thermometer to measure
the temperature of an area or an object, a food, a liquid or room
temperature. Measurement range for “Surface” mode is 0°C to 60°C
(32°F to 140°F).
Important: The area temperature differs from the internal body
temperature. To obtain the internal temperature always use the “Body”
mode.
Please make sure to select “Body” mode for an internal temperature
reading and “Surface” mode for an external area reading, such as
bottle, bath, room, etc.
11-7. Data Memory
The device will automatically hold in memory the last 32 temperature
measurements, which will display at the right corner of LCD. Press “UP”
or “DOWN” button to display the last temperature measurement.
In power off, press “UP” and “DOWN” button together for two seconds,
to display the last temperature measurement. In this mode, change the
order number to 0 and press the “MODE” button, to delete all memory.
11-8. Changing The Batteries
When the LCD screen displays “ ”, the battery is depleted. When
changing the batteries, be careful not to damage the apparatus, which
will compromise the guarantee of the device. Never use rechargeable
batteries. Use only single-use batteries. Remove old battery immediately
to avoid damage to the thermometer resulting from battery leakage.
11-9. Guarantee
The Non-Contact Body Infrared Thermometer was developed for
professional use, and is guaranteed for 40,000 measurements.

11
Non-Contact Forehead IR Thermometer User Manual Non-Contact Forehead IR Thermometer User Manual
12. Advice
• The protective glass covering the lens is the most fragile part of the
thermometer. Please take great care to prevent damage.
• Do not use rechargeable batteries.
• Do not place device in re.
• Do not expose the device to sunlight or water.
• Maintain measuring distance of 5 cm to 15 cm (2 in to 5.9 in).
13. Maintenance and Cleaning
• Prevent damage to the infrared sensor.
• Clean the device surface with a cotton pad lightly moistened with
70% Isopropyl alcohol.
• Do not clean the device with corrosive detergent.
• Keep the device away from water or other liquid.
• Store the device in a dry environment.
• Keep device away from dust and direct sunlight.

1212
Non-Contact Forehead IR Thermometer User Manual Non-Contact Forehead IR Thermometer User Manual
14. Troubleshooting
The following may be used to address common problems. If the
problem persists please contact customer service.
The screen displays the body temperature lower than 32°C
(89.6°F) If you’re on “Surface” mode the 32°C (89.6 °F) temperature
displayed shows the external body temperature.
The screen displays “HI”
The measurement is above the range
selected, either higher than 42.5°C (108 °F)
in “Body” mode or higher than 60°C (140°F)
in “Surface” mode.
The screen displays “LO”
The measurement is lower than the range
selected, either lower than 32°C (90°F) in
“Body” mode or lower than 0°C (32°F) in
“Surface” mode.

13
Non-Contact Forehead IR Thermometer User Manual Non-Contact Forehead IR Thermometer User Manual
Troubleshooting, Continued
Reasons for “LO” or “HI”
Message Display
See Section 8
Temperature reading hampered
by hair, perspiration, etc.
Make sure that there are no
obstructions prior to taking
temperature.
Temperature hampered by an
air ux.
Make sure there is no air ux, as
this could interfere with the infrared
system.
The measuring distance is too far. Maintain a distance of 5 cm to 15
cm (2 in to 5.9 in).
From high/low temperature
condition to room temperature.
Wait 10 minutes before taking the
body temperature.
Contact Tech Support at:
Phone: 925-249-2250 Ext 5120 Toll Free: 800-390-0804

1414
Non-Contact Forehead IR Thermometer User Manual
15. This appliance conforms to the following standards:
• EN12470-5: European Standard for Clinical Thermometer
Performance - Infrared Ear Thermometers
• ASTM E1965-1998:
ASTM (American Society of Testing and Methodology)-
Standard Specication of Infrared Thermometers for Intermittent
Determination of Patient Temperature
• EN 980: Graphical symbols for use in the labeling of medical devices
• EN 1041:Information supplied by the manufacturer with medical
devices
• EN 60601-1: Medical electrical equipment Part 1: General
requirements for safety (IEC:60601-1 :1998)
• EN 60601-1-2: Medical electrical equipment Part 1-2 : General
requirements for safety Collateral standard Electromagnetic
compatibility Requirements and test (IEC 60601-1-2:2001)
EMC Statement
This device has been tested and approved in accordance with
EN60601-1-2:2007 for EMC. This does not guarantee in any way that
the device will not be affected by electromagnetic interference. Avoid
using the device in high electromagnetic environments.
© 2020 DeltaTrak, Inc.

Non-Contact Forehead IR Thermometer User Manual
UG_15004_20C7
P.O. Box 398 Pleasanton CA 94566
(925) 249-2250 (800) 962-6776
www.deltatrak.com
DeltaTrak Corporate
Guadalajara, Mexico
+52-33-3188-3161 / 36712190
www.deltatrakmexico.com
DTI Mexico International
Shenzhen China
+86-755-8442-9388/2837-2741
2837-2664 | 8923-2778
www.deltatrakchina.com.cn
DTI China Limited
NT, Hong Kong
+852-3568-5538
www.dtiap.com
DTI Asia Pacic
Auckland, New Zealand
+64 9 5757 886
www.deltatraksouthpacific.com
DTI South Pacic
Osaka, Japan
+81-6-6616-5900
www.dtijapan.co.jp
DTI Japan Limited
Antwerp Belgium
+32 (0) 3-455-61-25
www.deltatrakeurope.be
DTI Europe bvba
DTI South America SA
Santiago, Chile
+562 2758 2866 +569 7477 1061
www.deltatraksouthamerica.com
Other manuals for DT-8806H
1
This manual suits for next models
1
Table of contents
Other DeltaTRAK Thermometer manuals

DeltaTRAK
DeltaTRAK DT-8806H User manual

DeltaTRAK
DeltaTRAK FlashCheck 11083 User manual

DeltaTRAK
DeltaTRAK ThermoTrace 15039 User manual

DeltaTRAK
DeltaTRAK FlashCheck 12237 User manual

DeltaTRAK
DeltaTRAK JXB-178 User manual

DeltaTRAK
DeltaTRAK CK-T1501 User manual

DeltaTRAK
DeltaTRAK ThermoTrace 15050 User manual

DeltaTRAK
DeltaTRAK FlashCheck 12217 User manual

DeltaTRAK
DeltaTRAK FlashLink 12215 User manual

DeltaTRAK
DeltaTRAK Essentials Q1000 User manual