manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Innovx
  6. •
  7. Measuring Instrument
  8. •
  9. Innovx SOIL User manual

Innovx SOIL User manual

INNOV‐XSOILQUICKSTARTGUIDE
INNOV‐XSystems,Inc.
10GillStreet,SuiteQ
Woburn,MA01801
THIS INSTRUMENT PRODUCES IONIZING RADIATION.
OnlyUserswithManufacturerAuthorizedTrainingshouldoperateINNOV‐XInstruments
Beforeuse,readandunderstandinstructionmanual
Reviewradiationsafety–Chapter3intheOperator’sManual
MaintainPossession&ControloftheAnalyzeratallTimes
NeverPointtheInstrumentatYourselforOthers
Start‐uppreparationfornewunits:
•ChargemainbatteriesforanalyzerusingInnov‐Xbatterycharger.
•ChargeiPAQ.TheiPAQconsumesbatterypowerevenwhenitisturnedoff.Iftheinstrumentisnotusedfora
periodofmanydays,theiPAQmaydischargecompletely,ormaynothaveenoughpowertorunatest.If
instrumentwillnotbeusedforseveraldays,chargetheiPAQduringdowntime.
DailyStartupprocedure:(MoredetailedinformationcanbefoundinSection5.1oftheUser’smanual)
1. Placeabatteryintheanalyzer.
2. PowerontheAnalyzerMainPower(On/Offswitchlocatedonbaseofhandle)
3. PowerontheiPAQ(ButtonlocatedinupperrighthandcornerofiPAQ)
4. SelectInnov‐XfromthestartmenulocatedintheupperlefthandcornerofiPAQscreen.
5. Readtheradiationsafetynoticeandacknowledgethatyouareacertifieduserbypressingstart
6. SelectSoilMode.
7. Analyzerwillundergoa60secondhardwareinitialization.
8. Placeastandardizationcliponthenoseoftheanalyzer.Tapthebuttononthescreentostandardize.
9. Whenstandardizationiscomplete,removethestandardizationclip.(Manualsection4.4Standardization)
10. Tapthe“lock”iconinthelowerrightcorneroftheiPAQscreentoreleasesoftwaretriggerlock
11. Testasoilstandardtoverifyinstrumentperformance.
12. Resultswilldisplayonscreen.SubsequenttestsmaybestartedfromeitherResultsorAnalysisscreens.
•Ifnecessary,taptheXintheupperrighthandcorneroftheResultsscreentoreturntothe“Readyto
test”Analysisscreenatanytime.
•TestinformationcanbeenteredfromtheanalysisscreenbytappingtheInfobutton.(Manual4.5.1)
•Testscanbestoppedbyeitherpullingthetrigger,ortappingStopontheiPAQscreen.
CompaqIPAQPocketPC®Basics:
•Startmenuisfoundintheupperleftcornerofscreen.If“Start”iconisnotvisible,tapthescreenin
upperleftcorner.
•TextisenteredbytappingthekeyboardiconinlowerrighttobringuptheVirtualKeyboardor
Characterrecognizer.
•TheFileToolbarislocatedatthebottomofthescreen.
•Pressingtherightmostbuttonbelowthescreenprovidesalistofapplicationscurrentlyrunning
•ToResettheiPAQ,stickthetipofthestylusintothesmallholelocatednexttotheconnector.
oInnov‐Xsoftwareusesthreemainscreens:Main,AnalysisandResults:Usersselectmodetouseviathe
MainScreen.MosteditstotestingmethodaremadeviatheAnalysisscreen,whichtypicallysays
“Readytotest.”TheResultsscreenshowsalldata.Alldataeditingfeaturesaredonefromhere.
oAlwaysexitsoftwarecompletelywhenfinishedusinganalyzer!Itmaybenecessarytocloseallthree
screens.UseFile→Exit.Softwarewillpromptforconfirmationbeforeclosing.
oRemovingiPAQ:SlideiPAQtowardsnoseofanalyzertoclearconnectorbeforeliftingupwards.
Innov-X Systems ●10 Gill Street. Suite Q ●Woburn, MA 01801
Tel 781-938-5005 ●Fax 781-938-0128
SamplePreparationConsiderations:
•Thedegreeofsamplepreparationwilllargelydeterminetheaccuracyobtainedwhenusingportable
XRF!
•EPAMethod6200forportableXRFiswrittenforin‐situtesting,asopposedtoprepared‐sampletesting.
•SamplePreparationwilldependonthegoaloftheanalysis.ForQuantitativeFieldScreening,itisoften
sufficienttoanalyzesamplesinplasticbags,ordirectlyontheground–in‐situtesting.In‐situtesting
providesahighnumberofdatapointsquicklyinordertoprofileasiterapidlyandfind“hotspots.”
•Formorequantitativeresults,orfordirectcomparisontoalaboratoryanalyticalmethodsuchasAAor
ICP,samplesshouldbedried,homogenizedandpreparedinsamplecups.Forthebestpossible
comparisons,itisbesttohavethelaboratoryanalyzethesamesamplesanalyzedwiththeXRF.
•SamplepreprecommendationscanbefoundinEPAMethod6200,Pg.8“SamplePreparation.”
•Forlaboratoryanalysis,besuretorequestatotaldigestionmethod.

Exportingdata:
Exporteddataissavedascommadelimited(.csv)files.
ThesefilescanbeopenedandformattedinMSExcel,or
anyotherspreadsheetprogram.Totransferdatatoa
PC,ActiveSyncmustbeinstalledonthePC.
FromResultsScreen
1. ChooseFile→ExportReadings
2. Selectreadingstobeexported.TapOK.
3. InputFilename,folder,tapOK
4. TheresultsfilewillbestoredintheMy
DocumentsFolderontheiPAQ.Itcanbe
copiedtoaPCusingActiveSync.IfActiveSync
issetuptoautomaticallysynchronizefiles,this
filewillbecopiedwhentheiPAQisconnected
tothePC.
Erasingdata:
Thesoftwareloadsstoreddatawhenfirstresultis
takenineachmodeaftersoftwareisstarted.Asmore
readingsarestored,theanalyzerwilltakelongerto
displaytheresultsofthefirstreading.Ifthisperiod
oftimebecomestoolong,erasethedatastoredin
memory.
FromResultsScreen:
1. SelectFile→EraseReadings
2. Enteradministrativepassword
3. SelectreadingstobedeletedandtapOK.
4. Select“Yes”toconfirmdeletion.

Manual:Section4.6.3Erasingreadings

ChangingTestSettingsfromAnalysisscreen:
Testtime:SelectOptions→SetupTesting
•MinTimemustelapseinareadingforresultsto
bereportedonscreen.Testsstoppedbefore
MinTimewillbeconsideredtobeabortedtests.
•MaxTesttimeismaximumlengthoftestifa
manualtestendconditionisselected.

Turnliveupdateson/off:Select View →LiveUpdates

Soilmodeoptionsandtestendconditions
Aredescribedinsection5.2oftheusersmanual
Testendconditions:
SelectOptions→SoilModeTestingSetup
•Manual:Testendswhentestisstopped
•MaximumTime:TestendswhenMaxtime
haselapsed
•Rel.Std.Dev.:Mustchooseelementand
standarddeviation.Testendswhenerroris
withinspecifiedlimits
•ActionLevel:Mustchooseelementand
desiredactionlevel.Testendswhenanalyzer
candetermineifelementsconcentrationis
aboveorbelowactionlevel.

Popular Measuring Instrument manuals by other brands

Johnson 1886-0000 instruction manual

Johnson

Johnson 1886-0000 instruction manual

mru VARIOluxx user manual

mru

mru VARIOluxx user manual

LaserLiner MultiWet-Finder Plus manual

LaserLiner

LaserLiner MultiWet-Finder Plus manual

Troxler 3440-L Manual of Operation and Instruction

Troxler

Troxler 3440-L Manual of Operation and Instruction

Samoa MP30 Parts and technical service guide

Samoa

Samoa MP30 Parts and technical service guide

JVC HR-S3500U instructions

JVC

JVC HR-S3500U instructions

Laserworld Ecoline Series manual

Laserworld

Laserworld Ecoline Series manual

PDK PDR100 user manual

PDK

PDK PDR100 user manual

Endress+Hauser Micropilot FMR62 technical information

Endress+Hauser

Endress+Hauser Micropilot FMR62 technical information

ABB LWT300 series quick start guide

ABB

ABB LWT300 series quick start guide

YSI 550 DO Operation manual

YSI

YSI 550 DO Operation manual

TFT 723 instruction manual

TFT

TFT 723 instruction manual

Hazet Brake Fluid Tester Application note

Hazet

Hazet Brake Fluid Tester Application note

Tektronix RSA6100A Series Service manual

Tektronix

Tektronix RSA6100A Series Service manual

ACCRETECH UF200R user manual

ACCRETECH

ACCRETECH UF200R user manual

Endress+Hauser Cerabar S PMC71 operating instructions

Endress+Hauser

Endress+Hauser Cerabar S PMC71 operating instructions

Bender PEM735 manual

Bender

Bender PEM735 manual

PCB Piezotronics IMI Sensors M648A00 Installation and operating manual

PCB Piezotronics

PCB Piezotronics IMI Sensors M648A00 Installation and operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.