Switch on ventilator
Press the «ON/OFF» button on the left-hand side of your bellavista ventilator.
It takes about one minute for the device to start.
During the power-up process bellavista automatically performs some
system tests. If any irregularities should arise, bellavista generates an alarm
signal and the power-up process is discontinued.
Start button
Start screen
When you have started bellavista, the start screen appears:
4a
1c
1b
1a 2
34b
76
6
5
4c
1a Ventilation with the same settings as previously 4a Invasive, non-invasive ventilation or HFOT
1b Ventilation for a new patient 4b ATC
1c Predefined profile 4c Breathing circuit
2 Select patient type 5 Ventilation adjustment parameters
3 Gender, height and lung condition 6 Start ventilation
4 Select breathing circuit 7 Main menu
Please make sure your
bellavista ventilator is plugged into a
power source before you press the
«ON/OFF» button.
Quick Start Manual
General

imtmedical ag . Gewerbestrasse 8 .9470 Buchs .Switzerland . www.vyaire.com
the art of ventilation
Connection Assist
After system start, press the info button («i») on the left or right-hand side
of bellavista.
On the screen you will see the side of the device with the info button being
pressed. Here you will find explanations of the various connection ports.
Info button
Depending on the model of
bellavista ventilator your connection
ports may be different. It may be that
your information screen differs from this
illustration.

imtmedical ag . Gewerbestrasse 8 .9470 Buchs .Switzerland . www.vyaire.com
the art of ventilation
Flow sensor and breathing circuits
Connect breathing circuit D to a flow sensor
Breathing circuit D, single breathing circuit, Adult/Pediatric
Connect breathing circuit E to a flow sensor
Breathing circuit E, double breathing circuit, Adult/Pediatric/Neonatal
Circuit test
Press the «Circuit test» button on the start screen. Then follow the instruc-
tions on the screen.
Breathing circuits D and E
can be used both for invasive and for
non-invasive ventilation types
Every time you have changed
a flow sensor or a breathing circuit,
perform a flow sensor and leak test. To
ensure that the test is performed properly
it is essential that you select the breath-
ing circuit used on the start screen.

imtmedical ag . Gewerbestrasse 8 .9470 Buchs .Switzerland . www.vyaire.com
the art of ventilation
Circuit test
As soon as you hear the signal, insert the blind plug in the flow sensor until
the leak test has been successfully completed.
When you have performed the leak test, turn the adapter as illustrated and
wait for further instructions.
As soon as indicated, turn and connect the sensor and the adapter in the
opposite direction.
Follow the instructions and turn the flow sensor around again. When the test
has also been completed in that direction,you will hear a signal after approx.
90 seconds.
The circuit test illustrated
relates to breathing circuits D and E,
which are used most frequently.
You will find information about the other
breathing circuits available in the detailed
User Manual for your bellavista ventilator.

imtmedical ag . Gewerbestrasse 8 .9470 Buchs .Switzerland . www.vyaire.com
the art of ventilation
Start ventilation
4
2
3
5
1
1 Choose between a new patient with default settings, a last patient with his/her personal settings and a
predefined profile.
2 If you have selected the «New patient» category, decide on one of the following patient groups: adult,
pediatric or neonatal.
3 Enter the patient's weight
4Determine whether invasive or non-invasive ventilation is to be performed.
5 Select the type of breathing circuit used.
Select ventilation modes and ventilation
settings by touching the current ventila-
tion mode.
Touch the individual parameters you
would like to adjust for ventilation.
Adjust the values by swiping up or down.
Confirm the values by pressing «Con-
firm».
When you have adjusted the settings
accordingly to suit the patient's needs,
press «Start ventilation» and bellavista
starts ventilating.

imtmedical ag . Gewerbestrasse 8 .9470 Buchs .Switzerland . www.vyaire.com
the art of ventilation
Cockpit
When ventilation has started you will
see the Cockpit view. You can get this
view at any time by tapping the Cockpit
button. In NIV modes, alarm signals are
suppressed for two minutes. To activate
them, press the blue window.
Keep the curve configuration bar
pressed until the menu appears. To the
right of the curves you can select various
icons to zoom in, have all curves scaled
automatically or bring curves to a halt.
Keep a monitoring setting pressed un-
til the context menu appears. You can
delete or change monitoring settings.
If values are deleted, you can use the
space for new settings.
Press ventilation and the context menu
appears. This is where you control venti-
lation practices such as manual breaths,
O2 suction and, if fitted, the nebuliser.
1Manoeuvres page
Page for inspiratory/expiratory stop. This is where parameters such as NIF and AutoPEEP can be deter-
mined.
2Cockpit view
Main display during the ventilation process.
3Settings
Page for the settings of ventilation modes and checking the timing parameters.
4Apnoea ventilation settings
Switching backup on/off, setting apnoea time and adjusting the backup mode.
You can access the entire
manual by tapping the «?» icon in the
main menu or in the context menu of all
the monitoring settings.
14
3
2

imtmedical ag . Gewerbestrasse 8 .9470 Buchs .Switzerland . www.vyaire.com
the art of ventilation
Alarms
1
2
1Alarm level Yellow/Red
Stop the alarm by pressing the bell icon. Then the alarm is interrupted for two minutes. To adjust alarm
settings, press either the coloured parameter (yellow or red) or the alarm menu at bottom right.
2Alarm menu
Adjust the relevant parameter by clicking it. With the «Cockpit» button you return to the Cockpit view.
Stopping ventilation
1
2
1 To stop ventilation, press «Stop ventilation» in the Cockpit view.
Then a window appears with an arrow which you have to drag to the right. You return to the start
screen. If you wish to discontinue the process, press «Cancel».
2 To switch bellavista off, briefly press (for 1 second) the «ON/OFF» button on the left-hand side of the
device. Then a window appears with an arrow which you have to drag to the right. To shut down com-
pletely, bellavista takes about 20 seconds. To discontinue the process, press «Cancel».
304.794.000_02, 2020-03
Table of contents
Popular Medical Equipment manuals by other brands

Stryker
Stryker Quantum 2000 operating manual

BHM Medical
BHM Medical MINISTAND 865 Series Technical manual

Simulab
Simulab ArteriaLineTrainer Easy setup

SCHÖLLY FIBEROPTIC
SCHÖLLY FIBEROPTIC LED Light Source Instructions for use

Synovis
Synovis Peri-Strips Dry Series Instructions for use

Jas
Jas EZ Shoulder Body Cuff Fitting instructions