inofab SPIROHOME User manual

Before using your SpiroHome®Personal device and mobile application, please ensure that you
have read this user manual and all product labelling. This user manual is available digitally or
for print through the Inofab Health Apps and Inofab Health websites.
User Manual
Personal
®

First Pub. Date: 05.03.2018 R.11-1 / 14.10.2020

CONTENTS
1. INTRODUCTION ...............................................................................................................................................................................................................4
1.1.PRODUCT DESCRIPTION................................................................................................................................................................................................................4
1.2.WHAT’S IN THE BOX.......................................................................................................................................................................................................................4
1.3.INTENDED USE................................................................................................................................................................................................................................4
1.4.RESTRICTIONS ON USE AND CONTRAINDICATIONS..............................................................................................................................................................4
1.5.PARAMETERS....................................................................................................................................................................................................................................5
2. OPERATION...........................................................................................................................................................................................................................6
2.1.STORAGE AND OPERATING ENVIRONMENT............................................................................................................................................................................6
2.2.SETTING UP YOUR DEVICE............................................................................................................................................................................................................6
2.3.DEVICE INDICATORS......................................................................................................................................................................................................................7
2.4.SPIROMETRY MEASUREMENT.......................................................................................................................................................................................................8
2.5.UNDERSTANDING THE TEST QUALITY.......................................................................................................................................................................................10
2.6.SIGNS AND SYMBOLS.................................................................................................................................................................................................................13
2.7.TECHNICAL FEATURES..................................................................................................................................................................................................................13
2.8.SAFETY WARNINGS AND PRECAUTIONS................................................................................................................................................................................14
3.MAINTENANCE..................................................................................................................................................................................................................15
3.1.CLEANING AND DISINFECTION................................................................................................................................................................................................15
3.2.CLEANING THE SPIROWAY®REUSABLE....................................................................................................................................................................................16
3.3.BATTERIES.......................................................................................................................................................................................................................................16
3.4.DISPOSAL OF SPIROHOME®.......................................................................................................................................................................................................17
4.TROUBLESHOOTING..................................................................................................................................................................................................17
5.ORDERABLE ACCESSORIES...................................................................................................................................................................................18
6.TERMS OF WARRANTY.............................................................................................................................................................................................18
7.ELECTROMAGNETIC COMPATIBILITY.........................................................................................................................................................19
8.MANUFACTURER INFORMATION...............................................................................................................................................................23

4
1.INTRODUCTION
1.1. PRODUCT DESCRIPTION
The SpiroHome®Personal is a portable
spirometer that pairs (via Bluetooth®) with smart
devices running iOS or Android SpiroHome
Personal®applications. The SpiroHome®
Personal measures and displays information
about the user’s lung function. The user performs
a spirometry test with the device as described
in the “Spirometry Measurement” section
of this user manual. The ultrasonic sensors
within the SpiroHome®Personal detect the
volume and speed of air moving through the
device and display this as spirometric data
on the SpiroHome®Personal app. The app
guides the user throughout the test session. The
SpiroHome®Personal app can be downloaded
from GooglePlay or App Store. The device
is powered by standard 2 x AAA Alkaline
batteries. SpiroHome®Personal is used with the
SpiroWay®Reusable mouthpiece.
1.2. WHAT’S IN THE BOX
The SpiroHome®Personal box contains:
• SpiroHome®Personal device (b)
• SpiroWay®Reusable mouthpiece (a)
• SpiroHome®Personal Cap©
• User Manual
• Carrying pouch (d)
CAUTION: Please check that there is no
visible damage on any device components. If any
damage is present, do not use or attempt to repair
the device but contact the manufacturer directly.
1.3. INTENDED USE
The SpiroHome®Personal is intended to be used
as a portable spirometer in the lung function
testing of:
• children (over the age of 5) and adults who
may have a chronic pulmonary disease
including, but not limited to, asthma, chronic
obstructive pulmonary disease and cystic
fibrosis.
NOTE:A competent adult should assist
patients (children or older patients) who may
need assistance.
1.4. RESTRICTIONS ON USE AND
CONTRAINDICATIONS
Diagnosis of medical conditions or prescription
of treatments can only be made by a qualified
healthcare professional who may use results
obtained with the SpiroHome®Personal as adjunct
information when performing a full medical
examination that has taken into consideration your
clinical history and other test results.
SpiroHome®Personal is a single-user device and
is to be used by a single user. If the device will be
used by a new user, ensure that the data of the
previous user is erased from the device memory,
their account is removed from the app and a new
user account is created for the new user. The device
must also be cleaned and disinfected according to
the information given in this user manual before use
by a new user.
A SpiroWay®Reusable mouthpiece must not be
shared between users, including family members.
A new mouthpiece must be used for a new user.
Spirometry tests should only be performed if you
are not experiencing any shortness of breath,
are in good health and capable of performing a
lung function test. Test results may otherwise be
unreliable.
Failure to perform the required breathing maneuver
correctly during a test may lead to inaccurate and
unacceptable results. More information about how
to perform a spirometry test correctly is described
in this user manual. The device should not be used
if test accuracy and/or reliability is jeopardized by
these or other external factors.
Spirometry tests can be physically demanding.
The forced expiratory maneuver used in spirometry
increases intrathoracic, intraabdominal, and
intracranial pressures. Potential risks of spirometry
are primarily related to maximal pressures
generated in the thorax and their impact on
abdominal and thoracic organs, venous return
and systemic blood pressure, and expansion of the
chest wall and lung. The physical effort required
can increase myocardial demand. Caution must
be used if you have medical conditions that could
be adversely affected by these physiological
consequences. Although such risks are likely to be
minimal for spirometry in most patients, the potential

5
Table 1: Parameters & Definitions
Parameters Definition
FVC (L)
Forced Vital Capacity — The volume of
air that can forcibly be blown out after
full inspiration
FEV0.75(L)
Forced Expiratory Volume after 0.75
seconds: The volume of air that can
forcibly be blown out in 0.75 seconds,
after full inspiration.
FEV1 (L) Forced Expiratory Volume within
1 second
FEV6 (L) Forced Expiratory Volume within
6 seconds
FEV0.75/FVC The ratio of FEV0.75 to FVC
FEV1/FVC The ratio of FEV1to FVC
PEF( L/s)
Peak Expiratory Flow — The maximal
flow rate achieved during the maxi-
mally forced expiration initiated at full
inspiration.
FEF25-75( L/s)
Forced Expiratory Flow from 25% to
75% of vital capacity — synonymous
with MMEF
FEV1/FEV6The ratio of FEV1to FEV6
FET(s) Forced Expiratory Time
BEV(L) Back extrapolated volume
The recommended number of trials per spirometry
session is 3, however, you may perform up to 8
trials. The best values obtained from the spirometry
trials performed in one session are displayed on
the app. You also have the option to view each
individual trial result of a spirometry session.
The device provides a reference value. This is
calculated from large epidemiological studies
and requires your height, weight, age, sex and
ethnicity information. Your results are compared to
the reference values as a percent predictive value
indicator of your respiratory health. Your personal
best value for a spirometry trial can be discussed
and with your healthcare provider for medical
interpretation.
CAUTION: Interpretation of spirometry results
or diagnosis of medical conditions, if any, is to
be made by a physician or allied health care
professional with sufficient training in spirometry.
risks associated with testing should always
be weighed against the benefit of obtaining
information about lung function. Spirometry
should be discontinued if you experience pain
during the maneuver. If you have any of these
potential contraindications please seek spirometry
testing in primary care settings or pulmonary
function laboratories where you will be under the
supervision of healthcare professionals and there
may be access to emergency care if needed.
Relative Contraindications for Spirometry;
Due to increases in myocardial demand or
changes in blood pressure;
•Acute myocardial infarction within 1 week
•Systemic hypotension or severe hypertension
•Significant atrial/ventricular arrhythmia
•Uncompensated heart failure
•Uncontrolled pulmonary hypertension
•Acute cor pulmonale
•Clinically unstable pulmonary embolism
•History of syncope related to forced
expiration/cough
Due to increases in intracranial/intraocular
pressure;
•Cerebral aneurysm
•Brain surgery within 4 weeks
•Recent concussion with continuing symptoms
•Eye surgery within 1 week
Due to increases in sinus and middle ear pressures;
•Sinus surgery or middle ear surgery or
infection within 1 week
Due to increases in intrathoracic and intra-
abdominal pressure;
•Presence of pneumothorax
•Thoracic surgery within 4 weeks
•Abdominal surgery within 4 weeks
•Late-term pregnancy
Infection control issues;
•Active or suspected transmissible respiratory
or systemic infection, including tuberculosis
•Physical conditions predisposing to the
transmission of infections, such as hemoptysis,
significant secretions, or oral lesions or oral
bleeding
If you have or suspect having any of the conditions
above, consult your healthcare professional
before using the SpiroHome®Personal.
1.5. PARAMETERS
The SpiroHome®Personal records and displays
the following spirometric data:

6
Enable Bluetooth®on your smart device
and pair the SpiroHome®Personal
with your smart device by following the
instructions on the app.
Login
Create Account
2.2. SETTING UP YOUR DEVICE
Download the SpiroHome®Personal
app from the App Store or Google Play
Store into your smart device.
Follow the steps given in the app to
create an account as a new user or
login to your existing account.
Slide open the battery cover, insert the AAA alkaline
batteries in the correct orientation, slide the battery
cover back to the closed position and press the power
button for 1 second to switch on the device as shown.
1
2
3
4
2. OPERATION
2.1. STORAGE AND OPERATING
ENVIRONMENT
The SpiroHome®Personal is intended to
be used in home settings. It is not intended
for use in clinical settings such as hospitals
or private clinics.
The operating conditions for the
SpiroHome®Personal are:
•Temperature: +15°C to +35°C
•Relative Humidity: 10% to 85%
The storage conditions for the
SpiroHome®Personal are:
•Temperature: -20°C to +60°C
•Relative Humidity: 0% to 85%
•Pressure: 500 hPa to 1060 hPa
The SpiroHome®Personal should not
be used in the presence of inflammable
liquids or detergents, nor in the presence
of inflammable anesthetic gases (oxygen
or nitrogen).
The device should not be used in direct
air currents (e.g. wind), sources of heat or
cold, direct sun rays, or other sources of
light or energy, dust, sand, or any other
chemical substances.

7
2.3. DEVICE INDICATORS
There are 3 LED lights located on the front of
the device. The LED lights may be turned on
or flashing various colors in various patterns.
The LED lights indicate the current status of the
device. Please see the following information
for guidance on LED light indications.
Table 2: Device Led Indicators
LED Display Indication/s
None of the LEDs are on The device is switched off
LED indicators are consecutively flashing green The device is switching on
LED number 3 is constantly flashing green The device is switched on
LED number 2 is fading blue The device is connected to the app. Bluetooth
connection has been established.
LED number 2 and LEDs 1 and 3 together are
flashing yellow in turn. The zero flow level adjustment is setting up.
LED number 1 is constantly blue. The device is ready for a test.
During a test, LED number 1 is constantly yellow. The test has timed-out (there has been no inhala-
tion/exhalation over a while)
During zero flow level adjustment setup LED
number 1 is constantly yellow.
The zero flow level adjustment setup has been
unsuccessful.
All LEDs are flashing red. There is a foreign object between the sensors.
(Check device error in troubleshooting section)
LEDs are consecutively flashing yellow. Over-the-air connection is being established.
LED number 3 flashes red three times. Battery low warning.
LEDs flash in reverse order and remain switched off. The device is switching off.

8
2.4. SPIROMETRY MEASUREMENT
Follow the instructions that
appear on the screen. The first
step will be to record a zero
flow level adjustment for the
device. You will need to leave
the device on a flat surface for
this process to be completed.
Sit upright with your back straight
and your feet flat on the ground.
Remove the SpiroWay®Reusable
mouthpiece from its plastic
packaging and insert it all
the way into the SpiroHome®
Personal in the correct orientation
(as shown). You will hear a click
when the mouthpiece is fully
inserted into the device.
Open the SpiroHome®App
on your smart device and
make sure you are signed in.
Tap the ‘+’ button to start the
test procedure.
Setting Up
Your Device
1 2 3 4
Place the mouthpiece in your
mouth, past your teeth, and
form a tight seal around the
mouthpiece with your lips.

9
5 6 7 8
You will now need to perform
a forced expiratory maneuver.
To ready yourself, inhale and
exhale normally a couple of
times, then take a fast and
deep breath, filling your lungs
completely. Do not hold breath
for longer than 2 seconds.
Without hesitation and with your lips
sealed tightly around the mouthpiece
blow out the air in your lungs as hard
and fast as you can. Keep blowing
until you feel like you have completely
emptied your lungs. You may use
a nose clip at the beginning of the
procedure to ensure that you are
exhaling only through your mouth.
Repeat these steps for
each successive trial,
ensuring that you rest
for at least 20 seconds
between each trial.
After completing your
spirometry session, switch
off the device by pressing
the power key. Use the cap
to protect the mouthpiece
from contamination when
the device is not in use.

10
IMPORTANT:This procedure describes how
to perform the forced expiratory maneuver when
‘Tidal Mode’ is selected on the SpiroHome®
application.
‘Tidal Mode’ requires you to breathe normally
several times into the mouthpiece at the beginning
of the test before a forceful expiration is
performed.
If you deselect ‘Tidal Mode’ from the app settings,
you do not need to breathe normally several times
into the mouthpiece at the beginning of the before
a forceful expiration, and the test begins with a
direct forceful expiration.
2.4.1. End of Forced Expiration (EOFE)
The end of a forced expiratory maneuver is
referred to as ‘End of Forced Expiration’ or ‘EOFE’.
The EOFE is important in recognizing when a true
FVC measurement has been achieved. Any of the
following three cases will indicate an EOFE:
1.Expiratory plateau (≤0.025 L in the last 1 s of
expiration)
2.Expiratory time ≥15 s
3.FVC is within the repeatability tolerance of or is
greater than the largest prior observed FVC *
*More information in section 2.5
Table 3: Summary of Acceptability, Usability, and Repeatability Criteria for FEV1and FVC
Acceptability and Usability Criterion
Required for
Acceptability
Required for
Usability
FEV1FVC FEV1FVC
Must have BEV ≤5% of FVC or 0.100 L, whichever is greater Yes Yes Ye s Yes
Must have no evidence of a faulty zero-flow setting Yes Yes Ye s Yes
Must have no cough in the first second of expiration Yes No Ye s No
Must have no glottic closure in the first second of expiration* Yes Yes Ye s Yes
Must have no glottic closure after 1 s of expiration No Yes No No
2.5. UNDERSTANDING THE TEST QUALITY
After each test session, the quality of your test will be graded based on how well you performed the breathing maneuver and whether your results are
acceptable, usable or neither. This grading relates to the consistency of your blows, not the health of your lungs.
The American Thoracic Society (ATS) and European Respiratory Society (ERS) grade FVC and FEV1parameters in children and adults as given in the table 3;

11
EOFE = end of forced expiration
Must achieve one of these three EOFE indicators:
1. Expiratory plateau (≤0.025 L in the last 1 s of expiration)
2. Expiratory time ≥15 s
3. FVC is within the repeatability tolerance of or is greater than the largest prior observed FVC *
No Yes No No
Must have no evidence of obstructed mouthpiece or spiromete Yes Yes No No
Must have no evidence of a leak Yes Yes No No
If the maximal inspiration after EOFE is greater than FVC, then
(FIVC — FVC) must be ≤0.100 L or 5% of FVC, whichever is greater ** Ye s Yes No No
Repeatability criteria (applied to acceptable FVC and FEV1values)
Age > 6 yr: The difference between the two largest FVC values must be ≤0.150 L, and the difference between the two largest FEV1values must be ≤0.150 L
Age ≤ 6 yr: The difference between the two largest FVC values must be ≤0.100 L or 10% of the highest value, whichever is greater, and the difference between
the two largest FEV1values must be ≤0.100 L or 10% of the highest value, whichever is greater
* Occurs when the patient cannot expire long enough to achieve a plateau (e.g., children with high elastic recoil or patients with restrictive lung disease) or when
the patient inspires or comes off the mouthpiece before a plateau. For within-maneuver acceptability, the FVC must be greater than or within the repeatability
tolerance of the largest FVC observed before this maneuver within the current prebronchodilator or the current post-bronchodilator testing set.
** This criterion is not used, as there is no parameter calculated using the inspiration maneuver in SpiroHome®Personal.
Although the performance of a maximal forced inspiration is strongly recommended, its absence does not preclude a maneuver from being judged acceptable,
unless the extrathoracic obstruction is specifically being investigated.

12
Table 4: Grading System for FEV1and FVC (Graded Separately)
Grade Number of Measurements Repeatability:
Age >6 yr
Repeatability:
Age ≤6 yr *
A ≥ 3 acceptable Within 0.150 L Within 0.100 L *
B 2 acceptable Within 0.150 L Within 0.100 L *
C ≥ 2 acceptable Within 0.200 L Within 0.150 L *
D ≥ 2 acceptable Within 0.250 L Within 0.200 L *
E
≥ 2 acceptable > 0.250 L > 0.200 L *
OR 1 acceptable NA NA
U 0 acceptable AND ≥ 1 usable NA NA
F 0 acceptable and 0 usable NA NA
* Or 10% of the highest value, whichever is greater; applies for age 6 years or younger only.
NA: Not Applicable

13
2.6. SIGNS AND SYMBOLS
The following signs and symbols are provided for the safe use and storage of your
SpiroHome®Personal. 2.7. TECHNICAL FEATURES
Table 5: Markings & Descriptions
Markings Descriptions Markings Descriptions
“Manufacturer”
This symbol accompanied by the name
and the address of the manufacturer
adjacent to the symbol
Single-use only
Sign of Conformity Type BF of Medical Electrical
Equipment
Disposal in Compliance with WEEE Serial Number
Temperature Limit Lot Number
Humidity Limit Ref Number
Atmospheric pressure limitation IP IP Number
Do not use if the package is damaged The device includes RF
transmitters
Keep away from sunlight The instruction manual/
booklet must be read.
Keep dry Caution
Explanatory information
Table 6: Technical Features
Flow / Volume measurement
method
Ultrasonic transducer
measurement
Power Supply 2 x 1.5V AAA
Alkaline batteries
Dimensions 110 x 63 x 41 mm
Weight (With batteries) 90 g
Weight (Without batteries) 67 g
Flow range 0 - 14 L/s
Maximum volume measured 10 L
Volume accuracy (Average) 2.00 %
Dynamic resistance at 14 L/s 86 Pa*s/L
Volume resolution 1 mL
Flow resolution 1 mL/s
Medical device class Class IIA
Wireless connection BLE 4.2

14
2.8. SAFETY WARNINGS AND
PRECAUTIONS
IMPORTANT:Please adhere to all safety
warnings, precautions and recommendations
given in this user manual as failure to comply may
result in measurement errors, display of incorrect
results or possible harm to the device or the
user. The manufacturer is not responsible for any
damage or harm to the device or user that is the
direct result of non-compliance to these warnings
and precautions other instructional materials
provided with the device.
•If any damage is present on the device or its
components upon initial unboxing of the product
then do not use the device and return it to the
supplier.
•Do not use this SpiroHome®Personal for any
other purpose than its intended use.
•Competent adults assisting pediatric, elderly or
differently-abled users should inform them about
the safety warning and precautions given in this
user manual before the use of the device.
•Regardless of the data presented on the
SpiroHome® Personal, if you feel unwell or have
respiratory distress symptoms cease device use
and contact your healthcare provider immediately.
•If there is an excessive decrease in your FEV1
value then cease use of the device and inform your
healthcare provider.
•Do not perform more than 8 spirometry trials
in one spirometry session. If you experience pain
during the maneuver, cease device use and rest.
•Maximal inflation is unnatural and if you have
not achieved it before it may seem somewhat
uncomfortable. If you feel sensations of dizziness
or giddiness during pulmonary function sessions,
then cease device use and inform your healthcare
provider.
•Do not walk or run whilst performing a lung
function test with the SpiroHome®Personal.
•Do not perform a spirometry test with food or
objects in your oral cavity as this may lead to risk
of choking.
•Do not share your SpiroHome®Personal with
any other users, including family members. The
SpiroHome®Personal and SpiroWay®Reusable is
to be used by a SINGLE user only.
•If the device is to be used by a new user:
Clean and disinfect device and cap according to
instructions in the Maintenance section of this user
manual AND use a new mouthpiece for the new
user AND create a new account for the new user
on the SpiroHome®Personal app.
•To prevent damage to the device due to battery
leakage or oxidation, remove all batteries if the
SpiroHome®device is not to be used or is to be
stored for a long time.
•Dispose of the device and/or the device
batteries responsibly as required by local
legislation.
•Check the device and components for foreign
bodies or surface impurities before each use as this
could lead to inaccuracies in test measurements.
Coughing or spitting into the device may cause
incorrect readings.
•Do not expose the device to liquids, prevent any
liquids from entering the device. In the event of a
liquid spill on or around the device, immediately
remove batteries and let the device dry thoroughly
before use.
•If you experience any adverse events using the
device, report immediately to your healthcare
provider and local authorities as required by
local legislation. Please seek to also report such
incidents to the manufacturer.
•Do not use the SpiroHome®Personal with a
charging smart device. Make sure the smart device
is unplugged from its charger before conducting a
spirometry test.
•Use only original accessories specified and
provided by the manufacturer. Accessories that
are not original will cause inaccurate readings, or
harm to the user and/or device.
•Do not hold the mouthpiece from the filters
located on the mouthpiece. Do not use the
mouthpiece if these filters have been physically
compromised.
•Store and operate the device only as
specified in this user manual (see Section 3.1)
to avoid device malfunction and/or incorrect
measurements.
•Do not use the device in the presence of strong
electromagnetic sources, such as electrosurgical
equipment, or the presence of computed
tomography (CT) equipment.
•Do not attempt to repair, modify or reconfigure
the device. Contact the manufacturer/distributor/
retailer directly if your SpiroHome®Personal is
damaged or malfunctioning or you encounter
data that you cannot make sense of. Unauthorised
repairs, modifications or reconfigurations of the
device will void the warranty of the product.

15
•Follow all data security warnings and
recommendations for your personal smart
device as per it’s manufacturer’s instructions to
protect your personal data.
•Do not share your SpiroHome®account
information with unauthorized parties.
•The SpiroHome®Personal conforms to EN
60601-1, EN 60601-1-11, EN 60601-1-2 and
EN 300 328. As this device operates with RF
technology, it must be used as only specified by
the manufacturer to avoid interference by radio
communications.
3. MAINTENANCE
Handle your SpiroHome®Personal and
SpiroWay®Reusable mouthpiece with care.
Store the SpiroHome®Personal and
SpiroWay®Reusable in dust/dirt- and
moisture-free conditions. You may utilize the
pouch provided with the product to device and
components.
Before each use, always check that the device
and components are free from contaminants
and do not have any visible damage.
NOTE: One ‘use’ of the spirometer is
defined as one complete spirometric testing
session (can include up to 8 individual
successive spirometry trials).
The SpiroHome®Personal does not require
routine calibration as it uses ultrasonic flow
measurement technology. If you suspect a
problem with the calibration of the device, cease
use and contact the manufacturer immediately.
3.1. CLEANING AND DISINFECTION
You should clean the SpiroHome®Personal body and cap at least once a week or whenever the device
is visibly contaminated. You must perform the cleaning step before performing the disinfection step.
Regular cleaning will prevent the physical buildup of contaminants on device surfaces. Disinfection kills
and destroys pathogens such as bacteria, viruses, or other microorganisms which might still be present on
device surfaces after initial cleaning.
Wash Hands
Before beginning the
procedure, wash hands
thoroughly with soap
and water.
Wash Hands
Wash hands thoroughly after
performing a cleaning or
disinfection procedure, and
before handling the cleaned or
disinfected components again
for packing and storage.
Perform disinfection
To disinfect the device and cap use a new high-level
disinfectant (Sodium Hypochlorite <0.6%, Sodium
hydroxide <0.02%, Sodium Chloride <4%)
wipe to wipe over all accessible surfaces again
using moderate pressure and for the contact time
recommended by the wipe manufacturer.
Medipal®Chlorine Disinfectant wipes (Pal
International Ltd.) are high-level disinfectant
(Sodium Hypochlorite <0.6%, Sodium hydroxide
<0.02%, Sodium Chloride <4%) wipes suitable for
cleaning and disinfecting SpiroHome®devices and
is available at;
https://www.palinternational.com/en/healthcare/
wipe-products/chlorine-disinfectant-wipes/
Perform cleaning
First, remove the SpiroWay®Reusable from the
SpiroHome®Personal device. Using a high-level
disinfectant (Sodium Hypochlorite <0.6%, Sodium
hydroxide <0.02%, Sodium Chloride <4%) wipe,
wipe all accessible surfaces of the device and cap
using moderate pressure for at least 30 seconds to
remove contaminants. Be gentle and use care when
wiping the sensors to avoid damaging them.
Wipe all accessible surfaces of the device and cap,
using moderate pressure, as shown.
CAUTION: Prevent any excess liquids contained
within the wipes from entering the components of the
SpiroHome®Personal. Never immerse the product
in water or any other liquid solution.
1
2
3
4

16
3.2. CLEANING THE SPIROWAY®
REUSABLE
To clean the SpiroWay® Reusable mouthpiece
once a week and whenever visibly soiled;
•Add dishwashing detergent (e.g. those
containing 5-15% anionic surfactant, 5%
nonionic surfactant) to warm water to create
a soapy solution.
•Shake the mouthpiece gently in the soapy
solution.
•Hold the mouthpiece under running tap
water to rinse, do not rub
•Leave the mouthpiece upright on a clean
lint-free cloth at room temperature until it is
completely dry.
CAUTION: Do not insert the SpiroWay®
Reusable mouthpiece into your SpiroHome®
Personal device until it is completely dry
The SpiroWay® Reusable should be replaced
every 3 months. The SpiroWay® Reusable
mouthpiece must be replaced if you used or
suspect having used the mouthpiece whilst
having a bacterial or viral infection. Replace
the SpiroWay® Reusable immediately if the
filters are damaged or whenever a risk of
contamination is suspected.
CAUTION:
Risk of Cross-Contamination
The SpiroWay® Reusable mouthpiece is
indicated for single-patient-use only to prevent
any potential of cross-contamination. Thorough
cleaning and disinfection of the device must be
performed prior to use by a new user. A new
mouthpiece must be used by the new user.
To purchase new mouthpieces, contact
authorized local distributors or, if there is no
local distributor, contact Inofab Health at
www.inofab.health
3.3.BATTERIES
The SpiroHome® device operates with 1.5V
AAA Alkaline batteries. The battery life of the
SpiroHome® is approximately 12-18 months,
assuming daily use of the device. The battery
charge level is continuously monitored by the
device. The device will not turn on if the battery
charge level is low and will make a beeping
sound to notify you.
WARNING: The batteries of the device
should be removed if the device is not going to
be used for more than a month.
Instructions for battery
replacement
Remove cap and
SpiroWay®Reusable
mouthpiece from
the device.
Slide battery cover
to open position.
Remove the empty batteries.
Insert new batteries in the
correct orientation.
Slide the battery
cover back to the
closed position.
Insert the SpiroWay®
Reusable mouthpiece in the
correct orientation. Your
device is now ready to use.
Place cap on device
to protect mouthpiece
from contamination
during storage.
1
2
3
4
5
7
6

17
4. TROUBLESHOOTING
Table 7: Troubleshooting
Problem Cause Solution
Device not turning on Multiple possible causes
Check battery orientation and correct polarities
Remove the AAA batteries, wait 30 seconds and reinstall AAA batteries
Replace AAA batteries
Check that battery cap is in the lock position, or if the cap is broken, contact manufacturer
SpiroHome®cannot connect to a
smart device via Bluetooth®
The smart device is out of range Bring your smart device closer to the SpiroHome®device
Smart device Bluetooth®is disabled Enable Bluetooth®on your smart device
Bluetooth®connection not working
properly
Your smart device will need Bluetooth®version 4.0 or higher. Find and select SpiroHome®
Personal from the list of detected devices
Test results are inconsistent
SpiroWay®mouthpiece is dirty Clean SpiroWay®to ensure that the lumen is not obstructed or replace with a new
mouthpiece
SpiroWay®mouthpiece is damaged Replace SpiroWay®
SpiroWay®mouthpiece is installed
incorrectly Refer to the user manual for proper installation of SpiroWay®
Spirometry test was performed incorrectly Refer to “Spirometry Measurement” in the user manual or refer video tutorial on app
The test does not start - Cannot set
up zero flow level adjustment Direct air current in the environment
Close the cap of the SpiroHome®to avoid effects of environmental flow
Place the device on a flat surface
Remove causes of direct air current e.g. air conditioner, opened window, fan, etc.
The test does not start - balloon
animation is not moving Multiple possible causes
Quit test and start a new test
Quit the application and start a new test
Switch the device off and turn on again to reset
Test starts before you start blowing Rough handling of the device Keep the device as stable as possible after starting a test
3.4. DISPOSAL OF SPIROHOME®
This product is not to be discarded as regular household waste and should be discarded as electronic waste in accordance with local regulations and
returned to a collection point of recycling for electric and electronic devices.
Used batteries should be disposed of in designated battery recycling containers in accordance with local laws and regulations.

18
Table 7: Troubleshooting
Device disconnected during test
The device is turned off accidentally or due
to rough handling during use Switch the device on again and proceed with a new test
Bluetooth®connection disrupted Reconnect the device and proceed with a new test
Test quality grade always low Not performing the test correctly Repeat the test following the rules and conditions specified in the “Spirometry Measurement”
section of this user manual
Measurement error screen
showed up
Flow limit exceeded This device is intended to measure 0-14 L/s.
SpiroWay®mouthpiece is dirty Clean SpiroWay®to ensure that the lumen is not obstructed or replace SpiroWay®
SpiroWay®mouthpiece is damaged Replace SpiroWay®
Device malfunction Contact manufacturer
Device error screen showed up
SpiroWay®mouthpiece is installed
incorrectly
Refer to the user manual for proper installation of SpiroWay®
There is a foreign object between the
sensors.
Check SpiroHome®device lumen and clean if necessary
SpiroWay®mouthpiece is dirty Clean SpiroWay®to ensure that the lumen is not obstructed or replace with a new
mouthpiece
SpiroWay®mouthpiece is damaged Replace SpiroWay®
For any other technical queries please call customer service on +90 312 988 03 08 or e-mail at [email protected]
5.ORDERABLE ACCESSORIES
•SpiroWay®Reusable Mouthpiece
(Reference number: 03000)
•SpiroHome®Personal Cap
(Reference number: 01104)
•SpiroHome®Pouch
(Reference number: 01509)
To purchase these accessories please contact
your local distributors or İnofab Health at
www.inofab.health
6. TERMS OF WARRANTY
SpiroHome®Personal, together with any
accessories provided, is guaranteed for 24
months, effective from the date of purchase, upon
the provision of an invoice or sales receipt. The
service life of the product is 5 years, effective from
the date of purchase.
The user is responsible for checking the product
for damage or missing components at the time
of purchase or delivery, and any claims must be
made in writing to the manufacturer.
The customer must return goods for replacement or
repair at the customer’s expense to the authorized
supplier or manufacturer. The product must be
returned with a clear written explanation of the
fault or problem.
This warranty does not apply, at the discretion of
the manufacturer, in the following cases:
•Improper installation or operation of the
device.
•Use of the product for purposes other than
those specified in this user manual.
•Damage due to failure to follow instructions.
(Continuation of page17)

19
•Damage due to unauthorized repair, modification or reconfiguration performed on the device.
•Damage caused by falls, hits, lack of proper care or maintenance.
•Damage caused by abnormal physical or electrical stress or defects of the electric supply (battery cell) or of equipment.
•If the serial number is altered, deleted, removed or rendered illegible.
Table 8: Emission table for IEC 60601-1-2
Guidance and manufacturer’s declaration – electromagnetic emissions
SpiroHome®battery-operated spirometer devices are intended for use in the electromagnetic environment specified below. Users of these devices should assure
that it is used in such an environment.
Emission Test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
Class B
The SpiroHome®battery-operated devices use RF
energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Harmonic emissions
IEC 61000-3-2 Not applicable Emissions are not applicable because SpiroHome®
does not connect to mains supply but operates with
AAA batteries.
Voltage fluctuations / flicker emissions IEC
61000-3-3 Not applicable
7.ELECTROMAGNETIC COMPATIBILITY
Meeting the requirements for EMC (electromagnetic compatibility) and preventing the unsafe use of the device, medical devices including SpiroHome®
manufactured by Inofab Health Technologies conform to the EN60601-1-2 standard which defines the levels of immunity to electromagnetic interference
as well as maximum levels of electromagnetic emissions for medical devices.
•For details, please see the following tables:

20
Table 9: Immunity (Stimulation mode) table according to IEC 60601-1-2
Guidance and manufacturer’s declaration – electromagnetic immunity
SpiroHome®battery-operated spirometer devices are intended for use in the electromagnetic environment specified below. Users of these devices should assure
that it is used in such an environment.
Immunity Test Standard IEC 60601 test level Compliance level Recommended separation distance
Electrostatic discharge (ESD)
IEC 61000-4-2
±2 kV
±4 kV
±6 kV
±8 kV
±15 kV
±8 kV contact
±2 kV air
±4 kV air
±8 kV air
±15 kV air
Floor should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Electrostatic fast transient / burst
IEC 61000-4-4 NA NA
Surge IEC 61000-4-5 NA NA
Voltage dips, short interruptions and voltage variations on power
supply lines IEC 61000-4-11 NA NA
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 30 A/m 30 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical domestic, commercial or hospital
environment.
Other manuals for SPIROHOME
1
Table of contents
Other inofab Personal Care Product manuals