USERS MANUAL
Contents:
1. Short description of signalling apparatus
2. Operating description
3. Technical data
4. Characteristic data of signalling apparatus
5. General data
6. Wiring of instrument
1. SHORT DESCRIPTION OF SIGNALER
This is electronic signalling instrument with analog display.
Electronic is mounted in plastic housing. On the signaller front the
analog instrument displays measuring value. Signalling LED-s MAX
ΙΙ and MIN Ιindicates when measuring value is greater or lover then
limit value. On the back side of the instrument are two potentiometers
for adjusting a lower limit (CHANNEL Ι) and upper limit
(CHANNEL ΙΙ) of signaller. Also there are potentiometers for delay
time adjusting (DELAY Ιand DELAY ΙΙ).
2. OPERATING DESCRIPTION
Versions of signalling instrument provide use in various applications.
Versions of inputs provide measuring and signalling of different
physical quantities. Relay outputs can control signalling or control
elements. Limit value for lower and upper limit can be adjusted on
rear of the instrument by potentiometers in range of 0 ... 100%.
Delay of activating of relay outputs can be adjusted on rear by
potentiometers in range 0,5 ... 30 s. Pointer on scale displays value of
measuring input. LED on scale illuminates when input value becomes
greater then MAX ΙΙ limit value or lower then MIN Ιlimit value.
After delay time interval if input value stays out of limit relay
energies. Relay ΙΙ deenergizes when input value is lover then MAX ΙΙ
limit value. Relay Ιdeenergizes when input value is greater then MIN
Ιlimit value.
3. TECHNICAL DATA
3.1.INPUTS
3.1.1 DC voltage and DC current inputs
Measuring system: Instrument with moving coil for
measuring DC current.
DC V - meter ranges: 40 mV ... 600V
Accuracy class: 1,5 %
Input resistance: > 10 MΩfor ranges 40 mV ... 4 V
> 1 MΩfor ranges 5 V ... 600 V
DC A - meter ranges: 25 µA ... 5 A
Accuracy class: 1,5 %
Input resistance is range dependent on ranges.
Shunt voltage is 100 mV
3.1.2 AC voltage and AC current inputs
Measuring system: Instrument with moving coil and
rectifier for measuring AC current and
voltages.
AC V - meter ranges: 100 mV ... 600 V
AC A - meter ranges: 1 mA ... 5 A
Accuracy class: 1,5 %
Frequency range: 40 ... 400 Hz
Input resistance: > 1 MΩ; < 50 pF
3.1.3 Measuring of AC effective values
Measuring system: Instrument with moving iron for
effective values measuring.
AC Vef - meter ranges: 6 V ... 600 V
AC Aef - meter ranges: 100 mA ... 5 A (5/10 A)
Accuracy class: 1,5 %
Frequency range: 30 ... 100 Hz
NOTE: The signalling electronic is calibrated on r.m.s. values, at
a sine shaped wave
3.1.4 Frequency meter
Measuring ranges: 45 ... 55 Hz Accuracy class: 0,5 %
48 ... 52 Hz Accuracy class: 0,5 %
45 ... 65 Hz Accuracy class: 1 %
55 ... 65 Hz Accuracy class: 0,5 %
58 ... 62 Hz Accuracy class: 0,5 %
Input voltage: 60 ... 500 V AC
3.1.5 Thermometer with thermocouple probe
Temperature probe: thermocouple - IEC 584-1
Measuring ranges: 0 ... 250 °C ( Fe - const) J
0 ... 600 °C ( Fe - const) J
0 ... 1200 °C ( NiCr - Ni) K
0 ... 1600 °C ( PtRh - Pt) S
Temperature compensation: built-in, reference temperature 20 °C
Accuracy class: 1,5 %
3.1.6 Thermometer with Pt 100 resistance probe
Temperature probe: Pt 100 - IEC 751
Measuring ranges: -200 ... 0 ...+ 50 °C
0 ... 250 °C
200 ... 450 °C
400 ... 650 °C
Connecting: Two wires
Accuracy class: 1,5 %
3.2 OUTPUTS
Signaller has two output relays.
Switching element: potential free alternating relay
contacts. Maximal switching power at
resistive load: ≤600 VA (≤3 A,
≤250 V)
- Output CHANNEL Ι: adjustable by MIN limit value
- Output CHANNEL ΙΙ: adjustable by MAX limit value
4. CHARACTERISTIC DATA OF SIGNALING APPARATUS
4.1 LIMIT VALUE AND DELAY TIME ADJUSTING
Adjusting elements are on rear side of housing:
Range of limit value adjusting (MIN,
MAX) 0 ... 100 % F.S.D.
Setting accuracy: ±5 % (25 ... 75 %), ±15 % (0...25 %,
75 ... 100 %)
Setting reproducibility: < 2 %
Hysteresis: < 1 % F.S.D
Delay time adjusting range: 0,5 ... 30 s
Setting accuracy: ±20 % ±2 s
Setting reproducibility: < 2 %
ATTENTION: Isolate supplies before adjusting % F.S.D. or DELAY
settings!
MI7350
ELECTRONIC SIGNALER