4.) Next, remove the (2) screws from the under side of the base (located near the front right corner).
This will free the motor bracket from the base. Carefully slide the motor and bracket away from
gear assembly and remove from unit. See balloon “G”.
5.) Using an allen wrench loosen, then remove the coupler from the motor shaft. Keep this part; you
will need to reattach it to the new motor. Next, using an allen wrench, remove the (4) screws
holding the motor to the bracket. After removing the screws, remove the motor and discard.
Align the new motor in the same fashion as the old one, and replace the (4) screws using the
allen wrench to tighten.
6.) Place the coupler onto the shaft of the motor, aligning the flat of the shaft with the set screw.
The inside of the coupler should be flush with the end of the motor shaft. Tighten set screw.
7.) Rotate the gear assembly so the gear assembly coupler and motor coupler properly align, and
slide motor/bracket assembly into place. Once coupler is aligned, replace (2) screws from under
the base of the unit to secure the motor bracket assembly. Once aligned, tighten screws.
8.) Now place the front cover in front of the base and locate the wire cable from the motor assembly.
Connect the cable to the board mounted under the top cover. To assemble slide the cable into the
connector (through the hole in sheet metal), once the cable is inserted, press the tabs on either
side of the connector inward to secure the cable. See balloon “E”.
9.) Next, replace the red/black wire harness from the board mounted to the underside of the top
cover to the board mounted on the left side of the base. See balloon “D”.
10.) Carefully rotate top cover assembly onto base and press into position. Use care not to bend or
crease the motor cable, cable is very fragile! Finally replace (2) screws into side of unit and (2)
screws into front of unit and tighten.
11.) Reconnect unit’s power cord to input entry module and standard power source. Turn control dial
on front of unit and ensure unit will now function properly.