
6
Thermo Ion Max—DART Interface Manual
Attaching the DART Flange to the Instrument
Attach the IonSense flange to the source by matching the lock bars with the slots on the
side and then locking them on the top by turning the knobs simultaneously until they are
facing at least 45 degrees from the side plane of the flange.
Insert the ceramic tube into the Swagelok nut on the end of the VAPUR attachment and
secure it with the graphite ferrule. Take care that there is a 2mm gap inside the VAPUR
between the end of the ceramic tube and the protruding end of the Agilent capillary.
This ensures that the excess helium gas will be evacuated and maintain proper vacuum in
the instrument.
Then, attach the DART either by 1) attaching the large front ring of the DART to the
holes on the front of the flange, or 2) by screwing the base of the DART into the two
holes on the bottom of the flange.
Attach the black silicone rubber
tubing to the port on the left side
of the flange.
Finally, attach the DART source
to the flange.
Figure: The DART is shown attached to the flange using the screws in the front ring.