MODE DIRECTIONS
CALI To calibrate the meter, enter CALI mode (CALI icon will be ashing), then press
ENTER. Completely cover the photosensitive portion of the probe, and hold ENTER
for 3 seconds to reset the zero point.
LOG To display your data history, enter LOG mode (LOG icon will be ashing). Press
ENTER. Use and to cycle through data records. Press ENTER to exit this mode.
RCFS To reset the device to factory settings, enter RcFS mode (RcFS icon will be ashing).
Press ENTER (“NO” will be ashing on LCD screen). Use or to select “YES,”and
then hold ENTER for 3 seconds. An audible beep will indicate when the device has
been reset. If you hold ENTER for 3 seconds while“NO” is ashing, the device will
not reset to factory settings. Important: This option erases all stored log data.
SPECIFICATIONS
Typical test conditions, unless otherwise specied: Ambient Temp =23+/-3°C, RH=50%–70%,
Altitude=0~100 meter
MEASUREMENT SPEC
Operating Temperature 32°F to 122°F (0°C to 50°C)
Storage Temperature -4°F to 140°F (-20°C to 60°C)
Operating & Storage RH 0–95%, non-condensing
PPFD MEASUREMENT
Repeatability ±1 umol/m-2/sec-1
Measurement Range 0– 3,999 umol/m-2/sec-1
Display Resolution 0.1 umol/m-2/sec-1 (0–999); 1 umol/m-2/sec-1 (1,000–3,999)
Cut-On Wavelength 400±10nm
Cut-O Wavelength 700±10nm
Power Requirements 2 x AAA batteries
Dimension Main Unit: 115 x 60 x 24mm Sensor Unit: 80 x 45.6 x 26mm
Weight 100g (without batteries and cable)
OPERATING INSTRUCTIONS
SYMPTOM POSSIBLE CAUSE SOLUTION
No display on LCD screen
No power Press the Power button to turn on PAR meter.
Bad Batteries Replace battery.
Cable error Check the cable. Maybe there is an electrical short.
Replace the sensor cable if needed.
LCD main display shows
”----”
Sensor unit is not
working Check sensor cable to make sure it is securely
connected to the PAR meter main body and
sensor unit.
Sensor cable is not
connected
Some foreign materials
are in the RJ11 port
Please check the RJ11 port and clean any foreign
material that could cause interference.
Battery icon is ashing
on the LCD screen Battery is running low Replace the batteries.
Reset to zero point failed
The light sensitive
portion of the probe is
not completely covered
during calibration
Completely cover the photosensitive portion of
the probe before calibrating
TROUBLESHOOTING