iReliev ET-1616 Parts list manual

Instruction & Operating Manual
Read Before Using
MODEL # ET-1616

IMPORTANT INFORMATION
iReliev®Conductive Back Wrap is particularly safe and user-friendly.
When the Wrap is fastened and the electrode pads are in contact
with the skin, the intensity of the iReliev®TENS device may be
adjusted.
iReliev®Conductive Back Wrap
Accessory to iReliev®TENS Model # ET-1313 and TENS + EMS
Model # ET-7070. Also serves as replacement for ET-9090 Back Pain
Relief System.
Table of Contents
Introduction 3
Safety Information 4-5
Package Contents 6
Parts Diagram/Features 7
Putting on the Wrap 8-9
Troubleshooting . 10
Care and Maintenance 11
Warranty . 12
2

3
The iReliev®CONDUCTIVE BACK WRAP has been designed for people who
suffer from muscular tension and pain in their back. It provides a natural,
drug-free alternative that treats back pain the natural way with no side ef-
fects.
Your iReliev®CONDUCTIVE BACK WRAP is an accessory to your Transcutane-
ous Electrical Nerve Stimulation (TENS) or TENS + EMS device. It delivers elec-
trotherapy treatment in the low frequency range. When using this method,
electrical impulses are dispersed onto the skin using two or four electrodes.
This stimulates muscles, nerve fibers and acupuncture meridians in order to
activate your body’s own pain relief mechanisms.
Chronic pain can be treated with different methods such as medication, in-
jections, surgical procedures or with alternative treatment methods such as
acupuncture or TENS.
Electrotherapy treatment can block pain signals by interrupting the trans-
mission of impulses through the skin by means of electrical impulses. As a
result, the signal interpreted as “pain” is not transmitted to the brain and thus
not felt or greatly reduced.
Furthermore, it will work with the body’s own pain inhibitors as gentle electric
currents from the device stimulates the release of endorphins that function
as a pain reducer. The pain is therefore reduced or eased.
Electrical stimulation cannot remove the causes of pain or cure a disease,
but it could help ease or block pain.
In order to achieve all the benefits from the iReliev®Conductive Back Wrap,
carefully read the following safety information and warnings before use.

Safety Information
Use this device only according to its intended purpose as specified in
this instruction manual. The warranty will be invalidated if the device is
used for other than its intended purpose.
Do not use the device if it is not operating correctly, or if it has fallen into
water or has been damaged.
Do not attempt to repair the device yourself in the event of a malfunc-
tion; this will invalidate the warranty.
Do not operate this device near high-frequency transmitters, such as
microwaves, short wave transmitters or radio transmitters. This may
impair the device’s ability to function.
Do not turn on the device until the wrap has been secured on the body.
Do not use the device while taking a bath or a shower.
Do not use this device while sleeping.
Do not use this device while driving a car or another kind of vehicle or
while operating machinery or heavy equipment.
Protect the device from moisture. If liquid should get into the device,
remove the battery immediately and discontinue use. Should this hap-
pen, contact iReliev®at 855-723-2582.
Never use the device in rooms where aerosols (sprays) are being used.
Do not use with any creams or lotions.
Please use iReliev conductive gel included. Replacement gel may be
purchased at your authorized iReliev®retailer or at iReliev.com.
Do not use with a heating pad or any other electrical device.
4

5
Caution, the device should not be used by patients with:
Pacemakers, defibrillators or extreme cardiac irregularities
Metal implants or electronic auxiliary devices
Diabetes or abnormally high blood pressure
A tendency toward internal bleeding
Pregnant women
Therapy with the iReliev®CONDUCTIVE BACK WRAP is a proven and natural
way of fighting pain; however, it does not replace medical diagnosis or treat-
ment. If you are not sure, consult your doctor before using this device.
Consult your doctor always before use, if you suffer from epilepsy or
certain disabilities.
People whose sensitivity is disturbed or impaired should only apply this
device under supervision of your physician.
Do not use this device if suffering from menstrual pains.
Do not use this device if your skin burns or is numb.
Operation of this device should only be carried out under the careful su-
pervision and instruction of a doctor or physical therapist.
Do not treat parts of the body that are swollen, burned, inflamed, peel-
ing skin, wounds or other sensitive areas.
In case of doubt, consult a doctor or medical professional for instruc-
tions and guidance or contact iReliev for additional operating instruction.
Keep out of reach of children.
Treatment should be monitored if used on or near children or persons
who are handicapped.
If you feel pain or have an unpleasant feeling, stop use of this treat-
ment and consult your doctor.

Package Contents
Before using this accessory, make sure that you have all components.
The iReliev®Conductive Back Wrap Includes:
1. 1 Conductive Back Wrap
2. 1 Conductive Back Wrap Extension
3. 2 Lead Wires
4. 1 Conductive Gel Tube
5. 1 Instruction/User Manual
The packaging can be reused or recycled. Please dispose of properly if no longer re-
quired. If you notice any damage from shipping, please contact ExcelHealth Inc. at
855-723-2582.
6
1
2
45
3

7
Parts Diagram/Features
Description:
1.) Lead Wire Cables (Coiled Type)
2.) Conductive Silicone Pads
3.) iReliev®Conductive Back Wrap
4.) Hook and Loop Closure

Putting on the iReliev®Conductive Back Wrap
1) Before putting on the Conductive Back Wrap, remove any residue on skin, cream or
ointment. Make sure that your skin is clean. Do not place the wrap on injured or inflamed
areas of the skin, such as wounds, sores, rashes or reddening.
2) Before use, apply iReliev®Conductive Gel. Or you may dampen the integrated silicone
electrodes with water but use extreme caution to not oversaturate.
3) Position the Conductive Back Wrap so that the electrode pads are on the area of the
body to be treated. If your lower back aches at waist level, position the wrap so that the
middle of the wrap sits approximately 1 to 2 inches below the waist. The wrap should fit
very snugly around your body as follows:
8

9
4) When the wrap has been placed in the correct position, bring the two ends of the neo-
prene straps together and fasten with the built-in hook and loop closure so that it is snug.
5) The wrap may be tightened or loosened as shown by adjusting the hook and loop clo-
sure. (See diagram below)
▲NOTE/CAUTION: Subjective perception of the electric current may alter with changing
frequencies or pulse widths. Lower the intensity as soon as the application becomes un-
pleasant or the pleasant prickling sensation is not felt for a long period of time.
Please ensure that iReliev®Conductive Gel has been applied over the silicone pads to al-
low for optimal conductive impulse dispersion. If this step is not completed, impulse sen-
sation may feel like a prickling sensation and may be uncomfortable.

PROBLEM
The device does
not turn on.
The device turns on
and then off again.
The device turns
on, but does not
generate electric
pulses.
The unit does not
turn on even though
new batteries have
been inserted.
REMEDY
Replace batteries.
Insert battery again.
Replace battery.
Contact ExcelHealth
at 855-723-2582.
Reconnect lead wires to
device & wrap.
Connect cable properly
and power on.
Securely fasten wrap
on body.
Contact ExcelHealth
at 855-723-2582.
CAUSE
No battery or battery life
expired.
Battery not inserted
properly.
Battery life expired.
Lead wires not connected.
Cable not connected
properly or treatment time
has expired.
Wrap not securely fastened.
Troubleshooting
Always check the unit and accessories before use to prevent damage and de-
fects; these are some of the simple checks:
1.) Make sure the battery on the iReliev®TENS Device has sufficient charge and
is not corroded.
2.) Make sure the cables fit tightly into the connection sockets of the device. The
table below shows some common defects. If you cannot remedy the defects
as described, contact iReliev®at 855-723-2582.
10
The sensation is
uncomfortable and feels
like a “prickling” sensation.
Electrode pads are not
moistened.
Apply iReliev®Conductive
Gel for optimal dispersion
of electrical current.

11
Care and Maintenance
Wrap:
If the wrap has become damp during use, let it dry naturally before us-
ing again.
Do not wash the wrap in water. Do not wash the wrap in the washing
machine.
Lay the wrap out flat or hang it up to dry naturally. Do not tumble dry.
The wrap may be cleaned using a lightly dampened sponge.
Do not use bleach when washing the wrap.
Do not dry clean your wrap. Do not dry it over anything hot (e.g. a radia-
tor) as it contains plastic parts. Ensure the wrap is completely dry before
using it again. The wrap should never be ironed.
Always detach the lead wires from snap and the unit from the lead
wires. Do not pull on the lead wire cables but on the connectors at-
tached to the ends of the cables.

Limited One (1) Year Warranty
This item is warranted to be free of defects in materials or workmanship for
(1) year from the date of purchase, provided the product has been used un-
der normal conditions of intended use. If this product is defective, warrantor
will replace the product or refund the purchase price of the product at the
place of purchase upon presentation of sales receipt and original UPC code
from packaging. This warranty extends only to the original purchaser, and
excludes any damage to the product resulting from accident, misuse, abuse,
damage caused by operation in ways not recommended or authorized by
these operating and safety instructions. This warranty is voided if the prod-
uct is ever used in a commercial or business environment.
There are no warranties beyond this limited warranty. This limited warran-
ty is your complete and exclusive legal means. Warrantor makes no other
warranties, express or implied, including but not limited to any implied war-
ranty of merchantability or of fitness for a particular purpose. Warrantor un-
dertakes no responsibility for the quality of the product except as otherwise
provided herein. Warrantor assumes no responsibility that the product will
be fit for any particular purpose for which you may be buying this product.
Warrantor expressly disclaims liability for any special, incidental, indirect, or
consequential damages arising out of the purchase or use of this product.
The amount of Warrantor’s liability under this warranty is limited to the
amount of the original purchase price paid for the product by the original re-
tail purchaser. If Warrantor cannot lawfully disclaim statutory or implied war-
ranties, then all such warranties shall be limited in duration to the duration
of this warranty and shall be limited to the amount of the original purchase
price paid for the product by the original retail purchaser.
This warranty gives you specific legal rights; you may also have other rights
that vary by jurisdiction. This warranty is valid only within the United States
of America (USA) and Canada.
Model # ET-1616
Return to:
ExcelHealth Inc.
1603 Hart St.
Southlake, TX 76092
©2017 iReliev®, All Rights Reserved
www.iReliev.com
855-723-2582
12


ExcelHealth Inc.
www.iReliev.com
1603 Hart St.
Southlake, TX 76092
1-855-723-2582