Humboldt H-1200 User manual

product manual
05.11
H-1200, H-1250, H-1252
Cone Penetrometer

Introduction
Universal Penetrometer— H-1200
Direct-reading instrument for precision penetration measurements of bituminous
materials, cement, petrolatum and waxes, as well as food, cosmetics and pharmaceutical
products. Unit has 5" diameter indicator dial, graduated in 400 divisions of 0.1mm,
corresponding to 40mm penetration. Zero preset to eliminate errors. Includes 47.5g
plunger with 3.2mm hole, and two loading weights (50g and 100g). Overall dim. 10-1/2"
x 13" x 22" (266.7 x 330.2 x 558.8mm). Meets ASTM D5, D217, D1168, D1191, D1321,
D1403, D1831, D1855; AASHTO T49, T187 and others.
Special Application Penetrometers
Special Application Penetrometers are available in electric versions, like the model
H-1240. Call 1-800-544-7220 for information.
Portable Penetrometer— H-1250
Lighter and smaller than H-1200 for field work, unit’s micrometer adjusts for accurate
settings as well as coarse adjustment for approximate settings. Only one additional 50g
loading weight is included. Overall dim. 7" x 7" x 16" (178 x 178 x 406mm). Shipping wt.
8 lbs. (3.6kg)
Special Application Penetrometer
Installation
Level the instrument carefully by means of the leveling screws on the base, sighting
the spirit level from above to obtain accurate alignment. After the instrument has been
leveled, avoid shifting its location. Otherwise it will be necessary to repeat the leveling
process.
Insert penetrating needle or cone into the chuck and tighten the chuck screw.
Set the zero reference. Grasp the gear rack knob and pull up to release the clutch trigger.
With the trigger released, raise the test plunger as high as possible. If the dial reading
is not exactly zero, adjust the reading to zero by means of the zero setting screw (which
is on the bottom of the dial housing) which is adjustable as required. Use a #8 Allen
Wrench.
Description Model Shpg. wt.
Grease Penetrometer: Model H-1200
penetrometer with H-2520 grease cone H-2510 30lbs.
(13.6kg)
Battery Paste Penetrometer: Model H-1200
penetrometer with H-1255 battery paste cone H-1202 30lbs.
(13.6kg)
Description Model Shpg. wt.
Portable Grease Penetrometer: Model H-1250
penetrometer with H-2520 grease cone H-1252 8lbs.
(3.6kg)

Procedure
Add the required weights to plunger head. In adding weights and in order to make
up the required load for the test, note that the weight of the test plunger itself is 47.5
grams. For example, when the specified load for grease penetrations involving the use
of a grease cone is 150 grams, no weights need be added to the plunger head since
the combined weight of the cone (102.5 grams) and test plunger (47.5 grams) makes up
required load as specified by ASTM. Place a prepared sample in position on base.
Adjust the height of the mechanism head via the control sleeve on the main post so as
to bring the point of the penetrating instrument exactly into contact with the surface of
the sample. To adjust the height of the mechanism head, release the lock screw. This
adjustment may be accomplished more easily by placing a weak light to one side of the
sample container and observing the shadow formed by the penetrating instrument on
the surface of sample, until light disappears from the point. BE SURE TO RELOCK THE
HEAD SECURELY BY MEANS OF THE LOCK SCREW AFTER THE HEIGHT ADJUSTMENT
IS COMPLETED.
Release the test plunger, allowing the penetrating instrument to descend into
the sample. To release the test plunger, depress and hold the clutch trigger
in its released position during the specified time required for the test. Use a
stopwatch to verify release duration.
CAUTION: In operating the clutch trigger, grasp the finger grip firmly with the
forefinger, and with the thumb depress the clutch trigger quickly as far in as it
will go. HOLD THE CLUTCH TRIGGER IN DURING THE SPECIFIED LENGTH OF
THE TEST. At the specified duration, let go of the clutch trigger quickly, allowing
it to lock the test plunger instantly.
Reading Penetration
To read the depth of penetration, push down the gear rack knob gently as far as it will go.
The dial reading now indicates the depth of penetration directly in tenths of millimeters.
For example: If the pointer comes to a rest at the fourth mark past the 270 point, the
depth of penetration is 274 tenths of a millimeter or 27.4 millimeters.
With the Universal Penetrometer, penetration measurements can be made to a total
depth of 40 mm on a single reading. On depths greater than 40 mm the dial pointer
makes a complete revolution and moves past the zero position for a fraction of another
revolution. Simply add the fractional revolution to determine the total depth of
penetration.
Return the dial pointer to zero, as directed in the Installation paragraph. If the original
zero setting has been accurate, the dial pointer will return exactly to zero and subsequent
readings will check against standard gauge blocks.
We also supply a 15-gram test plunger – readily installed or removed
CAUTION: Keep hands, clothing and other objects away from moving parts when the
machine is in operation.

Penetration Needles and Cones
Application Specification Model Description
Bituminous materials
ASTM D5; AASHTO
T49, IP49; ASA Std.,
A37.1; Fed Spec.
SS.R. 406C, Meth.
214.01
H-1280
Standard hardened stainless steel
needle, 40-45mm exposed needle
length. Wt. 2.5g.
H-1300
Same as H-1280. Certified to ASTM
accuracy by independent laboratory.
Wt. 2.5g.
H-1302
Same as H-1290. Certified to ASTM
accuracy by independent laboratory.
Wt. 2.5g.
H-1290
Long hardened stainless steel needle,
50-55mm exposed needle length.
Wt. 2.5g.
Waxes with 250 or
less penetration ASTM D1321
H-1310
Hardened stainless steel wax
penetration needles with tapered point,
blunt tip of truncated cone. Ferrule is
approx. 3.2mm dia. Wt. 2.5g.
H-1317 Same as H-1310. Certified to ASTM
accuracy by independent laboratory.
Battery paste
N/A H-1255
Hardened stainless steel tip with
special plunger. Total wt. 60g ± .050g.
Joint sealant for
asphalt & concrete
pavements
ASTM D5329 H-1320
Resilience ball penetration tool.
Total wt. 27.5g.
Grease-testing
penetrometers ASTM D217, D937
ASA Std. Z11.3
H-2520
Hollow 90° brass cone, highly polished
stainless steel tip. Removable nut and
stem. Wt. 102.5g.
H-2522 Same as H-2520, completely made of
stainless steel. Wt. 102.5g.
For all 90° cones H-2525 Stainless steel replacement tip, nut and
stem.
Grease-testing
penetrometers ASTM D2884 H-2524
Magnesium cone and plunger, same
dim. as H-2520. Total wt. cone and
plunger 30.0g.
Food products and
paste paints
U.S. Dept. of
Agriculture H-2529 Aluminum. Same dimension as H-2520.
Total wt. 35g.
(shortenings,
margarine, butter,
etc.) Measuring
firmness of solid and
plasticized fats
AOCS Cc 16-60 H-1270
10° aluminum cone, 3.2mm ferrule,
0.8mm stainless steel blunt tip.
Overall length 106mm. Wt. 45g.
(i.e., recovery of
used grease) Small
obtainable samples ASTM D1403 H-2519
1/4 scale. (Not considered a substitute
for full-size cone specified in ASTM
D217.) Wt. 9.38g.

Penetration Needles and Cones
H-2519
H-2520
H-2522
H-2529
H-1320
H-1270
H-1255
H-1310
H-1317
H-1280
H-1290
H-1300
H-1302


H-1250

www.humboldtmfg.com
HUMBOLDT
Testing Equipment for Construction Materials
Humboldt Mfg. Co.
3801 North 25th Avenue
Schiller Park, Illinois 60176 U.S.A.
U.S.A. Toll Free: 1.800.544.7220
Voice: 1.708.456.6300
Fax: 1.708.456.0137
Email: [email protected]
Warranty
Humboldt Mfg. Co. warrants its products to be free from defects in material or
workmanship. The exclusive remedy for this warranty is Humboldt Mfg. Co.,
factory replacement of any part or parts of such product, for the warranty of this
product please refer to Humboldt Mfg. Co. catalog on Terms and Conditions
of Sale. The purchaser is responsible for the transportation charges. Humboldt
Mfg. Co. shall not be responsible under this warranty if the goods have been
improperly maintained, installed, operated or the goods have been altered or
modified so as to adversely affect the operation, use performance or durability
or so as to change their intended use. The Humboldt Mfg. Co. liability under
the warranty contained in this clause is limited to the repair or replacement of
defective goods and making good, defective workmanship.
This manual suits for next models
2
Table of contents
Other Humboldt Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Endress+Hauser
Endress+Hauser Liquiline System CA80AL Brief operating instructions

Leica
Leica Lino L2 user manual

ETI
ETI SIGNALS operating instructions

KNX
KNX TH65-AP Technical Data and Installation Notes

DOLD
DOLD VARIMETER RCM IL 5882 Installation / monitoring technique

Bosch
Bosch PTD 1 Original instructions