DRAIN VESSEL
Remove arm cap, press and hold the DRAIN button until the uid has completely
drained. Do NOT remove the tissue until the vessel has completely drained and
lines are clear.
REMOVE TISSUE
When the vessel has been drained remove the hand cover and tissue.
REMOVE ARTERY FROM STANDS & DETACH CONNECTORS
Pull the artery up and out of the artery stands. TWIST each end of the connectors
counterclockwise to release the clasp at the coupler. A small CLICK will be
heard when the clasp is released. Repeat on the other end of the artery.
INSTALL NEW ARTERY INTO COUPLERS
Grab a new replaceable artery. To install, TWIST the couplers together clockwise
until you hear a small CLICK. This click will indicate that the connection has been
made. Repeat on the other end of the artery.
SECURE ARTERY IN STANDS
Replace and SECURE the connected artery into the artery stands.
REPLACE HAND COVER & ARM CAP
Replace the hand cover and the arm cap into their secured position before
replacing tissue.
PLACE ARTERY IN REPLACEABLE TISSUE
Place the replaceable tissue into the tissue cradle. Start by laying in the wider side
of the tissue. Ensure that the artery is COMPLETELY TUCKED into the cut-out in
the tissue before installing the rest of the tissue into the cradle. If the
tissue does not sit perfectly, lift the side of the tissue and make sure that the artery is
aligned properly.
PULL ARTERY UP AND OUT OF STANDS
TWIST UNTIL IT CLICKS
ARTERY REPLACEMENT